FAMILY OF SELECTA OPHTHALMIC LASER SYSTEMS (SELECTA 1064, SELECTA SLT, SELECTA DUO), DELIVERY DEVICE AND ACCESSORIES
Applicant
Lumenis, Inc.
Product Code
HQF · Ophthalmic
Decision Date
Sep 30, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4390
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
photodisruption of ocular tissue using light energy emitted by a Nd:YAG Selecta 1064: laser, including discission of the posterior capsule of the eye (posterior capsulotomy), and discission of pupillary membranes (pupillary membranectomy) in aphakic and pseudophakic patients, and iridotomy/iridectomy selective laser trabeculoplasty Selecta SLT: photodisruption of ocular tissue using light energy emitted by a Nd: Y AG Selecta Duo: laser, including discission of the posterior capsule of the eye (posterior capsulotomy), and discission of pupillary membranes (pupillary membranectomy) in aphakic and pseudophakic patients, and iridotomy/iridectomy; and selective laser trabeculoplasty The Selecta 1064, Selecta SLT and Selecta Duo Ophthalmic Lasers are also intended for use as a diagnostic slit lamp, specifically, An AC-powered slit lamp biomicroscope intended for use in eye examination of the anterior segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affects the structural properties of the anterior segment LaserLink S: laser delivery system for use by an ophthalmologist in the treatment of ocular tissue; laser delivery system indicated for use for a variety of ophthalmic uses, including the indications specified in the laser operator manual. Refer to the laser operator manual, Indications for Use section.
Device Story
Family of Selecta ophthalmic laser systems (1064, SLT, Duo) and LaserLink S delivery device; flash lamp pumped, solid-state Nd:YAG laser. Inputs: physician-controlled energy settings via remote display; activation via footswitch or joystick. Operation: 1064nm pulses for photodisruption (capsulotomy, membranectomy, iridotomy); 532nm frequency-doubled pulses for SLT. Beam shaping optics generate plasma micro-pulses; twin aiming beams assist focal precision. LaserLink S allows coupling to existing 532nm photocoagulators. Used in clinical settings by ophthalmologists for ocular pathology treatment and anterior segment diagnosis. Output: laser energy delivery to ocular tissue; visual slit lamp examination. Benefits: precise, adjustable tissue photodisruption and trabeculoplasty; integrated diagnostic capability.
Clinical Evidence
No clinical data provided. Substantial equivalence based on comparison of technical characteristics, system/software hazard analysis, and software verification/validation.
Technological Characteristics
Flash lamp pumped, solid-state Nd:YAG laser; 1064nm (photodisruption) and 532nm (SLT) wavelengths. Q-switched pulse delivery (4 nsec). Adjustable energy (0.1-10mJ). Twin aiming beam system. Integrated slit lamp biomicroscope. LaserLink S delivery adaptor. AC-powered. Software-controlled energy delivery.
Indications for Use
Indicated for aphakic and pseudophakic patients requiring posterior capsulotomy, pupillary membranectomy, or iridotomy/iridectomy via Nd:YAG photodisruption; indicated for patients requiring selective laser trabeculoplasty (SLT); indicated for anterior segment eye examination as a diagnostic slit lamp.
Regulatory Classification
Identification
An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.
Coherent Popeye Ophthalmic Laser and Coherent Popeye Ophthalmic Laser in Combination with the Coherent Ultima 2000 Laser Photocoagulator Coherent Popeye Nd: YAG Ophthalmic Laser Systems (K973470)
Selecta 7000 Frequency Doubled Q-Switched Nd: Y AG Ophthalmic Laser (K004006)
{0}------------------------------------------------
K 051944
## SEP 3 0 2005 Attachment 10 510(k) Summary Statement
#### I. General Information
Submitter:
Lumenis, Inc. 2400 Condensa Street Santa Clara, California, U. S. A. 95051-0901
Contact Persons:
Karen L. Baker
Summary Preparation Date: July 15, 2005
#### II. Names
Device Names:
Family of Selecta Ophthalmic Laser Systems, Delivery Device and Accessories Selecta 1064 Ophthalmic Laser System Selecta SLT Ophthalmic Laser System Selecta Duo Ophthalmic Laser System LaserLink S
Primary Classification Name: 79, General and Plastic Surgery Panel
GEX, Laser powered surgical instrument
#### III. Predicate Devices
- Lumenis Selecta Duet (K021550) ●
- Coherent Popeye Ophthalmic Laser and Coherent Popeye Ophthalmic Laser in . Combination with the Coherent Ultima 2000 Laser Photocoagulator Coherent Popeye Nd: YAG Ophthalmic Laser Systems (K973470)
- t Selecta 7000 Frequency Doubled Q-Switched Nd: Y AG Ophthalmic Laser (K004006)
- LaserLink Z-1000 (K022181) .
#### IV. Product Description
The Lumenis Selecta is a fully integrated flash lamp pumped, solid state, Nd: YAG ophthalmic surgical laser system intended for use in the treatment of ocular pathology and for use as a diagnostic slit lamp.
The Family of Selecta Ophthalmic Lasers, Delivery Device and Accessories consists of the following models:
- I) Selecta 1064° a Nd: YAG laser providing Q-switched laser pulses at a wavelength of 1064nm for use in photodisruption of ocular tissue (posterior capsulotomy, pupillary membranectomy, iridotomy). The 1064nm treatment beam delivers a 4 nsec, 0.3-10mJ adjustable and selectable single, double or triple pulse of energy. It is
{1}------------------------------------------------
conditioned through beam shaping optics to generate a photodisruption micro pulse of plasma at a precision adjustable location relative to the visual focal plane (located at slit lamp center of rotation) and along the slit lamp objective lens axis. A twin aiming beam is also focused by the slit lamp objective to a converging 20um spot located at the focal point of the lens. The focal point of photodisruption is adjustable 0-350um in the posterior direction by the physician relative to this convergence of the twin aiming beams.
- 2) Selecta SLT® -- a Nd: YAG laser providing Q-switched frequency doubled pulses at a wavelength of 532nm for use in selective laser trabeculoplasty. The treatment beam delivers a 4nsec, 0.1-2mJ adjustable single pulse of energy. The aiming and treatment beams are coaxial with each other and focused by the slit lamp objective to a 400um spot at the focal point of the lens.
- 3) Selecta Duo a Nd:YAG laser providing Q-switched laser pulses at a wavelength of 1064 nanometers for use in photodisruption or Q-switched frequency doubled pulses at a wavelength of 532nm for use in selective laser trabeculoplasty, depending upon the mode selected. The Selecta Duo contains two aiming beam modules that produce a single beam for the 532nm mode and a dual beam for the 1064nm mode, respectively.
For each Selecta model, the physician controls delivery of laser energy from the Selecta remote control display unit and activates the treatment laser beam with a footswitch or joystick pushbutton. In addition, a laser slit lamp delivery adaptor, the LaserLink S, may be coupled to each of the above Selecta models and connected to a currently cleared Lumenis 532nm photocoagulator to allow the physician to use the Selecta slit lamp to deliver 532nm continuous wave laser energy for photocoagulation.
#### V. Indications for Use
photodisruption of ocular tissue using light energy emitted by a Nd: Y AG Selecta 1064: laser, including discission of the posterior capsule of the eye (posterior capsulotomy), and discission of pupillary membranes (pupillary membranectomy) in aphakic and pseudophakic patients, and iridotomy
selective laser trabeculoplasty Selecta SLT:
- photodisruption of ocular tissue using light energy emitted by a Nd: YAG Selecta Duo: laser, including discission of the posterior capsule of the eye (posterior capsulotomy), and discission of pupillary membranes (pupillary membranectomy) in aphakic and pseudophakic patients, and iridotomy; and
selective laser trabeculoplasty
The Selecta 1064, Selecta SLT and Selecta Duo Ophthalmic Lasers are also intended for use as a diagnostic slit lamp, specifically,
An AC-powered slit lamp biomicroscope intended for use in eye examination of the anterior segment, from the cornea epithelium to the posterior capsule. It is used to aid in
{2}------------------------------------------------
the diagnosis of diseases or trauma which affects the structural properties of the anterior segment
- LaserLink S: laser delivery system for use by an ophthalmologist in the treatment of ocular tissue;
laser delivery system indicated for use for a variety of ophthalmic uses, including the indications specified in the laser operator manual. Refer to the laser operator manual, Indications for Use section.
#### VI. Rationale for Substantial Equivalence
The Family of Selecta Ophthalmic Laser Systems, Delivery Device and Accessories have the same intended uses and indications for use as the predicate devices, and havens imilar technological characteristics as the predicate devices - treatment wavelength, laser media, mode of operation, energy output, pulse duration, pulse settings, spot sizes, aiming beam, cooling system, laser energy delivery control (footswitch/ handswitch), and delivery system - as the predicate devices, and therefore are substantially equivalent to the predicate devices referenced in Section III.
#### Performance Data VII.
System and software hazard analysis information, software verification and validation information, and clinical literature were submitted in conjunction with this Premarket Notification submission. The determination of substantial equivalence was based uppor the comparison of the technical characteristics between the Family of Selecta Ophthalmic Laser Systems, Delivery Device and Accessories and the predicate laser systems and the evaluation of the performance data.
#### VIII. Conclusion
The Family of Selecta Ophthalmic Laser Systems, Delivery Device and Accessories are substantially equivalent to similar predicate laser devices, delivery systems and accessories. The Family of Selecta Ophthalmic Laser Systems, Delivery Device and Accessories share the same intended use, indications for use, and technological characteristics as the predicate laser systems.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a human figure with three arms or lines extending upwards, representing health, services, and the human aspect of the department's mission.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 3 0 2005
Ms. Karen L. Baker Manager, Regulatory Affairs Lumenis. Inc. 2400 Condensa Street Santa Clara, California 95051
Re: K051944
Trade/Device Name: Family of Selecta Ophthalmic Laser Systems (Selecta SLT, Selecta 1064, and Selecta Duo) and Delivery Device (LaserLink S) and Accessories Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic laser Regulatory Class: II Product Code: HQF, GEX Dated: July 15, 2005 Received: July 18, 2005
Dear Ms. Baker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2- Ms. Karen L. Baker
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark A. Millman
Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Attachment 2 Indications for Use Statement as Requested by FDA
510(K) Number (if Known):
K 051944
# Device Name: Family of Selecta Ophthalmic Laser Systems (Selecta SLT, Selecta 1064, and Selecta Duo) and Delivery Device (LaserLink S) and Accessories
Indications for Use:
photodisruption of ocular tissue using light energy emitted by a Nd:YAG Selecta 1064: laser, including discission of the posterior capsule of the eye (posterior capsulotomy), and discission of pupillary membranes (pupillary membranectomy) in aphakic and pseudophakic patients, and iridotomy/iridectomy
selective laser trabeculoplasty Selecta SLT:
- photodisruption of ocular tissue using light energy emitted by a Nd: Y AG Selecta Duo: laser, including discission of the posterior capsule of the eye (posterior capsulotomy), and discission of pupillary membranes (pupillary membranectomy) in aphakic and pseudophakic patients, and iridotomy/iridectomy; and
selective laser trabeculoplasty
The Selecta 1064, Selecta SLT and Selecta Duo Ophthalmic Lasers are also intended for use as a diagnostic slit lamp, specifically,
> An AC-powered slit lamp biomicroscope intended for use in eye examination of the anterior segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affects the structural properties of the anterior segment.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON AND THER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) |
|------------------------------------------------------------|
| Division of General, Restorative, and Neurological Devices |
| 510(k) Number | K051944 |
|---------------|---------|
|---------------|---------|
| Prescription Use: | <div> <span style="text-decoration: underline;">✓</span> </div> (Per 21 CFR 801.109)<br>Optional Format 1-2-96 | OR | Over-The-Counter Use: ________ |
|-------------------|----------------------------------------------------------------------------------------------------------------|----|--------------------------------|
|-------------------|----------------------------------------------------------------------------------------------------------------|----|--------------------------------|
{6}------------------------------------------------
### Attachment 2, Continued Indications for Use Statement as Requested by FDA
510(K) Number (if Known):
K051944
## Device Name: Family of Selecta Ophthalmic Laser Systems (Selecta SLT, Selecta 1064, and Selecta Duo) and Delivery Device (LaserLink S) and Accessories
Indications for Use:
LaserLink S: laser delivery system for use by an ophthalmologist in the treatment of ocular tissue;
> laser delivery system indicated for use for a variety of ophthalmic uses, including the indications specified in the laser operator manual. Refer to the laser operator manual, Indications for Use section.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use:
(Per 21 CFR 801.109) OR Over-The-Counter Use: **__**
Optional Format 1-2-96
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
510(k) Number **__** K051944
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.