UNIVERSAL LOCKING SYSTEM, 2.7MM PLATES & SCREWS

K063303 · Zimmer, Inc. · HRS · Nov 22, 2006 · Orthopedic

Device Facts

Record IDK063303
Device NameUNIVERSAL LOCKING SYSTEM, 2.7MM PLATES & SCREWS
ApplicantZimmer, Inc.
Product CodeHRS · Orthopedic
Decision DateNov 22, 2006
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Universal Locking System is indicated for temporary internal fixation and stabilization of osteotomies and fractures, including comminuted fractures, supracondylar fractures, extra-articular fractures, fractures in osteopenic bone, nonunions and malunions.

Device Story

Zimmer Universal Locking System is a metallic bone fixation plate and screw system. Plates feature figure-8 shaped holes to accommodate either standard or locking screws. Configurations include dual compression, reconstruction, tubular, straight T, straight L, and oblique L. Used by surgeons for internal fracture fixation and stabilization of osteotomies. Device provides mechanical stability to bone segments during healing process. No electronic or software components.

Clinical Evidence

No clinical data provided. Substantial equivalence supported by non-clinical laboratory performance testing.

Technological Characteristics

Metallic bone fixation plates and screws. Plate configurations: dual compression, reconstruction, tubular, straight T, straight L, oblique L. Features figure-8 shaped holes for standard or locking screw compatibility. No software, energy source, or connectivity.

Indications for Use

Indicated for temporary internal fixation and stabilization of osteotomies and fractures, including comminuted, supracondylar, and extra-articular fractures, fractures in osteopenic bone, nonunions, and malunions.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the Zimmer logo. The logo consists of a large, bold letter 'Z' enclosed in a circle, followed by the word 'zimmer' in a smaller, sans-serif font. The 'Z' in the circle is the most prominent element, drawing the eye with its size and enclosure. ## Summary of Safety and Effectiveness # K063303 NOV 2 2 2006 | Submitter: | Zimmer, Inc.<br>P.O. Box 708<br>Warsaw, IN 46581-0708 | |---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Anthony Francalancia<br>Senior Associate, Regulatory Affairs<br>Telephone: (574) 372-4570<br>Fax: (574) 372-4605 | | Date: | October 31, 2006 | | Trade Name: | Zimmer® Universal Locking System | | Common Name: | 2.7mm Locking Plate System | | Classification Name<br>and Reference: | Plate, Fixation, Bone (21 CFR § 888.3030)<br>Screw, Fixation, Bone (21 CFR § 888.3040) | | Predicate Device: | Zimmer® Universal Locking System 3.5mm Plates<br>and Screws, manufactured by Zimmer, Inc.,<br>K060710, cleared April 26, 2006. | | Device Description: | The Zimmer Universal Locking System is a plate<br>and screw system intended for internal fracture<br>fixation. The plate selection consists of dual<br>compression, reconstruction, tubular, straight "T",<br>straight "L" and oblique "L" configurations. Plates<br>accommodate either standard or locking screws via<br>figure-8 shaped holes. | | Intended Use: | The Universal Locking System is indicated for<br>temporary internal fixation and stabilization of<br>osteotomies and fractures, including comminuted<br>fractures, supracondylar fractures, extra-articular<br>fractures, fractures in osteopenic bone, nonunions<br>and malunions. | | Comparison to Predicate Device: | The Zimmer Universal Locking System 2.7mm<br>plates and screws have the same intended use,<br>similar performance characteristics, are | {2}------------------------------------------------ Image /page/1/Picture/0 description: The image contains the logo for Zimmer Biomet. The logo consists of a stylized letter "Z" enclosed in a circle, followed by the word "zimmer" in a bold, sans-serif font. The logo is black and white. manufactured from the same materials and are similar in design to the predicate device. Performance Data (Nonclinical and/or Clinical): Non-Clinical Performance and Conclusions: The results of non-clinical (laboratory) performance testing demonstrate that the device is safe and effective. Clinical Performance and Conclusions: Clinical data and conclusions were not needed for this device. {3}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure, with three overlapping profiles facing to the right. The profiles are rendered in black and are connected by flowing lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Zimmer. Inc. % Mr. Anthony Francalancia Senior Associate, Regulatory Affairs P.O. Box 708 Warsaw, Indiana 46581 NOV 2 2 2006 Re: K063303 Trade/Device Name: Universal Locking Plate System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Codc: HRS Dated: October 31, 2006 Received: November 1, 2006 Dear Mr. Francalancia: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ Page 2 – Mr. Anthony Francalancia This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0210 Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Barbara Buechner Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use ## 510(k) Number (if known): Device Name: Universal Locking System, 2.7mm Locking Plates and Screws Zimmer Universal Locking Plate System ### Indications for Use: The Universal Locking Plate System is indicated for temporary internal fixation and stabilization of osteotomies and fractures, including; - Comminuted fractures . - Supracondylar fractures . - Extra-articular fractures . - Fractures in osteopenic bone . - Nonunions . - Malunions . Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (Please do not write below this line - Continue on another page if needed) Concurrence of CDRH, Office of Device Evaluation (ODE) Starbare bruchn (Division Sign-Cif) Division of General, Restorative, and Neurological Devices **510(k) Number** K063303 Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...