K060307 · Ge Medical Systems Information Technologies · DRT · Aug 4, 2006 · Cardiovascular
Device Facts
Record ID
K060307
Device Name
12 RL ALGORITHM
Applicant
Ge Medical Systems Information Technologies
Product Code
DRT · Cardiovascular
Decision Date
Aug 4, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Intended Use
The GE 12RL program generates a 12-lead ECG report from a subset of the electrodes used to acquire a standard 12 Icad ECG. Four of the precordial channels of the 12-lead ECG (V2, V3, V4, V6) are not acquired from the patient; rather, they are reconstructed from information that is directly recorded in the other channels of the 12-lead ECG. The four signals generated by the GE 12RL program are similar but not identical to the standard 12-lead ECG. All ECG data generated via 12RL are clearly identified as to which channels have been synthesized. The GE 12RL program is intended for use in a monitoring environment. Computerized measurements may be generated from these data; however, a computerized interpretation will not. The product is intended for use in the general adult population ranging from healthy subjects to patients with cardiac and/or non-cardiac abnormalities. The product is to be used in conjunction with the patient's clinical history, symptoms, and other diagnostic tests for final clinical judgment.
Device Story
12RL program reconstructs 4 precordial ECG channels (V2, V3, V4, V6) from remaining recorded channels using linear equations; enables 12-lead ECG presentation from reduced electrode set. Used in clinical monitoring environments; operated by healthcare professionals. Outputs 10-second records or continuous recordings for ST segment trending. Synthesized channels clearly identified in reports. Computerized measurements supported; automated interpretation excluded. Used alongside clinical history and diagnostic tests for final clinical judgment. Benefits include reduced electrode burden while maintaining 12-lead diagnostic utility.
Clinical Evidence
Bench testing only. No clinical data provided. Compliance with voluntary standards and software verification/validation testing confirmed performance equivalent to predicate devices.
Technological Characteristics
Software-based algorithm using linear equations to reconstruct ECG signals. Operates on standard ECG monitoring hardware. Connectivity: Integrated into monitoring systems. Software verification/validation performed.
Indications for Use
Indicated for general adult population, including healthy subjects and patients with cardiac or non-cardiac abnormalities, in a monitoring environment to reconstruct 12-lead ECG reports using a subset of electrodes.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
K020456 — EASI 12 LEAD ALGORITHM · Philips Medical Systems · Jul 9, 2002
K990476 — HP VIRIDIA COMPONENT MONITORING SYSTEM, HP VIRIDIA CMS 24/26, MODEL REV. K WITH EASI-ECG OPTION · Hewlett-Packard GmbH · Jul 22, 1999
K030738 — SIEMENS MULTIVIEW INFINITY TELEMETRY SYSTEM · Siemens Medical Solutions USA, Inc. · Jun 6, 2003
K102378 — VECTRAPLEXECG SYSTEM WITH VECTRAPLEXAMI · Vectracor, Inc. · Nov 2, 2011
K052049 — PHILIPS 12-LEAD ALGORITHM · Philips Medical Systems · Aug 26, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
## Section 2 Summary of Safety and Effectiveness
| Date: | July 28, 2006 | | | |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|-------------------------------------------------------------------------------|--------------|
| Submitter: | GE Medical Systems Information Technologies<br>8200 West Tower Avenue<br>Milwaukee, WI 53223 USA | | | AUG - 4 2006 |
| Contact Person: | Adrienne Lenz<br>Quality Engineer<br>GE Medical Systems Information Technologies<br>Phone: 414-721-3245<br>Fax: 414-721-3863<br>Email: Adrienne.Lenz@ge.com | | | |
| Device:<br>Trade<br>Name: | Twelve Lead Transformation Function, "12RL" | | | |
| Common/Usual<br>Name: | 12RL | | | |
| Classification<br>Names: | 21 CFR 870.2300 Cardiac monitor | | | 74DRT |
| | 21 CFR 870.2340 Electrocardiograph | | | 74DPS |
| | 21 CFR 870.1025 Detector and Alarm, Arrhythmia | | | 74DSI |
| | 21 CFR 870.1425 Programmable diagnostic<br>computer | | | 74DQK |
| | 21 CFR 870.2920 Transmitters and Receivers,<br>Electrocardiograph, Telephone | | | 74DXH |
| Predicate Device: | 1 K992595<br>2 K030738<br>3 K964750 | 1 EASI<br>2 TruST<br>3 Marquette Eagle 4000 | 1 Philips<br>2 Siemens<br>3 GE Medical Systems<br>Information<br>Technologies | |
The 12RL feature is intended to reconstruct electrocardiograph Device Description: (ECG) signals such that a 12-lead ECG presentation can be made when at least six electrodes are attached to a patient. These 12-lead ECGs may be 10 second records or continuous recordings suitable for ST segment trending.
{1}------------------------------------------------
The GE 12RL program generates a 12-lead ECG report from a Intended Use: subset of the electrodes used to acquire a standard 12 lead ECG. Four of the precordial channels of the 12-lead ECG (V2, V3, V4, V6) are not acquired from the patient; rather, they are reconstructed from information that is directly recorded in the other channels of the 12-lead ECG.
> The four signals generated by the GE 12RL program are similar but not identical to the standard 12-lead ECG. All ECG data generated via 12RL are clearly identified as to which channels have been synthesized.
> The GE 12RL program is intended for use in a monitoring environment. Computerized measurements may be generated from these data; however, a computerized interpretation will not.
The product is intended for use in the general adult population ranging from healthy subjects to patients with cardiac and/or noncardiac abnormalities.
The product is to be used in conjunction with the patient's clinical history, symptoms, and other diagnostic tests for final clinical judgment.
- Technology: The 12RL technology is based on linear equations as is used by the predicate devices. Therefore there is no change in technology.
- The 12RL complies with the voluntary standards as detailed in Test Summary: Section 9 of this submission. The following quality assurance measures were applied to the development of the 12RL program:
- Risk Analysis .
- Requirements Specification Review .
- Code Inspections .
- Software Verification and Validation Testing .
- The results of these measurements demonstrated that the 12RL Conclusion: program is as safe, as effective, and performs as well as the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized eagle with three lines forming its body and two curved lines representing its legs. The seal is surrounded by the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG - 4 2006
GE Healthcare Information Technologies c/o Ms. Adrienne Lenz Regulatory Affairs Specialist 9900 Innovation Drive Wauwatosa, WI 53226
Re: K060307
Trade Name: 12 Lead Transformation Function, "12RL" Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II (two) Product Code: DRT Dated: July 28, 2006 Received: July 31, 2006
Dear Ms. Lenz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 – Ms. Adrienne Lenz
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bfremmima for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): K060307
Device Name: 12 RL Algorithm
Indications For Use:
The GE 12RL program generates a 12-lead ECG report from a subset of the electrodes used to acquire a standard 12 Icad ECG. Four of the precordial channels of the 12-lead ECG (V2, V3, V4, V6) are not acquired from the patient; rather, they are reconstructed from information that is directly recorded in the other channels of the 12-lead ECG.
The four signals generated by the GE 12RL program are similar but not identical to the standard 12-lead ECG. All ECG data generated via 12RL are clearly identified as to which channels have been synthesized.
The GE 12RL program is intended for use in a monitoring environment. Computerized measurements may be generated from these data; however, a computerized interpretation will not.
The product is intended for use in the general adult population ranging from healthy subjects to patients with cardiac and/or non-cardiac abnormalities.
The product is to be used in conjunction with the patient's clinical history, symptoms, and other diagnostic tests for final clinical judgment.
Prescription Use X (Per 21 CFR 801.109 Subpart D) OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE
Blummenauer
Division Sign Off
IDivision Sian-Off Division of Cardlovascular Devices 510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.