BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM FOR USE WITH THE ANTIMICROBIAL AGENT GATIFLOXACIN 0.25-8 UG/ML - GRAM-NEGATIVE
K060324 · Becton, Dickinson & CO · LON · May 25, 2006 · Microbiology
Device Facts
Record ID
K060324
Device Name
BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM FOR USE WITH THE ANTIMICROBIAL AGENT GATIFLOXACIN 0.25-8 UG/ML - GRAM-NEGATIVE
Applicant
Becton, Dickinson & CO
Product Code
LON · Microbiology
Decision Date
May 25, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1645
Device Class
Class 2
Indications for Use
The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.
Device Story
System performs rapid identification and antimicrobial susceptibility testing (AST) of bacterial isolates. Input: pure culture bacterial isolates prepared in Phoenix ID Broth (0.5 McFarland standard). Process: automated microdilution in sealed 136-well polystyrene panels; panels incubated at 35°C; instrument measures redox indicator changes and bacterial turbidity every 20 minutes. Output: identification of isolate, MIC values, and category interpretations (S, I, R). Used in clinical microbiology laboratories by trained personnel. Output assists clinicians in selecting appropriate antibiotic therapy for patients with bacterial infections.
Clinical Evidence
Performance evaluated using clinical, stock, and challenge isolates compared to CLSI reference broth microdilution method. Study included 2213 Gram-negative isolates for Gatifloxacin (0.25-8 µg/mL). Results: 98.8% Essential Agreement (EA) and 95.8% Category Agreement (CA). Reproducibility study at three sites showed >90% intra-site and >95% inter-site reproducibility.
Indicated for in vitro quantitative MIC determination of antimicrobial susceptibility for Gram-negative (Enterobacteriaceae, Non-Enterobacteriaceae) and Gram-positive (Staphylococcus, Enterococcus, Streptococcus) bacterial isolates from pure culture. Gatifloxacin results for Enterobacteriaceae are indicated only for urinary tract isolates.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
Predicate Devices
VITEK® System (PMA No. N50510)
BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321)
BD Phoenix™ Automated Microbiology System with Ofloxacin (K020323)
BD Phoenix™ Automated Microbiology System with Levofloxacin (K020322)
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(K) SUMMARY
| SUBMITTED BY: | Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152 Phone: (410) 316-4278 Fax: 410-316-4499 |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | MAY 25 2006 |
| CONTACT NAME: | Monica E Giguere Regulatory Affairs Specialist |
| DATE PREPARED: | May 11, 2006 |
| DEVICE TRADE NAME: | BD Phoenix™ Automated Microbiology System - Gatifloxacin 0.25-8 µg/mL |
| DEVICE COMMON NAME: | Antimicrobial susceptibility test system-short incubation |
| DEVICE CLASSIFICATION: | Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Device, 21 CFR 866.1645 |
| PREDICATE DEVICES: | VITEK® System (PMA No. N50510) and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322, March 27, 2002). |
| INTENDED USE: | The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram- positive bacteria of human origin. |
## DEVICE DESCRIPTION:
The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:
- BD Phoenix instrument and software.
- BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents . for AST determinations.
- BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. ●
- BD Phoenix AST Broth used for performing AST tests only. .
- BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth . determination. .
{1}------------------------------------------------
The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. For each isolation equivalent to a 0.5 McFarland standard is prepared in Phoenix ID Broth.
The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.
The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).
## DEVICE COMPARISON:
The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the CLSI reference broth microdilution method. This premarket notification provides data for additional organism groups with Gatifloxacin 0.25-8 ug/mL and the BD PhoenixTM Automated Microbiology System.
## SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:
The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the CLSI reference broth microdilution method (AST panels prepared according to NCCLS M7). The system has been evaluated as defined in the FDA Draft guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA," February 5, 2003.
## Site Reproducibility
Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was evaluated at three sites using a panel chosen isolates. Each site tested the isolates in triplicate on three different days using one lot of Phoenix panels containing the antimicrobial agents and associated reagents.
The results of the study demonstrate for each antimicrobial agent there was an overall intra-site reproducibility of greater than 90% and an overall inter-site reproducibility greater than 95% for the isolates tested.
{2}------------------------------------------------
# Clinical Studies
Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with the Phoenix panel formats containing antimicrobial agents. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the CLSI reference broth microdilution method.
The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within ± one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD PhoenixTM Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, and resistant).
Table 1 summarizes the performance for the isolates tested in this study.
| Antimicrobial | Concentration | EA (n) | EA (%) | CA (n) | CA (%) |
|-----------------|-------------------------|--------|--------|--------|--------|
| Gatifloxacin GN | 0.25-8 $ µg/mL $ | 2213 | 98.8 | 2213 | 95.8 |
#### Table 1: Performance of RD Phoenix System by Drug
# Conclusions Drawn from Substantial Equivalence Studies
The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA draft guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA," February 5, 2003. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322, March 27, 2002).
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird-like figure.
#### Public Health Service
Food and Drug Administration
2098 Gaither Road Rockville MD 20850
## MAY 2 5 2006
Ms. Monica E. Giguere Regulatory Affairs Specialist BD Diagnostics Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152
k060324 Re:
Trade/Device Name: BD Phoenix™ Automated Microbiology System Gatifloxacin 0.25-8 ug/mL - Gram-Negative ID/AST or AST Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: May 11, 2006 Received: May 16, 2006
Dear Ms. Giguere:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
### Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours,
Sally, a story
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Page 1 of 1
510(k) Number: K060324
Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent Gatifloxacin 0.25-8 ug/mL - Gram-negative ID/AST or AST only Phoenix panels.
### Indications for Use:
The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus, and Streptococcus.
This premarket notification is for additional Gram-negative organism groups and Gatifloxacin (0.25-8 ug/mL) on the BD Phoenix Automated Microbiology System. Gatifloxacin has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDAapproved package insert for this antimicrobial agent.
### Active In Vitro and in Clinical Infections Against:
Aerobic Gram-negative microorganisms Escherichia coli Klebsiella pneumoniae Proteus mirabilis
#### Active In Vitro Against:
Aerobic Gram-negative microorganisms Acinetobacter Iwoffii Citrobacter koseri Citrobacter freundii Enterobacter aerogenes Enterobacter cloacae Klebsiella oxytoca Morganella morganii Proteus vulgaris
Results for Enterobacteriaceae tested with Gatifloxacin should only be reported for isolates recovered from the urinary tract.
Prescription Use (Per 21 CFR 801.109) Over-the-Counter Use
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Luddite Lade
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K060324