K060705 · Draeger Medical B.V. · CBK · Jun 14, 2006 · Anesthesiology
Device Facts
Record ID
K060705
Device Name
CARINA HOME
Applicant
Draeger Medical B.V.
Product Code
CBK · Anesthesiology
Decision Date
Jun 14, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Intended use of the Canna home is round-the-clock pressure controlled and pressure regulated volume controlled ventilation of patients with respiratory insufficiency, using room air, also with additional oxygen. The Carina home is intended for use by people without medical qualifications (patients and caregivers) as well as qualified medical and technical personnel. The device is intended for use with patients with a tidal volume of 100 mL upwards. It can be used invasive and non-invasive, for both controlled and assisted ventilation. The device can be used in Home Care environments, in and out of hospital, in stationary position (lying or sitting) or in a vehicle (e.g. car). Federal law restricts this device to sale by or on the order of a physician.
Device Story
Carina home is a software-controlled ventilator providing pressure-controlled and pressure-regulated volume-controlled ventilation. It delivers respiratory support using room air, optionally supplemented with oxygen, to patients with respiratory insufficiency (tidal volume ≥100 mL). Operated by patients, caregivers, or medical personnel in home, hospital, or mobile environments. Device functions as a life-support system; provides both invasive and non-invasive ventilation. Healthcare providers use device settings to manage patient ventilation; output allows for continuous respiratory support. Benefits include round-the-clock respiratory assistance in various settings.
Clinical Evidence
Bench testing only. Verification and validation testing conducted per IEC60601-1, IEC60601-1-2, and ASTM F1246 standards. Testing included mechanical, electrical, and biocompatibility assessments, alongside hazard analysis and system-level qualification to confirm design requirements were met.
Technological Characteristics
Pressure-controlled and pressure-regulated volume-controlled ventilator. Features integrated HEPA filtration. Operates on room air with optional oxygen enrichment. Designed for stationary or mobile use. Software-controlled architecture. Complies with IEC60601-1 (electrical safety), IEC60601-1-2 (EMC), and ASTM F1246 (ventilator standards).
Indications for Use
Indicated for patients with respiratory insufficiency requiring pressure-controlled or pressure-regulated volume-controlled ventilation. Suitable for patients with tidal volumes of 100 mL or greater. Intended for invasive or non-invasive, controlled or assisted ventilation in home, hospital, or transport settings.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
510(k) Notification Carina ™home QM06_049
Kobo705
Image /page/0/Picture/2 description: The image shows the logo for Dräger Medical. The logo is in black and white, with the word "Dräger" in a bold, sans-serif font. The word "medical" is in a thinner, sans-serif font. Below the logo is the text "A Dräger and Siemens Company" in a smaller font.
510(k) Summary
acc. to 807.92
JUN 1 4 2006
| Submitter's Name and Address: | Dräger Medical b.v.<br>Kanaaldijk 29<br>5683CR Best<br>The Netherlands |
|-----------------------------------------|------------------------------------------------------------------------------|
| Contact Person: | Mr. Hans Venings<br>Vice President Processes, Quality and Regulatory Affairs |
| | Phone: +31 499 331331<br>Fax: +31 499 331222 |
| Applicants US Contact Person: | Ms Monica Ferrante<br>Director, Regulatory Affairs |
| | Phone: (215) 721-5400<br>Fax: (215) 721-5425 |
| Date submission was prepared: | 2006-03-10 |
| Device Name: | |
| Common Name: | Ventilator |
| Classification Name: | Ventilator, Continuous |
| Regulation Number: | 21 CRF 868.5895 |
| Class: | II |
| Legally Marketed Device Identification: | Carina home |
#### Device Description:
the Carina home is a pressure controlled and volume controlled (pressure regulated) ventilator. The device is intended for use with patients with a tidal volume of 100 mL upwards. It is used for patients with respiratory insufficiency, using room air (also with additional oxygen).
#### Intended Use:
The Intended use of the Canna home is round-the-clock pressure controlled and pressure regulated volume controlled ventilation of patients with respiratory insufficiency, using room air, also with additional oxygen. The Carina home is intended for use by people without medical qualifications (patients and caregivers) as well as qualified medical and technical personnel. The device is intended for use with patients with a tidal volume of 100 mL upwards. It can be used invasive and non-invasive, for both controlled and assisted ventilation. The device can be used in Home Care environments, in and out of hospital, in stationary position (lying or sitting) or in a vehicle (e.g. car).
type TEMPLATE release status RELEASED
effective date 30.09.2004
number DMS PQ2160 A4
{1}------------------------------------------------
510(k) Notification Carina™ home QM06 049
Image /page/1/Picture/1 description: The image shows the logo for Dräger medical. The word "Dräger" is in bold, and the word "medical" is in a thinner font. Below the logo, it says "A Dräger and Siemens Company". The logo is simple and professional.
#### Predicate Devices:
| 510(k) Number | Device Name |
|---------------|--------------------------------------|
| K052554 | iVent 201 Portable Ventilator |
| K040790 | LTV 1000 Ventilator |
| K033008 | Air Safety HEPA and non-HEPA Filters |
| K003068 | Savina |
| K970839 | Siemens Servo Ventilator 300A |
### Substantial Equivalence:
The Carina home is similar to the homecare features to he iVent 201 (K052554) and the LTV 1000 (K040790).
It incorporates a pressure regulated volume controlled ventilation mode like the PRCV mode of the Siemens Servo Ventilator 300A (K970839).
It uses an integrated HEPA filter similar to HEPA filters from Air Safety Limited (K033008).
The Carina home and the iVent 201 are both software controlled ventilators using the same operating principle and can be used for invasive as well as non invasive ventilation.
The performance data of both ventilators is comparable, with the exception of some parameters. However, for these parameters, performance data is comparable with the LTV 1000 ventilator (K040790)
#### Summary of Performance Testing:
Safety testing was conducted per IEC60601-1, IEC60601-1-2, ASTM F1246 and other applicable standards with respect to mechanical, electrical and biocompatibility.
The results of all verification and validation testing demonstrate that all system and design requirements for the Carina home device have been met.
Qualification included hazard analysis, system level qualification and verification / validation tests.
TEMPLATE
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like symbol, with three curved lines forming the body and wings. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 1 4 2006
Dräger Medical, B.V. C/O Ms. Monica Ferrante Director, Regulatory Affairs Dräger Medical, Incorporated 3135 Quarry Road Telford, Pennsylvania 18969
Re: K060705
Trade/Device Name: Carina™Home Regulation Number: 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: March 10, 2006 Received: March 16, 2006
Dear Ms. Ferrante:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
## Page 2 - Ms. Ferrante
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
Indications for Use
5.10(k) Number (if known):
Device Name: Carina™home
Indications for Use:
the Carina home is a pressure controlled and volume controlled (pressure regulated) ventilator. The device is intended for use with patients with a tidal volume of 100 mL upwards. It is used for patients with respiratory insufficiency, using room air (also with additional oxygen). Federal law restricts this device to sale by or on the order of a physician.
gn-CT)
on of Anesthesiology, General Hospital,
-don Control, Dental Devices
Number: K060
Prescription Use J (Part 21 CFR 801 Subpart D)
ರಿ
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
のです。 このため、それではないと思います。 このため、 には、 には、 には、 に、 に、 して、 して
| template / rev. | - / - |
|-----------------|-----------------------------------------|
| | CVI HC USA |
| | Indicated Use Statement Carina home USA |
| reference | STD321672 |
| revision | 2 |
| page / of | 3 / 3 |
:
..
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.