K052981 · Aspect Medical Systems, Inc. · OLW · Nov 1, 2005 · Neurology
Device Facts
Record ID
K052981
Device Name
BIS EEG MONITOR, MODEL BISX4
Applicant
Aspect Medical Systems, Inc.
Product Code
OLW · Neurology
Decision Date
Nov 1, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.1400
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The BISx4 is intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. The BISx4 is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals. The BIS may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.
Device Story
BISx4 is an EEG signal acquisition and processing device; houses digital signal converter, host interface software, and BIS algorithm. Inputs: up to 4 channels of referential EEG signals. Operation: processes EEG to compute BIS (Bispectral Index), SQI (Signal Quality Index), SR (Suppression Ratio), and EMG parameters. Output: processed EEG data for clinical monitoring. Used in hospitals/medical facilities; operated by licensed healthcare practitioners or trained personnel. No onboard display or user interface; connects to host system. Benefits: aids in monitoring anesthetic effects; helps guide anesthetic administration to potentially reduce awareness with recall during general anesthesia and sedation.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including software validation, electrical safety testing, and EMI analysis.
Technological Characteristics
4-channel EEG acquisition system; includes digital signal converter, host interface software, and BIS algorithm. Connectivity: connects to host system (no display/UI). Software: BIS algorithm for EEG parameter computation (BIS, SQI, SR, EMG). Sterilization: N/A. Materials: standard medical device enclosure components.
Indications for Use
Indicated for adult and pediatric patients in hospital or medical facilities to monitor brain state via EEG data acquisition. Used as an aid in monitoring effects of anesthetic agents and may assist in reducing incidence of awareness with recall in adults during general anesthesia and sedation.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
Predicate Devices
Aspect Medical Systems BISx
Submission Summary (Full Text)
{0}------------------------------------------------
# KOS2981
#### 12) 510(k) Summary
## NOV - 1 2005
Company name and address: Aspect Medical Systems, 141 Needham St., Newton, MA 02464
Device name: BISx4
Classification: Class II, classification code: GWO
Predicate device: Aspect Medical Systems BISx
#### Device Description
The BISx4 is a device that houses both the EEG digital signal converter, as well as the host interface software and BIS algorithm. The device will acquire up to four channels of referential EEG and compute BIS and other EEG parameters.
The BISx4 will have no display or user interface. Aspect Medical Systems is submitting this 510(k) in response to anesthesiologists/users who are interested in obtaining more than 2 channels of EEG.
Intended use: The BISx4 is intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. The BISx4 is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals.
The BIS may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.
### Summary of technological characteristics compared to predicate device (similarities/differences)
The BISx4 is similar to the BISx (predicate device).
Similarities: Same design, technology, indication for use, same enclosure, both possess host interface software, BIS algorithm and a digital signal converter (DSC), both have 2 channels, both process BIS, both process parameters such as EEG, SQI, BIS, EMG, Signal Quality Index , (SQI), Suppression Ratio (SR), electrode impedance.
Differences: Addition of 2 channels, with accompanying changes to connectors, software, cables and yoke to enable acquisition of 4 channels.
Testing: Software, electrical, hazard analysis, EMI statement
Conclusion: The BISx4 is substantially equivalent to the BISx predicate device, and is safe and effective for its intended use.
{1}------------------------------------------------
#### 30-Dec-2004 (Ver. 2) Page I of 1
#### Third Party Review Quality Assessment
| 510(k) No.: | K052981 |
|------------------------------------|------------------------------|
| Third Party Organization: | Intertek Testing Services NA |
| Third Party's Primary Reviewer(s): | Jay Kogoma |
| ODE/OIVD Division: | D6RND |
| Branch/Team: | GSDB |
Section 2 – 510(k) Decision
| Third party recommendation: | SE <span></span> X NSE ______ Other (specify): ______ |
|-----------------------------|-------------------------------------------------------|
| ODE/OIVD final decision: | SE <span></span> X NSE ______ Other (specify): ______ |
Section 3 - Assessment of Third Party Review
| Review Element | | Rating (check one) | | |
|------------------------------------------------------------------------------------------------------------------------------------------------|----------|--------------------|-------------------|--|
| | Adequate | Minor<br>Issue(s) | Major<br>Issue(s) | |
| a. Determination of device eligibility for third party review | ✓ | | | |
| b. Extent of pre-submission consultation with ODE/OIVD division | ✓ | | | |
| c. Organization and format of review documentation | ✓ | | | |
| d. Determination of 510(k) administrative completeness (screening<br>review) | ✓ | | | |
| e. Summary of device characteristics, intended use, and performance<br>(including accessories, if applicable) and reason for 510(k) submission | ✓ | | | |
| f. Comparison to legally marketed devices—identification and analysis of<br>key similarities and differences | ✓ | | | |
| g. Rationale for conclusions and recommendation | ✓ | | | |
| h. Use of guidance documents and standards | ✓ | | | |
| i. Resolution of 510(k) deficiencies and FDA requests for additional<br>information | | | | |
| j. Scope of reviewer expertise and use of consulting reviewers | ✓ | | | |
| k. Other (specify): | | | | |
Comments (explanation of ratings/issues): _ Well prepared review, No isSues
.....
______________________________________________________________________________________________________________________________________________________________________________
Section 4 – ODE/OIVD Assessor Information
| <strong>Labels</strong> | <strong>Values</strong> |
|-------------------------|-------------------------|
| Assessed by: | Michel Janda |
| Date: | 10/28 |
| Tel. No.: | (301) 594-1307 *137 |
Routing: Division -- Clip completed assessment (this page only) to inside front cover of 510(k). DMC -- Forward this page only to Eric Rechen, POS/ODE, Rm. 120J, Corp. Blvd. (HFZ-402).
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three curved lines representing its wings and body. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
NOV - 1 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Aspect Medical Systems, Inc. c/o Daniel W. Lehtonen Responsible Third Party Official Intertek Testing Services NA, Inc. 70 Codman Hill Road Boxborough, Massachusetts 01719
Re: K052981
Trade/Device Name: BISx4 Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: II Product Code: GWQ Dated: October 21, 2005 Received: October 24, 2005
Dear Mr. Lehtonen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2-Daniel W. Lehtonen
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours,
Barbara Bucchud
Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): _KOS298 |
Device Name: Aspect Medical Systems, Inc. BISx4
Indications for Use:
The BISx4 is intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. The BISx4 is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EFG nigrals.
The BIS may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Barbara Buchner
(Miss) Clara O'Brien
Division Sign-Of Division of General, Restorative, and Neurological Devices
Page i of 1
**510(k) Number** k052981
(Posted November 13, 2003)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.