K071164 · Amco International Manufacturing & Design, Inc. · MHX · Dec 7, 2007 · Cardiovascular
Device Facts
Record ID
K071164
Device Name
LIFE+CEL BATTERY
Applicant
Amco International Manufacturing & Design, Inc.
Product Code
MHX · Cardiovascular
Decision Date
Dec 7, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Indications for Use
To power the functions of various devices for which batteries or battery packs are configured. Since rechargeable batteries and battery packs are "device specific", and are designed to operate and fit into the equipment for which they were manufactured, only qualified personnel should evaluate, test, charge, or install these devices. This battery is shipped only to customers who request a replacement battery for a particular device or to replace a competitor's replacement battery. Biomedical equipment service professionals, EMT's, etc. therefore know that the intended use is as a replacement battery . Only replacement battery 2L226 in this submission (Patient Monitor) is provided Over-the-Counter (OTC). All other replacement batteries, specifically; 5L375, 6L577, 7L991, 9L977, 9L990, and 9L998 in this submission (Defib/AED's) are provided by presc ription only.
Device Story
Life+cel battery packs serve as primary, standby, or backup DC power sources for portable and stationary medical equipment. Input consists of electrical energy stored via chemical reaction; output is regulated DC power. Used in clinical settings (hospitals, EMS) by biomedical technicians or clinicians to replace OEM batteries. Operation involves installation into specific host devices (monitors, defibrillators). Benefits include continued device functionality during power loss or portability. Performance is verified against OEM benchmarks for voltage, capacity, and cycle life. Testing ensures compatibility with host device power requirements.
Clinical Evidence
Bench testing only. No clinical data. Testing included Cadex C7400ER analyzer evaluation (3 charge/discharge cycles, 90% capacity threshold), voltage/polarity verification, and NETECH Model Delta 2000 defibrillator analyzer testing to confirm shock delivery capacity. Environmental testing performed at 0, 25, and 40°C. Mechanical integrity verified via drop testing (2-3 feet).
Technological Characteristics
Rechargeable/non-rechargeable battery packs using Sealed-Lead Acid (SLA), Nickel-Cadmium (NiCad), Nickel-Metal Hydride (NiMH), or Lithium chemistries. Provides DC power. Form factor is device-specific to match OEM host equipment. Testing utilizes Cadex C7400ER analyzer and NETECH Delta 2000. No software algorithm class specified; testing uses software-based utility for parameter entry.
Indications for Use
Indicated for use as replacement power sources for specific medical devices, including patient monitors and defibrillators/AEDs. Intended for use by qualified personnel, biomedical service professionals, or EMTs. Model 2L226 is available OTC; models 5L375, 6L577, 7L991, 9L977, 9L990, and 9L998 are prescription-only.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left and consists of a stylized graphic element. The FDA logo is on the right and includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 16, 2021
Amco International Manufacturing & Design, Inc. Alex Henderson Technical Consultant 377 Zane Court Hillside Village Elizabeth, Colorado 80107
Re: K071164
Trade/Device Name: Life+cel Battery Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm) Regulatory Class: Class II Product Code: MHX
Dear Alex Henderson:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated December 7, 2007. Specifically, FDA is updating this SE Letter because FDA has better categorized your device technology under product code MHX.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Jennifer Shih, Office of Cardiovascular Devices, 301-796-5813, Jennifer.Shih@fda.hhs.gov.
Sincerely,
Jennifer W. Shih -S
Jennifer Shih Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OEC 7 2007
Amco International Manufacturing & Design, Inc. c/o Mr. Alex Henderson Phoenix Technology 377 Zane Court, Hillside Village Elizabeth, CO 80107
Re: K071164
Trade/Device Name: life+cel™ (or lifecel™) Battery Packs with Model Numbers: 2L226. 5L375. 6L577. 7L991. 9L977. 9L990 and 9L998 Regulation Number: Unclassified Regulation Name: Unclassified Regulatory Class: Unclassified Product Code: MOY, MOX Dated: December 4, 2007 Received: December 5, 2007
Dear Mr. Henderson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 - Mr. Alex Henderson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Blzimima fh
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Page 1 of 1
Image /page/3/Picture/1 description: The image shows the logo for Phoenix Technology. The logo features a phoenix bird with its wings spread, surrounded by a circular halo. The text "PHOENIX TECHNOLOGY" is written in a serif font to the right of the bird.
## INDICATIONS FOR USE STATEMENT
510(K) Number: K071164
Device Name: Box, Battery
Indications for Use:
To power the functions of various devices for which batteries or battery packs are configured.
Since rechargeable batteries and battery packs are "device specific", and are designed to operate and fit into the equipment for which they were manufactured, only qualified personnel should evaluate, test, charge, or install these devices.
This battery is shipped only to customers who request a replacement battery for a particular device or to replace a competitor's replacement battery. Biomedical equipment service professionals, EMT's, etc. therefore know that the intended use is as a replacement battery .
Only replacement battery 2L226 in this submission (Patient Monitor) is provided Over-the-Counter (OTC).
All other replacement batteries, specifically; 5L375, 6L577, 7L991, 9L977, 9L990, and 9L998 in this submission (Defib/AED's) are provided by presc ription only.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Blummo
(Division-Sign-Off)
Division of Cardiovascular Devices
510(k) Number K071164
377 Zane Court · Elizabeth, Colorado USA 80107 Telephone: 303-646-3715 · Email: alex henderson(@phoenixsolutions.org
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Phoenix Technology. The logo features a stylized phoenix bird with its wings spread, surrounded by a halo. To the right of the bird, the words "PHOENIX TECHNOLOGY" are written in a serif font, with "PHOENIX" in bold and "TECHNOLOGY" in a lighter, less prominent font.
OEC - 7 2007
# 510(k) Summary – K071164 -1 of 5-
Replacement Battery Packs 2L226, 5L375, 6L577, 7L991, 9L977, 9L990, and 9L998
| Submitter: | Amco International Manufacturing & Design, Inc.<br>Attn: Mr. Adam Milewski<br>69-81 108th Street, Suite 6G<br>Forest Hills, New York 11375 | |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------|
| Contact Person: | Alex Henderson<br>Phoenix Technology<br>377 Zane Court,<br>Hillside Village<br>Elizabeth, CO 80107<br>Tel / Fax: 303-646-3715<br>Email: alex_henderson@phoenixsolutions.org | |
| Date Prepared: | April 16, 2006 | |
| Device Name: | Trade/Proprietary Name: life+celTM or lifecelTM Battery Pack<br>Common/Generic Name: Box Battery OR Pack Battery<br>Classification Name: Box Battery, Rechargeable | |
| Classification: | Cardiovascular Panel | Class |
| 21CFR 870.1025 | Arrhythmia Detector and Alarm | II |
| 21CFR 870.1110 | Blood Pressure Computer | II |
| 21CFR 870.1130 | Systems, Measurement, Blood Pressure, Non-Invasive | II |
| 21CFR 870.2300 | Monitor, Cardiac (Including Cardio tachometer & Rate Alarm) | II |
| 21CFR 870.2340 | Electrocardiograph | II |
| No Regulatory Reference | Rechargeable Batteries for Class III Devices | III |
377 Zane Court • Elizabeth, Colorado USA 80107 Telephone: 303-646-3715 ● Email: alex_henderson@phoenixsolutio ns.org
{5}------------------------------------------------
# 510(k) Summary -2 of 5 -
#### Legally Marketed Predicate Devices:
This submission compares the specifications and functionality of various rechargeable battery packs with those of similar devices that were included as part of the following original predicate devices and submissions:
- 1. The AMCO 2L226 life+cel™ is equivalent to that manufactured by:
- I Alpha Source / Access Battery, cleared under K032051
- Unipower Corp., cleared under K940842 and K935111
- . R & D Battery, cleared under K951281
- I OEM, GE Marquette - Patient Monitor Dash 3000/4000 - Reference: K051367, K020290 (Battery included with device)
- 2. The AMCO 5L375 life+cel™ is equivalent to that manufactured by:
- प्र Alpha Source / Access Battery, cleared under K920964
- 트 Unipower Corp., cleared under K940842 and K935111
- I R & D Battery, cleared under K951281
- 프 OEM, Physio Control / Medtronic's - Monitor/Defibrillator (FastPak) LP5 - Reference: K890079, K931053 (Battery included with device)
- 3. The AMCO 6L577 life+cel™ is equivalent to that manufactured by:
- l Alpha Source / Access Battery, cleared under K0920964
- . Unipower Corp., cleared under K940842 and K935111
- 트 R & D Battery, cleared under K951281
- OEM, Physio Control - Monitor/Defibrillator (Non-FastPak) LP5 / LP10 - Reference: K890079, K931053 (Battery included with device)
- 4. The AMCO 7L991 life+cel™ is equivalent to that manufactured by:
- Alpha Source / Access Battery, cleared under K920964
- . Unipower Corp., cleared under K940842 and K935111
- L OEM, Physio Control / Medtronic's - Monitor/Defibrillator LP12 (NiCad) - Reference: K973486, K010918 (Battery included with device)
377 Zanc Court • Elizabeth, Colorado USA 80107 Telephone: 303-646-3715 ● Email: alex henderson@phoenixsolutions.org
{6}------------------------------------------------
# 510(k) Summary -3 of 5 -
## Legally Marketed Predicate Devices (continued):
- 5. The AMCO 9L977 life+cel™ is equivalent to that man ufactured by:
- Alpha Source / Access Battery, cleared under K920964 .
- r Unipower Corp., cleared under K940842 and K935111
- R & D Battery, cleared under K945869 프
- OEM, Health Dyne / Vickers Respironics Smart Monitor 900 Series 트
- Reference: K874695 (Battery included with device)
- 6. The AMCO 9L990 life+cel™ is equivalent to that manufactured by:
- · Alpha Source / Access Battery, cleared under K032051
- Unipower Corp., cleared under K940842 and K935111 .
- OEM, CAS Medical NIBP Monitor 740 .
- Reference: K033048 (Battery included with device)
- 7. The AMCO 9L998 life+cel™ is equivalent to that manufactured by:
- Alpha Source / Access Battery, cleared under K920964 .
- . Unipower Corp., cleared under K940842 and K935111
- OEM, Physio Control / Medtronic's Monitor/Defibrillator LP12 (SLA) - Reference: K973486. K010918 (Battery included with device)
## Description:
Amco rechargeable and non-rechargeable battery packs are utilized as a primary direct current (d-c) power source or as a standby or backup d -c power source for portable as well as stationary medical equipment.
## Statement of Intended Use:
To power the functions of the specific devices for which the batteries or battery packs are intended for back-up power or portable use.
## Comparison of Technological Characteristics
The design components and functionality of the various battery packs listed are similar to those of their predicate devices. All these devices provide a means of supplying electrical power through chemical reaction. The energy provided depends upon the voltage and capacity rating of a particular pack and the amount of current used by the device into which they are installed. The performance and life span of these batteries depends on operating conditions of temperature, current drain, and the charge/discharge method (if applicable). These parameters are taken into account in designing such batteries. The goal is to develop battery packs that maintain capacity for as high and as long as possible. Typical cell chemistries are Sealed -Lead Acid (SLA), Nickel-Cadmium (NiCad), Nickel-Metal Hydride (NiMH), and Lithium based.
> 377 Zane Court · Elizabeth, Colorado USA 80107 Telephone: 303-646-3715 · Email: alex henderson@phoenixsolutio ns.org
{7}------------------------------------------------
# 510(k) Summary -4 of 5 -
### BATTERY PACK TESTING - GENERAL PROTOCOL
#### INCOMING INSPECTION
All cells are inspected for correct specification, visible damage, and randomly voltage tested prior to acceptance. The lot numbers are recorded for tracking purposes should any fail during final assembly and inspection activities. Cases are also inspected for form, fit, function, and cosmetics.
#### CADEX TESTING
Voltage and capacity of battery cells and core packs are tested using a Cadex Electronics Battery Analyzer Model C7400ER (Extended Range) in the "Auto Mode" prior to installation into cases or shrink wrap. This exercises the batteries in order to identify performance characteristics by running them through three (3) full charge/discharge cycles. Tests typically take 12 hours for each battery pack.
All battery chemistries can be tested using custom test parameters, depending on Quality Control and customer requirements. This allows for various C -Rates, delta V (ΔV), and volts-per-cell to be entered into the test protocol through the Cadex Battery Shop Software utility (Reference Cadex Test Report Examples - Exhibit A).
Target capacity is the percentage of the battery capacity compared to nominal capacity and serves as a threshold, This threshold, or target capacity, can be set to any desired range (90 - 95% is typical).
Target capacity is a pass/fail mark and our batteries must meet or exceed a required threshold of 90%, or higher, prior to final Quality Control inspection. Any samples that do not meet the criteria are rejected, and subsequently, the entire lot is tested in this manner.
Battery packs are not shipped fully charged. There are specific DOT, FAA, and EPA requlations and quidelines that address these concerns.
#### VOLTAGE TESTING - Completed Packs
All Battery Packs are tested 100% for correct voltage / polarity prior to shipment. Those devices that fail are rejected and quarantined.
#### DEFIBRILLATOR TESTING
Independent testing (Beta Tests), as well as random tests on finished packs, are performed using NETECH Model Delta 2000 Defibrillator Analyzers to insure that they meet the expected number of shocks as specified by the OEM's.
> 377 Zane Court · Elizabeth, Colorado USA 80107 Telephone: 303-646-3715 · Email: alex henderson@phoenixsolutions.org
{8}------------------------------------------------
# 510(k) Summary -5 of 5 -
### BATTERY PACK TESTING (continued)
#### SAFETY and PERFORMANCE
Safety and performance testing of battery packs are performed to ensure that these devices meet all functional requirements and performance specifications.
In comparison analysis. OEM Battery Packs set the benchmark. Replacement devices must meet or exceed these benchmark results consistently.
Concerns that are addressed during bench test comparison analysis are:
- = Life cycle
The replacement battery must provide as many or more charge and discharge cycles as the original. This is an ongoing process and is not part of the standard QC final inspection protocol. Shelf life, on the other hand, is based on the original cell manufacturer's specification sheets and Certificates of Conformance.
#### I Temperature
The replacement battery must function correctly over the same temperature range as compared to the original. Testing is done at 0, 25, and 40°C (32, 77, and 104°F respectively).
#### . Mechanical & Electrical Component Integrity
Normal testing would involve drop tests from a predetermined height, usually 2-3 feet, onto a hard, uniform surface. Battery packs are inspected for case cracks, cell separation, and electrical/electronic component damaqe. Root cause analysis is performed should any damage occur.
- If there is no visible damage, the battery is tested for form, fit, and function.
- Active Safety devices are inspected and tested before use and after installation.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.