K220404 · Advanced Sterilization Products, Inc. · MLR · May 16, 2022 · General Hospital
Device Facts
Record ID
K220404
Device Name
STERRAD NX with ALLClear Technology
Applicant
Advanced Sterilization Products, Inc.
Product Code
MLR · General Hospital
Decision Date
May 16, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The STERRAD NX Sterilizer is designed for sterilization of both metal and nonmetal medical devices at low temperatures. The STERRAD Sterilization Process is a multiphase sterilization process that utilizes a combination of exposure to hydrogen peroxide vapor and plasma to affect sterilization. The STERRAD NX Sterilizer can sterilize instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors. Medical devices with the following materials and dimensions can be processed in the STERRAD NX Sterilizer STANDARD Cycle: · Single-channel stainless steel lumens with an inside diameter of 1 mm or larger and a length of 150 mm or shorter. † · Single-channel stainless steel lumens with an inside diameter of 2 mm or larger and a length of 400 mm or shorter. † The validation testing for this lumen size was conducted using a maximum of 10 lumens per load. Your loads should not exceed the maximum number of lumens validated by this testing. Medical devices, including most flexible endoscopes, with the following materials and dimensions can be processed in the STERRAD NX Sterilizer ADVANCED Cycle: · Single-channel stainless steel lumens with an inside diameter of 1 mm or larger and a length of 500 mm or shorter. † · Single-channel PE/PTFE flexible endoscopes with an inside diameter of 1 mm or larger and a length of 1065 mm or shorter.* Note: With the exception of the flexible endoscopes, the validation studies were performed using a validation load consisting of devices in one instrument tray weighing 10.7 lbs (4.9 kg). The flexible endoscopes were validated without any additional load. ** † The validation testing for this lumen size was conducted using a maximum of 10 lumens per load. Your loads should not exceed the maximum number of lumens validated by this testing. * Only one flexible endoscope up to 1065 mm long with or without APTIMAX Instrument Tray Holders can be used per cycle, or only one flexible endoscope up to 850 mm long per cycle with or without a STERRAD Instrument Tray Mat. No additional load. Check the medical device manufacturer's instructions for use prior to processing any scope in the STERRAD NX Sterilizer; and check if Instrument tray mats or only tray holders are validated for use with flexible endoscopes longer than 850 mm. ** The validation testing for flexible endoscopes up to 850 mm long was conducted with one endoscope per cycle using a STERRAD Instrument Tray Mat to help protect the endoscope no additional load. The validation testing for flexible endoscopes longer that 850 mm was conducted with one endoscope per cycle using APTIMAX Instrument Tray Holders to assist with device placement and help protect the endoscope: no additional load.
Device Story
STERRAD NX Sterilizer is a self-contained, stand-alone system for low-temperature sterilization of medical instruments. It utilizes a multiphase process: injection of aqueous hydrogen peroxide into a vaporizer, introduction of vapor into a chamber under sub-ambient pressure, and transformation into gas plasma via electrical energy. The system includes a hydrogen peroxide monitor that cancels cycles if concentration specifications are not met. Hardware includes an aluminum chamber, covered frame, and accessories like instrument trays and an optional cart. It is operated in clinical settings to sterilize heat- and moisture-sensitive devices. The healthcare provider loads instruments, selects a cycle (STANDARD or ADVANCED), and the device provides automated sterilization. This ensures effective decontamination of complex, diffusion-restricted instruments, reducing infection risk for patients.
Clinical Evidence
No clinical data was generated. Evidence consists of non-clinical performance testing: dose response studies and simulated use testing. Results demonstrated a spore log reduction > 6 for all test devices and successful achievement of SAL 10^-6 for single-channel flexible endoscopes (≥1mm diameter x ≤1065mm length) in the ADVANCED cycle.
Technological Characteristics
Sterilization via hydrogen peroxide gas plasma. Aluminum chamber. System uses thermistors for temperature monitoring and pressure transducers/switches for pressure monitoring. Disposable 59% hydrogen peroxide cassette. Stand-alone unit. Software-controlled cycle management with automated hydrogen peroxide concentration monitoring.
Indications for Use
Indicated for sterilization of metal and nonmetal medical devices, including single-channel stainless steel lumens and single-channel PE/PTFE flexible endoscopes (up to 1065mm length, ≥1mm diameter), at low temperatures using hydrogen peroxide gas plasma.
Regulatory Classification
Identification
An ethylene gas sterilizer is a nonportable device intended for use by a health care provider that uses ethylene oxide (ETO) to sterilize medical products.
Predicate Devices
STERRAD® NX Sterilizer with ALLClear® Technology (K160818)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
May 16, 2022
Advanced Sterilization Products, Inc. Katy Nennig Senior Regulatory Affairs Specialist 33 Technology Drive Irvine, California 92618
Re: K220404
Trade/Device Name: STERRAD® NX Sterilizer with ALLClear™ Technology Regulation Number: 21 CFR 880.6860 Regulation Name: Ethylene Oxide Gas Sterilizer Regulatory Class: Class II Product Code: MLR Dated: February 6, 2022 Received: February 14, 2022
Dear Katy Nennig:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220404
#### Device Name
STERRAD® NX™ Sterilizer with ALLClear Technology™
#### Indications for Use (Describe)
The STERRAD NX Sterilizer is designed for sterilization of both metal and nonmetal medical devices at low temperatures. The STERRAD Sterilization Process is a multiphase sterilization process that utilizes a combination of exposure to hydrogen peroxide vapor and plasma to affect sterilization. The STERRAD NX Sterilize instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors.
Medical devices with the following materials and dimensions can be processed in the STERRAD NX Sterilizer STANDARD Cycle:
· Single-channel stainless steel lumens with an inside diameter of 1 mm or larger and a length of 150 mm or shorter. ↑
· Single-channel stainless steel lumens with an inside diameter of 2 mm or larger and a length of 400 mm or shorter. ↑ 7 The validation testing for this lumen size was conducted using a maximum of 10 lumens per load. Your loads should not
exceed the maximum number of lumens validated by this testing.
Medical devices, including most flexible endoscopes, with the following materials and dimensions can be processed in the STERRAD NX Sterilizer ADVANCED Cycle:
· Single-channel stainless steel lumens with an inside diameter of 1 mm or larger and a length of 500 mm or shorter. ↑
· Single-channel PE/PTFE flexible endoscopes with an inside diameter of 1 mm or larger and a length of 1065 mm or shorter.*
Note: With the exception of the flexible endoscopes, the validation studies were performed using a validation load consisting of devices in one instrument tray weighing 10.7 lbs (4.9 kg). The flexible endoscopes were validated without any additional load. **
f The validation testing for this lumen size was conducted using a maximum of 10 lumens per load. Your loads should not exceed the maximum number of lumens validated by this testing.
* Only one flexible endoscope up to 1065 mm long with or without APTIMAX Instrument Tray Holders can be used per cycle, or only one flexible endoscope up to 850 mm long per cycle with or without a STERRAD Instrument Tray Mat. No additional load. Check the medical device manufacturer's instructions for use prior to processing any scope in the STERRAD NX Sterilizer; and check if Instrument tray mats or only tray holders are validated for use with flexible endoscopes longer than 850 mm.
** The validation testing for flexible endoscopes up to 850 mm long was conducted with one endoscope per cycle using a STERRAD Instrument Tray Mat to help protect the endoscope no additional load. The validation testing for flexible endoscopes longer that 850 mm was conducted with one endoscope per cycle using APTIMAX Instrument Tray Holders to assist with device placement and help protect the endoscope: no additional load.
Type of Use (Select one or both, as applicable)Prescription Use (Part 21 CFR 801 Subpart D)
|X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Advanced Sterilization Products (ASP). The logo consists of the letters "ASP" in a stylized, connected font, with the words "Advanced Sterilization Products" written to the right of the letters. The letters and words are in a dark blue color. The logo is simple and professional, and it is likely used to represent the company's brand.
# 510(k) Summary Advanced Sterilization Products, Inc. STERRAD® NX Sterilizer with ALLClear™ Technology ADVANCED Cycle Claims Expansion
This summary of 510(k) information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
#### General Information
| Submitter Name: | Advanced Sterilization Products, Inc. |
|-----------------|------------------------------------------------------------------------------------------------------------|
| Address: | 33 Technology Drive<br>Irvine, CA 92618 |
| Contact Person: | Katy Nennig<br>Senior Regulatory Affairs Specialist<br>Tel: (920)254-6370<br>Email: Katelyn.nennig@asp.com |
| Date Prepared: | May 11th, 2022 |
## Device Name
| Proprietary Name: | STERRAD® NX Sterilizer with ALLClearTM Technology |
|----------------------|---------------------------------------------------|
| Common Name: | Hydrogen Peroxide Gas Plasma Sterilization System |
| Classification Name: | Ethylene oxide gas sterilizer |
| Device Class: | Class II |
| Product Code: | MLR |
| CFR Section: | 21 CFR 880.6860 |
## Predicate Device
STERRAD® NX Sterilizer with ALLClear® Technology cleared via 510(k) K160818 on September 27, 2016.
#### Device Description
The STERRAD NX Sterilizer is a self-contained stand-alone system of hardware and software designed to sterilize medical instruments and devices using a patented hydrogen peroxide gas plasma process. Hydrogen peroxide vapor is generated by injecting aqueous hydrogen peroxide into the vaporizer where the solution is heated and vaporized, introducing the vapor into the chamber under sub-ambient pressure and transforming the vapor into a gas plasma using electrical energy. The STERRAD NX Sterilizer has two cleared sterilization cycles, the STANDARD and ADVANCED Cycles.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Advanced Sterilization Products (ASP). The logo is in a dark blue color. The letters "ASP" are in a bold, sans-serif font, with the "S" being larger than the "A" and "P". To the right of the letters, the words "Advanced Sterilization Products" are written in a smaller, sans-serif font.
The sterilizer uses a disposable sterilant cassette that contains the 59% nominal hydrogen peroxide solution in a plastic cell pack and cassette shells. The hydrogen peroxide is concentrated before into the sterilizer chamber and its concentration is monitored during the cycle. The sterilizer cancels the cycle if the hydrogen peroxide monitor data does not meet the specification.
The hardware for the STERRAD NX Sterilizer consists of a sterilizer chamber, constructed with aluminum, and a variety of instruments and components which are housed in a covered frame. The sterilizer also uses accessories such as reusable instrument trays, printer paper, and an optional movable cart. The STERRAD NX Sterilizer can be placed directly on a table, countertop, or on the movable cart.
An expansion of existing claims is being applied to the ADVANCED Cycle without affecting the technology, software, or other physical features of the subject device.
The STERRAD NX Sterilizer with ALLClear Technology described within this submission expands the indications of the ADVANCED Cycle of the Sterilizer to include single channel flexible endoscopes with extended dimensions of 1065mm in length when the inside diameter is ≥1mm. The previously cleared dimensions of the predicate device are 850mm in length when the inside diameter is ≥1mm. There are no other changes to the indications for use of the ADVANCED Cycle. The ADVANCED Cycle is compatible with single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscopes with an inside diameter of 1mm or larger and length of 1065mm or shorter2
## Intended Use/Indications For Use
The STERRAD NX Sterilizer is designed for sterilization of both metal and nonmetal medical devices at low temperatures. The STERRAD Sterilization Process is a multiphase sterilization process that utilizes a combination of exposure to hydrogen peroxide vapor and plasma to affect sterilization. The STERRAD NX Sterilizer can sterilize instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors.
Medical devices with the following materials and dimensions can be processed in the STERRAD NX Sterilizer STANDARD Cycle:
- . Single-channel stainless steel lumens with an inside diameter of 1 mm or larger and a length of 150 mm or shorter.t
- . Single-channel stainless steel lumens with an inside diameter of 2 mm or larger and a length of 400 mm or shorter.t
†The validation testing for this lumen size was conducted using a maximum of 10 lumens per load. Your loads should not exceed the maximum number of lumens validated by this testing.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Advanced Sterilization Products (ASP). The logo consists of the letters "ASP" in a stylized, bold, dark blue font. To the right of the letters, the words "Advanced Sterilization Products" are written in a smaller, lighter font. The logo is simple and professional, conveying a sense of trust and reliability.
Medical devices, including most flexible endoscopes, with the following materials and dimensions can be processed in the STERRAD NX Sterilizer ADVANCED Cycle:
- Single-channel stainless steel lumens with an inside diameter of 1 mm or larger and a length of 500 mm or shorter. †
- . Single-channel PE/PTFE flexible endoscopes with an inside diameter of 1 mm or larger and a length of 1065 mm or shorter. *
Note: With the exception of the flexible endoscopes, the validation studies were performed using a validation load consisting of devices in one instrument tray weighing 10.7 lbs (4.9 kg). The flexible endoscopes were validated without any additional load.**
†The validation testing for this lumen size was conducted using a maximum of 10 lumens per load. Your loads should not exceed the maximum number of lumens validated by this testing.
* Only one flexible endoscope up to 1065 mm long with or without APTIMAX Instrument Tray Holders can be used per cycle, or only one flexible endoscope up to 850 mm long per cycle with or without a STERRAD Instrument Tray Mat. No additional load. Check the medical device manufacturer's instructions for use prior to processing any scope in the STERRAD NX Sterilizer; and check if Instrument tray mats or only tray holders are validated for use with flexible endoscopes longer than 850 mm.
**The validation testing for flexible endoscopes up to 850 mm long was conducted with one endoscope per cycle using a STERRAD Instrument Tray Mat to help protect the endoscope no additional load. The validation testing for flexible endoscopes longer that 850 mm was conducted with one endoscope per cycle using APTIMAX Instrument Tray Holders to assist with device placement and help protect the endoscope: no additional load.
## Technological Characteristics Comparison
The following table provides a comparison of the technological characteristics associated with the sterilization process of the proposed STERRAD NX Sterilizer with ALLClear Technology with expanded indications for the ADVANCED Cycle as compared to those of the previously cleared STERRAD NX with ALLClear Technology.
| Device & Predicate Device(s): | K160818 | K220404 |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Designed for sterilization of both<br>metal and nonmetal medical<br>devices at low temperatures.<br>Because the cycle operates within a<br>dry environment and at low<br>temperatures, it is especially<br>suitable for instruments sensitive<br>to heat and moisture. | Designed for sterilization of both<br>metal and nonmetal medical<br>devices at low temperatures.<br>Because the cycle operates within a<br>dry environment and at low<br>temperatures, it is especially<br>suitable for instruments sensitive<br>to heat and moisture. |
| | Table 5.1 Technological Characteristics Comparison | |
|--|----------------------------------------------------|--|
| | | |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Advanced Sterilization Products (ASP). The logo consists of the letters "ASP" in a stylized, connected font, with the words "Advanced Sterilization Products" to the right of the letters. The letters are in a dark purple color, while the words are in a lighter gray color. The logo is simple and modern, and it effectively communicates the company's focus on sterilization products.
| Device & Predicate Device(s): | K160818 | K220404 |
|---------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sterilization Process | Hydrogen peroxide gas plasma | Hydrogen peroxide gas plasma |
| Principle of Operation | Combined use of hydrogen<br>peroxide vapor and plasma safely<br>and rapidly sterilizes most medical<br>instruments and materials with<br>leaving toxic residues. | Combined use of hydrogen<br>peroxide vapor and plasma safely<br>and rapidly sterilizes most medical<br>instruments and materials with<br>leaving toxic residues. |
| Pre-Programmed Sterilization<br>Cycles | Two Cycles:<br>STANDARD and ADVANCED | Two Cycles:<br>STANDARD and ADVANCED |
| Material Compatibility<br>(Recommended Materials) | Common materials found in<br>reusable medical devices. All<br>medical devices should be<br>processed in accordance with the<br>medical device manufacturer's<br>recommendations. | Common materials found in<br>reusable medical devices. All<br>medical devices should be<br>processed in accordance with the<br>medical device manufacturer's<br>recommendations. |
| Approximate Cycle Time | STANDARD Cycle: 28 Minutes<br>ADVANCED Cycle: 38 Minutes | STANDARD Cycle: 28 Minutes<br>ADVANCED Cycle: 38 Minutes |
| Hydrogen Peroxide Monitor | Cancels sterilization cycle if the<br>areas under the concentration-time<br>curve or rate constant do not meet<br>predetermined specifications. | Cancels sterilization cycle if the<br>areas under the concentration-time<br>curve or rate constant do not meet<br>predetermined specifications. |
| Temperature | Chamber, Chamber door and<br>vaporizer/condenser thermistors | Chamber, Chamber door and<br>vaporizer/condenser thermistors |
| Pressure | Chamber pressure transducers<br>Chamber atmospheric pressure<br>switch Vaporizer/condenser<br>pressure transducer | Chamber pressure transducers<br>Chamber atmospheric pressure<br>switch Vaporizer/condenser<br>pressure transducer |
## Summary of Non-Clinical Testing
Performance testing was conducted to verify that single channel flexible endoscopes with the expanded lumen claim of ≥ 1mm diameter x ≤ 1065mm length can be successfully sterilized by the ADVANCED Cycle of the STERRAD NX Sterilizer with ALLClear Technology. No clinical data was generated in support of this submission.
| Testing<br>Methodology | Description | Acceptance Criteria | Results<br>(Pass/Fail) |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Biocompatibility | The proposed changes to the indications for use<br>do not affect sterilization cycle parameters,<br>sterilant, sterilant injection volume, or types of<br>materials processed in the sterilizer. The<br>previously submitted biocompatibility data for<br>the predicate device (K160818) remains<br>applicable to the subject STERRAD NX Sterilizer<br>with ALLClear Technology. | Process residual levels remaining<br>on/in materials used in the<br>sterilizer shall be non-toxic when<br>evaluated by in vivo toxicity tests. | Pass |
#### Table 5.2 Performance Testing Results
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for Advanced Sterilization Products (ASP). The logo consists of the letters "ASP" in a stylized, bold, sans-serif font, with the letters connected. To the right of the letters, the words "Advanced Sterilization Products" are stacked vertically in a smaller, sans-serif font. The color of the logo is a dark blue or purple.
| Testing<br>Methodology | Description | Acceptance Criteria | Results<br>(Pass/Fail) |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Dose Response | Study demonstrated sterility assurance level of<br>(SAL) of 10-6 was reached when processing<br>single channel flexible endoscopes with<br>expanded claim (≥ 1 mm diameter X ≤ 1065<br>mm) in STERRAD NX ADVANCED half-cycle<br>conditions.<br>• Study demonstrated a total kill of BIs in the<br>half cycle condition at 1.5 mL of 53% hydrogen<br>peroxide injected.<br>• The results demonstrated an increasing<br>number of sterile BIs corresponding to an<br>increasing injection volume of hydrogen<br>peroxide.<br>• The positive control demonstrated growth and<br>the negative control demonstrated no growth. | • BI Samples from 1.5mL of 53%<br>hydrogen peroxide in half cycles<br>shall be sterile at the end of the<br>incubation period<br>• A trend of increasing number of<br>sterile BIs corresponding to the<br>increasing injection volume of<br>hydrogen peroxide must be seen.<br>• Positive control demonstrates<br>growth and negative control<br>demonstrates no growth | Pass |
| Simulated Use | Study demonstrated sterilization efficacy of<br>flexible endoscopes with proposed claims<br>expansion of ≥1 mm diameter x ≤1065 mm in<br>length when processed in the STERRAD NX<br>ADVANCED cycle.<br>• The spore log reduction was greater than 6<br>for all test devices.<br>• The positive controls ranged from 1.6 x 106 to<br>1.8 x 106 CFU | • Spore log reduction must be<br>greater than 6 for test devices.<br>• Positive control shall show a<br>combined exhaustive recovery of<br>at least 1 x 106 CFU | Pass |
The results of performance testing demonstrates that single channel flexible endoscopes with lumen dimensions of ≥ 1mm x ≤ 1065mm can be sterilized by the ADVANCED Cycle of the STERRAD NX Sterilizer.
#### Summary
The subject device, STERRAD NX Sterilizer with ALLClear Technology with expanded indications, and its predicate device utilize the same technology, sterilization cycles, and sterilization methods to sterilize medical devices. Based on the results of the performance testing, the indications of the ADVANCED Cycle does not raise any new questions of safety or effectiveness.
## Conclusion
The conclusions drawn from the non-clinical performance data demonstrate that the subject device, STERRAD NX with ALLClear Technology including the ADVANCED Cycle indications expansion, is as safe,
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for Advanced Sterilization Products (ASP). The logo consists of the letters "ASP" in a bold, sans-serif font, with the words "Advanced Sterilization Products" to the right of the letters. The letters are in a dark blue color, and the words are in a smaller, lighter font.
as effective, and performs as well or better than the legally marketed predicate device cleared via K160818.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.