STERRAD 100NX STERILIZER

K071385 · Advanced Sterilization Products · MLR · Nov 29, 2007 · General Hospital

Device Facts

Record IDK071385
Device NameSTERRAD 100NX STERILIZER
ApplicantAdvanced Sterilization Products
Product CodeMLR · General Hospital
Decision DateNov 29, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6860
Device ClassClass 2

Indications for Use

The STERRAD® 100NX™ Sterilizer is designed for sterilization of both metal and nonmetal medical devices at low temperatures. The STERRAD sterilization process is a multiphase medical "devices" at 2011" withing a combination of exposure to hydrogen peroxide vapor and plasma to affect sterilization. The STERRAD® 100NX™ Sterilizer can sterilize instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors. Medical devices with the following materials and dimensions can be processed in the STERRAD 100NX™ Sterilizer Standard cycle: Single channel stainless steel lumens with an inside diameter of 0.7 mm or larger and a length of 500 mm or shorter. * Medical devices, including most flexible endoscopes, with the following materials and Modical as nocessed in the STERRAD® 100NX™ Sterilizer Flex Scope cycle: Single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscopes with an inside diameter of 1 mm or larger and length of 850 mm or shorter. ** Note: With the exception of the 1 x 850 mm flexible endoscopes, the validation studies were performed using a validation load consisting of two instrument trays each weighing 10.7 Ibs. The 1 x 850 mm flexible endoscopes were validated without any additional load. * A maximum of ten single channel stainless steel lumens, five per tray per sterilization cycle. ** A maximum of two flexible endoscopes, one per tray per sterilization cycle. No additional load.

Device Story

STERRAD 100NX Sterilizer is a self-contained, stand-alone system for low-temperature sterilization of medical instruments. System utilizes a multiphase process: injection of aqueous hydrogen peroxide into a vaporizer; introduction of vapor into a chamber under sub-ambient pressure; transformation into gas-plasma via electrical energy. Device features two cycles: Standard and Flex Scope. Hardware includes sterilization chamber, covered frame, disposable sterilant cassettes, reusable instrument trays, and optional cart. Operated in clinical settings; provides automated sterilization to ensure instrument safety for reuse. Benefits include effective sterilization of heat-sensitive and diffusion-restricted devices without high-heat damage.

Clinical Evidence

Bench testing only. Validation performed using the "overkill" approach with G. stearothermophilus biological indicators to confirm sterilization efficacy across specified lumen dimensions and load configurations.

Technological Characteristics

Hydrogen peroxide gas plasma sterilization system. Components: sterilization chamber, vaporizer, electrical energy source for plasma generation. Materials: stainless steel, polyethylene, Teflon. Connectivity: standalone system. Sterilization: chemical vapor/plasma. Software: embedded system control.

Indications for Use

Indicated for sterilization of metal and nonmetal medical devices, including flexible endoscopes (polyethylene/Teflon) and instruments with diffusion-restricted spaces (hinged forceps/scissors), at low temperatures.

Regulatory Classification

Identification

An ethylene gas sterilizer is a nonportable device intended for use by a health care provider that uses ethylene oxide (ETO) to sterilize medical products.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ NOV 2 9 2007 ## (071385 510(k) Summary Applicant's Name, Address, Telephone, FAX, Contact Person Advanced Sterilization Products Division of Ethicon, Inc. 33 Technology Drive Irvine, CA 92618 ## Contact Person Reuben Lawson Regulatory Affairs Manager Tel: (949) 789-8545 Fax: (949) 789-3900 May 15, 2007 CLASSIFICATION, COMMON OR USUAL NAME, DEVICE NAME 1.0 Classification Name: Sterilizer, Class II Common/Usual Name: Hydrogen Peroxide Gas Plasma Sterilization System Product Classification: Sterilizer, Class II Proprietary Name: STERRAD® 100NXTM Sterilizer ### 2.0 PREDICATE DEVICES STERRAD® NX® Sterilization System [K042116] #### 3.0 INDICATIONS FOR USE The STERRAD 100NXTM Sterilizer is designed for sterilization of both metal and nonmetal medical devices at low temperatures. The STERRAD® sterilization process is a multiphase sterilization process that utilizes a combination of exposure to hydrogen peroxide vapor and plasma to affect sterilization. The STERRAD® 100NX™ Sterilizer can sterilize instruments which have diffusionrestricted spaces, such as the hinged portion of forceps and scissors. Medical devices with the following materials and dimensions can be processed in the STERRAD® 100NX™ Sterilizer Standard cycle: - > Single channel stainless steel lumens with an inside diameter of 0.7mm or larger and a length of 500mm or shorter* Flexible endoscopes, with the following materials and dimensions can be processed in the STERRAD® 100NX™ Sterilizer Flex Scope cycle: {1}------------------------------------------------ - A flexible endoscope with a polyethylene or Teflon Single channel (polytetrafluoroethylene) lumen with an inside diameter of 1mm or larger and length of 850mm or shorter** Note: With the exception of the 1 x 850 mm flexible endoscopes, the validation studies were performed using a validation load consisting of two instrument trays each weighing 10.7 Ibs. The 1 x 850 mm flexible endoscopes were validated without any additional load. * A maximum of ten lumens, five per tray per sterilization cycle. ** A maximum of two flexible endoscopes, one per tray per sterilization cycle. No additional load. #### 4.0 DESCRIPTION OF DEVICE The STERRAD® 100NXTM Sterilizer is a self-contained stand-alone system of hardware and software designed to sterilize medical instruments and devices using a patented hydrogen peroxide gas plasma process. Hydrogen peroxide vapor is generated by injecting aqueous hydrogen peroxide into the vaporizer where the solution is heated and vaporized. introducing the vapor into the chamber under sub-ambient pressure and transforming the vapor into a gas-plasma using electrical energy. The STERRAD® 100NX™ Sterilizer has two different sterilization cycles, the Standard cycle and the Flex cycle. The hardware for the STERRAD® 100NX™ Sterilizer consists of a sterilization chamber and a variety of instruments and components which are housed in a covered frame. The sterilizer system also uses accessories such as a disposable sterilant cassette, reusable instrument trays, printer paper, and an optional movable cart. The STERRAD® 100NXTM Sterilizer can be placed directly on a table, counter top, or on the movable cart. #### 5.0 SUMMARY OF NONCLINICAL TESTS #### 5.1 Validation Testing Testing was performed using the "overkill" approach utilizing G. stearothermophilus spores. Table 8-1 on the following page identifies the validation studies performed and the results obtained. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines representing its body and wings. The eagle is facing left. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 NOV 2 9 2007 Mr. Reuben Lawson Regulatory Affairs Manager Advanced Sterilization Products, Incorporated 33 Technology Drive Irvine, California 92618 Re: K071385 Trade/Device Name: STERRAD 100NX Regulation Number: 21 CFR 880.6860 Regulation Name: Ethylene Oxide Gas Sterilizer Regulatory Class: II Product Code: MLR Dated: November 12, 2007 Received: November 13, 2007 Dear Mr. Lawson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Lawson Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Clars Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use ## 510(k) Number (if known): K071385 Device Name: STERRAD 100NX Indications for Use: The STERRAD® 100NX™ Sterilizer is designed for sterilization of both metal and nonmetal medical devices at low temperatures. The STERRAD sterilization process is a multiphase medical "devices" at 2011" withing a combination of exposure to hydrogen peroxide vapor and plasma to affect sterilization. The STERRAD® 100NX™ Sterilizer can sterilize instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors. Medical devices with the following materials and dimensions can be processed in the STERRAD 100NX™ Sterilizer Standard cycle: Single channel stainless steel lumens with an inside diameter of 0.7 mm or larger and a length of 500 mm or shorter. * Medical devices, including most flexible endoscopes, with the following materials and Modical as nocessed in the STERRAD® 100NX™ Sterilizer Flex Scope cycle: Single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscopes with an inside diameter of 1 mm or larger and length of 850 mm or shorter. ** Note: With the exception of the 1 x 850 mm flexible endoscopes, the validation studies were performed using a validation load consisting of two instrument trays each weighing 10.7 Ibs. The 1 x 850 mm flexible endoscopes were validated without any additional load. * A maximum of ten single channel stainless steel lumens, five per tray per sterilization cycle. ** A maximum of two flexible endoscopes, one per tray per sterilization cycle. No additional load. Over-The-Counter Use _ X Prescription Use AND/OR I reseription Coo ____________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) > Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K071385
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...