NANOPIA WIDE RANGE C-REACTIVE PROTEIN (CRP) REAGENT KIT

K052591 · Clinical Data, Inc. · DCK · Feb 9, 2006 · Immunology

Device Facts

Record IDK052591
Device NameNANOPIA WIDE RANGE C-REACTIVE PROTEIN (CRP) REAGENT KIT
ApplicantClinical Data, Inc.
Product CodeDCK · Immunology
Decision DateFeb 9, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5270
Device ClassClass 2

Intended Use

The intended use of the Nanopia Wide Range CRP Reagent is for the in vitro quantitative determination of C-Reactive Protein in human serum and plasma. The assay is intended for use in the evaluation of infection, tissue injury and inflammatory disorders in combination with a complete clinical evaluation.

Device Story

Nanopia Wide Range CRP Reagent is a turbidimetric immunoassay for quantitative measurement of C-Reactive Protein (CRP) in serum or plasma. Device consists of two liquid reagents: buffer solution and latex beads coated with mouse monoclonal anti-human CRP antibodies. Used on clinical chemistry analyzers (e.g., Roche Hitachi 917) in clinical laboratory settings. Principle of operation: sample mixed with buffer and antibody-coated beads; CRP binds beads causing agglutination; light scattering measured is proportional to CRP concentration. Output is CRP concentration (mg/L). Results assist clinicians in evaluating inflammatory lesions and monitoring treatment. Increases in CRP are non-specific; must be interpreted with complete clinical history.

Clinical Evidence

Bench testing only. Precision evaluated per CLSI EP5-A (within-run %CV 0.61–1.83%). Linearity established 0.1–400 mg/L. Limit of quantitation 0.024 mg/L. Interference testing showed <3% change with bilirubin, hemoglobin, lipids, ascorbic acid, and rheumatoid factor. Method comparison with predicate (N=98) yielded R2=0.9992; low-range correlation (N=71) yielded R2=0.9981. Matrix comparison (serum vs plasma) showed high recovery (96.4–105.7%).

Technological Characteristics

Turbidimetric immunoassay; latex beads coated with mouse monoclonal anti-human CRP antibodies. Two-reagent liquid system. Measuring range 0.1-400 mg/L. Calibrated with 5-level calibrator set traceable to CRM470. Designed for use on automated clinical chemistry analyzers. Performance standards referenced: NCCLS EP9-A, EP5-A, EP6-A, EP7-A.

Indications for Use

Indicated for quantitative measurement of C-Reactive Protein (CRP) in serum or plasma to evaluate infection, tissue injury, and inflammatory disorders in conjunction with clinical history. For prescription use.

Regulatory Classification

Identification

A C-reactive protein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the C-reactive protein in serum and other body fluids. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Clinical Data, Inc. USA: One Gateway Center, Suite 415 Newton, MA 02158 1-617-527-9933, Ext. 22 Fax: 1-617-527-8230 ## 510(k) Summary ### Nanopia Wide Range CRP Koszsql The assigned 510(k) Number is:________________________________________________________________________________________________________________________________________________ - A. Analyte: C-Reactive Protein (CRP) - B. Type of Test: Turbidimetric immunoassay - C. Applicant: One Gateway Center, Suite 415 Newton, MA 02158 Phone: 617-527-9933, Ext. 41 Fax: 617-527-8230 #### D. Proprietary and Established Names: Nanopia Wide Range CRP Reagent Nanopia Wide Range CRP Calibrator #### E. Regulatory Information: - 1. Regulation section: 21 CFR § 866.5270, C-reactive protein immunological test system 21 CFR § 862.1150, Calibrator - 2. Classification: Class II - 3. Product Code: DCK, C-reactive protein, antigen, antiserum, and control JIS, Calibrator, primary - 4. Panel: Immunology (82) {1}------------------------------------------------ #### F. Intended Use: - 1. Intended use(s): The Nanopia Wide Range CRP Reagent is intended for the quantitative measurement of C-Reactive Protein (CRP) concentration in serum or plasma. Measurement of CRP is useful for determining the existence of inflammatory lesions and to monitor treatment. The Nanopia Wide Range CRP Calibrator is intended for the calibration of the Nanopia Wide Range CRP assay. - 2. Indication(s) for use: See Intended Use section. - 3. Special condition for use statement(s): For in vitro diagnostic use. Increases in CRP values are non-specific and should not be interpreted without a complete clinical history. - 4. Special instrument Reguirements: Clinical chemistry analyzers (testing performed on Roche Hitachi 917 analyzer) #### G. Device Description: The Nanopia Wide Range CRP assay consists of two liquid reagents. Reagent 1 is a buffering solution and Reagent 2 contains latex beads coated with mouse monoclonal anti-human CRP antibodies. The assay is for use on general clinical chemistry analyzers. #### H. Substantial Equivalence Information: - 1. Predicate device name(s): N-Geneous Wide Range CRP Reagent. - 2. Predicate K number(s): K040241 {2}------------------------------------------------ #### 3. Comparison with predicate: | Item | Device | Predicate | |--------------------------|-------------------------------------------------------|-------------------------------------------------------| | Intended Use | Quantitative measurement<br>of CRP in serum or plasma | Quantitative measurement<br>of CRP in serum or plasma | | Sample Matrix | Serum or plasma | Serum or plasma | | Antibody | Mouse monoclonal anti-<br>human CRP | Mouse monoclonal anti-<br>human CRP | | Assay Range | 0.1 to 400 mg/L | 0.04 to 320 mg/L | | Antibody substrate | Latex | Latex | | Number of<br>calibrators | 5 | 5 | The N-Geneous Wide Range CRP Reagent and the Nanopia Wide Range CRP are essentially the same product both produced by Daiichi Pure Chemicals Co., Ltd. for distribution by the respective companies_ # I. Standard/Guidance Document Referenced (if applicable): | NCCLS Guideline EP9-A - Method Comparison and Bias Estimation Using Patient<br>Samples | |----------------------------------------------------------------------------------------------| | NCCLS Guideline EP5-A - Evaluation of Precision Performance of Clinical Chemistry<br>Devices | | NCCLS Guideline EP6-A - Evaluation of the Linearity of Quantitative Analytical<br>Methods | | NCCLS Guideline EP7-A - Interference Testing in Clinical Chemistry | #### J. Test Principle: Sample is mixed with the buffer solution and the anti-CRP antibody-coated beads. CRP in the sample binds the antibody-coated beads and agglutinates. The light scattering caused by an increase in particle size is measured. The amount of light scattering is proportional to the concentration of CRP in the sample. {3}------------------------------------------------ #### K. Performance Characteristics (if/when applicable): - 1. Analytical performance: - a. Precision/Reproducibility: Device imprecision was evaluated according to NCCLS EP5-A. Spiked serum samples were run in triplicate daily for 20 days (units = mg/L). | Control | Mean | Within Run | | Total | | |---------|-------|------------|------|-------|------| | | | SD | % CV | SD | % CV | | 1 | 0.886 | 0.16 | 1.83 | 0.022 | 2.46 | | 2 | 6.71 | 0.05 | 0.76 | 0.089 | 1.31 | | 3 | 38.88 | 0.24 | 0.61 | 0.430 | 1.11 | #### b. Linearity/assay reportable range: The useable range of this device is 0.10 - 400 mg/L (the Limit of Quantification to the upper end of the linear range). Linearity was assessed using a sample with CRP values of 400 mg/L. Using a serum pool free of CRP dilutions were prepared to create samples with theoretical concentrations from 40 to 400 mg/L in increments of 40 mg/L. A sample from a serum pool free from CRP was included for a zero concentration. These samples were run in duplicate and the actual mean values were compared to the theoretical values. The % recovery from 0 to 400 mg/L ranged from 97.5 to 102.7%. Results are summarized below (units = mg/L). | | (mg/L) | | | | |--------------------------|-------------------|--------|--------|------------| | Theoretical<br>CRP value | measurement value | | | % Recovery | | | 1 | 2 | mean | | | 0 | 0.00 | 0.00 | 0.00 | --- | | 40 | 39.33 | 40.20 | 39.77 | 99.4 | | 80 | 79.01 | 78.42 | 78.72 | 98.4 | | 120 | 119.26 | 116.14 | 117.70 | 98.1 | | 160 | 155.44 | 156.48 | 155.96 | 97.5 | | 200 | 196.46 | 194.97 | 195.72 | 97.9 | | 240 | 236.89 | 236.10 | 236.50 | 98.5 | | 280 | 280.76 | 278.09 | 279.43 | 99.8 | | 320 | 322.15 | 323.21 | 322.68 | 100.8 | | 360 | 374.30 | 365.09 | 369.70 | 102.7 | | 400 | 401.12 | 406.17 | 403.65 | 100.9 | The acceptable criteria was 100% ± 5% The results above indicate that the assay is linear across the measuring range of the assay {4}------------------------------------------------ #### c. Traceability (controls, calibrators, or method): The device is calibrated by a set of 5 calibrators (sold separately) with targeted CRP concentrations of 3, 6, 30, 180 and 360 mg/L. The calibrator is traceable to CRM470. A master calibrator is prepared and value assigned by multiple measurements of multiple lots using the device. The calibrators are traceable to this master calibrator and are value assigned using the device. Stability testing is described. #### d. Detection limit: Analytical sensitivity, defined as the lowest concentration at which the assay performs with + 2 SD as assessed by diluting a serum sample with a concentration of 0.5 mg/L to 0.0 mg/L and measuring the mAbs in replicates of 10. By this method, the functional sensitivity of this assay is said to be 0.10 mg/L (derived by extrapolation of plot of theoretical concentration vs. recovered concentration. - e. Analytical specificity: Interference of endogenous substances was evaluated by testing for the effect of hemoglobin (up to 500 mg/dL), ascorbic acid (up to 100 mg/dL), free bilirubin (up to 50 mg/dL), conjugated bilirubin (up to 50 mg/dl, lipid emulsion (up to 5000 turbidity), Rheumatoid factor (up to 500 IU/mL), and ascorbic acid (up to 100 mg/dL) on serum samples containing a nominal concentration of 3.7 mg/L. No interference was noted on any of the above. Interference was defined as a result +/- 5 % of the control. This information is in the package insert. - f. Assay cut-off: Not applicable - 2. Comparison studies: - a. Method comparison with predicate device: Serum samples (n=98) ranging from 0.0 to 293 mg/L were measured using the Nanopia method and comparing those measurement to those made with the predicate method. Results are summarized below. Serum Samples: Nanopia = 1.015(Predicate) - 0.0349; R2 = 0.9992 - b. Matrix comparison: {5}------------------------------------------------ Paired samples of serum and EDTA plasma were collected concurrently from 50 individuals and measured using the Nanopia method. Results are summarized below: Plasma = 0.994(Serum) + 0.03; r = 0.999 Additional serum pools were spiked with 110 mM of sodium citrate, 100 mM sodium oxalate, 25 mM EDTA, and 100 U/mL of sodium heparin and assayed. There were no significant differences observed between sample types. - 3. Clinical studies: - a. Clinical sensitivity: Not applicable - b. Clinical specificity: Not applicable - c. Other clinical supportive data (when a and b are not applicable): Not applicable - 4. Clinical cut-off: Not applicable - 5. Expected values/Reference range: Less than 3.0 mg/L. #### L. Conclusion: Clinical Data concludes that the Nanopia Wide Range CRP assay has a similar intended use, a similar technological principle, and clinically acceptable performance comparable to similar devices currently in commercial distribution and is substantially equivalent to the predicate device. {6}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with three curved lines representing its wings or feathers. The bird is positioned above a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 2006 FEB 9 2006 Israel M. Stein, MD President Clinical Data, Inc. One Gateway Center Suite 411 Newton, MA 02146 k052591 Re: > Trade/Device Name: Nanopia Wide Range CRP Reagent Regulation Number: 21 CFR 866.5270 Regulation Name: C-reactive protein immunological test system Regulatory Class: Class II Product Code: DCK, JIS Dated: January 12, 2006 Received: January 13, 2006 Dear Dr. Stein: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {7}------------------------------------------------ Page 2 -- This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally president redicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the I ou may of annovaturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Alberto Gutierrez, Ph.D. Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ # Indications for Use 510(k) Number (if known): K052591 Nanopia Wide Range CRP Reagent Device Name: Indications For Use: The Nanopia Wide Range CRP Reagent is intended for the quantitative measurement of C-Reactive Protein (CRP) in serum or plasma. The assay is intended for use in the evaluation of infection, tissue injury, and inflammatory disorders in combination with a complete clinical evaluation. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Cinn Chapper Office of In Vitro Diagnostic Device Evaluation and Safety Page 1 of 1 FOSZSYI .
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...