ENZYMATIC CREATININE ASSAY (265 SERIES)

K070383 · Diagnostic Chemicals , Ltd. · JFY · Jun 22, 2007 · Clinical Chemistry

Device Facts

Record IDK070383
Device NameENZYMATIC CREATININE ASSAY (265 SERIES)
ApplicantDiagnostic Chemicals , Ltd.
Product CodeJFY · Clinical Chemistry
Decision DateJun 22, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1225
Device ClassClass 2

Intended Use

The Diagnostic Chemicals Limited Enzymatic Creatinine Assay is for the quantitative determination of creatinine in serum, plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. This device is intended for professional use and in vitro diagnostic use only.

Device Story

Two-reagent liquid enzymatic assay; measures creatinine in serum, plasma, and urine samples. Principle: creatinine amidohydrolase converts creatinine to creatine; creatine amidinohydrolase converts creatine to sarcosine and urea; sarcosine oxidase and peroxidase produce colored product measured via spectrophotometry at 545 nm. Used in clinical laboratories by professional staff on analyzers like Bayer Advia 1650. Output is quantitative creatinine concentration (mg/dL). Results assist clinicians in diagnosing/treating renal disease and monitoring dialysis.

Clinical Evidence

Bench testing only. Precision established via 20-day study (N=80) and within-run study (N=20) for serum and urine controls; CVs ranged from 0.2% to 2.7%. Linearity validated across reportable ranges (serum 0.04–30 mg/dL; urine 0.03–175 mg/dL). Method comparison against predicate (N=40 samples each for serum and urine) showed high correlation (r=1.0000 and r=0.9995 respectively). Interference testing confirmed no significant impact from icterus, lipemia, hemolysis, or ascorbic acid.

Technological Characteristics

Two-reagent liquid system containing enzymes, stabilizers, and buffers. Enzymatic colorimetric detection principle. Requires spectrophotometer/analyzer capable of 545 nm measurement at 37°C. Traceable to NIST SRM 914A. Performance evaluated per CLSI guidelines EP5-A2, EP6-A, and EP9-A2.

Indications for Use

Indicated for the quantitative determination of creatinine in serum, plasma, and urine for the diagnosis and treatment of renal diseases, monitoring of renal dialysis, and as a calculation basis for other urine analytes. Intended for professional in vitro diagnostic use.

Regulatory Classification

Identification

A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K070383 JUN 2 2 2007 Image /page/0/Picture/2 description: The image shows a logo or symbol that appears to be composed of stylized letters. The letters are bold and blocky, with a combination of straight lines and curves. The overall design is abstract and modern, with a focus on geometric shapes and negative space. The logo is likely used for branding or identification purposes. ## Enzymatic Creatinine Assay Product Cat. No. 265-30, 265-50 # 510(K) SUMMARY Introduction: This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. Submitter: Diagnostic Chemicals Limited 16 McCarville Street Charlottetown, P.E.I. Canada C1E 2A6 (Tel:) (902)566-1396 (Fax:) (902)566-2498 Contact Person: Debbie Murray Date Prepared: April 11, 2007 Device Name: Tradename: Enzymatic Creatinine Assay FDA Regulation Name: 862.1225 Creatinine Test System FDA Product Code: JFY Predicate Device: Roche Diagnostics Corp. Creatinine Plus (K003261) Device Description: The Enzymatic Creatinine Assay contains an in vitro diagnostic reagent system intended for the quantitative determination of creatinine in serum, plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. Reagent is a two-part liquid in plastic bottles packaged in the appropriate box. Intended Use For the IN VITRO quantitative determination of creatinine in serum, plasma and urine. {1}------------------------------------------------ ## Comparison to Predicate Device: Description of the Item Being Compared: Enzymatic in vitro assay for the direct quantitative determination of creatinine in human serum, plasma and urine using Roche automated clinical chemistry analyzers. Creatinine Plus Reagent: R1 TAPS buffer: 30 mmol/L, pH 8.1; creatinase > 333 ukat/L; sarcosine oxidase > 133 ukat/L; ascorbate oxidase > 33 ukat/L; HTIB: 5.9 mmol/L; detergents; preservative. Creatinine Plus Reagent: R2 TAPS buffer: 30 mmol/L, pH 8.0; creatininase > 500 µkat/L; peroxidase ≥ 16.7 µkat/L; 4-aminophenazone: 2.0 mmol/L; potassium hexacyanoferrate (II): 163 umol/L; detergent; preservative. ## Similaries: The submission device and the predicate device have the same intended use. The submission device and the predicate device are both provided in a ready to use liquid format. The submission device and the predicate device both use enzyme-linked steps using sarcosine oxidase and peroxidase to produce a colored product. #### Differences: There are no differences. # Comments on Substantial Equivalence: Testing results demonstrate that the Enzymatic Creatinine Assay is equivalent to the predicate device. Method comparison results provided the following: #### Serum Deming regression using NCCLS EP9-P. A comparison was made between this method and a similar enzymatic method using 40 samples ranging from 0.7 to 31.2 mg/dL. The correlation coefficient was 1.0000. Linear regression analysis gave the following equation: This method = 1.033 (reference method) - 0.13 mg/dL. {2}------------------------------------------------ Urine Deming regression using NCCLS EP9-P. A comparison was made between this method and a similar enzymatic method using 40 samples ranging from 13.5 to 141.7 mg/dL. The correlation coefficient was 0.9995. Linear regression analysis gave the following equation: This method = 1.041 (reference method) +1.06 mg/dL. Plasma Deming regression using NCCLS EP9-P. A comparison was made between plasma and serum using 33 samples ranging from 0.61-27.04 mg/dL. The correlation coefficient was 0.9997. Linear regression analysis gave the following equation: Plasma = 1.006 (serum) - 0.03 mg/dL. Conclusion: Diagnostic Chemicals Limited's Enzymatic Creatinine Assay is substantially equivalent in principle and performance to the predicate product. > Debbie Murray, External Regulatory Affairs Coordinator Diagnostic Chemicals Limited Diagnostic Division {3}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three overlapping wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 JUN 2 2 2007 Diagnostic Chemicals, Ltd. c/o Ms. Debbie Murray External Regulatory Affairs Coordinator 16 McCarville St. Charlottetown, Price Edward Island C1E 2A6 Canada k070383 Re: > Trade Name: Enzymatic Creatinine Assay (265 series) Regulation Number: 21 CFR 862.1225 Regulation Name: Creatinine Test System Regulatory Class: Class II Product Code: JFY Dated: April 11, 2007 Received: May 04, 2007 Dear Ms. Murray: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820). {4}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Jean M. Cooper, M.S., D.V.M. Yéan M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number: 16070383 Device Name: Enzymatic Creatinine Assay (265 Series) Indications For Use: Reagent The Diagnostic Chemicals Limited Enzymatic Creatinine Assay is for the quantitative determination of creatinine in serum, plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. This device is intended for professional use and IN VITRO diagnostic use only. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety K0703Y3
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...