ADVIA CHEMISTRY TOTAL BILIRUBIN_2

K063845 · Bayer Healthcare, LLC · JFM · Dec 7, 2007 · Clinical Chemistry

Device Facts

Record IDK063845
Device NameADVIA CHEMISTRY TOTAL BILIRUBIN_2
ApplicantBayer Healthcare, LLC
Product CodeJFM · Clinical Chemistry
Decision DateDec 7, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1110
Device ClassClass 2

Indications for Use

The ADVIA Chemistry Total Bilirubin_2 assay is for in vitro diagnostic use in the THE ADVIA Onemination of total bilirubin in human serum and plasma on the ADVIA® Chemistry Systems. Such measurements are used in the diagnosis and treatment of hemolytic, biliary, and liver disorders, including hepatitis and cirrhosis.

Device Story

Advia Chemistry Total Bilirubin 2 is a quantitative colorimetric assay for total bilirubin in human serum/plasma. Used on ADVIA Chemistry Systems (1200, 1650, 1800, 2400) in clinical laboratory settings. Principle: vanadate oxidation at pH 2.9 converts bilirubin to biliverdin; reaction causes decrease in optical density proportional to bilirubin concentration. Measured as endpoint reaction. Provides clinicians with quantitative bilirubin levels to assist in diagnosing/treating liver, biliary, and hemolytic disorders. Reagents are ready-to-use liquids.

Clinical Evidence

Bench testing only. Precision evaluated per CLSI EP5-A2 (5 levels, 10 days); total CV% ranged 1.1%–4.7% across platforms. Linearity confirmed 0.1–35 mg/dL. Analytical specificity showed no interference (<10% bias) with 50 mg/dL ascorbic acid, 1000 mg/dL hemoglobin, or 1000 mg/dL triglycerides. Method comparison (N=118-119) against predicate showed high correlation (r=0.998–1.000). Matrix equivalence (serum vs. lithium heparin plasma) confirmed (r=0.999–1.000).

Technological Characteristics

Quantitative colorimetric assay. Principle: Vanadate oxidation at pH 2.9. Two ready-to-use liquid reagents. Measuring range: 0.1–35 mg/dL. Platforms: ADVIA Chemistry Systems (1200, 1650, 1800, 2400). Traceable to AACC/NIST reference methods. Standards: CLSI EP5-A2, CLSI EP17-A.

Indications for Use

Indicated for quantitative determination of total bilirubin in human serum and plasma for adult patients to aid in diagnosis/treatment of hemolytic, biliary, and liver disorders. Not for use with neonates. Prescription use only.

Regulatory Classification

Identification

A bilirubin (total or direct) test system is a device intended to measure the levels of bilirubin (total or direct) in plasma or serum. Measurements of the levels of bilirubin, an organic compound formed during the normal and abnormal distruction of red blood cells, if used in the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block.

Predicate Devices

Submission Summary (Full Text)

{1} 1 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k063845 B. Purpose for Submission: New Device C. Measurand: Bilirubin D. Type of Test: Quantitative Colorimetric assay E. Applicant: Bayer Healthcare, LLC F. Proprietary and Established Names: Advia Chemistry Total Bilirubin 2 G. Regulatory Information: 1. Regulation section: 21 CFR 862.1110 Bilirubin (Total or Direct) Test System 2. Classification: Class II 3. Product code: JFM 4. Panel: 75, Chemistry {2} 2 H. Intended Use: 1. Intended use(s): See indications for use below. 2. Indication(s) for use: For in vitro diagnostic use in the quantitative determination of total bilirubin in human serum and plasma on the ADVIA Chemistry Systems. Such measurements are used in the diagnosis and treatment of hemolytic, biliary, and liver disorders, including hepatitis and cirrhosis. 3. Special conditions for use statement(s): For prescription use only. Not intended for use with neonates. 4. Special instrument requirements: ADVIA Chemistry System I. Device Description: The device consists of two ready to use reagents. Both reagents contain buffers and/or stabilizers. Calibrators were previously cleared under k030169 and controls under k883209. J. Substantial Equivalence Information: 1. Predicate device name(s): ADVIA IMS Total Bilirubin Assay 2. Predicate 510(k) number(s): k992399 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Specimen Type | Human serum or plasma (lithium heparin) | Human serum or plasma (lithium heparin) | | Reaction Type | Colorimetric endpoint | Colorimetric endpoint | | Calibration | Single point | Single point | {3} | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Principle | Vanadate oxidation | Diazotized sulfanic acid with blank | | Reagents | Two liquid reagents contained in system specific packaging | One lyophilized reagent, diluent, and one liquid reagent contained in system specific packaging | | Measuring range | 0.1 – 35 mg/dL | 0 – 40 mg/dL | | Intended use | For the quantitative determination of total bilirubin in adults | For the quantitative determination of total bilirubin | K. Standard/Guidance Document Referenced (if applicable): CLSI EP5-A2 Evaluation of Precision Performance of Quantitative Measurement Methods CLSI EP17-A Protocols for Determination of Limits of Detection and Limits of Quantitation L. Test Principle: The Total Bilirubin 2 assay is based on a vanadate oxidation method. Total bilirubin (conjugated and unconjugated) is oxidized by vanadate at about pH 2.9 to produce biliverdin. In the presence of detergent and vanadate, both conjugated and unconjugated bilirubin are oxidized. The oxidation reaction causes a decrease in the optical density which is proportional to the total bilirubin concentration in the sample. The concentration is measured as an endpoint reaction. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Within run and total imprecision were evaluated on each chemistry platform by testing five levels of serum pools. Each sample was assayed two times per run, 1 or 2 runs per day for at least 10 days. Precision estimates were computed according to CLSI EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Methods. The results are summarized below. {4} 4 ADVIA 1200 Precision | | | Within-run | | Total | | | --- | --- | --- | --- | --- | --- | | Specimen Type | Level mg/dL | SD | CV% | SD | CV% | | Serum | 1.1 | 0.00 | 0.4 | 0.04 | 3.6 | | Serum | 7.6 | 0.02 | 0.3 | 0.10 | 1.3 | | Serum | 14.7 | 0.08 | 0.5 | 0.21 | 1.4 | | Serum | 22.1 | 0.16 | 0.7 | 0.32 | 1.4 | | Serum | 27.6 | 0.32 | 1.2 | 0.47 | 1.7 | ADVIA 1650/1800 Precision | | | Within-run | | Total | | | --- | --- | --- | --- | --- | --- | | Specimen Type | Level mg/dL | SD | CV% | SD | CV% | | Serum | 1.0 | 0.01 | 1.3 | 0.02 | 2.0 | | Serum | 7.1 | 0.15 | 2.2 | 0.19 | 2.6 | | Serum | 15.4 | 0.06 | 0.4 | 0.49 | 3.2 | | Serum | 22.1 | 0.11 | 0.5 | 0.26 | 1.2 | | Serum | 27.5 | 0.14 | 0.5 | 0.30 | 1.1 | ADVIA 2400 Precision | | | Within-run | | Total | | | --- | --- | --- | --- | --- | --- | | Specimen Type | Level mg/dL | SD | CV% | SD | CV% | | Serum | 1.0 | 0.05 | 4.6 | 0.05 | 4.7 | | Serum | 7.2 | 0.15 | 2.0 | 0.16 | 2.3 | | Serum | 15.5 | 0.12 | 0.8 | 0.25 | 1.6 | | Serum | 21.8 | 0.09 | 0.4 | 0.21 | 1.0 | | Serum | 27.1 | 0.30 | 1.1 | 0.41 | 1.5 | b. Linearity/assay reportable range: Nine equally spaced levels of serum pools prepared using high and low linearity pools were tested on the ADVIA 1200, 1650, and 2400 platforms. The sample concentrations tested ranged from approximately $0.17\mathrm{mg/dL}$ to $37.5\mathrm{mg/dL}$. All results were within $\pm 5\%$ of the expected value except the lowest dilution level which had concentrations approaching the limits of detection of the assay. The results of the lowest dilution level were all with $\pm$ {5} 0.03 mg/dL of the expected value. The reportable range for all platforms is 0.1 mg/dL to 35 mg/dL. c. Traceability, Stability, Expected values (controls, calibrators, or methods): The sponsor states that the calibrators are traceable to the AACC reference method, which used reference materials from NIST, via patient correlation. Protocols and acceptance criteria were reviewed for reagent stability. The reagents are stable until their expiration dates when stored as instructed. Calibrators were previously cleared under k030169 and controls under k883209. d. Detection limit: The limit of detection and the limit of the blank were determined as outlined in CLSI EP17-A. A blank sample (saline) was evaluated 64 times on each platform to calculate the limit of the blank. A low level sample was also run 64 times on each platform and used to calculate the limit of detection. The observed limits of detection on the three platforms were all below the claimed low end of the measuring range of the assay (0.1 mg/dL). e. Analytical specificity: The interference study was carried out in the presence of approximately 1 mg/dL and 20 mg/dL total bilirubin. The samples were evaluated on the ADVIA 1200, 1650, and 2400. All samples were evaluated in duplicate with multiple concentrations of each interferant. Lack of interference was defined as a bias of &lt;10% compared to a sample without the interfering substance present. The assay showed no interference with up to 50 mg/dL ascorbic acid, 1000 mg/dL hemoglobin, and 1000 mg/dL triglycerides. f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: Correlation studies were performed by first comparing adult patient samples on the ADVIA 1650 to the ADVIA IMS predicate device. The same adult patient samples were then measured on the ADVIA 1200 and ADVIA 2400 and compared to the ADVIA 1650. The studies were performed over 5 days and the results are summarized below. {6} | Specimen | System | Comparison system | N | Regression equation | r | | --- | --- | --- | --- | --- | --- | | Serum | ADVIA 1650 | ADVIA IMS | 118 | Y=0.925x+0.12 | 0.998 | | Serum | ADVIA 1200 | ADVIA 1650 | 119 | Y=1.036x-0.05 | 1.000 | | Serum | ADVIA 2400 | ADVIA 1650 | 119 | Y=0.999x-0.02 | 1.000 | b. Matrix comparison: The serum/plasma equivalence study was carried out using lithium heparin plasma samples and the corresponding serum samples drawn from volunteers. Samples were spiked with unconjugated bilirubin in order to cover the measuring range. The samples, ranging from 0.3 mg/dL to 34.4 mg/dL, were run on all three platforms and the results are shown below. | Platform | N | Regression equation | r | | --- | --- | --- | --- | | ADVIA 1200 | 49 | Y=1.012x-0.00 | 1.000 | | ADVIA 1650 | 58 | Y=1.019x-0.02 | 0.999 | | ADVIA 2400 | 57 | Y=1.012x+0.01 | 0.999 | 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 1. Expected values/Reference range: Expected values from Tietz NW. Clinical Guide to Laboratory Tests. 3rd ed. Philadelphia, PA WB Saunders Company; 1995: 88-91 {7} | | mg/dL | | --- | --- | | Adult | 0.2 – 1.2 | N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 7
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...