DIAZYME CREATININE LIQUID REAGENTS ASSAY

K062105 · Diazyme Laboratories · JFY · Apr 23, 2007 · Clinical Chemistry

Device Facts

Record IDK062105
Device NameDIAZYME CREATININE LIQUID REAGENTS ASSAY
ApplicantDiazyme Laboratories
Product CodeJFY · Clinical Chemistry
Decision DateApr 23, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1225
Device ClassClass 2

Intended Use

Diazyme Creatinine Liquid Reagents Assay, in conjunction with Diazyme Creatinine Calibrator, is intended for the quantitative determination of creatinine in serum and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. For in vitro diagnostic use only. Diazyme Creatinine Control is an assayed QC material for use in quantitative in vitro diagnostic determination of creatinine in human serum and urine. It is intended as a reference sample for monitoring the Diazyme Creatinine Liquid Reagents Assay for gross systematic errors. For in vitro diagnostic use only.

Device Story

The Diazyme Creatinine Liquid Reagents Assay is an enzymatic, kinetic in vitro diagnostic test used to measure creatinine levels in human serum or urine. The device utilizes a two-reagent (R1/R2) system; R1 performs pre-incubation to eliminate endogenous creatine. The assay employs a coupled enzymatic reaction chain: creatininase converts creatinine to creatine; creatinase converts creatine to sarcosine; sarcosine oxidase (SOD) oxidizes sarcosine to produce hydrogen peroxide; peroxidase (POD) reacts with hydrogen peroxide to form a colored dye quantified at 550 nm. The test is performed on clinical laboratory instrumentation. Results are used by physicians to diagnose and manage renal disease and monitor dialysis. The system includes an assayed control material to monitor for systematic errors. The device provides rapid, quantitative results, assisting clinical decision-making regarding kidney function and patient treatment plans.

Clinical Evidence

Bench testing only. Precision evaluated per CLSI EP5-A (serum) and EP10 (urine); total CVs ranged 0.60%-3.0%. Linearity confirmed across 0.14-13.51 mg/dL (serum) and 0.14-141 mg/dL (urine). Interference testing showed no significant recovery variation (±10%) for triglycerides, ascorbic acid, bilirubin, and hemoglobin. Method comparison (n=55 serum, n=42 urine) against predicate showed high correlation (r²=0.9981 serum; r²=0.9969 urine).

Technological Characteristics

Enzymatic kinetic assay; two-reagent (R1/R2) system. Sensing principle: spectrophotometric quantification of colored dye at 550 nm. Reagents include creatininase, creatinase, sarcosine oxidase, and peroxidase. Operates on standard clinical chemistry analyzers. Calibrator is traceable to NIST SRM 914a. Liquid form reagents and controls.

Indications for Use

Indicated for quantitative determination of creatinine in serum and urine for diagnosis/treatment of renal diseases, monitoring renal dialysis, and as a calculation basis for other urine analytes. For prescription use only.

Regulatory Classification

Identification

A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Kol62105 Pre-market Notification Diazyme Creatinine Assay ## 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92 | Submitter's name: | Diazyme Laboratories | | |--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------| | Submitter's address: | 3550 General Atomics Court<br>San Diego, CA 92121<br>USA | APR 23 2007 | | Name of Contact Person: | Dr Abhijit Datta<br>Diazyme Laboratories<br>3550 General Atomics Court<br>San Diego, CA 92121<br>Phone: 858-455-4762<br>Fax: 858-455-2120 | | | Date the Summary was Prepared: | Revised on March 27, 2007 | | | Name of the Device | In Vitro Diagnostic Creatinine Test System | | | Trade Name: | Diazyme Creatinine Liquid Reagents Assay<br>Diazyme Creatinine Control | | | Common/Usual Name | Creatinine Test System | | | Device Classification Name | Enzymatic Method, Creatinine<br>Single (specified) analyte controls (assayed and unas-<br>sayed | | | Product code: | JFY, JJX | | | Submission Type | 510k | | | Regulation Number | 862.1225, 862.1660 | | | Device Class | II (Enzymatic Method, Creatinine)<br>I (Single (specified) analyte controls (assayed and unas-<br>sayed) | | | Predicate Device: | For the Creatinine test system, we are claiming equiva-<br>lence [807.92(a) (3) to CREATININE PLUS (k003261)<br>manufactured by Roche Diagnostics GmbH, D-68298<br>Mannheim, Germany.<br><br>For the Diazyme Creatinine control which is a single<br>(specified) analyte controls (assayed), the predicate is<br>k063206 | | R 3/29/07 {1}------------------------------------------------ # Substantial Equivalence Information #### 1. Predicate device name(s): Roche Creatinine Plus Dimension Vista Protein Control L, M, H #### 2. Predicate 510(k) number(s): k003261, k063206 #### 3. Comparison with predicate: ## Indications for Use | Diazyme Creatinine Liquid Reagents Assay Roche Creatinine plus test | Equivalency | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------| | Diazyme Creatinine Liquid Reagents Assay For the quantitative in vitro determination<br>in conjunction with Diazyme Creatinine of creatinine in human serum or urine.<br>Calibrator, is intended for the quantitative<br>determination of creatinine in serum and<br>urine. Creatinine measurements are used in<br>the diagnosis and treatment of renal<br>diseases, in monitoring renal dialysis, and as<br>a calculation basis for measuring other urine<br>analytes. For in vitro diagnostic use only. | Same | # Principle | Diazyme Creatinine Liquid Reagents Assay | Roche Creatinine plus test | Equivalency | |---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------| | The enzymatic assay for creatinine involves<br>a series of coupled enzymatic reactions in-<br>cluding creatininase enzymatic conversion<br>of creatinine into the product creatine<br>which itself is converted to sarcosine by<br>creatine amidinohydrolase (creatinase), fol-<br>lowed by oxidation of sarcosine by sarco-<br>sine oxidase (SOD), producing hydrogen<br>peroxide. In the presence of peroxidase<br>(POD), the hydrogen peroxide is quantified<br>at 550 nm by the formation of a colored<br>dye. Any endogenous creatine present in<br>the sample is removed by creatinase and<br>sarcosine oxidase during preincubation. | The enzymatic method is based on the es-<br>tablished determination of sarcosine after<br>conversion of creatinine with the aid of<br>creatininase, creatinase, and sarcosine oxi-<br>dase. The liberated hydrogen peroxide is<br>measured via a modified Trinder reaction.<br>Endogenous creatine is metabolized with<br>the addition of R1 to the sample. | Same | #### Test Objective | Diazyme Creatinine Liquid Reagents Assay | Roche Creatinine plus test | Equivalency | |------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|-------------| | For the quantitative in vitro determination of creatinine in human serum or urine. | For the quantitative in vitro determination of creatinine in human serum or urine. | Same | ### Type of Test | Diazyme Creatinine Liquid Reagents Assay | Roche Creatinine plus test | Equivalency | |------------------------------------------|----------------------------|-------------| | Quantitative | Quantitative | Same | {2}------------------------------------------------ # Specimen Type | Specimen Type | | | |------------------------------------------|----------------------------|-------------| | Diazyme Creatinine Liquid Reagents Assay | Roche Creatinine plus test | Equivalency | | Human serum or urine. | Human serum or urine. | Same | # Product Type | Diazyme Creatinine Liquid Reagents Assay | Roche Creatinine plus test | Equivalency | |-------------------------------------------|-------------------------------------------|-------------| | Calibrator, Controls, Reagent, Instrument | Calibrator, Controls, Reagent, Instrument | Same | ## Performance Comparison | Diazyme Creatinine Liquid Reagents Assay | Roche Creatinine plus test | Equivalency | |------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|-------------| | Recovery: Average of 102.6% | Recovery: Average of 105.0% | Similar | | Reportable Range:<br>Serum: 0.14-13.56mg/dL<br>(12-1200 µmol/L)<br>Urine: 0.14-141.25 mg/dL<br>(12-12500 µmol/L) | Reportable Range:<br>Serum/Plasma: 2.7-2652 µmol/L<br>Urine: 27-35360 µmol/L | | | Precision/Serum:<br>Within Run: 0.1%-21.1%<br>Total: 1.4%-3.0% | Precision/Serum:<br>Within Run: 0.7%-0.9%<br>Between Run: 1.6%-2.5% | | | Precision/Urine:<br>Within Run: 0.31%-0.46%<br>Total: 0.61%-2.64% | Precision/Urine:<br>Within Run: 0.8%-1.0%<br>Between Run: 2.1%-3.7% | | | Accuracy/Serum:<br>Correlation Coefficient: 0.9981<br>Slope/Intercept:<br>$y = 0.9467x + 0.0643$ | Accuracy/Serum:<br>Correlation Coefficient: 0.999<br>Slope/Intercept:<br>$y = 0.989x + 0.036$ | | | Accuracy/Urine:<br>Correlation Coefficient: 0.9969<br>Slope/Intercept:<br>$y = 1.005x - 0.2979$ | Accuracy/Urine:<br>Correlation Coefficient: 0.999<br>Slope/Intercept:<br>$y = 0.999x + 0.037$ | | # Control Comparison : | Diazyme Creatinine Control | Dimension Vista Protein Control L, M, H | Equivalency | |-------------------------------------------------------------|-----------------------------------------------------------------------------------------|-------------| | Liquid form | Liquid form | Same | | Analyte specific Creatinine constituent | Analyte specific, C3, C4, Homocysteine,<br>IGA, IGG, IGM and prealbumin<br>constituents | Same | | Traceable to purified Creatinine -NIST<br>SRM 914a standard | Traceable to purified S-adenosyl<br>homocysteine standard | Same | {3}------------------------------------------------ ## Rationale for Considering the Device Substantially Equivalent to Devices Approved for Inter-state Commerce Roche Creatinine plus test method (k003261) was selected for comparing serum and urine samples with to the results generated by Diazyme Creatinine Liquid Reagents Assay. Detailed performance characteristics and comparison analysis are given in this filing and demonstrate substantial equivalence to predicate device that is currently being legally marketed. The Diazyme Creatinine Liquid Reagents Assay is similar to the approved predicate test. The minor differences in the performances of the tests should not affect the safety and effectiveness of the Diazyme Creatinine Liquid Reagents Assay and offers users a rapid device to measure creatinine in human serum. In summary, the dissimilar features between the Diazyme Creatinine Liquid Reagents Assay and devices currently legally marketed do not affect the safety or effectiveness of the device. This is supported by the accuracy data comparing serum and urine sample values obtained using the Diazyme Creatinine Liquid Reagents Assay with those obtained using the predicate device, Roche Creatinine plus test (k003261). Comparison analysis presented in this 510k submission together with the stability data also demonstrates that the Diazyme Creatinine control is substantially similar to the legally marketed device Dimension Vista Control (k063206). {4}------------------------------------------------ ### Description of the Device This enzymatic assay for creatinine involves a series of coupled enzymatic reactions including creatininase enzymatic conversion of creatinine into the product creatine. which itself is converted to sarcosine by creatine amidinohydrolase (creatinase), followed by oxidation of sarcosine by sarcosine oxidase (SOD) producing hydrogen peroxide. In the presence of peroxidase (POD) the hydrogen peroxide is quantified at 550mm by the formation of a colored dye. Image /page/4/Figure/3 description: This image shows a series of chemical reactions. In the first reaction, Creatinine and H2O react with Creatininase to produce Creatine. In the second reaction, Creatine and H2O react with Creatinase to produce Sarcosine and Urea. In the third reaction, Sarcosine, H2O, and O2 react with SOD to produce Formaldehyde, Glycine, and H2O2. In the fourth reaction, 2H2O2, 4-AA, and TOOS react with POD to produce 4H2O and Quinone dye (λ max 556). #### Intended Use of the Device: Diazyme Creatinine Liquid Reagents Assay, in conjunction with Diazyme Creatinine Calibrator, is intended for the quantitative determination of creatinine in serum and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. For in vitro diagnostic use only Diazyme Creatinine Control is an assayed QC material for use in quantitative in vitro diagnostic determination of creatinine in human serum and urine. It is intended as a reference sample for monitoring the Diazyme Creatinine Liquid Reagents Assay for gross systematic errors. For in vitro diagnostic use only. ### Performance Characteristics Diazyme Creatinine Liquid Reagents Assay is a two-reagent (R1 / R2) based kinetic assay system. The results are obtained in 10 minutes by measuring absorbance at 550nm. On line pretreatment removes endogenous creatine. The assay has a wide measuring range of 0.14-13.56 mg/dL (12-1200 umol/L) for Serum samples and 0.14 -- 141 mg/dL (12 to12500 umol/L) for Urine samples. The assay offers excellent precision as shown in the tables below: | Serum Testing | 0.75 mg/dL<br>(66.3 μM) | 1.41 mg/dL<br>(125 μM) | 3.91 mg/dL<br>(346 μM) | 10.28 mg/dL<br>(908.7 μM) | |----------------------|-------------------------|--------------------------|----------------------------|---------------------------| | Within-Run Precision | $C_V\% = 2.1\%$ | $C_V\% = 1.1\%$ | $C_V\% = 0.7\%$ | $C_V\% = 0.1\%$ | | Total Precision | $C_V\% = 3.0\%$ | $C_V\% = 1.9\%$ | $C_V\% = 1.4\%$ | $C_V\% = 1.4\%$ | | Urine Testing | 30 mg/dL<br>(2652 μM) | 87.13 mg/dL<br>(7711 μM) | 196.70 mg/dL<br>(17407 μM) | | | Within-Run Precision | $C_V\% = 0.36\%$ | $C_V\% = 0.31\%$ | $C_V\% = 0.46\%$ | | | Total Precision | $C_V\% = 2.64\%$ | $C_V\% = 0.76\%$ | $C_V\% = 0.61\%$ | | {5}------------------------------------------------ Pre-market Notification Diazyme Creatinine Assay In method comparison studies, samples tested with Diazyme Creatinine Liquid Reagents Assay showed good correlation with Roche Creatinine plus (k003261) with correlation coefficients of 0.99 for both serum samples and urine samples. The average analytical recoveries were 97.8% to 109.4% for 11 creatinine serum samples and 97.0% to 104.3% for 22 creatinine urine samples. Serum Urine Substance Concentration Substance Concentration Triglycerides 1000 mg/dL Triglycerides 1000 mg/dL Ascorbic acid 10 mmol/L Ascorbic acid 10 mmol/L Bilirubin 40 mg/dL Bilirubin 40 mg/dL Bilirubin Conjugated Bilirubin Conjugated 30 mg/dL 40 mg/dL Hemoglobin 500 mg/dL Hemoglobin 1000 mg/dL We have conducted interference studies by spiking the substances to be tested to pooled human serum and urine, and found little interference at the indicated concentrations. Conclusion: Comparison analysis presented in this 510k submission filing in the comparison section, together with linearity, precision and interference and other detailed studies, demonstrates that the Diazyme Creatinine Liquid Reagents Assay has excellent accuracy and is safe and effective. There is no significant deviation between the results obtained by Diazyme Creatinine Liquid Reagents Assay and the legally marketed predicate device (k003261) when testing clinical patient samples and is therefore substantially similar. Comparison analysis presented in this 510k submission together with the stability data demonstrates that the Diazyme Creatinine control is substantially similar to legally marketed device (k063206). {6}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three overlapping wings, symbolizing health, services, and people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 ## APR 2 3 2007 Diazyme Laboratories c/o Dr. Abhijit Datta Associate Director 3550 General Atomic Court San Diego, CA 92121 Re: k062105 > Trade/Device Name: Diazyme Creatinine Liquid Reagents Assay Diazyme Creatinine Control Regulation Number: 21 CFR 862.1225 Regulation Name: Creatinine Test System Regulatory Class: Class II Product Code: JFY, JJX Dated: March 19, 2007 Received: March 21, 2007 Dear Dr. Datta: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {7}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Jean M. Cooper, M.S., D.V.M. Jean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ # Indications for Use 510(k) Number: k062105 Device Name: Indications for Use: Diazyme Creatinine Liquid Reagents Assay Diazyme Creatinine Control Diazyme Creatinine Liquid Reagents Assay, in conjunction with Diazyme Creatinine Calibrator, is intended for the quantitative determination of creatinine in serum and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. For in vitro diagnostic use only. Diazyme Creatinine Control is an assayed QC material for use in quantitative in vitro diagnostic determination of creatinine in human serum and urine. It is intended as a reference sample for monitoring the Diazyme Creatinine Liquid Reagents Assay for gross systematic errors. For in vitro diagnostic use only. Prescription Use X AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) | Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) | |-------------------------------------------------------------------| |-------------------------------------------------------------------| | Carol C Benson | |-------------------| | Division Sigh Off | Division Sigh-Off Office of In Vitro Diagnostic Device Evaluation and Safety K062105
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...