K041484 · Siemens Medical Solutions USA, Inc. · LLZ · Jul 30, 2004 · Radiology
Device Facts
Record ID
K041484
Device Name
UNDECIDED
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
LLZ · Radiology
Decision Date
Jul 30, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
syngo TrueD is a medical diagnostic application for viewing, image manipulation, 3D visualization and comparison of medical images. syngo TrueD is intended for visualization and comparison of modalities and/or multiple time-points into a single dataset that accurately represents the information from the initial data. The input datasets could be functional 3D volumes (e.g. PET and SPECT) and/or anatomic 3D volumes (e.g. CT and MR). This "fusion" of data may be viewed in a number of output formats including MIP and volume rendering. syngo TrueD enables visualization of information that would otherwise have to be visually compared unformally. syngo TrueD provides analytical tools to help the user assess, and document any morphological or functional activity at diagnostic examinations and therapy follow-up. syngo TrueD is designed to support the evaluation, quantification, follow-up and documentation of any such lesions. Note: The clinician retains the ultimate responsibility for making the pertinent clinical decisions and practices and visual comparison of the separate unregistered images. syngo TrueD is a complement to these standard procedures.
Device Story
syngo TrueD is a software-only image manipulation tool for 3D volume datasets. It takes existing DICOM-formatted imaging data from multiple modalities (e.g., PET, SPECT, CT, MR) or different time-points as input. The device performs data fusion, aligning and assembling these datasets into a single representation. It provides visualization tools, including Maximum Intensity Projection (MIP) and volume rendering, and allows for simple quantitative measurements. Used in clinical settings by trained professionals, the software assists in the visual assessment, quantification, and documentation of morphological or functional activity, such as oncological lesions. By enabling the comparison of registered datasets, it supports clinicians in therapy follow-up and diagnostic decision-making. The clinician retains ultimate responsibility for clinical interpretation.
Clinical Evidence
No clinical data provided. The submission relies on bench testing, hazard analysis, and software design descriptions to demonstrate safety and effectiveness, consistent with the FDA's Reviewers Guidance for Computer Controlled Medical Devices.
Technological Characteristics
Software-only solution; requires specific hardware. Operates on Windows XP. Supports DICOM standard for image data. Features include 3D volume fusion, MIP, volume rendering, and quantitative measurement tools. No patient-contacting materials.
Indications for Use
Indicated for clinicians to visualize, compare, and quantify morphological or functional medical images from multiple modalities (e.g., PET, SPECT, CT, MR) or multiple time-points. Used for oncological evaluation, therapy follow-up, and lesion assessment. No specific age or gender restrictions; intended for use by trained professionals.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
K041484
#### 510(k) Summary JUL 3 0 2004
This summary of 510(k) safety and effectiveness is provided in accordance with the requirements of SMDA 1990 and 21 CFR §807.92
### General Information
### Manufacturer Facility (Developer) Siemens Medical Solutions USA, Inc. 51 Valley Stream Pkwy Malvem, PA 19355 Establishment Registration Number: 2240869
## Manufacturer Facility (Contract Manufacturer)
Siemens Medical Solutions USA, Inc. 2501 North Barrington Rd. Hoffman Estates, IL 60195-7372 Establishment Registration Number: 1423253
#### Contact Person
Richard Andersen Manager of Quality Assurance Phone: (610) 448-4362 Fax: (610) 448-4274
### Device Name and Classification
| Trade Name: | syngo TrueD |
|----------------------|---------------------------------------------|
| Classification Name: | Picture Archiving and Communications System |
| CFR Section: | 21 CFR §892.2050 |
| Device Class: | Class II |
| Product Code: | LLZ |
## Safety and Effectiveness Information Supporting the Substantial Equivalence Determination
# Device Description and Intended Use
syngo TrueD is image manipulation software that is intended to assemble existing datasets from imaging modalities into a single dataset that accurately represents the information from the initial data. This "fusion" of data from multiple modalities will be performed on 3D volume datasets and may be represented in a number of output formats including MIP and volume rendering. Additionally, simple quantitative measurements may be made and subsequently compared to a dataset acquired at a different point in time.
More detailed information regarding the device can be found in the System Description included in Section 2 as well as in the technical information in Section 4.
### Technological Characteristics
syngo TrueD will be marketed as a software only solution for the end-user (with recommended hardware requirements) .It will be installed by Siemens service engineers.
{2}------------------------------------------------
syngo TrueD supports DICOM formatted images and information. It is based on the Windows XP operating system.
### Safety Information
Sarety Information
A summary of the software design description, hazard analysis, and technical and safety information can be found in the attached submission. The results of the hazard analysis, moonlined with the appropriate preventive measures taken indicate the device is of minor Contonious with the appropriate August 29, 1991 issue of the "Reviewers Guidance for Computer Controlled Medical Devices Undergoing 510(k) Review".
The device has no patient contacting materials and is utilized only by trained The device Thas no pations of the device is evaluated by trained professionals allowing sufficient review for identification and intervention in the event of a malfunction. Device sutten review for receinciente the appropriateness of a referral. The device does not impact the quality or status of the original acquired data.
#### Substantial Equivalence:
Substantial Equivalentially equivalent, both in intended use and technically, to the following devices:
| Predicate Device Name | FDA Clearance<br>Number |
|---------------------------------|-------------------------|
| syngo Multimodality Workstation | K010938 |
| Lung CARE CT Software Package | K033374 |
| Advantage Windows CT/PET Fusion | K010336 |
| Leonardo Workstation | K040970 (in review) |
In summary, Siemens is of the opinion that syngo TrueD does not introduce any new potential safety risks and is substantially equivalent to and performs as well as the predicate device.
{3}------------------------------------------------
Image /page/2/Picture/2 description: The image shows a circular logo with text around the perimeter and a symbol in the center. The text around the perimeter reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The symbol in the center appears to be three stylized, abstract shapes that could be interpreted as birds in flight or flowing lines.
JUL 3 0 2004
Food and Drug Administration 00 Corporate Boulevard deville MD 20050
Siemens Medical Solutions USA, Inc. c/o Mr. Glenn Luchen Staff Engineer Underwriters Laboratories, Inc. 1285 Walt Whitman Rd. MELVILLE NY 11747-3081
Re: K041484
Trade/Device Name: syngo TrueD Regulation Number: 21 CFR §892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: July 14, 2004 Received: July 15, 2004
Dear Mr. Luchen:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave teviewed your becamined the device is substantially equivalent (for the indications ferenced above and naro uses.
for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the enclosure) to regally actuated the Medical Device Amendments, or 10 commerce provision to May 20, 1976, the encordance with the provisions of the Federal Food, Drug, devices that have occh roomssing in avere approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, therefore, market the act include requirements for annual registration, listing of gencial controls provisions of the tice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it your device is elassified (500 above) ... Existing major regulations affecting your device can be it may be subject to adament. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I Tease oc accermination that your device complies with other requirements of the Act that I DA has made a doler minutions administered by other Federal agencies. You must of any I ederal statules and regulates.
comply with all the Act's requirements, including, but not limited to: registration and listing Comply with an the rioe Frequiremently (1) CFR Part 801); good manufacturing practice requirements as set (21 CFR Part 007), laboming (21 CFR Part 820); and if applicable, the electronic forth in the quality byevelib (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) I mo lotter notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincercly yours,
Nancy C. Brydon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications For Use
| 510(k) Number (if known): | K041484 |
|---------------------------|-------------|
| Device Name: | syngo TrueD |
## Indications For Use:
syngo TrueD is a medical diagnostic application for viewing, image manipulation, 3Dsyngo Trucib is a modical of medical images. syngo TrueD is intended for visualization and companson of modalities and/or multiple time-points into a single datasce that accurately represents the information from the initial data. The a single datasets could be functional 3D volumes (e.g. PET and SPECT) and/or anatomic Input uatascis court be runctional bis "fusion" of data may be viewed in a number of output formats including MIP and volume rendering.
syngo TrueD enables visualization of information that would otherwise have to be syngo Truct Chaires Visuanzation of TrueD provides analytical tools to help the user visually compared unformedialy. 97%8 morphological or functional activity at diagnostic assess, and document any eningions. syngo TrueD is designed to support the and therapy fonow-up examination the user to confirm the absence or presence of oncological working evaluation, quantification, follow-up and documentation of any such lesions.
Note: The clinician retains the ultimate responsibility for making the pertinent Note. The enmends returns and practices and visual comparison of the separate unregistered images. syngo TrueD is a complement to these standard procedures.
(Please do not write below this line - continue on another page if needed)
Concurrence of the CDRH, Office of Device Evaluation (ODE)
| Prescription Use | <div style="display:inline-block;"> <div style="display:inline-block;">✓</div> </div> | OR Over-The-Counter Use |
|------------------|---------------------------------------------------------------------------------------|-------------------------|
| | (Per 21 CFR 801.109) | |
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K041484
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.