K062386 · Medis Medical Imaging Systems BV · JAK · Oct 5, 2006 · Radiology
Device Facts
Record ID
K062386
Device Name
QANGIO CT
Applicant
Medis Medical Imaging Systems BV
Product Code
JAK · Radiology
Decision Date
Oct 5, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Software as a Medical Device
Intended Use
The QAngio CT software solution has been developed for the objective and reproducible analysis of vessels in CTA images. It enables the quantitative analysis of CT angiograms based on automated segmentation. More specifically, QAngio CT can be used to quantify a number of lesion characteristics. QAngio CT is intended for use as an auxiliary tool in assessing CTA studies in clinical practice and in clinical trials. The analysis results obtained with QAngio CT are to be interpreted by cardiologists or radiologists.
Device Story
QAngio CT is a software package for quantitative analysis of CTA images; reads DICOM CT images from major vendors. Device performs automated (or manual) vessel segmentation; detects stenosis; quantifies lesion characteristics. Used in clinical practice/trials on workstations in review rooms or integrated into PACS. Operated by radiologists, cardiologists, or technicians. Output consists of quantitative vessel analysis data saved in files; used by clinicians to support interpretation of CTA data, evaluate follow-up studies, and assess treatment effectiveness. Operator interprets objective values to accept or reject results; aids clinical decision-making by providing reproducible measurements.
Clinical Evidence
No clinical data provided. Safety and effectiveness supported by risk management plan, hazard/risk analyses, verification and validation testing, and hospital evaluations.
Technological Characteristics
Standalone PACS software for image post-processing. Utilizes automated segmentation algorithms for vessel analysis. Compatible with DICOM CT images. Software level of concern is 'Minor'.
Indications for Use
Indicated for use by cardiologists and radiologists as an auxiliary tool for the objective and reproducible quantitative analysis of vessels in CTA images, including quantification of lesion characteristics in clinical practice and clinical trials.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
K173637 — syngo.CT Coronary Analysis; syngo.CT Vascular Analysis · Siemens Medical Solutions USA, Inc. · Mar 30, 2018
K133993 — CAAS WORKSTATION · Pie Medical Imaging · Mar 25, 2014
K200274 — CariCloud · Caristo Diagnostics · May 21, 2020
K060779 — ADVANCE VESSEL ANALYSIS II · GE Healthcare · Apr 5, 2006
K122429 — VESSEL ANALYSIS AND AUTOPLAQUE FOR ORS VISUAL · Object Research Systems (Ors), Inc. · Nov 28, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
# 12. SUMMARY OF SAFETY AND EFFECTIVENESS
Submission in accordance with the requirements of 21 CFR Part 807.87(h)
| • Submitter | : Medis medical imaging systems bv | OCT -5 2006 |
|--------------------|--------------------------------------------------------|-------------|
| Address | : Schuttersveld 9<br>: 2316 XG Leiden, The Netherlands | |
| Telephone | : +31 71 522 3244 | |
| Fax | : +31 71 521 5617 | |
| Contact Person | : J.I. Hollander, Quality Coordinator | |
| Prepared | : August 10, 2006 | |
| • Device Name | : Automatic quantitative analysis of CTA images | |
| Common Name | : QAngio® CT | |
| Device Class. Name | : Class II; PACS software | |
| Regulation Number | : 21 CFR 892.1750 (90 JAK) | |
· Predicate Device(s)
## · Device Description and indications for Use
QAngio CT is able to read DICOM CT images from all major CT scanner vendors. Vessel analysis data, generated by automated (and/or manual) segmentation, detected stenosis, and quantitative results, can be saved in separate files enabling the comparison of results from different users.
Radiologists, cardiologists and technicians use the QAngio CT analytical software package to obtain objective and reproducible results. The obtained results may be used to support the interpretation of CTA data, or they are used in the evaluation of follow-up studies and the effectiveness of treatment.
In clinical practice the QAngio CT software is used on workstations in review rooms or integrated in a PACS environment ..
### · Intended use
062386
The QAngio CT software solution has been developed for the objective and reproducible analysis of vessels in CTA images. It enables the quantitative analysis of CT angiograms based on automated segmentation. More specifically, QAngio CT can be used to quantify a number of lesion characteristics.
QAngio CT is intended for use as an auxiliary tool in assessing CTA studies in clinical practice and in clinical trials. The analysis results obtained with QAngio CT are to be interpreted by cardiologists or radiologists.
### · Substantial equivalence Information
QAngio CT is substantially equivalent to the Predicate Device of MRA-CMS, K040746, using the same technique for the same intended use; only CT- instead of MR images.
### Conclusion
In Medis' opinion. QAngio CT is a safe medical device. During the development, potential hazards were controlled by a risk management plan, including hazard and risk analyses, verification and validation tests. Evaluations by hospitals and literature information support this statement. The software package QAngio CT itself will not have any adverse effects on health. The operator interprets the objective values of the analysis and chooses to accept or reject the results.
In current thinking, the level of concern for the standalone software in image postprocessing is "Minor' and the use of QAnqio CT does not change the intended use of CT scanners in practice, nor does the use result in any new potential hazard.
Based on the information supplied in this 510(k), Medis concludes that the subject device is safe, effective and substantially equivalent to the predicate device.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol. The logo is black and white and appears to be a scanned or printed image.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
OCT - 5 2006
Mr. J. I. Hollander Quality Coordinator Medis Medical Imaging Systems bv 9 Schuttersveld Leiden The Netherlands 2316XG
Re: K062386
Trade/Device Name: QAngio CT Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Dated: August 11, 2006 Received: August 16, 2006
Dear Mr. Hollander:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/1/Picture/10 description: The image is a black and white circular logo. The logo contains the text "1906-2006" at the top, followed by the letters "PA" in a large, bold font. Below the letters, the word "Centennial" is written in a cursive font. Three stars are located at the bottom of the logo. The logo is surrounded by a dotted border.
Protecting and Promoting Public Health
{2}------------------------------------------------
Page 2 -
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Or H in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) prematket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Number (if known): Ko 6 2 38 4
Device Name: QAngio CT
Indications For Use:
QAngio CT software solution has been developed for the objective and reproducible QAnglo CT sollware Solution has been Geven Groupstitative analysis of CT angiograms
analysis of vessels in CTA images. It enables the quantitative analysis of CT anyonify analysis of vessels in CTA images. It chable the qualing CT can be used to quantify a number of lesion characteristics.
a humber of lesion Characteristics.
QAngio CT is intended for use as an auxiliary tool in assessing CTA studies in clinical QAnglo CT is intended for use as an addidn'y tool in accession of QAngio CT are to be interpreted by cardiologists and radiologists.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDHR, Office of Device Evaluation (ODE)
*Prescription Use*
(Optional Format 3-10-98)
David A. Legron
(Division Sign-Off) Division of Reproductive, Ab and Radiological Devic 510(k) Num
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.