ETVIEW TRACHEOSCOPIC VENTILATION TUBE (TVT(TM))

K121028 · Etview , Ltd. · BTR · Jun 21, 2012 · Anesthesiology

Device Facts

Record IDK121028
Device NameETVIEW TRACHEOSCOPIC VENTILATION TUBE (TVT(TM))
ApplicantEtview , Ltd.
Product CodeBTR · Anesthesiology
Decision DateJun 21, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5730
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ETView Tracheoscopic Ventilation Tube (TVT™) is intended for intubation procedures. The ETView Tracheoscopic Ventilation Tube (TVTM) is indicated for use as a temporary artificial airway in adults requiring mechanical ventilation. It is intended for oral and nasal intubations. The TVT" System is indicated for viewing during non-difficult and difficult intubation procedures, for verifying endotracheal tube and endobronchial blocker placement and repositioning, for viewing during suctioning and for general inspection of the airway.

Device Story

Single-use, cuffed endotracheal tube (ETT) with embedded video imaging device in dedicated lumen. Provides real-time tracheal video feed to external monitor while tube is in situ. Used by clinicians during intubation, airway management, and suctioning. Facilitates visual confirmation of tube/blocker placement and general airway inspection. Benefits include improved visualization during intubation procedures and continuous monitoring of airway status.

Clinical Evidence

Bench testing only. Compliance with ISO 5361:1999, ANSI/AAMI/ISO 11135-1:2007, ISO 14971-1:2007, and ISO 10993-1:2003(E) demonstrated safety and effectiveness equivalent to the predicate device.

Technological Characteristics

Single-use cuffed tracheal tube; embedded video imaging sensor in dedicated lumen; video output to external monitor. Materials biocompatible per ISO 10993-1. Sterilization via ethylene oxide (ANSI/AAMI/ISO 11135-1:2007). Complies with ISO 5361:1999 for tracheal tubes.

Indications for Use

Indicated for adults requiring mechanical ventilation as a temporary artificial airway via oral or nasal intubation. Used for visualization during intubation (difficult or non-difficult), verification of tube/blocker placement, suctioning, and airway inspection.

Regulatory Classification

Identification

A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### K 121028 510(K) SUMMARY # ETVIEW TRACHEOSCOPIC VENTILATION TUBE SYSTEM (TVT™) # 510(k) Number K Applicant's Name: ETView Ltd. 17 Tchelet Street Misgav Business Park M.P Misgav 20174 Israel Tel: (972)72-260-7063 Fax: (972)72-260-7263 Contact Person: Yoram Levy, Qsite 31 Haavoda St. Binyamina, Israel 30500 Tel (972)4-638-8837 Fax (972)4-638-0510 · Yoram@gsitemed.com Trade Name: ETView Tracheoscopic Ventilation Tube System (TVTM) Device Type: Tracheal tube Preparation Date: April 01, 2012 Classification and Classification Name: Name: Tracheal tube Product Code: BTR Regulation No: 21 868.5730 Class: II Panel: Anesthesiology ## Indications for Use Statement: The ETView Tracheoscopic Ventilation Tube (TVT™) is intended for intubation procedures. The ETView Tracheoscopic Ventilation Tube (TVTM) is indicated for use as a temporary 2-2 ETView. | TVT™ 510k Notification {1}------------------------------------------------ artificial airway in adults requiring mechanical ventilation. It is intended for oral and nasal intubations. The TVT" System is indicated for viewing during non-difficult and difficult intubation procedures, for verifying endotracheal tube and endobronchial blocker placement and repositioning, for viewing during suctioning and for general inspection of the airway. ## Device Description: The ETView Tracheoscopic Ventilation Tube (TVT™) is a single use, cuffed device that functions as a standard endotracheal tube (ETT) and additionally has an embedded video imaging device embedded in a dedicated lumen. The system provides a video image of the patient's trachea, which is displayed on the monitor, for as long as the TVTTM is inside the patient's trachea. substantially equivalent to the following predicate devices: Predicate Device: | ETView Tracheoscopic Ventilation Tube (TVTTM) | K082420 | October 8,<br>2008 | |-----------------------------------------------|---------|--------------------| | Coopdech Endobronchial Blocker tube | K093888 | March 16.<br>2010 | #### Performance Standards ETView Tracheoscopic Ventilation Tube (TVT™) was tested and complies with the following standards: - ISO 5361:1999 Anesthetic and respiratory equipment -- Tracheal . tubes and connectors - ANSI/AAMI/ISO 11135-1:2007 Sterilization of health care . products - Ethylene oxide - ISO 14971-1:2007 Risk management for medical devices # 2-3 ETView. | TVT™ 510k Notification {2}------------------------------------------------ - ISO 10993-1:2003(E), Biological evaluation of medical devices -- Part 1: Evaluation and testing A detailed description appears in Section 14. ## Performance Testing Performance testing demonstrated that the ETView Tracheoscopic Ventilation Tube (TVT™) is as safe and effective as the cleared predicate device. The performance testing demonstrated that the minor addition to the indication for use is safe and effective. A detailed description appears in Section 14. # Technological Characteristics Both the proposed ETView Tracheoscopic Ventilation Tube (TVTTM) device and its predicate device (ETView Tracheoscopic Ventilation Tube k082420) function as a standard endotracheal tube (ETT) that additionally has an embedded video imaging device in a dedicated lumen. The system provides video image of patient's trachea that is displayed for as long as the ETT is inside the patient's trachea. # Biocompatibility: Materials of the ETView Tracheoscopic Ventilation Tube (TVTIM) system that are in contact with the human body, are identical to the predicate (K082420) and were tested and found to be biocompatible in accordance with ISO 10993-1. ### Conclusion: ETView Ltd. believes that, based on the information provided in this submission, the proposed ETView Tracheoscopic Ventilation Tube (TVTTM) System is substantially equivalent to its predicate device without raising any new safety and/or effectiveness concerns. 2-4 ETView. | TVT™ 510k Notification {3}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, a symbol often associated with medicine and healthcare, with three parallel lines forming the staff and a base resembling intertwined snakes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the caduceus. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 JUN 2 1 2012 Etview Limited C/O Mr. Yoram Levy Regulatory Consultant Osite 31 Haavoda Street Binyamina 30500 Israel Re: K121028 Trade/Device Name: ETView Tracheoscopic Ventilation Tube (TVT™) Regulation Number: 21 CFR 868.5730 Regulation Name: Tracheal Tube Regulatory Class: II Product Code: BTR Dated: April 1, 2012 Received: April 5, 2012 Dear Mr. Levy: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ # Page 2- Mr. Levy Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Nh foc Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # INDICATIONS FOR USE STATEMENT # 510(k) Number (if known): Device Name: ETView Tracheoscopic Ventilation Tube (TVTTM) Indications for Use: The ETView Tracheoscopic Ventilation Tube (TVTM) is intended for intubation procedures. The ETView Tracheoscopic Ventilation Tube (TVTM) is indicated for use as a temporary artificial airway in adults requiring mechanical ventilation. It is intended for oral and nasal intubations. The TVT 14 System is indicated for viewing during non-difficult and difficult intubation procedures, for verifying endotracheal tube and endobronchial blocker placement and repositioning, for viewing during suctioning and for general inspection of the airway. Prescription Use AND/OR X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-off) Division of Anesthesiology, General Hospital, Infection Control and Dental Devices 510(k) Number Schalk (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices K121028 ETView. | TVT™ 510k Notification 1-2 510(k) Number:
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...