The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.
Device Story
The BD Phoenix System is an automated microbiology platform for rapid identification and antimicrobial susceptibility testing (AST) of bacterial isolates. The system utilizes molded polystyrene panels containing 136 micro-wells with dried reagents. Pure culture bacterial isolates are inoculated into the panels and placed into the instrument, which maintains a constant 35°C incubation. The system employs a broth microdilution method, using a redox indicator and turbidity measurements to detect bacterial growth in the presence of antimicrobial agents. The instrument performs readings every 20 minutes. Software interprets these readings to provide identification, minimum inhibitory concentration (MIC) values, and categorical interpretations (S, I, R, or N). The system is intended for use in clinical laboratory settings by trained personnel. Results assist healthcare providers in selecting appropriate antibiotic therapy for patients with bacterial infections.
Clinical Evidence
Performance evaluated using clinical, stock, and challenge isolates across multiple U.S. sites. Comparison against CLSI reference broth microdilution method. For amoxicillin-clavulanate, study included 871 isolates. Results: 94.1% Essential Agreement (EA) and 96.7% Category Agreement (CA). Reproducibility testing demonstrated >95% agreement within ±1 dilution.
Technological Characteristics
System uses colorimetric oxidation-reduction (redox indicator) for growth detection. Materials: molded polystyrene trays with 136 micro-wells. Energy: 35°C incubation. Connectivity: automated instrument with barcode scanning. Software: expert system for result interpretation. AST broth: cation-adjusted Mueller-Hinton with 0.01% Tween 80. Inoculum: 0.5 McFarland standard.
Indications for Use
Indicated for testing Staphylococcus spp. and Enterococcus spp. using amoxicillin-clavulanate at concentrations of 0.25/0.12 – 32/16 µg/mL on Gram-positive ID/AST or AST panels.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
Predicate Devices
VITEK® System (PMA No. N50510)
BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, K060324)
BD Phoenix™ Automated Microbiology System with Ofloxacin (K020323)
BD Phoenix™ Automated Microbiology System with Levofloxacin (K020322)
Related Devices
K050982 — BD PHONIX AUTOMATED MICROBIOLOGY SYSTEM - CEFOTAXIME STREP (0.0625-4UG/ML) · Becton, Dickinson & CO · May 24, 2005
{0}------------------------------------------------
K061673
CONFIDENTIAL AND PROPRIETARY
# 510(k) SUMMARY
| SUBMITTED BY: | JUL 2 8 2006<br>Becton, Dickinson and Company<br>7 Loveton Circle<br>Sparks, MD 21152<br>Phone: 410-316-4161<br>Fax: 410-316-4499 | | | |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|
| CONTACT NAME: | Vicki L. Kennedy,<br>Regulatory Affairs Specialist | | | |
| DATE PREPARED: | June 13, 2006 | | | |
| DEVICE TRADE NAME: | BD Phoenix™ Automated Microbiology System -<br>Amoxicillin-clavulanate 0.25/0.12-32/16 ug/mL | | | |
| DEVICE COMMON NAME: | Antimicrobial susceptibility test system-short incubation | | | |
| DEVICE CLASSIFICATION: | Fully Automated Short-Term Incubation Cycle Antimicrobial<br>Susceptibility Device, 21 CFR 866.1645 | | | |
| PREDICATE DEVICES: | VITEK® System (PMA No. N50510) and BD Phoenix™<br>Automated Microbiology System with Gatifloxacin (K020321,<br>May 23, 2002 and K060324, May 25, 2006), Ofloxacin<br>(K020323, April 14, 2002), and Levofloxacin (K020322, March<br>27, 2002). | | | |
| INTENDED USE: | The BD Phoenix™ Automated Microbiology System is<br>intended for the rapid identification and in vitro antimicrobial<br>susceptibility testing of isolates from pure culture of most<br>aerobic and facultative anaerobic Gram-negative and Gram-<br>positive bacteria of human origin. | | | |
# DEVICE DESCRIPTION:
The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:
- BD Phoenix instrument and software. ●
- BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents . or AST determinations.
- BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. .
- BD Phoenix AST Broth used for performing AST tests only. .
- BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth . determination.
{1}------------------------------------------------
The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. Phoenix panels are inoculated with a specified organism density and placed into the instrument.
The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.
The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, R or N (susceptible, intermediate, resistant or not susceptible).
# DEVICE COMPARISON:
The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the CLSI reference broth microdilution method. This premarket notification provides data supporting the use of the BD Phoenix™ Automated Microbiology System Gram positive ID/AST or AST only Phoenix panels with this antimicrobial agent.
# SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:
The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the CLSI reference broth microdilution method (AST panels prepared according to NCCLS M7). The system has been evaluated as defined in the FDA Draft guidance document, "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", March 8, 2000.
### Site Reproducibility
Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was evaluated at three sites using a panel of Gram-positive isolates. Each site tested the isolates in triplicate on three different days using one lot of Gram Positive Phoenix panels containing this antimicrobial agent and associated reagents.
The results of the study demonstrate for the this antimicrobial agent there was an overall intra-site reproducibility of greater than 90% and an overall inter-site reproducibility greater than 95% for the Gram-positive isolates tested.
{2}------------------------------------------------
# Clinical Studies
Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with the Gram Positive Phoenix Panel format containing this antimicrobial agent. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the CLSI reference broth microdilution method.
The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within + one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, resistant or not susceptible).
Table 1 summarizes the performance for the isolates tested in this study.
#### Table 1: Performance of BD Phoenix System for Gram-Positive Organisms by Drug
| Antimicrobial | Concentration | EA (n) | EA (%) | CA (n) | CA (%) |
|-------------------------|----------------------------|--------|--------|--------|--------|
| Amoxicillin-clavulanate | 0.25/0.12-32/16 $\mu$ g/mL | 871 | 94.1 | 871 | 96.7 |
### Conclusions Drawn from Substantial Equivalence Studies
The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA draft guidance document, "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", March 8, 2000. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002 and K060324, May 25, 2006), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322, March 27, 2002).
{3}------------------------------------------------
# SUMMARY INFORMATION FOR AMOXICILLIN-CLAVULANATE Available Range 0.25/0.12-32/16 µg/mL
## Performance
#### Accuracy
| Antimicrobial | Concentration | EA (n) | EA (%) | CA (n) | CA (%) |
|-------------------------|------------------------|--------|--------|--------|--------|
| Amoxicillin-clavulanate | 0.25/0.12-32/16 µg/mL* | 871 | 94.1 | 871 | 96.7 |
* Performance excludes methicillin-resistant staphylococci
## Reproducibility
Testing performed at multiple clinical sites demonstrated 95% reproducibility or greater within ± 1 dilution.
### Breakpoints - CLSI
| Organism | S | I | R |
|---------------------|-----------|----|------------|
| Staphylococcus spp. | $\leq4/2$ | -- | $\geq8/4$ |
| Enterococcus spp. | $\leq8/4$ | -- | $\geq16/8$ |
# Recommended Quality Control Organisms
| Quality Control Strain | MIC Range | Source |
|-----------------------------------------|---------------------------------------|--------|
| <i>Staphylococcus aureus</i> ATCC 29213 | 0.12/0.06 - 0.5/0.25 $ µg/mL $ | CLSI |
| <i>Enterococcus faecalis</i> ATCC 29212 | 0.25/0.12 – 1.0/0.5 $ µg/mL $ | CLSI |
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird-like figure, composed of three curved lines or strokes.
Public Health Service
JUL 2 8 2006
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Vicki L. Kennedy Regulatory Affairs Specialist Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152
k061673 Re: Trade/Device Name: BD Phoenix™ Automated Microbiology System - Amoxicillinclavulanate 0.25/0.12-32/16 ug/mL Regulation Number: 21 CFR § 866.1645 Regulation Name: Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System Regulatory Class: II Product Code: LON Dated: June 13, 2006 Received: June 14, 2006
Dear Ms. Kennedy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Sally, anthony
Sally A. Hojvat. M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Page 1 of 1
510(k) Number: ¥ 0616 13
Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent amoxicillin-clavulanate (0.25/0.12-32/16 µg/mL) - Gram-positive ID/AST or AST only Phoenix Panels.
Indications for Use:
The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus and Enterococcus.
This premarket notification is for the antimicrobial agent amoxicillin-clavulanate at concentrations of 0.25/0.12-32/16 ug/mL to Gram-positive ID/AST or AST only Phoenix panels. Amoxicillin-clavulanate has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.
### Active In Vitro and in Clinical Infections Against:
Staphylococcus aureus (beta-lactamase and non-beta-lactamase producing)
#### Active In Vitro Against:
Enterococcus faccalis Staphylococccus epidermidis (beta-lactamase and non-beta-lactamase producing) Staphylococcus saprophyticus (beta-lactamase and non-beta-lactamase producing)
Prescription Use (Per 21 CFR 801.109) Over-the-Counter Use
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Luddi tu. Paola
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K221673
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.