AQUANOVA SUPER-ABSORBENT GELLING DRESSING, OTC

K070175 · Medtrade Products , Ltd. · QSY · Jul 25, 2007 · SU

Device Facts

Record IDK070175
Device NameAQUANOVA SUPER-ABSORBENT GELLING DRESSING, OTC
ApplicantMedtrade Products , Ltd.
Product CodeQSY · SU
Decision DateJul 25, 2007
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

MedTrade Products AQUANOVA Super-Absorbent OTC is indicated for minor burns, superficial cuts, lacerations and abrasions, and minor irritations of the skin. Under the supervision of a healthcare professional AQUANOVA may be used for wounds such as leg ulcers (Stages I-IV), diabetic ulcers, surgical wounds (postoperative, donor sites, dermatological), burns (first and second degree), and the management of surgical or traumatic wounds which have been left to heal by secondary intention. AQUANOVA may also be used for the local management of wounds that are prone to bleeding such as wounds that have been surgically or mechanically debrided, donor sites, and traumatic wounds. AQUANOVA can be used in the control of minor bleeding. Aditionally, AQUANOVA may be used for exudate absorption in oncology wounds (e.g. fungating cutaneous tumours, cutaneous metastases and Kaposi's sarcomas).

Device Story

AQUANOVA is a soft, woven, super-absorbent pad composed of chitosan and chitosan derivatives; intended for wound management and minor bleeding control. Device functions by absorbing wound exudate to form a gel, creating a moist healing environment. Used in clinical settings under professional supervision or as an OTC product for minor skin injuries. Applied topically to the wound site; remains intact for easy removal. Benefits include exudate management, hemostasis, and biocompatibility due to chitosan's non-toxic, biodegradable nature.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on descriptive comparisons of intended use, material composition, and physical characteristics. Safety of chitosan is supported by literature and its status as a GRAS food ingredient.

Technological Characteristics

Soft woven pad dressing; material: chitosan (poly-N-acetylglucosamine) and structural materials. Biocompatible, biodegradable. Sterilization: Gamma irradiation. Form factor: sterile single-use pouch.

Indications for Use

Indicated for minor burns, superficial cuts, lacerations, abrasions, and minor skin irritations (OTC). Under professional supervision, indicated for leg ulcers (Stages I-IV), diabetic ulcers, surgical/postoperative wounds, donor sites, dermatological wounds, first/second-degree burns, wounds healing by secondary intention, bleeding-prone wounds, and oncology wounds (fungating tumors, metastases, Kaposi's sarcomas).

Regulatory Classification

Identification

To temporarily control bleeding and cover external wounds.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is a stylized depiction of a human figure, while the FDA logo features the agency's acronym in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text. The logos are placed side by side. June 11, 2023 MedTrade Products, Ltd. c/o Jonathan D. Ranfield Director, Quality Assurance & Regulatory Affairs Crewe Business Park Crewe, Cheshire, CW1 6GL United Kingdom Re: K070175 Trade/Device Name: MedTrade Products AQUANOVA Super-Absorbent Dressing Regulatory Class: Unclassified Product Code: QSY Dear Jonathan D. Ranfield: The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated July 25, 2007. Specifically, FDA is updating this SE Letter because FDA has better categorized your device technology under product code QSY. Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Julie Morabito, OHT4: Office of Surgical and Infection Control Devices, 240-402-3839, Julie.Morabito@fda.hhs.gov. Image /page/0/Picture/8 description: The image shows the closing of a letter from Julie A. Morabito, Ph.D., Assistant Director at the Center for Devices and Radiological Health. Morabito is the Assistant Director of DHT4B: Division of Infection Control and Plastic Surgery Devices. She also works at OHT4: Office of Surgical and Infection Control Devices. The letter is from the Office of Product Evaluation and Quality. {2}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular fashion around the symbol. The caduceus is depicted in black, and the text is also in black, set against a white background. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # JUL 2 5 2007 Medtrade Products, Ltd. % Mr. Jonathan D. Ranfield Director, QA/RA Crewe Business Park Crewe, Cheshire CW1 6GL United Kingdom Re: K070175 Trade/Device Name: MedTrade Products AQUANOVA Super-Absorbent Dressing Regulatory Class: Unclassified Product Code: FRO Dated: July 13, 2007 Received: July 16, 2007 Dear Mr. Ranfield: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 -- Mr. Jonathan D. Ranfield This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a logally marketed predicate device results in a classification for your device and thus, perceits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html Sincerely yours, Mark A. Milherson Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # 3/: E Page 1 of 2 ### Indications for Use 510(k) Number K070175 Device Name: MedTrade Products AQUANOVA Super-Absorbent Dressing MedTrade Products AQUANOVA Super-Absorbent OTC is indicated for minor burns, superficial cuts, lacerations and abrasions, and minor irritations of the skin. Prescription Use (Per 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE, CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Mark A. Milliken (Division Sign-Of (Division Sign-On) Restorative, Division of Neurological Devices 510(k) Number {5}------------------------------------------------ Page 2 of 2 2 #### Indications for Use 510(k) Number K070175 Device Name: MedTrade Products AQUANOVA Super-Absorbent Dressing Indications for Use: Under the supervision of a healthcare professional AQUANOVA may be used for wounds such as leg ulcers (Stages I-IV), diabetic ulcers, surgical wounds (postoperative, donor sites, dermatological), burns (first and second degree), and the management of surgical or traumatic wounds which have been left to heal by secondary intention. AQUANOVA may also be used for the local management of wounds that are prone to bleeding such as wounds that have been surgically or mechanically debrided, donor sites, and traumatic wounds. AQUANOVÀ can be used in the control of minor bleeding. Aditionally, AQUANOVA may be used for exudate absorption in oncology wounds (e.g. fungating cutaneous tumours, cutaneous metastases and Kaposi's sarcomas). Prescription Use (Per 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE, CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) {6}------------------------------------------------ Safety & Effectiveness MedTrade Products AQUANOVA Super-Absorbent Dressing JUL 2 5 2007 Classification Name: 878 -- General and Plastic Surgery - FRO (Dressing) Contact_ Jonathan Ranfield - Director, Quality Assurance & Regulatory Affairs Prepared: January 31st 2007 Description: MedTrade Products AQUANOVA Super-Absorbent is a mixture of chitosan, chitosan derivatives and structural materials to produce a soft pad woven dressing. Chitosan is a material consisting of cellulostic polymer, poly-N-acetylglucosamine. A similar chitosan material has been self-affirmed as a GRAS (Generally Recognised As Safe) food ingredient in accordance with 21 CFR s 170.30. The GRAS report refers to safety studies in human beings and several apecies of animals. The studies sited represent research on the safety and use of chitosan, which have been published over a period of decades by scientists from around the world. This large body of scientific literature satisfies the requirement in 21 CFR s 170.30 (a), that a general recognition of safety requires common knowledge about the substance throughout the scientific community. Several bornedical applications of chitosan have already been reported. Chitosan has many advantages due to its non-toxicity and biodegradability without damaging the environment. It is a biocompatible material that breaks down slowly in to a harmless product, glucosamine that can be absorbed completely by the body. However, no product will be available to break down as with integrity studies have demonstrated that the dressing remains intact for removal from the wound. Med Trade Products AQUANOVA Super-Absorbent Dressings are provided in sterile single use pouches. Pouches will be provided in a carton with an information leaflet. MedTrade Products AQUANOVA Super-Absorbent Dressing is substantially equivalent to: - . Scion Cardio-Vascular Inc. Clo-sur P.A.D. K032986. It is similar to AQUANOVA in that it is a soft pad that provides an optimal wound-healing environment with exudates management. The manufacturers of Clo-sur P.A.D. also make antibacterial claims that MedTrade do not wish to make for AQUANOVA. - Convated's Aquacel Hydrofibre Wound Dressing K982116. It is similar in intended use to . AQUANOVA, both products gel in the presence of fluids to absorb large quantities of exudate and produce a moist wound healing environment. Both dressings have the same indications for use. Aquacel is based on sodium carboxymethycellulose technology. - Med Trade Products Celox Topical Hemostatic Granules. It is similar to AQUANOVA in absorbing large quantities of fluid to form a gel. It is also manufactured from chitosan. The product issues in the product is diffirent from AQUANOVA in its intended use for the professional market and the physical form (granules compared to a soft pad for AQUANOVA). A table of comparative features may be found below. l 그 {7}------------------------------------------------ # K070175 Page 2 og 3 ## COMPARATIVE FEATURES | Characteristic | MedTrade<br>Products<br>AQUANOVA<br>Super-Absorbent<br>Dressing | Scion Cardio-<br>Vascular Inc.<br>Clo-sur P.A.D. | Convatecs<br>Aquacel<br>Hydrofibre<br>Wound Dressing | MedTrade<br>Products<br>CELOX Topical<br>Hemostatic<br>Granules | |-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Chemistry | Absorbent<br>chitosan, a<br>material consisting<br>of cellulostic<br>polymer, poly-N-<br>acetylglucosamine | Absorbent<br>chitosan. | N/A - As sodium<br>carboxymethylcell<br>ulose based. | Absorbent<br>chitosan, a<br>material consisting<br>of cellulostic<br>polymer, poly-N-<br>acetylglucosamine | | Physical<br>Composition | Soft absorbent<br>pad | Soft absorbent<br>pad | Soft Absorbent<br>pad | N/A - Absorbent<br>granules | | Indications For<br>Use | Under the<br>supervision of a<br>healthcare<br>professional<br>AQUANOVA may<br>be used for<br>wounds such as<br>leg ulcers (Stages<br>I-IV), diabetic<br>ulcers, surgical<br>wounds (post-<br>operative, donor<br>sites,<br>dermatological),<br>burns (first and<br>second degree),<br>and the<br>management of<br>surgical or<br>traumatic wounds<br>which have been<br>left to heal by<br>secondary<br>intention.<br>AQUANOVA may<br>also be used for<br>the local<br>management of<br>wounds that are<br>prone to bleeding<br>such as wounds<br>that have been<br>surgically or<br>mechanically<br>debrided, donor<br>sites, and<br>traumatic wounds.<br>AQUANOVA can<br>be used in the<br>control of minor<br>bleeding. | A topical<br>haemostat that<br>provides an<br>optimal wound<br>healing<br>environment with<br>exudates<br>management. | Under the<br>supervision of a<br>healthcare<br>professional<br>Aquacel may be<br>used for wounds<br>such as leg ulcers<br>(Stages I-IV),<br>diabetic ulcers,<br>surgical wounds<br>(post-operative,<br>donor sites,<br>dermatological),<br>burns (first and<br>second degree),<br>and the<br>management of<br>surgical or<br>traumatic wounds<br>which have been<br>left to heal by<br>secondary<br>intention. Aquacel<br>may also be used<br>for the local<br>management of<br>wounds that are<br>prone to bleeding<br>such as wounds<br>that have been<br>surgically or<br>mechanically<br>debrided, donor<br>sites, and<br>traumatic wounds.<br>Aquacel can be<br>used in the control<br>of minor bleeding. | N/A for<br>professional use<br>as used as an<br>emergency<br>haemostat.<br><br>However, OTC<br>version is for<br>lacerations, minor<br>cuts and minor<br>bleeding the same<br>as AQUANOVA. | | Characteristic | MedTrade<br>Products<br>AQUANOVA<br>Super-Absorbent<br>Dressing | Scion Cardio-<br>Vascular Inc.<br>Clo-sur P.A.D. | Convatecs<br>Aquacel<br>Hydrofibre<br>Wound Dressing | MedTrade<br>Products<br>CELOX Topical<br>Hemostatic<br>Granules | | | AQUANOVA<br>Super-Absorbent<br>OTC is indicated<br>for minor burns,<br>superficial cuts,<br>lacerations and<br>abrasions, and<br>minor irritations of<br>the skin. | | Aquacel OTC is<br>indicated for the<br>exudates<br>management of<br>abrasions,<br>lacerations, minor<br>cuts, minor scalds<br>and minor burns –<br>taken from 510K<br>as no OTC<br>commercial<br>packaging could<br>be found. | | | Packaging | Paper Pouch | Unknown from<br>510K summary or<br>internet<br>information | Paper/Poly Pouch | Foil Pouch | | Sterilisation<br>Method | Gamma Irradiation | Unknown from<br>510K summary or<br>internet<br>information | Gamma Irradiation | Gamma Irradiation | {8}------------------------------------------------ # 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 / 6 Page 3 y 3 . . :
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...