CITOP10 GUIDEWIRE, 0.014(1.4MM TIP), CITOP6 GUIDEWIRE, 0.014(1.0MM TIP), CITOP SUPPORT ACCESSORY, 0.023(126CM)

K070212 · Ovalum , Ltd. · DQX · Oct 12, 2007 · Cardiovascular

Device Facts

Record IDK070212
Device NameCITOP10 GUIDEWIRE, 0.014(1.4MM TIP), CITOP6 GUIDEWIRE, 0.014(1.0MM TIP), CITOP SUPPORT ACCESSORY, 0.023(126CM)
ApplicantOvalum , Ltd.
Product CodeDQX · Cardiovascular
Decision DateOct 12, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1330
Device ClassClass 2

Indications for Use

The CiTop™ Guidewire is intended for use in angiographic procedures to facilitate the intra-luminal placement of the wire beyond stenotic lesions (including chronic total occlusions) in the peripheral vasculature prior to further percutaneous intervention.

Device Story

CiTop™ 0.014" guidewire; 280 cm length; nitinol shaped distal tip; platinum marker coil for fluoroscopic visualization. Used in peripheral vasculature to navigate stenotic lesions and chronic total occlusions. Operator-controlled handle facilitates torque and tip shaping; distal tip can be arched to release wedged wire. Supported by stainless steel helical cut accessory shaft (0.023" OD, 126 cm) to aid advancement and adjust distal flexibility. Single-use, disposable device. Used by physicians in clinical settings during angiographic procedures to gain access and position interventional devices. Output is physical navigation through vasculature; clinical benefit is successful crossing of difficult occlusions to enable further percutaneous intervention.

Clinical Evidence

Bench testing only. Evaluation included in vitro tensile, torque strength, torqueability, tip flexibility, and catheter compatibility tests per FDA's 1995 Coronary and Cerebrovascular Guidewire Guidance Document. Biocompatibility verified per ISO 10993-1 for external communicating, blood contact, short duration (<24 hours) devices. All results met specifications and predicate performance expectations.

Technological Characteristics

0.014" diameter guidewire; nitinol distal tip; platinum marker coil; stainless steel helical cut accessory shaft (0.023" OD). Manual operation; torque-controlled via handle. Sterilized, single-use, disposable. Biocompatibility per ISO 10993-1.

Indications for Use

Indicated for patients undergoing angiographic procedures requiring intra-luminal wire placement beyond stenotic lesions, including chronic total occlusions, in the peripheral vasculature to facilitate subsequent percutaneous intervention.

Regulatory Classification

Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary CiTop™ 0.014" Guidewire ## Date: August 29, 2007 ## Submitter Information: JCI 1 2 /007 OVALUM Ltd. 10. Plaut St. Rabin Industrial Park, Rehovot Israel 76706 Tel: (972)-8-9315679 Fax: (972)-8-9318107 ## Contact Person: Naáma Oren Director of QA and Regulatory Affairs #### Device Identification | Trade Name: | CiTop™ 0.014" Guidewire | |----------------------|----------------------------------| | Device Common Name: | Catheter Guide Wire | | Classification Name: | Catheter Guide Wire | | Classification: | Class II (74 DQX/21CFR 870.1330) | ## Predicate devices: The CiTop™ 0.014" Guidewire is substantially equivalent to the following predicate devices: - · Asahi PTCA & PTA guidewire Confianza Pro cleared under K041531 - · Cordis STEER-IT Guidewire cleared under K040592; - · LuMend Frontrunner CTO Catheter and Accessories K013284 #### Device Description: The CiTop™ Guidewire has an outside diameter of 0.014" (0.36 mm) and a length of 280 cm. The CiTop™ 0.014" Guidewire can be torqued to facilitate navigation through the vasculature. A handle control is supplied with the guidewire to facilitate torque and tip shaping for deployment and positioning according to standard practices. The distal tip can be momentarily shaped into an arched formation by the operator to aid in releasing a wedged guidewire and allow its advancement through difficult occlusions. A platinum marker coil at the distal tip aids visualization under fluoroscopy. The CiTop™ Guidewire is supported by a stainless steel helical cut accessory shaft with an outside diameter of 0.023" and length of 126cm. The support accessory may be independently torqued and its distal end shaped. The support accessory operates {1}------------------------------------------------ Image /page/1/Picture/15 description: The image shows a collection of symbols and numbers scattered across a white background. The symbols appear to be punctuation marks, such as colons and periods. There are also numbers, including "12" and "1", present in the image. The arrangement of these elements is somewhat random and sparse. as an aid in the advancement of the CiTop™ Guidewire and as an adjustment to the flexibility of the distal end of the CiTop™ Guidewire . The CiTop™ 0.014" Guidewire is supplied in a protective dispenser, that is packaged in a single-use sterilizeable pouch. The Pouch is labeled, and placed in an individual protective carton for sterilization. The CiTop™ 0.014" Guidewire is a single use/disposable product with a nitinol shaped tip. The Guidewire is used to gain intravascular access to and facilitate the positioning and exchange of interventional devices in small-diameter, tortuous vasculature. # Intended Use The CiTop™ Guidewire is intended for use in angiographic procedures to facilitate the intra-luminal placement of the wire beyond stenotic lesions (including chronic total occlusions) in the peripheral vasculature prior to further percutaneous intervention. # Technological Characteristics The components of the CiTop™ 0.014" Guidewire are similar in basic materials. design, construction and performance to the predicate devices. The distal tip has been designed to improve performance characteristics in passing total occlusions (i.e. tip flexibility, tip shaping and shape retention / relaxation). # Safety and Performance Testing Biocompatibility of the CiTop™ 0.014" Guidewire materials has been verified in accordance with ISO 10993-1. Biological evaluation of Medical Devices - Part 1. Materials test results confirmed biocompatibility of the subject device when tested as an external communicating, blood contact, short duration (<24 hours) device. Design analysis, in vitro and in vivo data confirm the safety and effectiveness of the device and that the basic functional characteristics are substantially equivalent to the predicate devices cited. Device evaluation based on testing specified in the FDA's Coronary and Cerebrovascular Guidewire Guidance Document (January 1995) and included in vitro tensile, torque strength, torqueability, tip flexibility, biocompatibility and catheter compatibility tests. All data was within device specification requirements, as well as standard requirements and predicate performance expectations. Subject product testing has yielded acceptable safety & performance outcomes. In addition, verification performance testing also yielded acceptable results. The results of these tests, in conjunction with the substantial equivalence claims as outlined in the premarket notification, effectively demonstrate that the CiTop™ 0.014" Guidewire is substantially equivalent to the cited predicate devices. ## Summary of Substantial Equivalence Based on the indications for use, technological characteristics, and safety and performance testing, the subject CiTop™ 0.014" Guidewire meets the minimum requirements that are considered adequate for its intended use and is substantially equivalent in design, materials, sterilization, principles of operation and indications for use to current commercially available guidewires/cited predicates. {2}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 307 1 2 2007 Ovalum, Ltd. c/o Mr. Jonathan Kahan Hogan & Hartson, LLP Columbia Square 555 Thirteenth Street, NW Washington, DC 20004 Re: K070212 Trade/Device Name: CiTop Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II (two) Product Code: DQX Dated: August 31, 2007 Received: August 31, 2007 Dear Mr. Kahan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 – Mr. Jonathan Kahan Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21) CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Bram D. Zuckerman, M.D. Directo Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # INDICATIONS FOR USE | 510(k) Number<br>(if known): | K070212 | | | |------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----|----------------------| | Device Name: | CiTop™ 0.014" Guidewire | | | | Indications for Use: | The CiTop™ Guidewire is intended for use in angiographic procedures to facilitate the intra-luminal placement of the wire beyond stenotic lesions (including chronic total occlusions) in the peripheral vasculature prior to further percutaneous intervention. | | | | Prescription Use | X | OR | Over the Counter Use | (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE Concurrence of CDRH, Office of Device Evaluation (ODE) (Per 21 CFR 801.109 subpart D) IF NEEDED) Page 1 of 1 (Division Sign-Off) Division of Cardiovasculur Des 510(k) Number_for 0212
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...