The ETView Tracheoscopic Ventilation Tube (TVT) is intended for oral and nasal intubation. The ETView Tracheoscopic Ventilation Tube (TVT) System has the following specific indications: - Temporary artificial airway in adults requiring mechanical ventilation, for oral and nasal intubation. - Viewing during non-difficult and difficult intubation procedures, for verifying tube placement and repositioning, for viewing during suctioning and for general inspection of the airway.
Device Story
The TVT System is a visualized endotracheal tube used in clinical settings for airway management. It integrates a CMOS video camera and light source directly into the tube, allowing real-time visualization of the airway during intubation, suctioning, and tube placement verification. Unlike fiberoptic-based predecessors, the TVT uses digital imaging technology to provide higher resolution video to an external monitor. The device features two Murphy eyes and a dedicated lens-clearing lumen to maintain visibility. It is operated by clinicians to assist in both routine and difficult intubation procedures. By providing direct visual feedback, the device aids in accurate tube placement and repositioning, potentially reducing complications associated with blind intubation or the need for separate bronchoscopic equipment.
Clinical Evidence
No clinical data. Bench testing only. Performance testing included cuff resting diameter, resistance to tube collapse, resistance to cuff herniation, cuff symmetry, air flow resistance, thermal safety, imaging performance, cuff leak resistance, mechanical integrity (fatigue, burst, cuff compliance), efficacy of selective suctioning, and software validation.
Technological Characteristics
Visualized endotracheal tube with integrated CMOS camera and light source. Conforms to ISO 5361 for tracheal tubes and connectors. Features two Murphy eyes and a lens-clearing lumen. Radiopaque material construction. Sterilized via ethylene oxide per ANSI/AAMI/ISO 11135. System includes external monitor for video output.
Indications for Use
Indicated for adults requiring mechanical ventilation via oral or nasal intubation. Used for temporary artificial airway maintenance, visualization during intubation (non-difficult and difficult), verification of tube placement/repositioning, suctioning, and general airway inspection.
Regulatory Classification
Identification
A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.
Parker Medical Flex -Tip Endotracheal Tube (K984528)
Submission Summary (Full Text)
{0}------------------------------------------------
DEC 13 2005
K052233
Image /page/0/Picture/2 description: The image shows the logo for "ETView Endotracheal Management". The logo is black and white. The "ET" is in a bold, blocky font, and the "View" is in a cursive font with a circle of dots above the "i". "Endotracheal Management" is written in a smaller, sans-serif font to the right of "View".
## 510(k) Summary
| Applicant's Name | Yoram Levy<br>Consultant |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | ETView Ltd. |
| | Misgav Technological Center |
| | 6 Kahol Street, |
| | Teradion Industrial Park |
| | ISRAEL 20179 |
| | Telephone: (972)4-999-1996 |
| | Fax: (972)4-999-1901 |
| Contact Person: | Yoram Levy, QSite |
| | 31 Haavoda St. |
| | Binyamina, Israel 30500 |
| | Tel (972)4-638-8837; Fax (972)4-638-0510 |
| | Yoram@qsitemed.com |
| A. Trade Name: | Tracheoscopic Ventilation Tube (TVT) System |
| B. Manufacturer: | ETView Ltd. |
| | Misgav Technological Center |
| | 6 Kahol Street, |
| | Teradion Industrial Park |
| | ISRAEL 20179 |
| | Telephone: (972)4-999-1996 |
| | Fax: (972)4-999-1901 |
| C. Classification: | Classification name: (1) Tracheal Tube, adult, cuffed<br>(2) Bronchoscope, flexible |
| | Common/usual name: (1) Tracheal Tube (or Endotracheal Tube) with<br>(2) image transmission, (3) integral video imager (4) external monitor, (5) Integral light source |
| | Class II device |
| | Product Code: BTR |
| | Regulation No.: 868.5730 |
| D. Contract Sterilizer: | Mediplast Israel Ltd. |
| | 7 Hayarkon St., |
| | P.O.B. 13214 Yavne 81227 |
9-2
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for "ETView Endotracheal Management". The logo is black and white and features the letters "ET" in a stylized font, with a graphic of a circle with smaller circles inside it. The words "Endotracheal Management" are written in a smaller font to the right of the logo.
Tel (972)8-932-7766 Fax (972)8-932-7992
## E. Reason for the Premarket Notification Submission:
#### New device
ETV ev intends to market the Tracheoscopic Ventilation Tube (TVT) System as an endotracheal tube that enables intubation and at the same time viewing of the airways during non-difficult and difficult intubation procedures, for verifying tube placement and repositioning, for viewing during suctioning and for general inspection of the airway. The TVT System provides higher video resolution and thus accuracy compared to the existing cleared device.
#### F. Identification of Legally Marketed Predicate Devices:
Pulmonx Visualized Endotracheal Tube (VETT) (K973191) Parker Medical Flex -Tip Endotracheal Tube (K984528)
#### G. Performance Standards or Special Controls:
The ETView TVT conforms to ISO 5361 (Dated December 15, 1999) and to Anaesthetic and Respiratory Equipment - Tracheal Tubes and Connectors. This standard replaces ASTM F 1242-96, "Standard Specification for Cuffed and Uncuffed Tracheal Tubes for the same devices, except where stated
The TVT conforms to ANSI/AAMI/ISO 11135 (Medical devices - Validation and routine control of ethylene oxide sterilization).
#### H. Indications for Use:
The ETView Tracheoscopic Ventilation Tube (TVT) is intended for oral and nasal intubation.
The ETView Tracheoscopic Ventilation Tube (TVT) System has the following specific indications:
- Temporary artificial airway in adults requiring mechanical ventilation, for oral and . nasal intubation.
- Viewing during non-difficult and difficult intubation procedures, for verifying tube . placement and repositioning, for viewing during suctioning and for general inspection of the airway.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for "ETview Endotracheal Management". The logo is black and white and features the text "ETview" in a stylized font. To the right of the text is the phrase "Endotracheal Management" in a smaller, simpler font. The logo is likely used to represent a company or product related to endotracheal management.
## I. Technological Characteristics:
The ETView Tracheoscopic Ventilation Tube (TVT) System is a visualized endotracheal tube The ETV10 Pacelleosooped Pulmonx VETT System (K973191), except that the ETView System same as tiny CMOS video camera and light source and the Pulmonx device uses fiberoptic components.
#### J. Performance Testing
Bench testing demonstrated that the Tracheoscopic Ventilation Tube (TVT) System is at least as Donor loomig demonstrations and Pulmonx VETT System for oral and nasal intubation and in providing visualization of the airway.
The following performance tests have been done and the results are attached to this submission: Determination of Cuff Resting Diameter, Resistance to Tube Collapse, Resistance to Cuff Herniation, Cuff Symmetry, Air Flow Resistance, Thermal Safety, Imaging Performance, Cuff Leak Resistance Integrity, Mechanical Integrity ( Fatigue, Burst, and Cuff Compliance as part of the Mechanical Integrity test), Efficacy of Selective Suctioning, Software Validation.
#### K. Comparison to the Predicate Device
The ETView Tracheoscopic Ventilation Tube (TVT) System has the same intended use, general and specific indications and principles of operation as the cleared Pulmonx Visualized Endotracheal Tube (VETT) System.
In addition, the TVT employs similar technology as the cleared VETT except for the different type of camera and illumination.
The TVT has two Murphy eyes like the Parker Medical Flex -Tip Endotracheal Tube (K984528), and an additional lens clearing lumen. The inherent radiopacity of the TVT obviates the need for a radiopaque strip.
The minor differences between the TVT and the VETT do not raise any new questions of safety or efficacy. Moreover, bench testing of the TVT System (bench testing is provided in Attachments 6B, and 6C) demonstrated that the TVT is as safe and effective as the VETT. Thus, the ETView TVT System is substantially equivalent to the already cleared VETT System.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo features a stylized eagle with three lines representing its wings and head. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 3 2005
ETView Limited Mr. Yoram Levy QSite 31 Haavoda Street Binyamina, ISRAEL 30500
Re: K052233
Trade/Device Name: Tracheoscopic Ventilation Tube (TVT) Regulation Number: 21 CFR 868.5730 Regulation Name: Tracheal Tube Regulatory Class: II Product Code: BTR Dated: November 28, 2005 Received: December 7, 2005
Dear Mr. Levy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Mr. Levy
Please be advised that FDA's issuance of a substantial equivalence determination does not Please be advised that FDA Sissuance of a sabstanter and busether Fodarel agencies mean that FDA has made a decemmance in administered by other Federal agencies.
of the Act or any Federal statutes and regulations and regulations negistration of the Act of ally rederal statutes and regirements, including, but not limited to: registration
You must comply with all the Act's requirements, including and recurse You must comply with an the Act 3 requirements (100); good manufacturing practice
and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing (21 CFR Part 8 and listing (21 CFR Part 807), laocing (21 OFF Parvals on (21 CFR Patt 820); and if
requirements as set forth in the quality systems (QS) regulation (21 CFR Patt 820); and if requirements as set form in the quality systems (QD) regained (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k)
the success of the consisted on the first of the spinsions of your device to a I his letter will anow you to begin maliceant 5 your device of your device to a
premarket notification. The FDA finding of substantial equivalence and thus, pa premarket notification. The PDA miding of substants of Submitted on the spermits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Patt 801), If you desire specific advice for your device of our our and on please note the regulation please contact the Office or Comphanes an (210) - 11 CFR Part 807.97). You
entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the may obtain of Small Manufacturers, International and Consumer Assistance at its toll-free DIVIsion of Simall Manufactar (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sylte y. Michael Oms.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for ETView Endotracheal Management. The logo is in black and white and features the letters "ET" in a stylized font. To the right of the letters is the word "View" with a graphic of a circle with dots inside. The words "Endotracheal Management" are to the right of the logo.
# 510(k) Number (if known): イインS223
Tracheoscopic Ventilation Tube (TVT) Device Name: for . Indications The ETView Tracheoscopic Ventilation Tube (TVT) is intended Use: for intubation procedures. The ETView Tracheoscopic Ventilation Tube (TVT) is indicated for use as a temporary artificial airway in adults requiring mechanical ventilation. It is intended for oral and nasal intubations. The TVT System is indicated for viewing during non-difficult and difficult intubation procedures, for verifying tube placement and repositioning, for viewing during suctioning and for general inspection of the airway.
AND/OR Use X ________________________________________________________________________________________________________________________________________________________________________ Prescription (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-off) Division of Cardiovascular, Respiratory and Neurological Devices 510(k) Number
ાયન mon dign-Of ) trian of Aheathesicingy, General Hospital, John Control, Dantal Devices - Nomber
10-2
ETView Ltd., Misgav Technological Center, Teradion Industrial Zone, 20179 israel
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.