CALCANEAL TRAUMA SYSTEM

K061400 · Orthohelix Surgical Designs, Inc. · HRS · Jun 29, 2006 · Orthopedic

Device Facts

Record IDK061400
Device NameCALCANEAL TRAUMA SYSTEM
ApplicantOrthohelix Surgical Designs, Inc.
Product CodeHRS · Orthopedic
Decision DateJun 29, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

To stabilize fractures of the calcaneus when internal fixation of either intra-articular or extra-articular fractures is required.

Device Story

The Calcaneal Trauma System consists of metallic bone fixation plates and screws used to stabilize calcaneal fractures. The system includes four fixation plates, forty-five screws, and associated surgical instruments (drill guides, drill bits, drivers). Used by surgeons in an operating room setting to provide internal fixation for intra-articular or extra-articular fractures. The device provides mechanical stabilization of bone fragments to facilitate healing.

Clinical Evidence

Bench testing only. No clinical data provided. Substantial equivalence supported by FEA and hand calculations.

Technological Characteristics

Materials: Titanium alloy (Ti 6Al 4V ELI per ASTM F 136). Components: Fixation plates, screws, and surgical instruments. Mechanical fixation principle.

Indications for Use

Indicated for the repair of intra and extra articular fractures of the calcaneus.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 510(k) Premarket Notification Calcaneal Trauma System ## 510(k) SUMMARY # Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared. OrthoHelix Surgical Designs, Inc. 1815 West Market St. Suite #205 Akron, Ohio 44313 Phone: (330) 869-9563 Fax: (330) 869-9583 Contact Person: Lee A. Strad Director of Operations & Development OrthoHelix Surgical Designs, Inc. #### Date Prepared: May 4, 2006 #### Name of Device Calcaneal Trauma System ## Common or Usual Name Fixation Plates and Screws ## Classification Name Single/Multiple Component Metallic Fixation Appliances and Accessories #### Predicate Devices DePuy Ace Calcaneal Peri-articular Plate #### Intended Use To stabilize fractures of the calcaneus when internal fixation of either intra-articular or extra-articular fractures is required. #### Device Description The OrthoHelix Calcaneal Trauma System is a set of metallic (titanium alloy), implantable, bone fixation plates and screws. Its' intended function use is to stabilize JUN 2 9 2006 {1}------------------------------------------------ KC0614000 Pg 2 of 2 # 510(k) Premarket Notification ### Calcaneal Trauma System fractures of the calcaneus when internal fixation of either intra-articular or extra-articular fractures is required. The System includes four (4) fixation plates and forty five (45) screws. It also includes various surgical instruments such as drill guides, drill bits and drivers. All screws and plates are made from implant grade titanium, Ti 6Al type 4V ELI per ASTM F 136. # Performance Data FEA's and hand calculations confirm that the Calcaneal Trauma System is substantially equivalent to its' predicate devices, and that it meets specified requirements for its' intended use. {2}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 2 9 2006 Orthohelix Surgical Designs, Inc. c/o Lee A. Strnad Director of Operations & Development 1815 West Market Street, Suite 205 Akron, Ohio 44313 Re: K061400 Trade/Device Name: Calcaneal Trauma System Regulation Number: 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories. Regulatory Class: II Product Code: HRS, HWC Dated: May 15. 2006 Received: May 25, 2006 Dear Mr. Strnad: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality {3}------------------------------------------------ # Page 2 - Mr. Strnad systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120 . Also, please note the regulation ent thed, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Bartare Bruckum Mark Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): TBD Device Name: Calcancal Trauma System Indications for Use: The Calcaneal Trauma System is indicated for the repair of intra and extra articular fractures of the calcaneus Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ Concurrence of CDRH, Office of Device Evaluation (ODE) Page / of /_ Martina Buchwald for Men (Division Sign-Off) Division of General, Restorative, and Neurological Devices **510(k) Number** K061406
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...