VARIANT II TURBO HEMOGLOBIN A1C PROGRAM, HEMOGLOBIN TESTING SYSTEM WITH CDM 4.0

K063400 · Bio-Rad Laboratories, Inc. · LCP · Dec 1, 2006 · Hematology

Device Facts

Record IDK063400
Device NameVARIANT II TURBO HEMOGLOBIN A1C PROGRAM, HEMOGLOBIN TESTING SYSTEM WITH CDM 4.0
ApplicantBio-Rad Laboratories, Inc.
Product CodeLCP · Hematology
Decision DateDec 1, 2006
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 864.7470
Device ClassClass 2

Intended Use

The Variant II Turbo Hemoglobin A1C Program, Hemoglobin Testing System with CDM 4.0 is intended for the measurement of Hemoglobin A1c in human whole blood.

Device Story

Variant II Turbo Hemoglobin A1C Program; automated hemoglobin testing system; measures Hemoglobin A1c in human whole blood; modification involves software upgrade from Windows NT (ver. 3.6T) to Windows XP (ver. 4.0) for Clinical Data Management (CDM); data analysis algorithms moved to external software module; system used in clinical laboratory settings; provides quantitative A1c results for diabetes management; modification verified via design control activities and risk analysis (FMEA).

Clinical Evidence

Bench testing only. Validation protocols were performed to address hazards identified in the risk analysis (FMEA). Results demonstrated that predetermined acceptance criteria were met.

Technological Characteristics

Cation exchange HPLC; utilizes cation exchange cartridge for hemoglobin separation. Software: CDM 4.0 running on Microsoft Windows XP. Database management allows direct data deletion. Connectivity: CD-R for data backup. Professional use only.

Indications for Use

Indicated for monitoring long-term glucose control in individuals with diabetes mellitus. For professional use only.

Regulatory Classification

Identification

A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Summary of Safety and Effectiveness DEC - 1 2006 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is: K_()(23 Y CO_ | Submitter: | Bio-Rad Laboratories, Inc.<br>Clinical System Division<br>4000 Alfred Nobel Drive,<br>Hercules, California 94547<br>Phone: (510) 741-5309<br>FAX: (510) 741-6471 | |------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Jackie Buckley<br>Regulatory Affairs Representative | | Date of Summary Preparation: | Nov. 8, 2006 | | Device Name: | VARIANT™ II TURBO Hemoglobin A1c Program run<br>on the VARIANT II TURBO Hemoglobin Testing<br>System using Clinical Management System (CDM) 4.0 | | Classification Name: | HbA1c: Assay, Glycosylated Hemoglobin<br>[21CFR 864.7470 / Prod. Code LCP] | | Predicate Devices: | VARIANT II TURBO Hemoglobin A1c<br>Bio-Rad Laboratories, Inc.<br>[K040872, April 15, 2004] | {1}------------------------------------------------ #### Intended Use: The Bio-Rad VARIANT™ II TURBO Hemoglobin Alc Program is intended for the percent determination of hemoglobin A1c in human whole blood using ion-exchange high performance liquid chromatography (HPLC). The VARIANT II TURBO Hemoglobin A1c Program is intended for Professional Use Only. For In Vitro Diagnostic Use. #### Indications for Use: Measurement of percent hemoglobin A1c is effective in monitoring long-term glucose control in individuals with diabetes mellitus. #### New Device Description The VARIANT II TURBO Hemoglobin Testing System uses the principles of high performance liquid chromatography (HPLC). The VARIANT II TURBO Hemoglobin A1c Program is based on chromatographic separation of Hemoglobin A1c on a cation exchange cartridge. The new feature in this submission is the upgrade in CDM software. The current software (CDM 3.6T) requires Windows NT. This product is nearing the end of its lifecycle. CDM 4.0 software is needed to transfer the CDM software to Microsoft XP Operating System. #### Technical Characteristics Compared to Predicate The VARIANT™ II TURBO Hemoglobin A1c Program run on the VARIANT™ II TURBO Hemoglobin Testing System with CDM 4.0 has the same basic technical characteristics as the predicate VARIANT II TURBO Hemoglobin A1c Program (k) 040872. The technical characteristics between the two submissions are summarized in the following tables: {2}------------------------------------------------ | Summary of Technological Characteristic Similarities in Comparison to Predicate Device | | | |----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Characteristics | New Device:<br>VARIANT II TURBO Hemoglobin A1c<br>Program run on the VARIANT II TURBO<br>Hemoglobin Testing System using CDM 4.0 | Predicate Device: (k) 040872<br>VARIANT II TURBO Hemoglobin A1c<br>Program run on the VARIANT II TURBO<br>Hemoglobin Testing System using CDM 3.6T | | Intended Use(s) | The VARIANT II TURBO Hemoglobin A1c is<br>intended for the percent determination of<br>hemoglobin A1c in human whole blood using<br>ion-exchange high-performance liquid<br>chromatography (HPLC).<br>The Bio-Rad VARIANT II TURBO<br>Hemoglobin A1c Program is intended for<br>Professional Use Only.<br>For In Vitro Diagnostic Use. | The VARIANT II TURBO Hemoglobin A1c is<br>intended for the percent determination of<br>hemoglobin A1c in human whole blood using<br>ion-exchange high-performance liquid<br>chromatography (HPLC).<br>The Bio-Rad VARIANT II TURBO<br>Hemoglobin A1c Program is intended for<br>Professional Use Only.<br>For In Vitro Diagnostic Use. | | Indication(s) for Use | Measurement of the percent hemoglobin A1c is<br>effective in monitoring long-term glucose<br>control in individuals with diabetes mellitus. | Measurement of the percent hemoglobin A1c is<br>effective in monitoring long-term glucose<br>control in individuals with diabetes mellitus. | | Assay Principle | Cation exchange high performance liquid<br>chromatography | Cation exchange high performance liquid<br>chromatography | | CDM Software<br>version | 4.0 | 3.6T | | Microsoft software | Windows XP | Windows NT | | Object Store version | 6.0 | 4.0 | | Backup and Restore | Use Windows operation to write only data to<br>CD-R | Used Easy CD writer Read/Write | | Database<br>Management | Delete data directly from database | Substitute database with a blank database | # VARIANT™ II TURBOHemoglobin A1c (k)040872 {3}------------------------------------------------ # Testing To Establish Substantial Equivalence: #### Accuracy: #### VARIANT II TURBO Hemoglobin HbA1c Program (3 minute) Method correlation between Bio-Rad VARIANT II TURBO Hemoglobin A1c Program with CDM 4.0 and VARIANT II TURBO Hemoglobin A1c Program with CDM 3.6T was evaluated using 40 EDTA whole blood samples ranging from (4.4% -11.6%) HbA1c. The results are presented in the following table: | VARIANT II TURBO Hemoglobin A1c Program Correlation | | | | | |-----------------------------------------------------|----|--------|--------|-----------| | Regression Method | n | r2 | Slope | Intercept | | Least Squares | 40 | 0.9991 | 1.0174 | 0.0559 | #### VARIANT II TUDDO Uomoclobin 41a Drogram ---------- #### Precision: ### VARIANT II TURBO Hemoglobin A1c Program The following precision table provides comparison data on the precision between VARIANT II TURBO Hemoglobin A1c Program with CDM 4.0 vs. CDM 3.6T each utilizing EDTA whole blood patient samples. Method precision was performed using a protocol based on the NCCLS Evaluation protocol, EP5-A for the VARIANT II TURBO Hemoglobin A1c Program with CDM 4.0 and 3.6T. The protocols for both VARIANT II TURBO Hemoglobin A1c Program with CDM 4.0 and 3.6T are similar. In each duplicate daily run for both verification studies, duplicate aliquots of normal HbAlc and diabetic HbA1c patient samples and controls were each analyzed per run. The position of the precision specimens in each run was randomized to simulate normal laboratory conditions. The precision data for the VARIANT II TURBO with CDM 3.6T was over 20 working days while the data for VARIANT II TURBO with CDM 4.0 was over 10 working days. Although precision samples are different, since they were run at different time periods, the precision results between the VARIANT II TURBO Hemoglobin A1c Program with CDM 4.0 and CDM 3.6T are equivalent. A summary of combined comparative precision results are presented in the following precision table. #### VARIANT II TURBO HbA1c with CDM 4.0 vs. VARIANT II TURBO HbA1c with CDM 3.6T - Precision | | VII TURBO HbA1c<br>with CDM 4.0 | | VII TURBO HbA1c<br>with CDM 3.6T | | |------------------------|---------------------------------|-----------------|----------------------------------|-----------------| | | Normal Sample | Diabetic Sample | Normal Sample | Diabetic Sample | | n= (number of samples) | 40 | 40 | 80 | 80 | | Mean (%HbA1c) | 5.4 | 9.7 | 6.2 | 12.5 | | Within run (%CV) | 0.9 | 0.9 | 0.8 | 0.5 | | Total Precision (%CV) | 1.2 | 1.9 | 1.9 | 2.6 | {4}------------------------------------------------ ## Conclusion: The similarities of the intended use and the general performance characteristics and results of the newly described and evaluated VARIANT II TURBO Hemoglobin A1c Program run on the VARIANT II TURBO Hemoglobin Testing System with CDM 4.0 are nearly identical to or logical extensions of those for cleared predicate program systems [i.e., VARIANT II TURBO Hemoglobin Alc Program run on the VARIANT II TURBO Hemoglobin Testing System with CDM 3.6T (k)040872]. Thus, one may conclude, based on the use of the same HPLC technology, and the nearly equivalent results obtained for the correlation and precision versus the corresponding results obtained with the predicate system that the new VARIANT II TURBO Hemoglobin A1c Program run on the VARIANT II TURBO Hemoglobin Testing System with CDM 4.0 is substantially equivalent to the cleared and currently marketed predicate system. {5}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing protection and care. The eagle is depicted in a simple, abstract design with bold lines. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Jackic Buckley Bio-Rad Laboratories, Inc. Clinical System Division 4000 Alfred Nobel Drive Hercules, California 94547 DEC - 1 2006 Re: k063400 Trade/Device Name: Varient II Turbo Hemoglobin Alc Program, Hemoglobin Testing System With CDM 4.0 Regulation Number: 21 CFR 864.7470 Regulation Name: Glycosylated hemoglobin assay Regulatory Class: Class II Product Code: LCP, LDM Dated: November 8, 2006 Received: November 9, 2006 Dear Ms. Buckley: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {6}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Alberto Gutt Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ # Indications for Use 510(k) Number (if known):________k063400 Variant II Turbo Hemoglobin A1c Program, Hemoglobin Testing Device Name:_ System With CDM 4.0 Indications For Use: The Bio-Rad VARIANT II TURBO Hemoglobin A1C Program with CDM 4.0 is intended THE DIO Rad VARIANT PROGED in A1c in human whole blood using ionexchange high-performance liquid chromatography (HPLC). The Bio-Rad VARIANT II TURBO Hemoglobin A1C Program with CDM 4.0 is intended for Professional Use Only. For In Vitro Diagnostic Use. Measurement of percent hemoglobin A1C is effective in monitoring long-term glucose control in individuals with diabetes mellitus Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Carol Benson Division Sign-Off Division Sign-Off Page 1 of 1 Office of In Vitro Diagnostic Device Evaluation and Safety K06340
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...