Two and three dimensional image review, manipulation, analysis and therapy planning capabilities that support image management display needs in the medical environment from multiple locations within and outside the hospital. Diagnostic Review Workstations - Assists medical professionals including but not limited to radiologists, surgeons, oncologists, neurologists, and cardiologists in conducting primary diagnostic review and planning through the flexible and interactive manipulation of multi-modality softcopy images. Two and three-dimensional image coregistration (fusion) and segmentation of multiple imaging modality data including but not limited to functional MRI, PET, SPECT, and CT for user-selected regions of interest taken at different time points or generated using different scanning protocols. Detailed measurement and reporting features assist clinicians in assessing and documenting changes in regions of interest such as pathologies (e.g., tumors, abscesses, AVMs), or other anatomical structures. The software can help track changes in morphology, functional activity, or other responses that may occur as a result of treatment therapy or disease progression.
Device Story
Software workstation for medical image processing; facilitates oncology and clinical specialty workflows. Inputs: multi-modality imaging data (MRI, PET, SPECT, CT) from different time points or scanning protocols. Processing: coregistration (fusion) of datasets; segmentation of Regions of Interest (ROI); calculation of quantitative metrics (ADC, SUV). Outputs: displayed images, measurements, and reports tracking morphological or functional changes. Used in clinical environments (hospitals, clinics) by physicians (radiologists, surgeons, oncologists, neurologists, cardiologists). Standalone or integrated as a plug-in to existing workstations. Assists clinicians in diagnostic review, therapy planning, and monitoring disease progression or treatment response.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Medical image processing software; supports multi-modality data (MRI, PET, SPECT, CT). Features include 2D/3D image review, coregistration (fusion), segmentation, and quantitative measurement (ADC, SUV). Operates as standalone application or plug-in for existing workstations.
Indications for Use
Indicated for medical professionals (radiologists, surgeons, oncologists, neurologists, cardiologists) for primary diagnostic review, therapy planning, and documentation of anatomical/pathological changes (tumors, abscesses, AVMs) using multi-modality imaging (MRI, PET, SPECT, CT).
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
K053301
November 25, 2005
Image /page/0/Picture/2 description: The image shows a logo with the word "CEDARA" in block letters below a circular graphic. The graphic is composed of several horizontal lines that curve slightly to suggest a sphere or globe. The lines are thicker in the middle and taper off towards the edges, creating a sense of depth.
## 5. 510(k) Summary
| | | FEB 3 | 2006 |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------|------|
| Submitter: | Cedara Software Corp. | | |
| Address: | 6509 Airport Road<br>Mississauga, Ontario<br>Canada L4V 1S7 | | |
| Contact: | Carol Nakagawa | | |
| Telephone: | (905) 672-2100 | | |
| Date: | November 25, 2005 | | |
| Trade Names: | Cedara I-Response™; Cedara PET/CT™ | | |
| Common Name: | Medical image processing software | | |
| | Classification Name: Picture archiving and communications system | | |
| Predicate Devices: | Cedara I-SoftView, 510(k) No. K040468<br>Siemens syngo TrueD, 510(k) No. K041484 | | |
| Device Description: | The Cedara I-Response and Cedara PET/CT software products are<br>line extensions to Cedara's medical image processing workstation<br>product, Cedara I-SoftView™. The products are designed to<br>facilitate the oncology or other clinical specialty workflow by<br>comparing medical imaging data from different modalities and/or<br>from different time points. The products are able to coregister<br>functional MRI, PET, SPECT, CT, or other modality datasets.<br>segment out Regions of Interest (ROI), and calculate, display, and<br>report relative differences in Apparent Diffusion Coefficient<br>(ADC). Standardized Update Value (SUV), or other values within<br>those regions. | | |
| | The products can be used as standalone applications or can be<br>"plugged in" and launched from within other applications such as<br>Cedara I-SoftView, or from another workstation or console. | | |
| Indications for Use: | Two and three dimensional image review, manipulation, analysis<br>and therapy planning capabilities that support image management<br>display needs in the medical environment from multiple locations<br>within and outside the hospital. | | |
| | Diagnostic Review Workstations - Assists medical professionals | | |
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo with a stylized globe made of horizontal lines at the top. The lines vary in length, creating a sense of depth and curvature. Below the globe, the word "CCDANA" is written in a simple, sans-serif font. The logo appears to be for an organization or company with the name CCDANA.
516(k) Premarket Notification Coura I Kesponse 1 " and Cedara PF F C F P Submiter Cedara Software Corp November 25, 2005
including but not limited to radiologists, surgeons, oncologists, neurologists, and cardiologists in conducting primary diagnostic review and planning through the flexible and interactive manipulation of multi-modality softcopy images.
Two and three-dimensional image coregistration (fusion) and segmentation of multiple imaging modality data including but not limited to functional MRI. PET. SPECT. and CT for user-selected regions of interest taken at different time points or generated using different scanning protocols.
Detailed measurement and reporting features assist clinicians in assessing and documenting changes in regions of interest such as pathologies (c.g., tumors, abscesses, AVMs). or other anatomical structures. The software can help track changes in morphology. functional activity, or other responses that may occur as a result of treatment therapy or disease progression.
Comparison to Predicate:
The intended use and technological characteristics of Cedara I-Response and Cedara PET/CT software are substantially equivalent. in the opinion of Cedara Software Corp.. to those of the predicate devices and do not pose any new issues of safety and effectiveness.
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines above two wavy lines.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Carol Nakagawa, B.Sc., RAC Sr. Manager, Quality and Regulatory Cedara Software Corp. 6509 Airport Road Mississauga, Ontario, L4V 1S7 CANADA
Re: K053301
FEB 3
2006
Trade/Device Name: Cedara I-Response™ and Cedara PET/CTTM Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications systems Regulatory Class: II Product Code: LLZ Dated: November 25, 2005 Received: November 28, 2005
Dear Ms. Nakagawa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Brigdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name:
K053301 Cedara I-ResponseTM and Cedara PET/CTTM
Indications For Use:
Two and three dimensional image review, manipulation, analysis and therapy planning capabilities that support image management display needs in the medical environment from multiple locations within and outside the hospital.
Diagnostic Review Workstations - Assists mcdical professionals including but not limited to radiologists, surgeons, oncologists, neurologists, and cardiologists in conducting primary diagnostic review and planning through the flexible and interactive manipelation of multi-modality softcopy images.
Two and three-dimensional image coregistration (fusion) and segmentation of multiple imaging modality data including but not limited to functional MRI, PET, SPFCT, and CT for user-selected regions of interest taken at different time points or generated using different scanning protocols.
Detailed measurement and reporting features assist clinicians in assessing and documenting changes in regions of interest such as pathologics (c.g., tumors, abscesses, A VMs), or other anatomical structures. The software can help track changes in morphology, functional activity, or other responses that may occur as a result of treatment therapy or disease progression.
区 Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Aarongl. Bragdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K053301
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.