SELECTA FAMILY OF OPHITHALMIC LASER SYSTEMS

K081704 · Lumenis, Inc. · GEX · Sep 11, 2008 · General, Plastic Surgery

Device Facts

Record IDK081704
Device NameSELECTA FAMILY OF OPHITHALMIC LASER SYSTEMS
ApplicantLumenis, Inc.
Product CodeGEX · General, Plastic Surgery
Decision DateSep 11, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Selecta 1064: Photodisruption of ocular tissue using light energy emitted by a Nd:YAG laser, includinq discission of the posterior capsule of the eye (posterior capsulotomy), and discission of pupillary membranes (pupillary membranectomy) in aphakic and pseudophakic patients, and iridotomy/iridectomy. Selecta SLT: Selective laser trabeculoplasty. Selecta Duet: Photodisruption of ocular tissue using light energy emitted by a Nd: YAG laser, including discission of the posterior capsule of the eye (posterior capsulotomy), and discission of pupillary membranes (pupillary membranectomy) in aphakic and pseudophakic patients, and iridotomy/iridectomy and selective laser trabeculoplasty. LaserLink S: Laser delivery system for use by an ophthalmologist in the treatment of ocular tissue. The laser delivery system is intended for a variety of ophthalmic uses, including the indications specified in the laser operator manual. Refer to the laser operator manual, General Intended Use section. Selecta Duo: Photodisruption of ocular tissue using light energy emitted by a Nd:YAG laser, including discission of the posterior capsule of the eye (posterior capsulotomy), and discission of pupillary membranes (pupillary membranectomy) in aphakic and pseudophakic patients, and iridotomy/iridectorny and retinal photocoaqulation. Selecta Trio: Photodisruption of ocular tissue using light energy emitted by a Nd:YAG laser, including discission of the posterior capsule of the eye (posterior capsulotomy), and discission of pupillary membranes (pupillary membranectomy) in aphakic and pseudophakic patients, and iridotomy/iridectomy, retinal photocoaqulation and selective laser trabeculoplasty.

Device Story

Selecta family consists of integrated diagnostic slit lamps and therapeutic ophthalmic lasers; utilizes Nd:YAG laser source to emit 1064 nm or 532 nm pulses. Photodisruption mode (1064 nm) uses twin-aiming beams to focus energy for plasma formation and acoustic wave tissue disruption. SLT mode (532 nm) uses coaxial aiming beams to deliver low-energy pulses to pigmented trabecular meshwork cells. LaserLink S adapter enables connection to external 532 nm retinal photocoagulators. Operated by ophthalmologists in clinical settings; slit lamp microscope used for targeting. Output allows precise ocular tissue treatment; benefits include non-invasive management of posterior capsule opacification, pupillary membranes, glaucoma (via trabeculoplasty), and retinal conditions. Laser functions are disabled when LaserLink S is attached to external photocoagulators.

Clinical Evidence

No clinical data provided; substantial equivalence supported by bench, safety, and functional performance testing.

Technological Characteristics

Integrated slit lamp and Nd:YAG laser system. Wavelengths: 1064 nm (photodisruption) and 532 nm (SLT/retinal photocoagulation). Q-switched pulse delivery. Aiming beams: twin-beam (1064 nm) or coaxial (532 nm). Connectivity: optional LaserLink S adapter for external photocoagulator integration. Class II device.

Indications for Use

Indicated for aphakic and pseudophakic patients requiring ocular tissue photodisruption (posterior capsulotomy, pupillary membranectomy, iridotomy/iridectomy), selective laser trabeculoplasty, or retinal photocoagulation, depending on the specific Selecta model used.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ KO81704 # SEP 1 1 2008 ### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990. #### A. SUBMITTER INFORMATION: | Submitter's Name: | Lumenis, Inc. | |----------------------|---------------------------------------------| | Address: | 3959 W. 1820 S.<br>Salt Lake City, UT 84104 | | Contact: | Tina Mayer, RAC | | Phone: | (801) 656-2543 | | Fax: | (801) 656-2321 | | Date of Preparation: | June 13, 2008 | #### B. DEVICE NAME: | Trade Name(s): | Selecta Family of Ophthalmic Laser Systems | |-----------------------|------------------------------------------------------------------------------------------------------------| | Common/Usual Name: | Ophthalmic Laser, Surgical Laser | | Classification Names: | 79 GEX, Laser Powered Surgical Instrument | | | 86 HQF, Laser, Ophthalmic | | CFR Reference: | 21 CFR 878.4810, Laser surgical instrument for<br>use in general and plastic surgery and in<br>dermatology | | | 21 CFR 886.4390, Ophthalmic laser | #### C. PREDICATE DEVICE NAMES: | Trade Name(s): | 1) Lumenis Selecta Duet (K021550) | |----------------|------------------------------------------------------------------------------| | | 2) Coherent Popeye Ophthalmic Laser System (EPIC I<br>and EPIC IV), (K97340) | | | 3) LaserLink Z-1000 (Z022181) | #### D. Device Description: The Lumenis Selecta is a fully integrated, high-performance diagnostic slit lamp and therapeutic laser delivery system. Selecta has all of the standard controls and functions of a diagnostic slit lamp and is intended for use in eve examination of the anterior segment, from the cornea epithelium to the posterior capsule. The Lumenis Selecta is also an ophthalmic surqical laser designed for performing photodisruption of ocular tissue using laser energy emitted by a Nd:YAG laser including discission of the posterior capsule of the eye (posterior capsulotomy), discission of pupillary membranes (pupillary membranectomy), and iridotomy/iridectomy; and selective laser trabeculoplasty. The Selecta Family of Ophthalmic Lasers is comprised of: Selecta 1064: A Nd:YAG laser providing Q-switched laser pulses at a {2}------------------------------------------------ wavelength of 1064 nm for use in photodisruption of ocular tissue (posterior capsulotomy, papillary membranectomy, and iridotomy). Selecta SLT: A Nd:YAG laser providing Q-switched laser pulses at a wavelength of 532 nm for use in selective laser trabeculoplasty. Selecta Duet: A Nd:YAG laser providing Q-switched laser pulses at a wavelength of 1064 nm for use in photodisruption or Q-switched frequency doubled laser pulses at a wavelength of 532 nm for use in selective laser trabeculoplasty, depending on the mode selected. LaserLink S: Laser delivery adapter that may be coupled to each of the above Selecta models and connected to a currently cleared Lumenis 532 nm retinal photocoaqulator (Novus Spectra, Novus Varia) to allow use of the slit lamp to deliver 532 nm continuous wave laser energy for retinal photocoagulation. Selecta Duo: A Nd:YAG laser providing Q-switched laser pulses at a wavelength of 1064 nm for use in photodisruption or Q-switched 532 nm continuous wave laser energy for retinal photocoagulation. Selecta Trio: A Nd:YAG laser providing Q-switched laser pulses at a wavelength of 1064 nm for use in photodisruption, Q-switched frequency doubled laser pulses at a wavelength of 532 nm for use in selective laser trabeculoplasty, or Q-switched 532 nm continuous wave laser energy for retinal photocoagulation, depending on the mode selected. The Selecta produces short, individual pulses of focused laser light with wavelengths of either 1064 nm or 532 nm. depending on the selected operational mode. Using a slit lamp microscope and aiming beam, the pulsed light is accurately targeted on a structure within the patient's eye. When the photodisruptor mode is selected, the treatment wavelength is 1064 nm. A twin-aiming beam targets the area of tissue disruption. The energy contained within a single short pulse is concentrated by focusing to a very small spot size so that plasma formation occurs at the focal point. This creates an acoustic wave which disrupts nearby tissue. When the SLT mode is selected, the treatment wavelength is 532 nm. A coaxial aiming beam targets the trabecular meshwork via a contact lens. The SLT treatment laser provides a low energy, short pulse of laser light that produces a thermal effect in pigmented cells in the trabecular meshwork. If the optional LaserLink S adapter is attached to the Selecta system and the compatible Lumenis Spectra 532 nm retinal photocoagulator laser, the Selecta works strictly as a diagnostic slit lamp - all photodisruptor and SLT laser functions are disabled. The Laser delivery adaptor is used for treatments specifically cleared for the compatible laser retinal photocoaqulator. {3}------------------------------------------------ The principles of operation and fundamental scientific technology are the same for the subject devices and predicate devices. #### ய் INTENDED USE: The intended use for the Selecta Family of Ophthalmic Laser Systems (Selecta 1064, Selecta SLT, Selecta Duet, LaserLink S, Selecta Duo, and Selecta Trio) and Accessories are as follows: - Selecta 1064: Photodisruption of ocular tissue using light energy emitted by a Nd:YAG laser, includinq discission of the posterior capsule of the eye (posterior capsulotomy), and discission of pupillary membranes (pupillary membranectomy) in aphakic and pseudophakic patients, and iridotomy/iridectomy. - Selecta SLT: Selective laser trabeculoplasty. - Selecta Duet: Photodisruption of ocular tissue using light energy emitted by a Nd: YAG laser, including discission of the posterior capsule of the eye (posterior capsulotomy), and discission of pupillary membranes (pupillary membranectomy) in aphakic and pseudophakic patients, and iridotomy/iridectomy and selective laser trabeculoplasty. - LaserLink S: Laser delivery system for use by an ophthalmologist in the treatment of ocular tissue. The laser delivery system is intended for a variety of ophthalmic uses, including the indications specified in the laser operator manual. Refer to the laser operator manual, General Intended Use section. - Selecta Duo: Photodisruption of ocular tissue using light energy emitted by a Nd:YAG laser, including discission of the posterior capsule of the eye (posterior capsulotomy), and discission of pupillary membranes (pupillary membranectomy) in aphakic and pseudophakic patients, and iridotomy/iridectorny and retinal photocoaqulation. - Selecta Trio: Photodisruption of ocular tissue using light energy emitted by a Nd:YAG laser, including discission of the posterior capsule of the eye (posterior capsulotomy), and discission of pupillary membranes (pupillary membranectomy) in aphakic and pseudophakic patients, and iridotomy/iridectomy, retinal photocoaqulation and selective laser trabeculoplasty. The intended use has not changed from the predicate devices (K021550, K97340, K022181). {4}------------------------------------------------ - F. TECHNOLOGICAL CHARACTERISTICS SUMMARY & SUBSTANTIAL EQUIVALENCE The subject device, the Selecta Family of Laser Systems, has the same intended use, general design and fundamental scientific technology as the predicate devices (K021550, K97340, K022181). The Selecta Family of Laser Systems uses technology substantially equivalent to the Lumenis Selecta Duet (K021550), Coherent Popeye Ophthalmic Laser System (EPIC I and EPIC IV), (K97340), and the LaserLink Z-1000 (Z022181). There are no new hazards introduced by the Selecta Family of Laser Systems as compared with the predicate devices. #### G. PERFORMANCE DATA SUMMARY: The appropriate testing including safety, performance and functional testing to determine substantial equivalence of the Selecta Family of Laser Systems has been conducted. {5}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its wing. The eagle is positioned to the right of the text, which is arranged in a circular pattern around the left side of the logo. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA". ### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP 1 1 2008 Lumenis, Inc. % Ms. Tina Mayer Regulatory Affairs Specialist 3959 West 1820 South Salt Lake City, Utah 84104 Re: K081704 Trade/Device Name: Selecta Family of Laser Systems Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic laser. Regulatory Class: II Product Code: GEX, HQF Dated: June 13, 2008 Received: June 17, 2008 Dcar Ms. Mayer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {6}------------------------------------------------ Page 2 Ms. Tina Mayer This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Survcillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Mark M. Millman Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ 1.3 Indications for Use Statement 510(k) Number (if known): Device Name: Selecta Family of Laser Systems Indications for Use: The Indications for use for the Selecta Family of Ophthalmic Laser Systems (Selecta 1064, Selecta SLT, Selecta Duet, LaserLink S, Selecta Duo, and Selecta Trio) and Accessories are as follows: - Selecta 1064: Photodisruption of ocular tissue using light energy emitted by a Nd: Y AG laser, including discussion of the posterior capsule of the eye (posterior capsulotomy), and discission of pupillary membranes (pupillary membranectomy) in aphakic and pseudophakic patients, and iridotomy/iridectomy. Selecta SLT: Selective laser trabeculoplasty. - Selecta Duet: Photodisruption of ocular tissue using light energy emitted by a Nd: Y AG laser, including discission of the posterior capsule of the eye (posterior capsulotomy), and discission of pupillary membranes (pupillary membranectomy) in aphakic and pseudophakic patients, and iridotomy/iridectomy and selective laser trabeculoplasty. - LaserLink S: Laser delivery system for use by an ophthalmologist in the treatment of ocular tissue. The laser delivery system is intended for a variety of ophthalmic uses, including the indications specified in the laser operator manual. Refer to the laser operator manual, General Intended Use section. - Selecta Duo: Photodisruption of ocular tissue using light energy emitted by a Nd: Y AG laser, including discission of the posterior capsule of the eve (posterior capsulotomy), and discission of pupillary membranes (pupillary membranectomy) in aphakic and pseudophakic patients, and iridotomy/iridectorny and retinal photocoagulation. (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED) {8}------------------------------------------------ ## CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE) Selecta Trio: Photodisruption of ocular tissue using light energy emitted by a Nd: YAG laser, including discission of the posterior capsule of the eye (posterior capsulotomy), and discission of pupillary membranes (pupillary membranectomy) in aphakic and pseudophakic patients, and iridotomy/iridectorny, retinal photocoagulation and selective laser trabeculoplasty. x Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ Db D (Division Sign-Off) Division of General, Restorative, and Neurological Devices **510(k) Number** k081704
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...