POWER RIGHT ANGLE LINEAR CUTTER DIGITAL LOADING UNIT WITH RELOADS
K052910 · Power Medical Interventions, Inc. · GDW · Nov 7, 2005 · General, Plastic Surgery
Device Facts
| Record ID | K052910 |
| Device Name | POWER RIGHT ANGLE LINEAR CUTTER DIGITAL LOADING UNIT WITH RELOADS |
| Applicant | Power Medical Interventions, Inc. |
| Product Code | GDW · General, Plastic Surgery |
| Decision Date | Nov 7, 2005 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 878.4750 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Power Right Angle Linear Cutter Digital Loading Units® have applications in gastrointestinal, gynecological, and general abdominal and thoracic surgical procedures for resection, transection, and creation of anastomoses.
Device Story
Power Right Angle Linear Cutter Digital Loading Unit (DLU) is a surgical stapling device used for resection, transection, and anastomosis in gastrointestinal, gynecological, thoracic, and abdominal surgeries. Device delivers two staggered rows of titanium staples on each side of a transection. Modification allows anvil to swivel 90 degrees when DLU is opened beyond 16 mm, facilitating use in confined anatomical spaces. Operated by surgeons in clinical settings; device functions as a mechanical tool for tissue management.
Clinical Evidence
No clinical data provided; substantial equivalence is based on design and functional comparison to predicate devices.
Technological Characteristics
Mechanical surgical stapler; utilizes titanium staples; features a modified anvil mechanism allowing 90-degree swivel for confined space access; manual operation; no electronic or software components described.
Indications for Use
Indicated for patients undergoing gastrointestinal, gynecological, abdominal, or thoracic surgical procedures requiring resection, transection, or anastomosis creation.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
- SurgASSIST® Right Angle Linear Cutter Digital Loading Units® (K012809)
- SurgASSIST® Right Angle Linear Cutter Digital Loading Units® with Reloads (K021701)
Related Devices
- K022313 — SURGASSIST RIGHT ANGLE LINEAR CUTTER DLU 30MM, 45MM & 60MM - VASCULAR WITH RELOADS, MODELS RALC30V, RALC45V & RALC60V · Power Medical Interventions, Inc. · Dec 9, 2002
- K021701 — SURGASSIST RIGHT ANGLE LINEAR CUTTER 30MM, 45MM & 60MM, MODELS RALC30, RALC45 & RALC60 · Power Medical Interventions, Inc. · Jun 21, 2002
- K063075 — NATURAL ORIFICE LINEAR CUTTER DIGITAL LOADING UNITS · Power Medical Interventions, Inc. · Nov 2, 2006
- K061649 — POWER CIRCULAR STAPLER DIGITAL LOADING UNIT WITH AND WITHOUT SHEATH · Power Medical Interventions, Inc. · Sep 1, 2006
- K083519 — AUTOSUTURE ENDO GIA STAPLERS WITH ENDO GIA SINGLE USE LOADING UNITS · Covidien Lp, Formerly Registered AS United States · Apr 10, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
## NOV -- 7 2005
Power Medical Interventions, Inc. Power Right Anale Linear Cutter Diaital Loadina Units® Special 510(k) Device Modification PreMarket Notification - October 14, 2005
#### SECTION E - Special 510(k) Summary
In Accordance with 21 CFR Section 807.92 Power Medical Interventions® is submitting the following safety and effectiveness summary.
1) Submitter Information:
Power Medical Interventions, Inc. 2021 Cabot Blvd. Langhorne, PA 19047 267-775-8151 Ph 267-775-8123 Fax
Applicant: Barbara J. Whitman
Date of Notification: October 14, 2005
- 2) Name of Device:
| Trade Name: | Power Right Angle Linear Cutter Digital Loading Unit® |
|----------------------|-------------------------------------------------------|
| Common Name: | Linear Staplers with Implantable Staples |
| Classification Name: | Staple, Implantable, GDW |
- 3) Predicate Devices:
SurgASSIST® Right Angle Linear Cutter Digital Loading Units®, Power Medical Interventions, Inc., K012809.
SurgASSIST® Right Angle Linear Cutter Digital Loading Units® with Reloads, Power Medical Interventions, Inc., K021701.
- 4) Device Description
The device described here is a Power Right Angle Linear Cutter Digital Loading Unit® used in gastrointestinal, gynecological, thoracic, bariatric and other surgeries for resections, transactions and the creation of anastomoses.
- 5) Device Modification
The Power Right Angle Linear Cutter Digital Loading Unit® cuts and staples identically to the predicate devices (K021701 and K012809). The Power Right Angle Linear Cutter Digital Loading Unit® was modified to enhance ease of
## 00000
{1}------------------------------------------------
K052910 page 2/2
Power Medical Interventions, Inc. Power Right Angle Linear Cutter Digital Loading Units® Special 510(k) Device Modification PreMarket Notification - October 14, 2005
> use. This new design enables the anvil to swivel 90 degrees once the DLU is opened beyond 16 mm, allowing the user to work in confined anatomical areas.
- 6) Indications For Use
The Power Right Angle Linear Cutter Digital Loading Units® have applications in gastrointestinal, gynecological, and general abdominal and thoracic surgical procedures for resection, transection, and creation of anastomoses.
- 7) Comparison to Predicate Devices
The Power Right Angle Linear Cutter Digital Loading Units® have the same indications for use and the same functions as the previously cleared predicate Right Angle Linear Cutter Digital Loading Units® (K012089 and K021701). Both the Power Right Angle Linear Cutter Digital Loading Units® and the Right Anale Linear Cutter Digital Loading Units® deliver two staggered rows of fitanium staples on each side of a transection. For further details, please see the Predicate Comparison Chart in Section J of this submission.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined snakes around a staff. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA" is arranged in a circular pattern around the caduceus. The seal is black and white.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV - 7 2005
Barbara J. Whitman Regulatory Affairs Manager Power Medical Interventions 2021 Cabot Boulevard Langhorne, Pennsylvania 19047
Re: K052910
Trade/Device Name: Power Right Angle Linear Cutter Digital Loading Units® Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: II Product Code: GDW Dated: October 14, 2005 Received: October 17, 2005
Dear Ms. Whitman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Codc of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not niean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2- Barbara J. Whitman
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely vours.
Sincerely, yours,
Jentare Buend
Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Power Medical Interventions, Inc. Power Right Angle Linear Cutter Digital Loading Units® Special 510(k) Device Modification PreMarket Notification - October 14, 2005
### SECTION D -- Statement of Indications for Use
# Indications for Use
052910 510(k) Number (if known):
Device Name: Power Right Angle Linear Cutter Digital Loading Units®
Indications For Use:
The Power Right Angle Linear Cutter Digital Loading Units® have applications in
gastrointestinal, gynecological, and general abdominal and thoracic surgical
procedures for resection, transection, and creation of anastomoses.
X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR