ACE HEMOGLOBIN A1C REAGENT, MODEL ACI-21, HEMOGLOBIN A1C CALIBRATORS, MODEL S2-72, HEMOGLOBIN A1C CONTROLS, MODEL C2-72
Applicant
Alfa Wassermann, Inc.
Product Code
LCP · Hematology
Decision Date
Apr 17, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.7470
Device Class
Class 2
Intended Use
ACE® Hemoglobin A1c (HbA1c) Reagent is intended for the quantitative determination of hemoglobin A1c (μmol/L) and total hemoglobin (g/dL) in human whole blood (EDTA) for the calculation of percent hemoglobin A1c using the ACE clinical chemistry system. This test is intended for use in clinical laboratories or physician office laboratories to monitor long term blood glucose control in individuals with diabetes mellitus. For in vitro diagnostic use only. The μmol HbA1c and total hemoglobin (THb) values generated are intended for use in the calculation of the HbA1c / THb ratio, and cannot be used individually for diagnostic purposes. Hemoglobin A1c Calibrators are intended for use in the performance of both a multi-point calibration for hemoglobin A1c and a single-point calibration of total hemoglobin on the ACE® clinical chemistry system, for the quantitative determination of percent (%) hemoglobin A1c in whole blood. For in vitro diagnostic use only. Hemoglobin A1c Controls are intended to reliably monitor the accuracy and precision of quantitative determinations of hemoglobin A1c on the ACE® clinical chemistry system. For in vitro diagnostic use only.
Device Story
The ACE Hemoglobin A1c system is an in vitro diagnostic assay for clinical chemistry analyzers. It processes human EDTA whole blood samples. The process involves a pretreatment step where red blood cells are lysed and hemoglobin chains hydrolyzed. HbA1c is measured via a latex agglutination inhibition assay; the presence of HbA1c in the sample competes with agglutinator reagents for antibody binding sites, inhibiting agglutination. Absorbance is monitored monochromatically at 592 nm; the increase in absorbance is inversely proportional to HbA1c concentration. Total hemoglobin is measured bichromatically at 573 nm/692 nm after conversion to alkaline hematin, where color intensity is directly proportional to concentration. The system calculates the HbA1c/THb ratio to report percent HbA1c. Used in clinical labs or physician offices by laboratory personnel to monitor long-term glucose control in diabetic patients. Results assist healthcare providers in assessing glycemic management.
Clinical Evidence
Bench testing only. Precision studies (CLSI EP5-A2) showed within-run CVs of 2.9-3.0% and total CVs of 5.9-7.0%. Physician Office Laboratory (POL) studies showed within-run and total CVs <5.4%. Accuracy was established via method comparison (CLSI EP9-T) against Tosoh Bioscience A1c 2.2 Plus HPLC, yielding correlation coefficients (r) of 0.975 to 0.984 across multiple sites.
Technological Characteristics
Reagent kit includes hemoglobin denaturant, total hemoglobin reagent, HbA1c agglutinator, and HbA1c antibody reagent. Measurement via latex agglutination inhibition (HbA1c) and alkaline hematin (total Hb). Monochromatic (592 nm) and bichromatic (573/692 nm) optical detection. Calibrators are liquid; controls are lyophilized human whole blood. System is an automated clinical chemistry analyzer.
Indications for Use
Indicated for the quantitative determination of HbA1c and total hemoglobin in human EDTA whole blood to monitor long-term blood glucose control in individuals with diabetes mellitus. For use in clinical laboratories or physician office laboratories.
Regulatory Classification
Identification
A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.
Predicate Devices
Tosoh Bioscience, Inc. A1c 2.2 Plus Automated Glycohemoglobin Assay
General Atomics, Inc. Hemoglobin A1c Enzymatic Assay
K081895 — ADVIA CHEMISTRY HEMOGLOBIN A1C, ADVIA CHEMISTRY A1C CALIBRATORS · Siemens Medical Solutions Diagnostics · Jan 8, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text "ALFA WASSERMANN" in all caps. The text is in a bold, sans-serif font. The letters are black and the background is white.
K063306
#### 510(k) SUMMARY
# Alfa Wassermann, Inc. ACE® Hemoglobin A1c Reagent, Hemoglobin A1c Calibrators and Hemoglobin A1c Controls
#### Submitter
Alfa Wassermann, Inc. 4 Henderson Drive West Caldwell, NJ 07006 Phone: 973-852-0120 Facsimile: 973-852-0121
Contact Person
Murray Rosenthal, Ph.D.
Date Prepared
October 31, 2006
### Product Trade and Common Name
ACE® Hemoglobin A1c (HbA1c) Reagent Hemoglobin A1c Calibrators Hemoglobin Alc Controls
#### Classification Name
Class II, Product Code LCP CFR § 864.7470: Glycosylated Hemoglobin Assay
Class II, Product Code JIT CFR § 862.1150: Calibrator, Secondary
Class I, Product Code JJX CFR § 862.1660: Quality Control Material (assayed and unassayed)
### Predicate Devices
Tosoh Bioscience, Inc. A1c 2.2 Plus Automated Glycohemoglobin Assay
General Atomics, Inc. Hemoglobin A1c Enzymatic Assay
ALFA WASSERMANN, INC. Henderson Drive st Caldwell, NJ 07006 3) 882-8630 (973) 227-6700 faWassermannUS.com APR 1 7 2007
{1}------------------------------------------------
#### Description and Intended Use of the Device
ACE® Hemoglobin A1c (HbA1c) Reagent is intended for the quantitative determination of hemoglobin A1c (umol/L) and total hemoglobin (g/dL) in human EDTA whole blood for the calculation of percent hemoglobin A1c using the ACE clinical chemistry system. This test is intended for use in clinical laboratories or physician office laboratories to monitor long term blood glucose control in individuals with diabetes mellitus. For in vitro diagnostic use only.
The umol HbAlc and total hemoglobin (THb) values generated are intended for use in the calculation of the HbA1c/THb ratio and cannot be used individually for diagnostic purposes.
Hemoglobin A1c Calibrators are intended for use in the performance of both a multi-point calibration for hemoglobin A1c and a single-point calibration of total hemoglobin on the ACE clinical chemistry system for the quantitative determination of percent (%) hemoglobin A1c in whole blood. For in vitro diagnostic use only.
Hemoglobin A1c Controls are intended to reliably monitor the accuracy and precision of quantitative determinations of hemoglobin A1c on the ACE clinical chemistry system. For in vitro diagnostic use only.
#### Technological Characteristics
The ACE Hemoglobin A1c Reagent is provided as a single kit and consists of four bottles containing a hemoglobin denaturant, a total hemoglobin reagent, a HbAlc agglutinator reagent and a HbA1c antibody reagent.
The ACE Hemoglobin Alc Calibrators are provided as a single kit and contain ready-to-use liquid calibrators, one of each of six levels.
The ACE Hemoglobin A1c Controls are provided as a single kit and contain lyophilized controls with normal and elevated levels of HbA1c and a reconstitution fluid. The controls are prepared from human whole blood, which has been tested and found negative for antibody to Human Immunodeficiency Virus (anti-HIV) Types 1 and 2, antibody to Hepatitis C (anti-HCV) and for Hepatitis B Surface Antigen (HBsAg) by FDA recommended (approved/licensed) tests.
#### Principles of Operation
Prior to assay, whole blood samples require a pretreatment step. The red blood cells in the sample are lysed by the Hemoglobin Denaturant and the hemoglobin chain hydrolyzed.
{2}------------------------------------------------
For determination of HbAlc, a latex agglutination inhibition assay is used. In the absence of HbA 1c in the sample, the agglutinator (synthetic polymer containing the immunoreactive portion of HbA1c) in the HbAIc Agglutinator Reagent and the antibody-coated microparticles in the HbAIc Antibody Reagent will agglutinate. The presence of HbA1c in the sample competes for the antibody binding sites and inhibits agglutination. The increase in absorbance, monitored monochromatically at 592 nm on the ACE clinical chemistry system, is inversely proportional to the HbA1c present in the sample.
For the determination of total hemoglobin, all hemoglobin derivatives in the sample are converted to alkaline hematin. The reaction produces a green colored solution, which is measured bichromatically at 573 nm/692 nm on the ACE clinical chemistry system. The intensity of color produced is directly proportional to the total hemoglobin concentration in the sample.
Both the concentration of HbA1c and total hemoglobin are measured; then the ratio is calculated and the result reported as percent HbA1c.
The ACE Hemoglobin A1c Calibrators are used to calibrate the ACE clinical chemistry system for both the Hemoglobin HbA1c and Total Hemoglobin assays.
The ACE Hemoglobin A 1c Controls are run in the same way as patient samples.
### Performance Data
The performance of the ACE Hemoglobin A 1 c kit was evaluated for precision, measuring range and accuracy.
The precision studies were done according to CLSI/NCCLS EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline - Second Edition. Using this protocol, precision was measured on the ACE clinical chemistry system using the ACE Hemoglobin A1c reagents, calibrated with the ACE Hemoglobin A1c Calibrators on a normal and elevated patient. The within-run CV for the normal patient was 2.9% and for the elevated patient was 3.0%. The total CV for the normal patient was 7.0% and for the elevated patient was 5.9%.
Further precision studies were done in three separate Physician Office Laboratory (POL) sites. These studies were conducted by personnel without formal medical technology education. The studies consisted of running three EDTA whole blood samples with varying levels of HbA1c in triplicate on five different days. The within-run and total CVs were all less than 5.4%.
The measuring range for total hemoglobin was determined using a whole blood sample, by serial dilution of the packed cells with plasma from the blood sample, HbAlc levels were measured on the ACE clinical chemistry system using the
{3}------------------------------------------------
ACE Hemoglobin Alc reagents, calibrated with the ACE Hemoglobin A1c Calibrators. The acceptable range was determined by using 95% confidence intervals from linear regression analysis with ±10% limits. The acceptable range for total hemoglobin was determined to be 10-21 g/dL.
A correlation study, to determine the accuracy of the ACE Hemoglobin A1c kit. calibrated with the ACE Hemoglobin A1c calibrators, was done against the Tosoh Bioscience's A1c 2.2 Plus Automated Glycohemoglobin Assay. The study followed the protocol in CLSI/NCCLS EP9-T, Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline - Second Edition. Linear regression analysis was performed: ACE clinical chemistry system (y) and Tosoh HPLC (x). The analysis yielded the following results:
| Regression<br>Equation | Correlation<br>Coefficient | Standard Error of<br>the Estimate | Confidence<br>Interval Slope | Confidence<br>Interval<br>Intercept |
|------------------------|----------------------------|-----------------------------------|------------------------------|-------------------------------------|
| y = 1.052x+0.23 | 0.975 | 0.49 | 1.000 to 1.104 | -0.20 to 0.67 |
Further accuracy studies were done in three separate POL labs. These studies were conducted by personnel without formal medical technology education. At each lab, these studies consisted of running the same 20 EDTA whole blood samples with varying levels of HbA1c (determined using the Tosoh Bioscience's A 1 c 2.2 Plus Automated Glycohemoglobin Assay) in duplicate in four different runs of ten samples each. Linear regression analysis was performed: ACE clinical chemistry system (y) and Tosoh HPLC (x). The analysis yielded the following results:
| Lab | Regression<br>Equation | Correlation<br>Coefficient | Standard Error<br>of the<br>Estimate | Confidence<br>Interval Slope | Confidence<br>Interval<br>Intercept |
|-----|------------------------|----------------------------|--------------------------------------|------------------------------|-------------------------------------|
| A | $y = 0.940x+0.41$ | 0.9762 | 0.46 | 0.871 to 1.009 | -0.16 to 0.99 |
| B | $y = 1.102x-0.04$ | 0.9818 | 0.47 | 1.032 to 1.172 | -0.63 to 0.55 |
| C | $y = 1.035x+0.38$ | 0.9844 | 0.41 | 0.974 to 1.095 | -0.89 to 0.13 |
In all instances, the ACE Hemoglobin A1c kit run on the ACE clinical chemistry analyzer functioned as intended.
{4}------------------------------------------------
## Substantial Equivalence
The evaluations and comparative analysis indicated substantial equivalence between the results obtained for the ACE Hemoglobin A1c Reagent, ACE Hemoglobin Alc Calibrators and ACE Hemoglobin A1c Controls and the predicate devices (Tosoh Bioscience's A1c 2.2 Plus Automated Glycohemoglobin Assay and General Atomics Enzymatic HbA1c Assay). The ACE Hemoglobin Alc Reagent, ACE Hemoglobin A1c Calibrators and ACE Hemoglobin A1c Controls have the same intended use and similar indications, technological characteristics, performance characteristics, and similar principles of operation as its predicate devices. Therefore, the ACE Hemoglobin A1c Reagent, ACE Hemoglobin A1c Calibrators and ACE Hemoglobin A1c Controls are substantially equivalent to commercially available products that measure HbA1c in whole blood samples.
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The seal is black and white and appears to be a simple graphic representation.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Alfa Wassermann Diagnostic Technologies, LLC c/o Dr. Murray Rosenthal Manager, Reagent Technology 4 Henderson Drive West Caldwell, NJ 07006
APR 1 7 2007
Re: k063306
Trade/Device Name: ACE® Hemoglobin A1c (HbA1c) Reagent Hemoglobin Alc Calibrators Hemoglobin Alc Controls
Regulation Number: 21 CFR 864.7470 Regulation Name: Glycosylated Hemoglobin assay. Regulatory Class: Class II Product Code: LCP, JIT, JJX Dated: March 2, 2007 Received: March 5, 2007
Dear Dr. Rosenthal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{6}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tollil the (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Jean M. Cooper, M.S., D.V.M.
Yean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### Section 4
#### Section 1.01 Indications for Use Statement
510(k) Number (if known): k063306
Device Name:
ACE® Hemoglobin A1c (HbA1c) Reagent Hemoglobin A1c Calibrators Hemoglobin Alc Controls
Indications for Use:
ACE® Hemoglobin A1c (HbA1c) Reagent is intended for the quantitative determination of hemoglobin Alc (umol/L) and total hemoglobin (g/dL) in human EDTA whole blood for the calculation of percent hemoglobin A 1 c using the ACE clinical chemistry system. This test is intended for use in clinical laboratories or physician office laboratories to monitor long term blood glucose control in individuals with diabetes mellitus. For in vitro diagnostic use only.
The umol HbA1c and total hemoglobin (THb) values generated are intended for use in the calculation of the HbA1c/THb ratio and cannot be used individually for diagnostic purposes.
Hemoglobin A1c Calibrators are intended for use in the performance of both a multi-point calibration for hemoglobin A1c and a single-point calibration of total hemoglobin on the ACE® clinical chemistry system, for the quantitative determination of percent (%) hemoglobin A1c in whole blood. For in vitro diagnostic use only.
Hemoglobin Alc Controls are intended to reliably monitor the accuracy and precision of quantitative determinations of hemoglobin A1c on the ACE® clinical chemistry system. For in vitro diagnostic use only,
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (OD
Page of
Carol C. Benson
Division Sign-Off
Division Sign-
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K063306
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.