SURGASSIST RIGHT ANGLE LINEAR CUTTER DLU 30MM, 45MM & 60MM - VASCULAR WITH RELOADS, MODELS RALC30V, RALC45V & RALC60V
K022313 · Power Medical Interventions, Inc. · GDW · Dec 9, 2002 · General, Plastic Surgery
Device Facts
Record ID
K022313
Device Name
SURGASSIST RIGHT ANGLE LINEAR CUTTER DLU 30MM, 45MM & 60MM - VASCULAR WITH RELOADS, MODELS RALC30V, RALC45V & RALC60V
Applicant
Power Medical Interventions, Inc.
Product Code
GDW · General, Plastic Surgery
Decision Date
Dec 9, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The SurgASSIST™ Right Angle Linear Cutter Digital Loading Unit™, 30 mm, 45 mm and 60 mm, Vascular with Reloads has applications for general and endoscopic surgery including multiple open or minimally invasive general, gynecological, urologic (including radical prostatectomy), thoracic, and pediatric surgical procedures for transection, resection, and/or creation of anastomoses. They can be used with staple line or tissue buttressing material such as bovine pericardium.
Device Story
SurgASSIST™ Right Angle Linear Cutter Digital Loading Unit™ (DLU) is a single-patient-use surgical stapler component. It attaches to a flexible shaft (FS) connected to a power console (PC). The DLU features a staple-forming anvil and a cutting blade. During operation, the PC provides mechanical power to drive the stapling and cutting mechanism. The surgeon controls the device via a handheld electronic Remote Control Unit (RCU). The DLU compresses tissue, sequentially forms staples, and simultaneously severs the tissue. Each DLU includes an integral electronic memory module for identification of reload type and size. The device is used in OR settings by surgeons to facilitate tissue anastomosis and resection. It benefits patients by providing a powered, steerable, and precise method for tissue transection and stapling in both open and endoscopic procedures.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Single-patient-use DLU with staple-forming anvil and cutting blade. Powered by external console via flexible shaft. Includes electronic memory module for reload identification. Sterilized via Ethylene Oxide (ETO). Dimensions: 30mm, 45mm, 60mm variants. Staple configuration: 6 rows, 2.4mm crown, 2.3mm leg, 0.20mm thickness.
Indications for Use
Indicated for patients undergoing general, gynecological, urologic (including radical prostatectomy), thoracic, or pediatric surgical procedures requiring transection, resection, or creation of anastomoses via open or minimally invasive techniques.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
SurgASSIST™ Right Angle Linear Cutter Digital Loading Unit™, 30 mm, 45 mm and 60 mm with Titanium Implantable Staples and Reloads (K021701)
ETS Compact-Flex45 Articulating Endoscopic Linear Cutter, Vascular/Thin (K002398)
Submission Summary (Full Text)
{0}------------------------------------------------
KO22313 page
Power Medical Interventions, Inc. Fower Medical Interventions, Inc.
SurgASSIST™ Right Angle Linear Cutter DLUs with Reloads, 30 mm, 45 mm, 60 mm Vascular Special 510(k) Device Modification PreMarket Notification, July 16, 2002
DEC 0 9 2002
# Special 510(k) Device Modification PREMARKET NOTIFICATION SAFETY AND EFFECTIVENESS SUMMARY
SurgASSIST™ Right Angle Linear Cutter Digital Loading Unit™ 30mm, 45mm, 60mm - Vascular with Reloads
In Accordance with 21 CFR Section 807.92 Power Medical Interventions, Inc., is submitting the following Safety and Effectiveness summary.
1) Submitter Information:
Power Medical Interventions, Inc. 110 Union Square Drive New Hope, PA 18938 267-775-8100 Ph 267-775-8123 Fax
Applicant:
Barbara J. Whitman
July 12, 2002 Date of Notification:
- 2) Name of Device:
| Trade Name: | SurgASSIST™<br>Right Angle Linear Cutter DLU<br>30 mm, 45 mm and 60 mm - Vascular<br>with Reloads |
|--------------|---------------------------------------------------------------------------------------------------|
| Common Name: | Linear Cutter with Implantable Staples<br>and Reloads |
Classification Name: Staple, Implantable, GDW
{1}------------------------------------------------
3) Predicate Devices:
- a) SurgASSIST™ Right Angle Linear Cutter Digital Loading Unit™, 30 mm. 45 mm and 60 mm with Titanium Implantable Staples and Reloads, Power Medical Interventions, Inc., New Hope, PA. REF RALC30, RALC45, RALC60, RALCR30, RALCR45, RALCR60 (K021701).
- b) ETS Compact-Flex45 Articulating Endoscopic Linear Cutter, Vascular/Thin. Ethicon Endo-Surgery, Inc., Cincinnati, Ohio. REF SCW45 (K002398).
- 4) Device Description:
The devices described here are Right Angle Linear Cutter Digital Loading Units™(DLUs), 30 mm, 45 mm and 60 mm, Vascular, with Reloads for single patient use. All have a maximum diameter of 3.1". Both DLUs and Reloads are supplied pre-sterilized and ready for use upon removal from their packaging.
The 30 mm Vascular DLU can only be used with the 30 mm Vascular Reload.
The 45 mm Vascular DLU can only be used with the 45 mm Vascular Reload.
The 60 mm Vascular DLU can only be used with the 60 mm Vascular Reload.
The DLU contains a staple-forming anvil. The anvil acts with the staple cartridge to compress and position layers of tissue in readiness for stapling and cutting. At the same time, the anvil provides support and a means for correctly forming staples while they are closed sequentially along the tissue, followed by the cutting blade. The Right Angle Linear Cutter Vascular DLU is perpendicular to the Flex Shaft, forming an extension to the Flex Shaft, to which they are connected.
A loaded DLU is used to anastomose tubular structures by applying staples through the tissue and forming the staples to a controlled closed condition to secure the layers of tissue together. It also severs the tubular structure.
Right Angle Linear Cutter Vascular Reloads contain staples, a cutting blade. and the means to simultaneously force staples toward the anvil. The cutting blade is advanced in conjunction with the staple pushers so that tissue is simultaneously stapled and cut.
The DLUs are attached to the end of the FlexShaft (FS), which contains a pair of flexible rotary drive shafts within an overall flexible shaft. The other end of the FS is connected to the Power Console (PC), which applies mechanical power to the drive shafts. DLUs have all functions powered by the PC. The
{2}------------------------------------------------
K022313 page
3/6
FS has a short steerable section at the distal end (near the attached DLU) so that the angle of attack (attitude) of the DLU can be adjusted by the surgeon to optimize patient accessibility.
> The surgeon operates a DLU via a hand held electronic Remote Control Unit (RCU).
> The DLUs have quick attach and release means for coupling to the FS. No tools are required. DLUs are pushed onto the FS end, snapping and locking into place. To remove a DLU from the FS, a sleeve on the DLU at the junction with the FS is rotated by hand.
DLU designs shall allow for attachment of their corresponding Reloads, but shall inhibit attachment of incompatible Reloads. Each Reload has an integral electronic memory module. This identifies the type and size of the Reload being used.
- 5) Indications For Use -
The SurgASSIST™ Right Angle Linear Cutter Digital Loading Unit™, 30 mm, 45 mm and 60 mm, Vascular with Reloads has applications for general and endoscopic surgery including multiple open or minimally invasive general, gynecological, urologic (including radical prostatectomy), thoracic, and pediatric surgical procedures for transection, resection, and/or creation of anastomoses. They can be used with staple line or tissue buttressing material such as bovine pericardium.
- 6) Comparison to Predicate Devices
The following table compares the subject Right Angle Linear Cutter DLU, 30 mm. 45 mm and 60mm, Vascular, with Reloads to the previously cleared predicate Right Angle Linear Cutter DLU (K012809) device and the Ethicon ETS Compact-Flex45 Articulating Endoscopic Linear Cutter, Vascular/Thin.
{3}------------------------------------------------
Power Medical Interventions, Inc.
SurgASSIST™ Right Angle Linear Cutter DLUs with Reloads, 30 mm, 45 mm, 60 mm Vascal
Special 510(k) Device Modification PreMarket Notificatio
ight Angle Linear Cutter DLU Product Features Comparison Ch
| Features & Description | Name | Manufacturer of Record | Contract Manufacturer | 510(k) Clearance Numbers | Product Codes | Predicate | Intended use |
|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|-----------------------------------|---------------------------------------|------------------------------|-----------------------------------------------------------|---------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SurgASSISTTM<br>Right Angle Linear Cutter DLU<br>30mm, 45mm, 60mm - Vascular<br>with Reloads | SurgASSISTTM<br>Right Angle Linear Cutter DLU<br>Vascular<br>with Reloads | Power Medical Interventions, Inc. | Lacey Manufacturing<br>Bridgeport, CT | Subject of this Notification | RALC30V, RALC45V, RALC60V<br>RALCR30V, RALCR45V, RALCR60V | Ethicon ETS Compact-Flex45<br>Articulating Endoscopic Linear<br>Cutter, Vascular/Thin | Has applications for general and<br>endoscopic surgery including<br>multiple open or minimally invasive<br>general, gynecological, urologic<br>(including radical prostatectomy),<br>thoracic and pediatric surgical<br>procedures for transection,<br>resection, and/or creation of<br>anastomoses. They can be used<br>with staple line or tissue buttressing<br>material such as bovine<br>pericardium |
| SurgASSISTTM<br>Right Angle Linear Cutter DLU<br>30mm, 45mm, 60mm<br>with Reloads | SurgASSISTTM<br>Right Angle Linear Cutter DLU<br>with Reloads | Power Medical Interventions, Inc. | Lacey Manufacturing<br>Bridgeport, CT | K021701 | RALC30, RALC45, RALC60<br>RALCR30, RALCR45, RALCR60 | ETS Compact-Flex45 Articulating<br>Endoscopic Linear Cutter, Vascular/Thin | Has applications in gastrointestinal,<br>gynecological, general abdominal<br>and thoracic surgical procedures for<br>resection, transection, and creation<br>of anastomoses. |
| | | | | K002398 | | Ethicon Endo-Surgery, Inc. | Application in gastroenterology for<br>transection, resection, and/or<br>creation of anastomoses and can<br>be used in multiple open or<br>minimally invasive surgical<br>procedures, including radical<br>prostatectomy, and can be used<br>with staple line or tissue buttressing<br>materials, such as bovine<br>pericardium. |
| | | | | | | Ethicon Endo-Surgery, Inc. | |
| | | | | | | SCW45<br>TR45W | |
K022313 page 4/6
{4}------------------------------------------------
ower Medical Interventions, Inc.
urgASSIST™ Right Angle Linear Cutter DLUs with Reloads, 30 mm, 45 mm, 60 mm Vasca
pesial 510(K) Device Modification PreMarket Notification, J
# ight Angle Linear Cutter DLU Product Features Comparison Chris
| Features & Description | FDA Class (System) | | Physical Characteristics | Predicate |
|------------------------------------------------------------------------------------------------|--------------------|--|---------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| SurgASSISTTM<br>Right Angle Linear Cutter DLU<br>30 mm, 45 mm, 60mm - Vascular<br>with Reloads | II | | | Ethicon ETS Compact-<br>Flex45 Articulating<br>Endoscopic Linear Cutter,<br>Vascular/Thin |
| SurgASSISTTM<br>Right Angle Linear Cutter DLU<br>with Reloads<br>30 mm, 45 mm, 60 mm | II | | | II |
| | | | 30mm Vascular- 46 staples<br>45mm Vascular - 69 staples<br>60mm Vascular - 92 staples | 30mm - 22 staples |
| | | | Number of Staples | 45mm - 32 staples |
| | | | | 60mm - 46 staples |
| | | | 30mm Vascular -6 rows<br>45mm Vascular - 6 rows<br>60mm Vascular - 6 rows | 30mm- 4 rows |
| | | | Rows of Staples | 45mm- 4 rows |
| | | | | 60mm- 4 rows |
| | | | 30mm Vascular - 2.4 mm<br>45mm Vascular - 2.4 mm<br>60mm Vascular - 2.4 mm | 30mm- 4.0 mm |
| | | | Staple Crown Dimension | 45mm- 4.0 mm |
| | | | | 60mm- 4.0 mm |
| | | | 30mm Vascular - 2.3 mm<br>45mm Vascular - 2.3 mm<br>60mm Vascular - 2.3 mm | 30mm- 4.4 mm |
| | | | Staple Leg Dimension | 45mm- 4.4 mm |
| | | | | 60mm- 4.4 mm |
| | | | 30mm Vascular - 0.20 mm<br>45mm Vascular - 0.20 mm<br>60mm Vascular - 0.20 mm | 30mm- 0.23mm |
| | | | Staple Thickness | 45mm- 0.23mm |
| | | | | 60mm- 0.23mm |
| | | | 30mm Vascular - 1.0 mm<br>45mm Vascular - 1.0 mm<br>60mm Vascular - 1.0 mm | 30mm- 1.2 / 2.0mm |
| | | | Staple Closed Range | 45mm- 1.2 / 2.0mm |
| | | | | 60mm- 1.2 / 2.0mm |
| | | | None | None |
| | | | DLU Internal Power | |
| | | | Memory module containing digital data<br>for identification, etc. | Memory module containing digital data for<br>identification, etc. |
| | | | Digital Information | |
| | | | Sterile - Single Patient Use | Sterile - Single Patient Use |
| | | | How Supplied | |
| | | | | 66 |
| | | | | 6 rows |
| | | | | 2.7 mm |
| | | | | 2.5 mm |
| | | | | 0.20 mm |
| | | | | 1.0 mm |
| | | | | None |
| | | | | Sterile - Single Patient Use |
{5}------------------------------------------------
Power Medical Interventions, Inc.
SurgASSIST™ Right Angle Linear Cutter DLUs with Reloads, 30 mm, 45 mm, 60 mm Vascul
Special 510(k) Device Modification PreMarket Notificatio
ght Angle Linear Cutter DLU Product Features Comparison Cha
| | AND AND & BLAND |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | . |
| | |
| | |
| | |
| | |
| Features & Description | Method of Sterilization | Packaging | |
|--------------------------------------------------------------------------------------------------------|--------------------------|-----------------------------|-----------------------------|
| SurgASSIST™<br>Right Angle Linear Cutter DLU<br>30 mm, 45 mm, 60 mm – Vascular<br>with Reloads | Ethylene Oxide Gas (ETO) | Blister Tray with Tyvek Lid | Blister Tray with Tyvek Lid |
| SurgASSIST™<br>Right Angle Linear Cutter DLU<br>with Reloads<br>30 mm, 45 mm, 60 mm | Ethylene Oxide Gas (ETO) | Blister Tray with Tyvek Lid | Blister Tray with Tyvek Lid |
| Predicate<br>Ethicon ETS Compact-<br>Flex45 Articulating<br>Endoscopic Linear Cutter,<br>Vascular/Thin | Irradiation | Blister Tray with Tyvek Lid | Blister Tray with Tyvek Lid |
000000
{6}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of three curved lines that resemble a human figure.
### Public Health Service
DEC 0 9 2002
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Power Medical Interventions, Inc. Barbara J. Whitman Regulatory Affairs Manager 110 Union Square Drive New Hope, Pennsylvania 18938-1364
Re: K022313
Trade/Device Name: SurgASSIST™ Right Angle Linear Cutter DLU™ 30mm, 45mm & 60mm- Vascular with Reloads
Regulation Number: 878.4750 Regulation Name: Implantable staple Regulatory Class: Class II Product Code: GDW Dated: October 30, 2002 Received: October 31, 2002
Dear Ms. Whitman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the
{7}------------------------------------------------
Page 2 - Ms. Barbara J. Whitman
quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product. radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Miriam C. Provost
برار Celia M. Witten, Ph.D., MD Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Power Medical Interventions, Inc. New Hope, PA 18938
510(k) No. K022313
Device Name:
SuraASSIST™ Right Angle Linear Cutter Digital Loading Unit™ 30mm, 45mm, 60mm - Vascular with Reloads
## INDICATIONS FOR USE:
The SuraASSIST™ Right Angle Linear Cutter Digital Loading Unit™, 30 mm, 45 mm and 60 mm, Vascular with Reloads has applications for general and endoscopic surgery including multiple open or minimally invasive general, gynecological, urologic (including radical prostatectomy), thoracic, and pediatric surgical procedures for transection, resection, and/or creation of anastomoses. They can be used with staple line or tissue buttressing material such as bovine pericardium.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Prescription Use Per 21CFR §801.109
OR Over-The-Counter Use
Miriam C. Provost
vision of General, Restorative Jeurological Devices
Ko 22313
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.