BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM MINOCYCLINE (1-32 UG/ML) GRAM NEGATIVE ID/AST OR AST ONLY
K063486 · Becton, Dickinson & CO · LON · Mar 7, 2007 · Microbiology
Device Facts
Record ID
K063486
Device Name
BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM MINOCYCLINE (1-32 UG/ML) GRAM NEGATIVE ID/AST OR AST ONLY
Applicant
Becton, Dickinson & CO
Product Code
LON · Microbiology
Decision Date
Mar 7, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1645
Device Class
Class 2
Intended Use
The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic gram-negative and gram-positive bacteria of human origin.
Device Story
BD Phoenix™ Automated Microbiology System performs antimicrobial susceptibility testing (AST) on bacterial isolates. System utilizes molded polystyrene trays with 136 micro-wells containing dried reagents and specific inoculum broth. Pure culture colonies suspended in broth, equated to 0.5 McFarland standard, and inoculated into AST broth containing redox indicator. Panels barcode-scanned and incubated at 35°C in instrument. Redox indicator changes color (blue to pink to colorless) based on organism growth in presence of serial two-fold dilutions of minocycline (1-32 µg/mL). Instrument reads color changes every 20 minutes; software interprets results to determine MIC and categorical interpretation (SIR). BDXpert software applies CLSI-derived rules to interpret results; automatically suppresses Shigella species susceptibility criteria. System provides automated, rapid AST results (4-16 hours) to clinicians to guide antimicrobial therapy decisions.
Clinical Evidence
Performance evaluated using clinical, stock, and challenge isolates compared to CLSI reference broth microdilution method. Metrics include Essential Agreement (EA) and Category Agreement (CA). Reproducibility evaluated at three sites using gram-negative isolates in triplicate over three days; intra-site reproducibility >90%, inter-site >95%.
Indicated for rapid identification and in vitro antimicrobial susceptibility testing of pure culture isolates of aerobic and facultative anaerobic gram-negative and gram-positive bacteria, specifically Acinetobacter species, Enterobacter aerogenes, Escherichia coli, Klebsiella species, and Shigella species.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
Predicate Devices
VITEK® System (PMA No. N50510)
BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, K060324)
BD Phoenix™ Automated Microbiology System with Ofloxacin (K020323)
BD Phoenix™ Automated Microbiology System with Levofloxacin (K020322)
Related Devices
K032275 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-IMIPENEM 1-16 UG/ML GRAM-NEGATIVE ID/AST OR AST ONLY PHOENIX PANELS · Becton, Dickinson & CO · Sep 29, 2003
K033784 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - ERYTHROMYCIN 0.0625-8 UG/ML · Becton, Dickinson & CO · Feb 3, 2004
K063300 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - CEFPODOXIME (GN) 0.125-8UG/ML · Becton, Dickinson & CO · Jan 29, 2007
K052269 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM -CIPROFLOXACIN GN · Becton, Dickinson & CO · Oct 3, 2005
K023858 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM · Becton, Dickinson & CO · Dec 20, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(K) SUMMARY
| SUBMITTED BY: | Becton, Dickinson and Company<br>7 Loveton Circle<br>Sparks, MD 21152<br>Phone: (410)-316-4287<br>Fax: (410)-316-4499 | | |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|--|
| CONTACT NAME: | Monica Giguere<br>Regulatory Affairs Specialist | MAR 07 2007 | |
| DATE PREPARED: | February 12, 2007 | | |
| DEVICE TRADE NAME: | BD Phoenix™ Automated Microbiology System –<br>Minocycline 1-32 µg/mL | | |
| DEVICE COMMON NAME: | Antimicrobial susceptibility test system-short incubation | | |
| DEVICE CLASSIFICATION: | Fully Automated Short-Term Incubation Cycle Antimicrobial<br>Susceptibility Device, 21 CFR 866.1645 | | |
| PREDICATE DEVICES: | VITEK® System (PMA No. N50510) and BD Phoenix™<br>Automated Microbiology System with Gatifloxacin (K020321,<br>May 23, 2002, and K060324, May 25, 2006), Ofloxacin (K020323,<br>April 14, 2002), and Levofloxacin (K020322, March 27, 2002). | | |
| INTENDED USE: | The BD Phoenix™ Automated Microbiology System is intended<br>for the rapid identification and in vitro antimicrobial susceptibility<br>testing of isolates from pure culture of most aerobic and facultative<br>anaerobic gram-negative and gram-positive bacteria of human<br>origin. | | |
## DEVICE DESCRIPTION:
The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:
- BD Phoenix instrument and software. .
- BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents for . AST determinations.
- BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. .
- BD Phoenix AST Broth used for performing AST tests only. .
- BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth . determination.
{1}------------------------------------------------
The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. Phoenix panels are inoculated with a specified organism density and placed into the instrument.
The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of twofold doubling dilution concentrations.
The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).
#### DEVICE COMPARISON:
The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the CLSI reference broth microdilution method. This premarket notification provides data supporting the use of the BD Phoenix™ Automated Microbiology System gram-negative ID/AST or AST only Phoenix panels with this antimicrobial agent.
#### SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:
The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the CLSI reference broth microdilution method (AST panels prepared according to NCCLS M7). The system has been evaluated as defined in the FDA Draft guidance document. "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", March 8, 2000.
#### Site Reproducibility
Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was evaluated at three sites using a panel of gram-negative isolates. Each site tested the isolates in triplicate on three different days using one lot of gram-negative Phoenix panels containing this antimicrobial agent and associated reagents.
The results of the study demonstrate for the this antimicrobial agent there was an overall intra-site reproducibility of greater than 90% and an overall inter-site reproducibility greater than 95% for the gram-negative isolates tested.
BD Diagnostic Systems Becton, Dickinson and Company
Page 6
{2}------------------------------------------------
## Clinical Studies
Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with the gram-negative Phoenix panel format containing this antimicrobial agent. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the CLSI reference broth microdilution method.
The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within ± one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, and resistant).
Table 1 summarizes the performance for the isolates tested in this study.
#### Performance of BD Phoenix System for Gram-Negative Organisms by Drug Table 1:
| THE REPORTERIAL FOR IN AN AND<br>THE LA LA CONSUL BOOK CHANAL CO WEEK SHOW SHOW SHOW FLAND LEASE<br>Antimicrobial | and and the first for the first from the many of the first for the first for the first and<br>Free of Canadian In and Callery<br>Concentration | September 1980 - And Article<br>Comments of Children a more the couple the couple of concess concession<br>יהו<br>n<br>Print | 10/<br>11.00 | Comments of the comments of the survey a response of the later in the larger of the larges of the larges of the larges of the larges of the larges of the larges of the lands<br>-A | 0<br>114<br>C |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| -<br>I . I . LEBERT SENSION OF A NO AND IN WANTED OF A NOT A WARRENT OND ANY AND THE WATER<br>Minocycline<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | (<br>ug/m<br>Carrely Controller Charles And All A | 000<br>200 - | 01 0<br><br>Carrelation Collection Collection Collection Comments and | 1 | C<br>/ Hol |
#### Conclusions Drawn from Substantial Equivalence Studies
The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA', February 5, 2003. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002, and K060324, May 25, 2006), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322, March 27, 2002).
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Monica Giguere Regulatory Affairs Specialist BD Diagnostics Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152
MAR 0 7 2007
k063486 Trade/Device Name: BD Phoenix TM Automated Microbiology System Minocycline (1-32 ug/mL) - Gram-Negative ID/AST or AST only Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: November 15, 2006 Received: November 17, 2006
Dear Ms. Giguere:
Re:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Sally, attorn
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
510(k) Number: KO63486
Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent Minocycline (1-32 ug/mL) - Gram-negative ID/AST or AST only Phoenix panels.
Indications for Use:
The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus, and Streptococcus.
This premarket notification is for the addition of the antimicrobial agent Minocycline at concentrations of 1-32 µg/mL to Gram-negative ID/AST or AST only Phoenix panels. Minocycline has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.
#### Active In Vitro and in Clinical Infections Against:
Acinetobacter species Enterobacter aerogenes Escherichia coli
Klebsiella species Shigella species
Prescription Use (Per 21 CFR 801.109) Over-the-Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Luddie Toole
Division Sign-Off
Sion Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k)_K063486
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.