K030738 · Siemens Medical Solutions USA, Inc. · MHX · Jun 6, 2003 · Cardiovascular
Device Facts
Record ID
K030738
Device Name
SIEMENS MULTIVIEW INFINITY TELEMETRY SYSTEM
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
MHX · Cardiovascular
Decision Date
Jun 6, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The INFINITY MultiView WorkStation (MVWS) Telemetry System is intended to measure and produce visual and audible alarms for one or more physiological parameters. The INFINITY MVWS telemetry System with TruST is intended for 12-Lead ECG monitoring with a reduced set of electrodes. Reconstructed leads are intended for real-time assessment of ST segment changes. This device is intended for use in an environment where patient care is provided by Healthcare Professionals, i.e. Physicians, Nurses, and Technicians, who will determine when use of the device is indicated, based upon their professional assessment of the patient's medical condition.
Device Story
The Siemens INFINITY MultiView WorkStation Telemetry System with TruST enables continuous 12-lead ECG monitoring using a reduced set of 6 electrodes. The system utilizes the TruST matrix, which applies general reconstruction coefficients to compute 4 precordial leads from the reduced electrode input. The device processes these signals to provide real-time assessment of ST segment changes and generates visual and audible alarms for physiological parameters including heart rate, arrhythmia, SpO2, and pulse rate. Used in hospital or clinical environments, the system is operated by healthcare professionals (physicians, nurses, technicians) to monitor patient status. The output is displayed on the workstation, allowing clinicians to assess ST segment changes and make informed clinical decisions regarding patient care.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and performance comparisons to predicate devices.
Technological Characteristics
The system uses a 6-electrode configuration to reconstruct a 12-lead ECG. It employs a TruST matrix with general reconstruction coefficients to compute 4 precordial leads. The device is a physiological monitor for ECG, SpO2, and blood pressure. It is a networked telemetry system. No specific materials or sterilization methods are listed as applicable.
Indications for Use
Indicated for adult and pediatric patients in clinical settings under professional care (physicians, nurses, technicians) requiring monitoring of heart rate, ECG arrhythmia, arterial oxygen saturation, pulse rate, or ST segment analysis. TruST feature indicated for 12-lead ECG monitoring using a reduced electrode set for real-time ST segment change assessment. Not MRI compatible.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
Philip's (HP) Viridia Component Monitoring System with EASI™ ST Segment (K992595)
INFINITY MultiView WorkStation Telemetry System Modification (K024108)
{0}------------------------------------------------
'JUN - 6 2003
# 510(k) SUMMARY
as required per 807.92(c)
## 1. Submitters Name, Address:
Siemens Medical Solutions, Inc. Electromedical Systems Group, PCS Danvers, MA 01923 Tel: (978) 907-7500 Fax: (978) 750-6879 Official Correspondent: Connie Hertel, Director Quality Assurance & Regulatory Affairs Contact person for this submission: Penelope Greco Date submission was prepared: March 7, 2003
#### 2. Trade Name, Common Name and Classification Name:
### A. Trade Name:
SIEMENS INFINITY MultiView WorkStation™ Telemetry System with TruST™
| Common Name | Classification<br>Number | Class | Regulation Number |
|--------------------------------------------------------------------------|--------------------------|-------|-------------------|
| Monitor, Physiological, Patient (with arrhythmia<br>detection or alarms) | MHX | III | 21 CFR 870.1025 |
| Radiofrequency physiological signal transmitter<br>and receiver | 74DRG | II | 21 CFR 870.2910 |
| Cardiac monitor | 74DRT | II | 21 CFR 870.2300 |
| Pulse rate monitor | 74BWS | II | 21 CFR 870.2300 |
| Pulse oximeter | 74DQA | II | 21 CFR 870.2700 |
| Breathing Frequency Monitor | 73BZQ | II | 21 CFR 868.2375 |
| Clinical Electronic Thermometer | 80BWX | II | 21 CFR 880.2910 |
| Indwelling Blood Pressure Monitor | 74CAA | II | 21 CFR 870.1110 |
| Heart Rate Monitor, Neonatal | 74FLO | II | 21 CFR 870.2300 |
| Ventilatory Effort Monitor (Apnea Detector) | 73FLS | II | 21 CFR 868.2375 |
| Monitor Blood Pressure, Neonatal, Invasive | 74FLP | II | 21 CFR 870.1110 |
| Arrhythmia detector & Alarm | 74DSI | III | 21 CFR 870.1025 |
| Medical Cathode-Ray Tube Display | 74DXJ | II | 21 CFR 870.2450 |
| ST Segment Monitor with Alarm | 74MLD | III | 21 CFR 870.1025 |
| Non-indwelling Blood Pressure Monitor | 74DXN | II | 21 CFR 870.1130 |
| End-tidal Carbon-Dioxide Monitor | 73CCK | II | 21 CFR 868.1400 |
### B. Common Name, Classification Name, Class and Regulation Number:
Page 1 of 4
## COMPANY CONFIDENTIAL
Siemens Medical Solutions, Inc.
Electromedical Systems Group, PCS
16 Electronics Avenue Danvers, MA 01923
Tel: (978) 907-7500 Fax: (978) 750-6879
{1}------------------------------------------------
### 3. Predicate Device Identification:
Philip's (HP) Viridia Component Monitoring System with EASI™ ST Segment- 510(k) K992595
INFINITY MultiView WorkStation Telemetry System Modification - K024108 INFINITY MultiView WorkStation Modifications - K022889 SC 9000 / SC 9015 Enhanced with 12-Lead ST Segment Analysis - 510(k) K974698 MultiView WorkStation Enhanced with Diagnostic Statement (Rest ECG) - 510(k) K980625 SC 7000 / SC 9000XL Infinity Modular Bedside Monitors - 510(k) K982730 SC 8000 w/Advanced Communication Option - 510(k) K990563
#### Device Description: 4.
Siemens' INFINITY MultiView WorkStation Telemetry with TruST enables continuous 12-lead ECG monitoring with a reduced number of electrodes. With TruST only 6 lead wires are required to acquire 12 leads of ECG. TruST relies on conventional lead placement and the INFINITY TruST telemetry transmitters. The TruST matrix uses general reconstruction coefficients to compute 4 precordial leads.
#### 5. Intended Use:
The INFINITY MultiView WorkStation (MVWS) Telemetry System is intended to measure and produce visual and audible alarms for one or more physiological parameters
The INFINITY MVWS telemetry System with TruST is intended for 12-Lead ECG monitoring with a reduced set of electrodes. Reconstructed leads are intended for real-time assessment of ST segment changes.
This device is intended for use in an environment where patient care is provided by Healthcare Professionals, i.e. Physicians, Nurses, and Technicians, who will determine when use of the device is indicated, based upon their professional assessment of the patient's medical condition.
Page 2 of 4
### COMPANY CONFIDENTIAL
Siemens Medical Solutions, Inc.
Electromedical Systems Group, PCS
16 Electronics Avenue Danvers, MA 01923
Tel: (978) 907-7500 Fax: (978) 750-6879 050-38
: Of
{2}------------------------------------------------
030738 50F )
510(k) Notification
SIEMENS INFINITY MultiView WorkStation™ Telemetry with TruST™
| Table of Device Similarities and differences to predicate device | | | | |
|------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Substantial Equivalent Device | Substantial Equivalent Device | Applicant | Explanation of<br>Differences |
| (k) Number | HP Viridia Component<br>Monitoring System with EASITM<br>ST Segment Measurement<br>Philips (HP) | SC 9000 / SC 9015 Bedside<br>Monitoring System Enhanced with<br>12-Lead ST Segment Analysis<br>Siemens Medical Solutions | Siemens Medical Systems Infinity<br>MultiView WorkStation Telemetry<br>System with TruST<br>Siemens Medical Solutions | |
| | K992595 | K974698 | To be assigned | |
| Intended Use | Assessment of real time ST<br>segment analysis in adult<br>patients. | To determine the ST Segment of the<br>ECG signal and to compute the<br>deviation of this ST Segment from<br>the iso-electric point (baseline). | The INFINITY MultiView<br>WorkStation (MVWS) Telemetry<br>System is intended to measure and<br>produce visual and audible alarms<br>for one or more physiological<br>parameters | Similar to Philips' Easi<br>in that the MVWS with<br>TruST monitors 12-lead<br>ECG with a reduced set<br>of electrodes. |
| Intended Population | Adult | Adult/Pediatric | The INFINITY MVWS telemetry<br>System with TruST is intended<br>for 12-Lead ECG monitoring with a<br>reduced set of electrodes.<br>Reconstructed leads are intended for<br>real-time assessment of ST segment<br>changes. | |
| Intended<br>Environment | In the medical clinic or hospital<br>environment for use by<br>physicians, nurses and ECG<br>technicians. | Adult/Pediatric | Same | |
| Diagnostic<br>ST Analysis | No | Yes | No | |
| | Up to 12 Derived Complexes | Up to 12 Measured ST Complexes | Up to 8 Measured ST Complexes<br>Up to 4 Derived ST Complexes | dV1, dV3, dV4, dV6 |
| | 5 | 12 | 6 | |
Page 3 of 4
COMPANY CONFIDENTIA
16 Electronics Avenue
Danvers, MA 01923
Tel: (978) 907-7500
Fax: (978) 750-6879
Siemens Medical Solutions,
Electromedical Systems Group, PCS
{3}------------------------------------------------
- 7. Assessment of non-clinical performance data for equivalence: Currently there are no FDA standards for this device.
- 8. Assessment of clinical performance data for equivalence: Section J
- 9. Biocompatability: Not applicable
- 10. Sterilization: Not applicable
- 11. Standards and Guidances: Section P
Page 4 of 4
## COMPANY CONFIDENTIAL
Siemens Medical Solutions, Inc. Electromedical Systems Group, PCS
16 Electronics Avenue Danvers, MA 01923
Tel: (978) 907-7500 Fax: (978) 750-6879
K 0 30 7 38 by afy
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, overlapping each other.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN - 6 2003
Ms. Penelope H. Greco Regulatory Submissions Manager Siemens Medical Solutions, Inc. Electromedical Systems Group, PCS 16 Electronics Avenue Danvers, MA 01923
Re: K030738
Trade Name: MultiView WorkStation Telemetry System with TruSTTM Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm Regulatory Class: Class III (three) Product Code: MHX Dated: March 7, 2003 Received: March 10, 2003
Dear Ms. Greco:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{5}------------------------------------------------
Page 2 - Ms. Penelope H. Greco
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Bram D. Zuckerman, M.D.
Director Division of Cardiovascular Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number (if known): K030738
### Siemens Infinity MultiView WorkStation Telemetry System Device Name: Enhanced with TruST™
Indications for Use:
Use of the INFINITY MultiView WorkStation Telemetry System is indicated for adult and pediatric patient populations in an environment where patient care is provided by Healthcare Professionals (Physicians, Nurses, Technicians) when the professional determines that a device is required to measure and produce visual and audible alarms for any one or more of the following parameters:
- . Heart rate
- ECG Arrhythmia Analysis ●
- Arterial oxygen saturation ●
- Pulse rate
- ST segment analysis .
The INFINITY MultiView WorkStation (MVWS) Telemetry System with TruST is indicated for use when 12-Lead ECG monitoring with a reduced set of electrodes is desired. Reconstructed leads are intended for real-time assessment of ST segment changes.
### MRI Compatibility Statement:
The Infinity MultiView WorkStation Telemetry System is not compatible for use in a MRI magnetic field.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
N. Carter
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.