K062619 · CAO Group, Inc. · GEX · Nov 14, 2006 · General, Plastic Surgery
Device Facts
Record ID
K062619
Device Name
DENLASER 800 PLUS
Applicant
CAO Group, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Nov 14, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DenLaser 800 Plus is indicated for the procedures of removal of lesions, excision, incision, vaporization, ablation, hemostasis, and photocoagulation on soft tissue in the fields of otolaryngology (ear, nose, and throat), dentistry and oral surgery, arthroscopy, gastroenterology, dermatology, podiatry, plastic surgery, urology, gynecology, and plastic surgery.
Device Story
DenLaser 800 Plus is a surgical diode laser system. Input: electrical power; Output: 810nm laser energy (max 5W). Principle: solid-state diodes generate laser energy transmitted via proprietary optical fiber to a metal handpiece with disposable single-use tips. Operation: foot-actuated switch; user-definable settings include continuous or pulse delivery modes and a 630nm aiming beam. Used in clinical settings by healthcare professionals for soft tissue surgical procedures. Laser energy enables precise cutting, ablation, and hemostasis. Benefits include controlled tissue interaction and reduced collateral damage compared to traditional methods.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and performance standards compliance.
Technological Characteristics
Solid-state diode laser; 810 ± 20nm wavelength; max 5W output. Features 630nm aiming beam, fiber delivery system, and foot-actuated switch. Complies with 21 CFR 1040.10/1040.11, IEC 60601-1:1998+A1, IEC 60601-2-22:1995, IEC 60825-1:1993+A1+A2, 47 CFR 15/18, and ETSI 301-489-1. Handpiece uses disposable single-use tips.
Indications for Use
Indicated for soft tissue surgical procedures including removal of lesions, excision, incision, vaporization, ablation, hemostasis, and photocoagulation across various medical specialties including otolaryngology, dentistry, oral surgery, arthroscopy, gastroenterology, dermatology, podiatry, plastic surgery, urology, and gynecology.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
### 510(k) Summary of Safety and Effectiveness
CAO Group, Inc. 4628 West Skyhawk Drive West Jordan. UT 84084 Tel: 801-256-9282 Fax: 801-256-9287
Robert Larsen, Preparation Date: August 15, 2006
NUV 1 4 2006
KO62619
#### Device Name:
| Trade Name: | DenLaser 800 Plus |
|-------------------------|----------------------------|
| Common Name: | 810nm Diode Laser |
| Product Classification: | Laser Instrument, Surgical |
# Legally Marketed Predicate Devices for Substantial Equivalence:
Odyssey 2.4G Diode Laser, manufactured by Ivoclar Vivadent, Inc. (K050453)
Ceralas D 810 Diode Laser, manufactured by Biolitec, Inc. (K032864)
Aurora SL Diode Laser, manufactured by Premier Laser Systems, Inc. (K993285)
### Rationale for Substantial Equivalence:
The aforementioned laser devices and their accompanying delivery systems share similar indications for use on soft tissue with the submitted device for cutting soft tissue, affecting lesions, and photocoagulation. The predicate devices and submitted device share similar design features including wavelength, operating controls, and laser delivery method. The devices share similar methods of control systems, safety features, and performance monitoring. The devices share similar performance specifications including power output and energy type. The devices share similar applications to soft tissue.
#### Description of Submitted Device:
The DenLaser 800 Plus is a device for delivering laser energy to human soft tissue for a variety of surgical procedures. This energy is generated by solid-state diodes, which provide a consistent and reliable generation of laser energy at 810 ± 20nm for a maximum of 5 watts of energy output. The laser energy is delivered to surgical site by means of a proprietary optical fiber system, which allows for the safe transmission of laser energy to the site without creating undue risk to the patient or operatory staff by
{1}------------------------------------------------
errant or collateral laser emissions. The device features some user definable settings, including a switchable 630nm aiming beam, adjustable power output for both the working beam and aiming beam, and continuous delivery or pulse delivery option.
The working end of the delivery fiber is contained within metal handpiece with a disposable single-use tip. This handpiece system is provided with the device. The activation of the working beam diodes is completed by use of a foot-actuated switch.
# Intended Uses of the DenLaser 800 Plus diode laser system:
The device is intended to be used for a variety of surgical procedures on soft tissue. See Indications for Use on page 2-1.
# Technological Characteristics and Substantial Equivalence:
The Odyssey 2.4G Diode Laser uses solid state diodes to generate laser energy in the 810nm range. This system uses a fiber delivery system to transmit laser energy to the surgical site. The system also features a 630nm aiming beam and features controls that allow for adjusting the output of the working beam, and switching between a continuous or pulsed-mode laser emission. The device features a wireless foot switch for actuating the working beam. The maximum output of the working beam is 5 watts.
The Ceralas D 810 Diode Laser uses solid state diodes to generate laser energy in the 810nm range. This system uses a fiber delivery system to transmit laser energy to the surgical site. The system also features a 635nm aiming beam and features controls that allow for adjusting the output of the working beam, and switching between a continuous or pulsed-mode laser emission. The maximum output of the unit is 15 watts.
The Aurora SL Diode Laser uses solid state diodes to generate laser energy in the 810nm range. This system uses a fiber delivery system to transmit laser energy to the surgical site. The system also features a 630-680nm aiming beam and features controls that allow for adjusting the output of the working beam. The maximum output of the unit is 6 watts.
#### Performance Standards:
The DenLaser 800 Plus diode laser complies with the performance requirements of 21 CFR 1040.10 and 1040.11, with permissible deviations relative to Laser Notice 50, dated July 26, 2001. The device also complies with IEC 60601-1:1998+A1, IEC 60601-2-22:1995, IEC 60825-1:1993+A1+A2, 47 CFR 15 and 18, and ETSI 301-489-1.
### Clinical Performance Data
{2}------------------------------------------------
# Conclusion
The DenLaser 800 Plus is substantially equivalent to the listed laser surgical devices without raising any issues of safety or effectiveness. This device shares similar intended uses, and similar functional and performance characteristics. The device is designed to comply with relevant federal and international safety and performance standards.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird symbol with three curved lines representing its body and wings. The bird is enclosed within a circular border, and the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is written around the border of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
CAO Group, Inc. % Mr. Robert K. Larsen Operations Director 4628 West Skyhawk Drive West Jordan, Utah 84084
NOV 1 4 2006
Re: K062619 Trade/Device Name: DenLaser 800 Plus Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: August 15, 2006 Received: September 6, 2006
Dear Mr. Larsen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been rcclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Mr. Robert K. Larsen
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sincerely yours,
Mark N. Melkerson
Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
| 510(k) Number (if known): | K062619 |
|---------------------------|---------|
|---------------------------|---------|
Device Name: DenLaser 800 Plus
Indications For Use:
The DenLaser 800 Plus is indicated for the procedures of
removal of lesions, excision, incision, vaporization, ablation, hemostasis, and photocoagulation
on soft tissue in the fields of
otolaryngology (ear, nose, and throat), dentistry and oral surgery, arthroscopy, gastroenterology, dermatology, podiatry, plastic surgery, urology, gynecology, and plastic surgery.
| Prescription Use | X | AND/OR Over-The-Counter Use | |
|----------------------------|---|-----------------------------|--|
| (Per 21 CFR 801 Subpart D) | | (21 CFR 801 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of General, Restorative, and Neurological Devices
| 510(k) Number | L06219 | Page 1 of 1 |
|---------------|--------|-------------|
|---------------|--------|-------------|
2-1
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.