MODIFICATION TO VARIANT II: HEMOGLOBIN A1C PROGRAM, BETA-THALASSEMIA SHORT PROGRAM AND TOTAL GHB PROGRAM

K063643 · Bio-Rad Laboratories Inc., Clinical Systems Divisi · LCP · Dec 27, 2006 · Hematology

Device Facts

Record IDK063643
Device NameMODIFICATION TO VARIANT II: HEMOGLOBIN A1C PROGRAM, BETA-THALASSEMIA SHORT PROGRAM AND TOTAL GHB PROGRAM
ApplicantBio-Rad Laboratories Inc., Clinical Systems Divisi
Product CodeLCP · Hematology
Decision DateDec 27, 2006
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 864.7470
Device ClassClass 2

Intended Use

The VARIANT II Hemoglobin A1c Program is intended for the percent determination of hemoglobin A1c in human whole blood using ion-exchange high-performance liquid chromatography (HPLC). The VARIANT II Hemoglobin Alc Program is intended for use only with the Bio-Rad VARIANT II Hemoglobin Testing System. For in vitro diagnostic use. The VARIANT II Beta thalassemia Short Program is intended for the separation and area percent determinations of hemoglobin A2 and F, and as an aid in the identification of abnormal hemoglobins in human whole blood using ionexchange high-performance liquid chromatography (HPLC). The VARIANT II Beta-thalassemia Short Program is intended for use only with the Bio-Rad VARIANT II Hemoglobin Testing System. For in vitro diagnostic use. The VARIANT II Total GHb Program is intended for the separation and area percent determination of total glycated hemoglobin (GHb) in human whole blood using ion-exchange high-performance liquid chromatography (HPLC). The VARIANT II Total GHb Program with is intended for use only with the Bio-Rad VARIANT II Hemoglobin Testing System. For in vitro diagnostic use.

Device Story

Variant II system performs hemoglobin A1c, beta-thalassemia, and total GHB analysis; utilizes ion-exchange high-performance liquid chromatography (HPLC). Modification involves software upgrade for Clinical Data Management (CDM) from Windows NT (v3.6T) to Windows XP (v4.0). Data analysis algorithms transitioned from base CDM software integration to external software module. System used in clinical laboratory settings by trained technicians. Output provides quantitative analysis of hemoglobin variants/GHB for clinical diagnostic decision-making.

Clinical Evidence

Bench testing only. Validation protocols performed to address hazards identified in FMEA risk analysis; results demonstrated that predetermined acceptance criteria were met.

Technological Characteristics

Ion-exchange HPLC system. Software-controlled data management. Modification: CDM software migration from Windows NT (v3.6T) to Windows XP (v4.0). Data analysis algorithms moved to external software module.

Indications for Use

Indicated for use in the Variant II: Hemoglobin A1c Program, Beta-Thalassemia Short Program, and Total GHB Program for clinical laboratory analysis.

Regulatory Classification

Identification

A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Summary of Safety and Effectiveness This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is: K_06 3643 | Submitter: | Bio-Rad Laboratories, Inc.<br>Clinical System Division<br>4000 Alfred Nobel Drive,<br>Hercules, California 94547<br>Phone: (510) 741-5309<br>FAX: (510) 741-6471<br>DEC 27 2006 | |------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Jackie Buckley<br>Regulatory Affairs Representative III | | Date of Summary Preparation: | December 4, 2006 | | Device Name: | VARIANTTM II Hemoglobin A1c, VARIANTTM II Beta-<br>thalassemia, and VARIANTTM II Total GHb Programs<br>run on the VARIANT II Hemoglobin Testing System<br>using Clinical Management System (CDM) 4.0 | | Classification Name: | HbA1c: Assay, Glycosylated Hemoglobin<br>[21CFR 864.7470 / Prod. Code LCP]<br>HbA2: Hemoglobin A2 Quantitation<br>[21CFR 864.7400/Prod. Code :JPD] | | Predicate Devices: | VARIANTTM II Hemoglobin A1c Program<br>Bio-Rad Laboratories, Inc.<br>[K984268, December 17, 1998] | | | VARIANTTM II Beta thalassemia Short Program<br>Bio-Rad Laboratories, Inc.<br>[K991127, June 10, 1999] | | | VARIANTTM II Total GHb Program<br>Bio-Rad Laboratories, Inc.<br>[K003280, December 19, 2000] | {1}------------------------------------------------ #### Intended Use: This software submission covers three assays with three different intended uses: #### Hemoglobin A1c: The VARIANT II Hemoglobin A1c Program is intended for the percent determination of hemoglobin A1c in human whole blood using ion-exchange high-performance liquid chromatography (HPLC). The VARIANT II Hemoglobin Alc Program is intended for use only with the Bio-Rad VARIANT II Hemoglobin Testing System. For in vitro diagnostic use. #### Beta-thalassemia: The VARIANT II Beta thalassemia Short Program is intended for the separation and area percent determinations of hemoglobin A2 and F, and as an aid in the identification of abnormal hemoglobins in human whole blood using ionexchange high-performance liquid chromatography (HPLC). The VARIANT II Beta-thalassemia Short Program is intended for use only with the Bio-Rad VARIANT II Hemoglobin Testing System. For in vitro diagnostic use. #### Total GHb: The VARIANT II Total GHb Program is intended for the separation and area percent determination of total glycated hemoglobin (GHb) in human whole blood using ion-exchange high-performance liquid chromatography (HPLC). The VARIANT II Total GHb Program with is intended for use only with the Bio-Rad VARIANT II Hemoglobin Testing System. For in vitro diagnostic use. #### Indications for Use: #### This software submission covers three assays and has two different Indications for Use: | HbA1c & GHb: | Measurement of percent hemoglobin A1c is effective in monitoring long-term glucose control in individuals with diabetes mellitus. | |---------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | HbA2 and HbF: | Measurement of the percent hemoglobin A2 and F are effective in screening of β-thalassemia (i.e. hereditary hemolytic anemias characterized by decreased synthesis of one or more types of abnormal hemoglobin polypeptide chains). | #### New Device Description The VARIANT II Hemoglobin Testing System uses the principles of high performance liquid chromatography (HPLC). The VARIANT II Hemoglobin A1c and Beta-thalassemia Short Program are based on chromatographic separation of hemoglobins on a cation exchange cartridge. The Total GHb Program is based on principles of boronate affinity high pressure liquid chromatography to separate glycated hemoglobin from non-glycated hemoglobin. Bio-Rad Laboratories, Inc. 4-A-2 {2}------------------------------------------------ The new feature in this submission is the upgrade in CDM software. The current software (3.5) requires Windows NT and Object Store N.T. These products are nearing the end of their lifecycle. CDM 4.0 software is needed to transfer the CDM software to Microsoft XP Operating System and Object Store version 4.0. #### Technical Characteristics Compared to Predicate The VARIANT™ II Hemoglobin A1c Program run on the VARIANT II Hemoglobin Testing System with CDM 4.0 has the same basic technical characteristics as the predicate VARIANT II Hemoglobin A1c Program (k) 984268. The technical characteristics between the two submissions are summarized in the following tables on pages 4-A-3 to 4-A-6: {3}------------------------------------------------ # VARIANT™ II Hemoglobin A1c (k) 984268 | Summary of Technological Characteristic Similarities in Comparison to Predicate Device | | | |----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Characteristics | New Device:<br>VARIANT II Hemoglobin A1c Program run<br>on the VARIANT II Hemoglobin Testing<br>System using CDM 4.0 | Predicate Device:(k)984268<br>VARIANT II Hemoglobin A1c Program run<br>on the VARIANT II Hemoglobin Testing<br>System using CDM 3.5 | | Intended Use(s) | The VARIANT II Hemoglobin A1c is intended<br>for the percent determination of hemoglobin<br>A1c in human whole blood using ion-exchange<br>high-performance liquid chromatography<br>(HPLC).<br>For In Vitro Diagnostic Use. | The VARIANT II Hemoglobin A1c is intended<br>for the percent determination of hemoglobin<br>A1c in human whole blood using ion-exchange<br>high-performance liquid chromatography<br>(HPLC).<br>For In Vitro Diagnostic Use. | | Indication(s) for Use | Measurement of the percent hemoglobin A1c is<br>effective in monitoring long-term glucose<br>control in individuals with diabetes mellitus. | Measurement of the percent hemoglobin A1c is<br>effective in monitoring long-term glucose<br>control in individuals with diabetes mellitus. | | Assay Principle | Cation exchange high performance liquid<br>chromatography | Cation exchange high performance liquid<br>chromatography | | CDM Software<br>version | 4.0 | 3.5 | | Microsoft software | Windows XP | Windows NT | | Object Store version | 6.0 | 4.0 | | Backup and Restore | Use Windows operation to write only data to<br>CD-R | Used Easy CD writer Read/Write | | Database<br>Management | Delete data directly from database | Substitute database with a blank database | | Standardization for<br>HbA1c | Traceable to the Diabetes Control and<br>Complications Trial (DCCT) reference method<br>and IFCC. Certified via the National<br>Glycohemoglobin Standardization Program<br>(NGSP) for HbA1c. | Traceable to the Diabetes Control and<br>Complications Trial (DCCT) reference method<br>and IFCC. Certified via the National<br>Glycohemoglobin Standardization Program<br>(NGSP) for HbA1c. | | Firmware upgrade<br>to VARIANT II<br>VCS and VSS | VCS 41.300 VSS 51.381 | VCS 41.295 VSS 51.373 | | Summary of Technological Characteristic Similarities in Comparison to Predicate Device | | | | Characteristics | New Device:<br>VARIANT II Beta thalassemia Short<br>Program run on the VARIANT II<br>Hemoglobin Testing System using CDM 4.0 | Predicate Device: (k)991127<br>VARIANT II Beta thalassemai Short<br>Program run on the VARIANT II<br>Hemoglobin Testing System using CDM 3.5 | | Intended Use(s) | The VARIANT II Beta thalassemia Short<br>Program is intended for the separation and area<br>percent determinations of hemoglobin A2 and<br>F, and as an aid in the identification of<br>abnormal hemoglobins in human whole blood<br>using ion-exchange high-performance liquid<br>chromatography (HPLC).<br>The VARIANT II B-thalassemia Short<br>Program is intended for use only with the Bio-<br>Rad VARIANT II Hemoglobin Testing<br>System.<br>For In Vitro Diagnostic Use. | The VARIANT II Beta thalassemia Short<br>Program is intended for the separation and area<br>percent determinations of hemoglobin A2 and F,<br>and as an aid in the identification of abnormal<br>hemoglobins in human whole blood using ion-<br>exchange high-performance liquid<br>chromatography (HPLC).<br>The VARIANT II B-thalassemia Short Program<br>is intended for use only with the Bio-Rad<br>VARIANT II Hemoglobin Testing System.<br>For In Vitro Diagnostic Use. | | Indication(s) for Use | Measurement of the percent hemoglobin A2 and<br>F are effective in screening of β-thalassemia<br>(i.e. hereditary hemolytic anemias<br>characterized by decreased synthesis of one or<br>more types of abnormal hemoglobins<br>polypeptide chains). | Measurement of the percent hemoglobin A2 and<br>F are effective in screening of β-thalassemia (i.e.<br>hereditary hemolytic anemias characterized by<br>decreased synthesis of one or more types of<br>abnormal hemoglobins polypeptide chains). | | Assay Principle | Cation exchange high performance liquid<br>chromatography | Cation exchange high performance liquid<br>chromatography | | CDM Software<br>version | 4.0 | 3.5 | | Microsoft software | Windows XP | Windows NT | | Object Store version | 6.0 | 4.0 | | Backup and Restore | Use Windows operation to write only data to<br>CD-R | Used Easy CD writer Read/Write | | Database<br>Management | Delete data directly from database | Substitute database with a blank database | | Firmware upgrade<br>to VARIANT II<br>VCS and VSS | VCS 41.300 VSS 51.381 | VCS 41.295 VSS 51.373 | | Summary of Technological Characteristic Similarities in Comparison to Predicate Device | | | | Characteristics | New Device:<br>VARIANT II Total GHb Program run on<br>the VARIANT II Hemoglobin Testing<br>System using CDM 4.0 | Predicate Device: (k) 003280<br>VARIANT II Total GHb Program run on the<br>VARIANT II Hemoglobin Testing System<br>using CDM 3.5 | | Intended Use(s) | The VARIANT II Total GHb Program is<br>intended for the separation and area percent<br>determination of total glycated hemoglobin<br>(GHb) in whole blood using boronate affinity<br>high-performance liquid chromatography<br>(HPLC).<br><br>For In Vitro Diagnostic Use. | The VARIANT II Total GHb Program is<br>intended for the separation and area percent<br>determination of total glycated hemoglobin<br>(GHb) in whole blood using boronate affinity<br>high-performance liquid chromatography<br>(HPLC).<br><br>For In Vitro Diagnostic Use. | | Indication(s) for Use | Measurement of the percent hemoglobin A1c is<br>effective in monitoring long-term glucose<br>control in individuals with diabetes mellitus. | Measurement of the percent hemoglobin A1c is<br>effective in monitoring long-term glucose<br>control in individuals with diabetes mellitus. | | Assay Principle | Affinity high performance liquid<br>chromatography | Affinity high performance liquid<br>chromatography | | CDM Software<br>version | 4.0 | 3.5 | | Microsoft software | Windows XP | Windows NT | | Object Store version | 6.0 | 4.0 | | Backup and Restore | Use Windows operation to write only data to<br>CD-R | Used Easy CD writer Read/Write | | Database<br>Management | Delete data directly from database | Substitute database with a blank database | | Standardization | Traceable to the Diabetes Control and<br>Complications Trial (DCCT) reference method<br>and IFCC. Certified via the National<br>Glycohemoglobin Standardization Program<br>(NGSP) for HbA1c. | Traceable to the Diabetes Control and<br>Complications Trial (DCCT) reference method<br>and IFCC. Certified via the National<br>Glycohemoglobin Standardization Program<br>(NGSP) for HbA1c. | | Firmware upgrade<br>to VARIANT II<br>VCS and VSS | VCS 41.300 VSS 51.381 | VCS 41.295 VSS 51.373 | {4}------------------------------------------------ # VARIANTIM II Beta thalassemia Short Program (k)991127 {5}------------------------------------------------ ## VARIANT™ II GHb (k)003280 {6}------------------------------------------------ # Testing To Establish Substantial Equivalence: ## Accuracy: ## VARIANT™ II Hemoglobin HbA1c Program Method correlation between Bio-Rad VARIANT II Hemoglobin A1c Program with CDM 4.0 and V ARIANT II Hemoglobin A1c Program with CDM 3.5 was evaluated using 40 EDTA whole blood samples ranging from (4.7% -12.9%) HbA1c. The results are presented in the following table: | VARIANT II Hemoglobin A1c Program Correlation | | | | | |-----------------------------------------------|----|--------|--------|-----------| | Regression Method | n | r2 | Slope | Intercept | | Least Squares | 40 | 0.9978 | 1.0119 | 0.0002 | # VADIANT II Hamaclahin A1a Buaguam Carrelatio ## Precision: ## VARIANT II Hemoglobin A1c Program The following precision table provides comparison data on the precision between VARIANT II Hemoglobin A1c Program with CDM 4.0 vs. CDM 3.5 each utilizing EDTA whole blood patient samples. Method precision was performed using a protocol based on the NCCLS Evaluation protocol, EP5-A for the VARIANT II Hemoglobin A1c Program with CDM 4.0 and 3.5. The protocols for both VARIANT II Hemoglobin A1c Program with CDM 4.0 and 3.5 are similar. In each duplicate daily run for both verification studies, duplicate aliquots of normal HbA1c and diabetic HbA1c patient samples and controls were each analyzed per run. The position of the precision specimens in each run was randomized to simulate normal laboratory conditions. The precision data for the VARIANT II with CDM 3.5 was over 20 working days while the data for VARIANT II with CDM 4.0 was over 10 working days. Although precision samples are different, since they were run at different time periods, the precision results between the VARIANT II Hemoglobin A1c Program with CDM 4.0 and CDM 3.5 are equivalent. A summary of combined comparative precision results are presented in the following precision table. ### VARIANT II HbA1c with CDM 4.0 vs. VARIANT II HbA1c with CDM 3.5 - Precision | | VII HbA1c with CDM 4.0 | | VII HbA1c with CDM 3.5T | | |------------------------|------------------------|-----------------|-------------------------|-----------------| | | Normal Sample | Diabetic Sample | Normal Sample | Diabetic Sample | | n= (number of samples) | 40 | 40 | 80 | 80 | | Mean (%HbA1c) | 5.4 | 10.4 | 5.4 | 13.7 | | Within run (%CV) | 1.1 | 1.2 | 1.5 | 0.7 | | Total Precision (%CV) | 2.6 | 2.7 | 2.1 | 1.7 | {7}------------------------------------------------ ## Accuracy: ## VARIANT™ II Beta thalassemia Short Program Method correlation between Bio-Rad VARIANT II Beta thalassemia Short with CDM 4.0 and V ARIANT II Beta thalassemia Short with CDM 3.5 was evaluated using 40 EDTA whole blood samples ranging from (2.0 - 7.0% %) HbA2 and (0.2 - 11.1%) HbF. The results are presented in the following tables: | VARIANT II Beta thalassemia Short Correlation (HbA2) | | | | |------------------------------------------------------|--|--|--| |------------------------------------------------------|--|--|--| | Regression Method | n | r2 | Slope | Intercept | |-------------------|----|--------|--------|-----------| | Least Squares | 40 | 0.9924 | 1.0070 | -0.0034 | #### VARIANT II Beta thalassemia Short Correlation (HbF) | Regression Method | n | r² | Slope | Intercept | |-------------------|----|--------|--------|-----------| | Least Squares | 40 | 0.9991 | 0.9806 | 0.0192 | #### Precision: ## VARIANT II Beta thalassemia Short Program The following precision table provides comparison data on the precision between VARIANT II Beta thalassemia Short with CDM 4.0 vs. CDM 3.5 each utilizing EDTA whole blood patient samples. Method precision was performed using a protocol based on the NCCLS Evaluation protocol, EP5-A for the VARIANT II Beta thalassemia Short with CDM 4.0 and 3.5. The protocols for both VARIANT II Beta thalassemia Short with CDM 4.0 and 3.5 are similar. In each duplicate daily run for both verification studies, duplicate aliquots of normal HbA1c and diabetic HbA1c patient samples and controls were each analyzed per run. The position of the precision specimens in each run was randomized to simulate normal laboratory conditions. The precision data for the VARIANT II with CDM 3.5 was over 20 working days while the data for VARIANT II with CDM 4.0 was over 10 working days. Although precision samples are different, since they were run at different time periods, the precision results between the VARIANT II Beta thalassemia Short with CDM 4.0 and CDM 3.5 are equivalent. A summary of combined comparative precision results are presented in the following precision tables. #### VARIANT II Beta thalassemia Short with CDM 4.0 vs. VARIANT II Beta thalassemia Short with CDM 3.5 - Precision | plant of the country of the couple in the program and<br>.<br>.<br>Company of Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children C<br><br>сения, непанскием пининания делагания и<br>a property of the superior compression in compression in a first a<br>на внес книжен с седавенность в «Массии «спорт «с «России»<br>. I a price a may a control in in interest a spendent a spendent a control a partify of the final<br>below and a be a mater of all a first find a for a promote controlled<br><br>само свъ по предваря в ки в полни и тимания свенни<br>manager and the many of the first of the Research and the many commended to a<br>с про компонинаниями когда и вест подательные в создания и в подать<br>photostanting a fine a status a some a management and management<br>CALLERS DIE LE COULT MELIAN (PLANS (A (C) (A) (C) (ANGC 335)<br> | .<br>VII Beta thalassemia Short<br>седания продавания и в полности и въ композительно в истории и в поли-<br>THERE THE INSTITUTE ADDRESS OF ANY OF THE THE WARRET HERE<br><br>with CDM 4<br> | <br><br> | VII Beta thalacean<br>.<br><br>дада перва палин и как и базан веча<br> | with CDM 3 5 | |------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|-------------------------------------------------------| | НА С П. К С С Р С Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р Р<br>control concerners of the count of the contribution come of<br>to as cannon house in come in country a society and<br>ста поставляется в составляется в подредния в подательным в по-<br>HIBHLE HE HONNEY . And<br>A P<br>(Max. AN 12 12) - Intellection of minim suppression a market the<br>11 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1<br>Charles, No 2 parat personal program<br>серия подата и продательного по подательного подарения<br>DIE FREE BEACH (ARACS 1853) (PANETHER S.S. (1932) (43) (43) (41) (2000) (1) (2000) (1)<br>саль мы выбарь с при с депиратив прини понинальн<br><br>дет подредерения в дел в полническог вечения и в вестика и в оби-<br>**** State A. A.S. and Cases of Cases of C. PAP<br>. | <br><br>.<br>Normal A2<br>.<br>species in a mini shalling add to provi in the<br>.<br>sample sample | <br>report and a status that and proposition and any<br>State (Section the School (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1)<br>the fine a sence as as an affer as<br><br>Call and Children and Children and Charles and<br><br> | .<br><br>Sampl | · CD (1935) 1995 (1992) 1999 (1992)<br>Chick Nob Affa | | | | | | | | n= {number of sampl | | | | | | Mean (%HbA2) | | | | | | Within run (%CV) | | | | | Bio-Rad Laboratories, Inc. {8}------------------------------------------------ | .<br><br>Carl Beta thalassemia Short<br>ден призначальное собразования по стерия помость (120 г.<br>" Collection('', Was (Colleges "Sidel ( + ) = ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) (<br>production and contraction of the manage of the complete of<br>画出版社会が発生した。 100with CDM 4.0 1 10<br>minted and the production in the program a many of the comments<br>a manager see a comments of the comments of the security of | | | VII Beta thalassemia Short<br>Bearing with CDM 3.5 The | | |----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|-------------------------------------------------------------------------|--------------------------------------------------------|---------| | managing same more of more of manufacturer and the superior<br>management of the cases and 1980 to the Market is<br>comprehensive in the modifies the reason of the contraction of<br>productions within in problem in the program and<br>and and are and some and some province and the season<br>с редини серед продоблава для сетиранно<br>and in as a promotive and considered to the comparis<br>the receive magnet canding the count of the concess on "hometer the more of<br>while a management and an any are a strange compression<br>у з 1,800 рабіцца 2,200 радару Мідарамары. Радаем «Ада ід 1,7<br>a, min ra ragangenji sem y qui turnji dagajidi xabar 2004 10:30 PM P<br>production of the list and consistence the research | Low HbF sample<br>the country and concession in the production of the many of | Low HbF<br>Canadian Caracterial Career of Children<br><br>sample sample | Low HbF<br>samnle min | LOW HOH | | n= (number of samples | | | | | | Mean (%HbF) | | 0.0 | | | | Within run (%CV) | | | | | | Total Precision (%CV) | | | | | #### VARIANT II Beta thalassemia Short with CDM 4.0 vs. VARIANT II HbA1c with CDM 3.5 - Precision (continue) ### Accuracy: ## VARIANT™ II Total GHb Program Method correlation between Bio-Rad VARIANT II Total GHb Program with CDM 4.0 and VARIANT II Total GHb Program with CDM 3.5 was evaluated using 40 EDTA whole blood samples ranging from (4.5 - 16.2%) GHb. The results are presented in the following table: | 1 | | | VARIANT II Total GHb Program Correlation | | |---|--|--|------------------------------------------|--| |---|--|--|------------------------------------------|--| | Regression Method | n | r2 | Slope | Intercept | |-------------------|----|--------|--------|-----------| | Least Squares | 40 | 0.9991 | 1.0054 | 0.042 | ## Precision: ### VARIANT II Total GHb Program The following precision table provides comparison data on the precision between VARIANT II Total GHb Program with CDM 4.0 vs. CDM 3.5 each utilizing EDTA whole blood patient samples. Method precision was performed using a protocol based on the NCCLS Evaluation protocol. EP5-A for the VARIANT II Total GHb Program with CDM 4.0 and 3.5. The protocol poth V ARIANT II Total GHb Program with CDM 4.0 and 3.5 are similar. In each duplicate daily run for both verification studies, duplicate aliquots of normal HbA lc and diabetic HbA lc patient samples and controls were each analyzed per run. The position of the precision specialisms in each run was randomized to simulate normal laboratory conditions. The precision data for the V ARIANT II with CDM 3.5 was over 20 working days while the data for VARIANT II with CDM 4.0 was over 10 working days. {9}------------------------------------------------ Although precision samples are different, since they were run at different time periods, the precision results between the VARIANT II Total GHb Program with CDM 4.0 and CDM 3.5 are equivalent. A summary of combined comparative precision results are presented in the following precision table. | an spessadia alle comments of the season VII Total (GHI<br>del Chinners provident and considered to the finition of the Researching<br>consistence in the children the school in the biological in the com-<br>· · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·<br>comments of considered to the submit model group of province in consideration of the comments<br>S State of the Castle with CDM 4.0 Twith CDM 4.0 The<br>· · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·<br>. | a madeling manager and the material comments of the may<br>да се поведения принасти с мах пательные компания в составляет<br>. | . | 100 1131 11 11 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1<br>and and a second station and the many comments of the first<br>CARDE STATE PART POST POST PART PART AND AND ANNO BERRENDER AND AND<br>In annually be have the state and successful and the programment<br>A for a comprehensive the children the management<br>with CDM 3. | | |---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--| | and and and a formation of the commend of the comments of the company of<br>.<br>and the county and can and a senses continue consistent was and<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>CONSULT BE INNOVEMBER 1997 The Sell Children and Children Come<br>and a parties an an an an angle and the<br>I was and service and success and security and any<br>CANTAL PART CONTRACTORIO COLLECTION COLLEGION CONTRACTORS COLLEGION<br>стания полния стр. «Рессия создавания подательн<br>ниям кинов в праведениями вини<br>. | ANARIAAWARANAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA<br><br>be<br>Normal | Dishetic Samnle | Normal Samnle | | | n= (number of samples | | | | | | Mean (%GHb) | | | | | | Within run (%CV) | | | | | | Total Precision ' | | | | | VARIANT II Total GHb with CDM 4.0 vs. VARIANT II Total GHb with CDM 3.5 -Precision ## Conclusion: The similarities of the intended use and the general performance characteristics and results of the newly described and evaluated VARIANT II Hemoglobin A1c , VARIANT II Betathalassemia and VARIANT II Total GHb Programs run on the VARIANT II Hemoglobin Testing System with CDM 4.0 are nearly identical to or logical extensions of those for cleared predicate program systems [i.e., VARIANT II Hemoglobin A1c, VARIANT II Beta-thalassemia and VARIANT II Total GHb run on CDM 3.5]. Thus, one may conclude, based on the use of the same HPLC technology, and the nearly equivalent results obtained for the correlation and precision versus the corresponding results obtained with the predicate system that the new VARIANT II Hemoglobin A1c, VARIANT II Beta-thalassemia and VARIANT II Total GHb Program runs on the VARIANT II Hemoglobin Testing System with CDM 4.0 is substantially equivalent to the cleared and currently marketed predicate system. {10}------------------------------------------------ Image /page/10/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. The eagle is black and white, and the text is in a simple, sans-serif font. The logo is simple and recognizable. Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Jackie Buckley Bio-Rad Laboratories, Inc. Clinical System Division 4000 Alfred Nobel Drive Hercules, California 94547 DEC 2 7 2006 Re: k063643 > Trade/Device Name: Variant II TM Hemoglobin A1c Program, Variant™ Betathalassemia and Variant ™ II Total Ghb run on the Variant ™ II Hemoglobin Testing System with CDM 4.0 Regulation Number: 21 CFR 864.7400, 21 CFR 864.7470 Regulation Name: Glycosylated hemoglobin assay; Hemoglobin A2 assay Regulatory Class: Class II Product Code: LCP, JPD Dated: December 7, 2006 Received: December 8, 2006 Dear Ms. Buckley: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a sub…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...