CAS 740 SERIES MONITOR, MODELS 740-1, 740M-1, 740-2MS, 740M-2MS, 740-2NL, 740M-2NL, 740-2NN, 740M-2NN, 740-3MS, 740M-3MS
K033048 · Cas Medical Systems, Inc. · MHX · Jun 1, 2004 · Cardiovascular
Device Facts
Record ID
K033048
Device Name
CAS 740 SERIES MONITOR, MODELS 740-1, 740M-1, 740-2MS, 740M-2MS, 740-2NL, 740M-2NL, 740-2NN, 740M-2NN, 740-3MS, 740M-3MS
Applicant
Cas Medical Systems, Inc.
Product Code
MHX · Cardiovascular
Decision Date
Jun 1, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The CAS 740 Series Vital Signs Monitor is indicated for use for non-invasive monitoring of blood pressure, oxygen saturation, pulse and temperature of adult, pediatric and neonatal patients, in the care of health professionals.
Device Story
The CAS 740 Series Monitor is a portable, multi-parameter vital signs monitor designed for spot-checking or continuous monitoring in clinical settings (hospitals, clinics, emergency vehicles). It processes inputs from non-invasive blood pressure (NIBP) cuffs (oscillometric technology), pulse oximetry sensors (Masimo, Nellcor, or Nonin modules), and predictive temperature probes (Welch Allyn SureTemp). The device features an ND Plus NIBP module with a secondary safety processor. Outputs are displayed on the unit for clinician review to support clinical decision-making. The device is powered by AC mains or a rechargeable NiMH battery. It includes wireless infrared printer communication and is designed for various mounting schemes. It replaces previous CAS 9303/9001/9002 models with improved portability, battery management, and human factors design.
Clinical Evidence
Clinical studies were conducted in accordance with AAMI SP10:2002 for non-invasive blood pressure measurement. Results met the requirements of the standard. Bench testing demonstrated that the ND Plus NIBP module performance is equivalent to the predicate NB module.
Indicated for non-invasive monitoring of blood pressure, oxygen saturation, pulse, and temperature in adult, pediatric, and neonatal patients by healthcare professionals.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
CAS 9303 Neonatal / Adult Vital Signs Monitor (K982776)
Masimo Set 2000 Pulse Oximeter / accessories (K990966)
Submission Summary (Full Text)
{0}------------------------------------------------
# MEDICAL SYSTEMS, INC. TECHNOLOGY APPLIED TO MEDICINE
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
| Ref.: | K033048 |
|--------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Submitter: | CAS Medical Systems, Inc. |
| Address: | 44 East Industrial Rd. Branford CT. 06405 USA |
| Contact: | Ron Jeffrey - Director, Regulatory Affairs<br>Phone - (203) 488-6056 Fax - (203) 488-9438<br>Email - rjeffrey@casmed.com |
| Prepared: | September 25, 2003 Revised January 30, 2004 / June 1, 2004 |
| Trade Name: | CAS 740 Series Monitor |
| Common Name: | Vital Signs Monitor |
| Classification Name: Noninvasive Blood Pressure Measurement System | |
{1}------------------------------------------------
#### EQUIVALENCE
The CAS 740 Monitor is equivalent to the following devices:
- * CAS 9303 Neonatal / Adult Vital Signs Monitor (K982776);
- * CAS 9001 EMS NIBP Monitor (K982135);
- * CAS 9002 EMS NIBP / Sp02 Monitor (K980879);
- * Nellcor Puritan Bennett N-550 Pulse Oximeter (K021090);
- * Nellcor Puritan Bennett N-595 Pulse Oximeter (K012891);
- * Masimo Set 2000 Pulse Oximeter / accessories (K990966)
#### DESCRIPTION
The CAS 740 Monitor is a replacement monitor based on the existing 9303 Vital Signs Monitor, and the CAS 9001 or 9002 EMS monitors. The 740 are primarily a repackaging of these monitors. The primary parameter is non-invasive blood pressure, which is included in all models (see below). Blood pressure measurement is based on oscillometric technology. The additional two parameters are pulse oximetry and temperature.
A purchaser has pulse oximetry choices that include modules manufactured by Masimo, Nellcor or Nonin.
The temperature parameter utilizes predictive technology from SureTemp™ (Welch Allyn Inc.)
The monitor is a rugged, portable and lightweight unit widely adaptable for many applications and mounting schemes. Used for spot-checking or continuous monitoring, features include an easily replaccable Nickel Metal Hydride rechargeable battery pack, wireless infrared printer communication,
and many more. The table below defines the different models within the 740 series and the relationship to the CAS predicate devices
{2}------------------------------------------------
The monitor, its equivalent and parameters:
| Model(s) | CAS Equiv. | Parameters (Variations) |
|----------------------|------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| 740-3MS,<br>740M-3MS | 9303, 9001 | ND NIBP, Masimo Sp02 & W/A Temp - domestic and international versions<br>100 – 240 VAC, and RS 232; (12VDC version; mounting versions available) |
| 740-3NL,<br>740M-3NL | 9303, 9001 | ND NIBP, Nellcor Sp02 & W/A Temp - domestic and international versions<br>100 – 240 VAC, and RS 232; (12VDC version; mounting versions available) |
| 740-3NN,<br>740M-3NN | 9303, 9001, 9002 | ND NIBP, Nonin Sp02 & W/A Temp - domestic and international versions<br>100 – 240 VAC, and RS232; (12VDC version; mounting versions available) |
### 740 Series Intended Use
The CAS 740 Series Vital Signs Monitor is indicated for use for non-invasive monitoring of blood pressure, oxygen saturation, pulse and temperature of adult, pediatric and neonatal patients, in the care of health professionals.
{3}------------------------------------------------
# SUBSTANTIAL EQUIVALANCE COMPARISON – PREDICATE(S) VS. 740
| CATEGORY | CAS 9303, 9001 AND 9002<br>(predicate) | CAS 740 SERIES |
|-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>use | The CAS predicate devices are<br>indicated for use for non-invasive<br>monitoring of blood pressure,<br>oxygen saturation, pulse and<br>temperature of adult, pediatric and<br>neonatal patients in the care of<br>health professionals. The 9001 and<br>9002 monitors are further indicated<br>for use with the EMS / transport<br>user, relying on a rechargeable<br>battery and charger. These models<br>have no direct mains AC input. | The CAS 740 Series devices are<br>indicated for use for non-invasive<br>monitoring of blood pressure, oxygen<br>saturation, pulse and temperature of<br>adult, pediatric and neonatal patients<br>in the care of health professionals.<br>In summary, the 740 series monitor<br>has all of the same indications for use<br>as its predicate device. This includes:<br>condition or disease monitored,<br>prescription vs. over the counter, part<br>of the body applied to, frequency of<br>use, physiological purpose and patient |
| Target<br>population | 9303 family - adults, pediatric and<br>neonatal patients. 9001 and 9002 -<br>adults and pediatric patients only. | population.<br>740 series - adults, pediatric, infant<br>and neonatal patients. |
| Design | 9303 family is comprised of three<br>models: 9301 – non-invasive blood<br>pressure only; 9302 - non-invasive<br>blood pressure and pulse oximeter<br>only; 9303 - non-invasive blood<br>pressure, pulse oximeter and<br>temperature. The 9001 is similar<br>to the 9301 and the 9002 is similar<br>to the 9302. All predicate CAS<br>monitors are small lightweight and<br>portable.<br>The entire line of predicate<br>monitors utilizes modular<br>components for non-invasive blood<br>pressure, pulse oximetry and<br>temperature, if included.<br>In all cases the non-invasive blood<br>pressure module is a CAS design<br>and the same item; (NB non-<br>invasive blood pressure module). | The 740 series is designed to be a<br>multi-parameter monitor. The 740<br>series will replace the 9301, 9302,<br>9303, 9001 and 9002 monitors. The<br>740 series is small lightweight and<br>portable.<br>Like its predicate, the 740 utilize<br>modular parameters for non-invasive<br>blood pressure, pulse oximetry and<br>temperature.<br>CAS designs the blood pressure<br>component, called the ND Plus. The<br>primary difference with the ND Plus,<br>as compared to the predicates NB<br>board, is its smaller size and its lower<br>power use. Also, there is the<br>additional of a secondary safety<br>processor. The boards are similar in<br>that ND Plus has the same front-end<br>control and the same basic algorithm<br>as the NB board. |
| CATEGORY | CAS 9303, 9001 AND 9002 | CAS 740 SERIES |
| | (predicate) | |
| | Where a pulse oximeter is included<br>it is a purchased OEM module<br>from Nonin Medical Systems. | Where a pulse oximeter is included,<br>CAS provides three possible OEM<br>options. The predicate device<br>provided only one. The customer has<br>the option of selecting the pulse<br>oximeter of their choice. Available<br>are the Masimo, Nellcor and Nonin<br>(same as predicate) pulse oximeters.<br>The FDA in the manufacturers own<br>finished medical devices has cleared<br>all three of these pulse oximeters. |
| | The temperature function is only<br>available in the 9303 monitors and<br>is an OEM purchased predictive<br>temperature module manufactured<br>by Welch Allyn. | The temperature function is available<br>if desired. Like the predicate, it is an<br>OEM purchased predictive<br>temperature module manufactured by<br>Welch Allyn. The FDA has cleared<br>this module, in the manufacturers<br>monitor, for marketing.<br><br>In summary, with the exceptions<br>noted above, and the new look of its<br>case design, the 740 is essentially the<br>same as the predicate devices. |
| Materials | Monitor case – ABS U.L.<br>approved material; | Monitor case – ABS U.L. approved<br>injection molded material. This<br>monitor is a custom industrial design |
| | Monitor interior – Series of PC<br>boards (Control / Display, interface<br>and non-invasive blood pressure) If<br>ordered with pulse oximeter or<br>temperature, these would be<br>additional PC board modules. | Monitor interior – Series of PC boards<br>(Control / Display, interface and non-<br>invasive blood pressure) If ordered<br>with pulse oximeter or temperature,<br>these would be additional PC board<br>modules. |
| | Power center – 9303 family has<br>AC mains input and a rechargeable<br>lead acid type battery. 9001/9002<br>monitors have no AC mains and<br>get power from a rechargeable<br>power supply (or direct DC from<br>transport vehicle) and lead acid<br>battery. | Power center – 740 series has AC<br>mains and a rechargeable Nickel<br>Metal Hydride customer replaceable<br>battery. A DC version is part of the<br>series where the power is obtained<br>from the transport vehicle. |
| | Non-invasive blood pressure | Non-invasive blood pressure module<br>(NIBP) – Of the three possible |
| CATEGORY | CAS 9303, 9001 AND 9002<br>(predicate) | CAS 740 SERIES |
| | module (NIBP) -- Of the three<br>possible parameters offered, this<br>module is the only one designed<br>and manufactured by CAS. All<br>monitors in this group contain the<br>NB NIBP module version which<br>consists of microprocessor,<br>circuitry, software, valves, pump<br>transducer, and hardware. | parameters offered, this module is the<br>only one designed and manufactured<br>by CAS. All monitors in this group<br>contain the ND Plus NIBP module<br>version, which consists of<br>microprocessor, circuitry, software,<br>valves, pump transducer, and<br>hardware.<br>The "design" section contains<br>reference to the subtle differences. |
| | Pulse oximeter module - This<br>parameter is an option in the 9302<br>and 9002 only. It is the same<br>module in both cases and is<br>manufactured by Nonin Medical<br>Systems. | Pulse oximeter module - As with the<br>predicate monitors, the same Nonin<br>module is one of the three types<br>offered. The other two types of pulse<br>oximeters are Masimo and Nellcor<br>(Tyco Healthcare). These are actively<br>marketed OEM devices designed for<br>use by other manufacturers as well as<br>the company's own branded devices.<br>All three types of pulse oximeter<br>occupy the same footprint and<br>consume the same power. |
| | Temperature module (9303 only) -<br>See design section. | Temperature module - The 740 has<br>this parameter. See "Design" section. |
| | Components / Accessories -<br>Included are blood pressure cuffs,<br>inflation tube, power cord, users<br>manual, at a minimum. For<br>monitors with the options of pulse<br>oximeter and temperature, the<br>manufacturers recommended<br>accessories are included. | Components / Accessories - Included<br>are blood pressure cuffs, inflation<br>tube, power cord, users manual, at a<br>minimum. These are the same as the<br>predicate devices. For pulse oximetry<br>and temperature, the manufacturers<br>recommended accessories are<br>included. |
| CATEGORY | CAS 9303, 9001 AND 9002<br>(predicate) | CAS 740 SERIES |
| Performance<br>(Clinical and non-clinical) | With regard to non-invasive blood pressure, CAS has tested and passed performance qualification to the SP10 requirements.<br><br>There is historical acceptance of these monitors from an accuracy viewpoint. CAS has very few complaints related to accuracy issues. | With regard to non-invasive blood pressure, CAS has tested and demonstrated ND Plus performance equivalent to the NB module in bench test simulations. No new issues of safety and effectiveness were raised.<br><br>Note: Predicate devices use NB NIBP module. 740 Series uses the ND Plus NIBP module.<br><br>In addition, the 740 monitor has undergone a series of clinical studies in accordance with AAMI SP10:2002, with results meeting the requirements of the standard<br>In summary, performance equivalence has been demonstrated. |
| Sterility | None of the equipment or its accessories is sterilized. | None of the equipment or its accessories is sterilized. |
| Biocompatibility | All patient contact equipment and accessories have been tested, at some level, for biocompatibility. | All patient contact equipment and accessories have been tested for biocompatibility. Where the accessories are provided "ready for sale" by the manufacturer of the pulse oximeter and temperature components, there is documentation showing that the manufacturer has obtained adequate biocompatibility results. |
| Mechanical Safety | These predicate devices have undergone mechanical safety testing as part of the successful, and ongoing evaluation by Underwriters Laboratories as well as testing to the medical equipment safety standard, IEC60601-1 and its applicable collateral and particular standards. | The 740 series monitors have been evaluated to the following standards: UL2601-1, IEC 60601-1, IEC 60601-2-30, IEC 60601-2-49, AAMI SP10 and other collateral and particular safety standards. A follow-up services program is planned for quarterly inspection for these standards.<br><br>In summary, the 740 has received greater testing for mechanical safety than the predicate devices. |
| Chemical Safety | Not applicable to these products | Not applicable to these products |
| CATEGORY | CAS 9303, 9001 AND 9002<br>(predicate) | CAS 740 SERIES |
| Anatomical Sites | Blood pressure - arms and thigh<br>(Neonates – arm and leg)<br><br>Sp02 – fingers and toes<br><br>Temperature – Orally or rectally<br>(neonates – axillary) | Blood pressure – arms and thigh<br>(Neonates – arm and leg)<br><br>Sp02 – fingers and toes<br><br>Temperature – Orally or rectally<br>(neonates – axillary)<br><br>In summary, there is no difference in<br>anatomical sites. |
| Human Factors | With these monitors human factors<br>issues were not monumental from<br>the beginning. Human factors<br>were considered during the<br>development of the monitors.<br>Some of the issues that did persist<br>were designed out during the<br>product lifetime | With the 740 series monitor additional<br>human factors were employed.<br>Examples of these are:<br>✓ Easier battery access for<br>replacement;<br>✓ Visible battery state of<br>condition;<br>✓ Auto off to save battery life;<br>✓ Attached label explaining<br>symbols used;<br>✓ All cables and cords better<br>located to avoid tangling;<br>✓ Designed for ease of<br>assembly;<br>✓ Numerous mounting<br>opportunities - tables, pole,<br>wall or rail;<br>✓ Wireless communication to<br>printer;<br>✓ More rugged than predicates;<br>✓ Additional languages offered. |
| | | |
| CATEGORY | CAS 9303, 9001 AND 9002<br>(predicate) | CAS 740 SERIES |
| Energy used | The 9303 monitor contains a<br>switching power supply which<br>handles input of 100 to 250VAC<br>50/60 Hz at 0.3A. This power runs<br>the monitor and charges a 6VDC<br>lead acid battery. | The 740 series contains the same<br>power supply, essentially the same as<br>the 9303 with the exception of the<br>battery, which in this case is a 7.2 volt<br>Nickel Metal Hydride (NiMH) battery<br>pack. |
| | The 9001 & 9002 monitors do not<br>have internal power supplies.<br>These monitors, designed with<br>portability EMS conditions in<br>mind, contain a 6VDC Lead Acid<br>battery, which is charged by a<br>120VAC power supply. | A version of the monitor is available<br>for transport use where its power is<br>derived from a vehicles 12VDC<br>power. In this case there is no internal<br>power supply or a mains connection.<br>This version also uses the same<br>battery pack. |
| | | In summary, the move to NiMH<br>greatly improves capacity and charge<br>time. |
| Compatibility<br>with<br>environment and<br>other devices | In general these devices do not<br>represent environmental<br>compatibility challenges.<br>Compatibility with other devices is<br>clearly specified by the<br>manufacturer in the user manuals.<br>Product safety and EMC rules<br>require that limits be established<br>by testing and documented. | In general these devices do not<br>represent environmental compatibility<br>challenges. Compatibility with other<br>devices is clearly specified by the<br>manufacturer in the user manuals.<br>Product safety and EMC rules require<br>that limits be established by testing<br>and documented. |
| | | In general there is equivalence here,<br>however the safety and EMC testing<br>authority has required that tighter<br>controls be placed on connection to<br>other devices. |
| | Where used | Hospitals, intra-hospital and inter-<br>hospital transport, clinics, doctor's<br>office, dental offices, emergency<br>vehicles, and first responders. Not<br>indicated for home use |
| | | In summary, there is no difference<br>here. |
| CATEGORY | CAS 9303, 9001 AND 9002 (predicate) | CAS 740 SERIES |
| Standards met | UL2601 or 544, EN 60601-1, EN<br>60601-1-2, EN 60601-2-30, EN<br>1061 and SP10 (accuracy) | UL2601-1, IEC60601-1, IEC60601-1-<br>1, IEC60601-1-2, IEC60601-1-4,<br>IEC60601-2-30, IEC60601-2-49, EN<br>865, and AAMI SP10:2002.<br>In summary, There were more<br>standards met and in some cases they |
| Electrical Safety | These predicate devices have<br>undergone mechanical safety<br>testing as part of the successful,<br>and ongoing evaluation by<br>Underwriters Laboratories as well<br>as testing to the medical equipment<br>safety standard, IEC60601-1 and<br>its applicable collateral and<br>particular standards. | The 740 monitor has undergone<br>electrical safety testing as part of the<br>successful, and ongoing evaluation by<br>Underwriters Laboratories as well as<br>testing to the medical equipment<br>safety standard, IEC60601-1 and its<br>applicable collateral and particular<br>standards.<br>In summary, the 740 has received<br>greater testing for electrical safety<br>than the predicate devices. |
| Thermal Safety | These devices do not present<br>thermal safety risks. Some aspects<br>of over-temperature are evaluated<br>during safety testing. | These 740 series does not present<br>thermal safety risks. Some aspects of<br>over-temperature are evaluated during<br>safety testing.<br>In summary, there are no differences. |
| Radiation safety | Not a risk | Not a risk |
{4}------------------------------------------------
{5}------------------------------------------------
:
{6}------------------------------------------------
{7}------------------------------------------------
:
{8}------------------------------------------------
:
:
{9}------------------------------------------------
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN - 1 2004
Mr. Ron Jeffrey Director, Regulatory Affairs CAS Medical Systems, Incorporated 44 East Industrial Road Branford, Connecticut 06405
Re: K033048
Trade/Device Name: CAS 740 Series Monitor Models 740-3MS, 740M-3MS, 740-3NL, 740M-3NL, 740-3NN, 740M-3NN Regulation Number: 870.1025 Regulation Name: Patient Physiologic Monitor Regulatory Class: II Product Code: MHX Dated: May 13, 2004 Received: May 14, 2004
Dear Mr. Jeffrey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{11}------------------------------------------------
Page 2 - Mr. Jeffrey
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if requirely as as tronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Chiu Lin, Ph.D
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{12}------------------------------------------------
## Indications for Use
510(k) Number: K033048
Device Name: 740 Series monitor
The CAS 740 Series Vital Signs Monitor is indicated for use for non-Indications for use: invasive monitoring of blood pressure, oxygen saturation, pulse and temperature of adult, pediatric and neonatal patients, in the care of health professionals.
Prescription Use Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
K033048
(Division Sign-Off)
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices
510(k) Number: Jum
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.