POCKET TRANSCRANIAL AND VASCULAR DOPPLER SPECTRUM ANALYZER, MODEL 500P; PW DOPPLER ULTRAOUND PROBE, MODEL 2MPB.
Applicant
Multigon Industries, Inc.
Product Code
IYN · Radiology
Decision Date
Jul 28, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Indications for Use
The 500P Transcranial and Vascular Doppler is to be used for the assessment of circulation in the adult cephalic (intracranial and extracranial) and peripheral vascular vessels. It is to be used in hospitals, clinics by trained medical personnel by prescriptions or doctors orders. It is not to be used for Obstetrics.
Device Story
The 500P is a battery-operated (or AC-powered) pocket transcranial and vascular Doppler spectrum analyzer. It utilizes 2 MHz (pulsed) and 4, 5, or 8 MHz (continuous wave) transducers to insonate adult intracranial and peripheral arteries. The device processes Doppler signals to calculate hemodynamic parameters including peak, mean, pulsatility index, resistance index, and systolic/diastolic ratio. It features a monitoring mode and High Intensity Transient Signals (HITS) detection for emboli. The system includes a display unit, memory for spectral data storage, and a USB interface for external computer connectivity. Healthcare professionals use the device to assess blood flow status; output includes real-time waveforms and printed patient reports. The device aids clinical decision-making by providing diagnostic information on vascular circulation.
Clinical Evidence
Bench testing only. The device underwent extensive safety and performance testing to ensure compliance with functional specifications and industry standards (UL 2601-1, AIUM/NEMA UD 2, AIUM/NEMA UD 3, IEC60601-1-2, IEC60601-2-37). No clinical data was required or presented.
Technological Characteristics
Battery or AC-powered Doppler system. Transducers: 2 MHz (pulsed), 4/5/8 MHz (continuous wave). Quick disconnect connectors. Features: spectral analysis, HITS detection, USB interface. Standards: UL 2601-1, AIUM/NEMA UD 2, AIUM/NEMA UD 3, IEC60601-1-2, IEC60601-2-37. Connectivity: USB for external computer.
Indications for Use
Indicated for assessment of circulation in adult cephalic (intracranial and extracranial) and peripheral vascular vessels. For use by trained medical personnel in hospitals and clinics via prescription. Contraindications: None known. Not for use in obstetrics.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Model 500V Transcranial and Vascular Doppler Analyzer (K882755/A)
{1}------------------------------------------------
K051739
JUL 2 8 2005
500P 510k Submission
500P Transcranial and Vascular Doppler Multigon Industries, Inc.
### SUMMARY
This summary of 510k safety and effectiveness information is being submitted in accordance with 21CFR part 807.92
1. Submitters name, address, phone number, contact person and preparation date:
Name: Multigon Industries, Inc. 1 Odell Plaza Yonkers, N.Y. 10701 Phone: 914 376 5200 ext. 27 Fax: 914 376 6111 Email: wstern@multigon.com Responsible person: William Stern
Official correspondent:
William Stern Multigon Industries, Inc. I Odell Plaza Yonkers, N.Y. 10701 Phone: 914 376 5200 ext. 27 Fax: 914 376 6111 Email: wstern@multigon.com
Date of Preparation: 6/16/05
2. Proprietary Name: Model 500P Pocket Transcranial and Vascular Doppler Spectrum Analyzer
Common /Usual Name:
Transcranial and Vascular Doppler
Diagnostic Ultrasound Transducer
Classification Name:
21 CFR892.1550 System, Imaging, Pulsed Doppler, Ultrasonic
21 CFR892.1570 Diagnostic Ultrasound Transducer
{2}------------------------------------------------
## 500P 510k Submission
#### Classification Number: 90IYN 9011X
3. Substantially Equivalent Devices
Multigon Industries, Inc. believes that the Model 500P Transcranial and Vascular Doppler Spectrum Analyzer and transducers is substantially equivalent to the following cleared devices:
| Trade or Proprietary Name | Manufacturer | S10(k) Number |
|---------------------------------------------------------------|-----------------------|---------------|
| Model 500V Transcranial<br>and Vascular Doppler Analyzer | Multigon Industries. | K882755/A |
| Intra-View Transcranial Doppler | Rimed LTD | K974588 |
| Explorer CVS Transcranial Doppler Diagnostic Medical Sys. | | K990517 |
| Nourogard Transcranial Dopplor | Mcdasonics, Inc. | K962796 |
| Transpect Transcranial Doppler | Medasonics, Inc. | K872292 |
| Model TC4040 Pioneer Transcranial Eden EMS-Nicolet<br>Doppler | | K874684 |
| Multidon Transcranial Donnler | DWI Flectronische Svs | K930458 |
Comparison of Multigon Model 500P Transcranial and Vascular Doppler Spectrum Analyzer and Transducers with the predicate device Multigon Model 500V Transcranial and Vascular Doppler Spectrum Analyzer.
SPECIFICATION
| | Model 500V | Model 500P |
|---------|------------|------------|
| PROBES: | 2 mHz PW | 2 mHz PW |
| | 5 mHz CW | 5 mHz CW |
| | | 4 mHz CW |
| | | 8 mHz CW |
The 4 mHz CW probe and 8 mHz CW probe substantial equivalence to predicate devices is claimed by comparison to the Rimed Intraview TCD k974588 and to the DMS CVS System 1990517.
{3}------------------------------------------------
500P Transcranial and 500P 510k Submission Vascular Doppler Multigon Industries, Inc. Bi- Directional DOPPLER TYPE Bi- Directional FREQUENCY RANGES 1,2,5,10,20,40 kHz 1,2,5,10,20 kHz FREQUENCY OFFSET 0,20,40% 0,20,40,50 % TIME AND FREQUENCY YES YES CALIPERS STORAGE OF SPECTRAL 104 seconds 2 hours DATA CALCULATIONS Peak, Mean, Pulsatility Index Same Parameters Resistance Index, Systolic/Diastolic Ratio ANNOTATION Date, Vessel I.D., Patient I.D. Same plus printed Patient Report Report of waveforms
। ર .. _
The printed report with the waveforms is substantially equivalent to that provided by the Explorer CVS device of Diagnostic Medical systems, cleared with k 990517. The printed report is also substantially equivalent to that provided on the Intraview TCD manufantured by Rimed LTD and cleared with k974588.
| MONITORING MODE | NO | YES |
|-----------------|----|-----|
|-----------------|----|-----|
For the Monitoring Mode for the 500P substantial equivalence is claimed by predicate devices as follows and previously cleared in the Pioneer model TC 4040 510k clearers k 874684 manufactured by Eden Medizinische Elektronik GmBH (EME) a subsidiary of Nicolet Biomedical, now called Viasys. Substantial Equivalence is also claimed by a predicate device manufactured by Rimed LTD, the model Intraview, 510 to use of a predicate are rice Mode of the 500P is substantially equivalent in form and function to the above predicated vices.
{4}------------------------------------------------
500D Trenaoranial and Vaocular Doppler Multigon Industries,Inc.
#### NO
YES
### HITS (HIGH INTENSITY TRANSIENT SIGNALS EMBOLI DETECTION)
The HITS (High Intensity Transient Signals) detection has been cleared in the Pioneer (TC4040)
Children and Children to the existent aredicate device manufactured by Eden The HITS (High Intensity Transient Siguals) contines device manufactured by Eden
( S10k number L874684) substantially equivalent prodicate device manufactured by Eden ( 510k number k874684) subsidiary of Nicolet Biometical, page 430 in their 510(k)
Medizinische Elektronik Ombildiry of Nicolet Biometical, page 430 in their 510(k)
Medizinis Medizinische Elektronik (imbl, a subsidiary of Neone Prististan the indication and seving of
submission described as follows: HITs indications in usingsisms in the indicatio submission described as follows (H.I.T.S.) by allowing physicians to customize the indication
High Intensity Transient signals (H.I.T.S.) by allowing physicians to customize High Intensity Transient signing (11:11:11) of and their personal preferences.
The study the study requirements and their personal preferences.
The HITS function is also present in a predicate Multi-Dop system (510k clearance 1930468)
States (1970) The HITS function is also present in a presented on presented on page 477 of the original 510k submission.
The Multigon 500P HITS detection is substantially equivalent in form and function to the HITS detection in the two predicate devices above.
Conclusion: The Model 500P Transcranial and Vascular Doppler are substantially equivalent to
r and the contraction of the consistion, soccifications, Conclusion: The Model 3001 Transcranial and 1 specifications, specifications, performance, safety and effectiveness.
# 4. DEVICE DESCRIPTION
The 500P device consists of a battery operated (also can be used with a medical grade AC)
Called of the consists of a battery of the collection of the AmHz of 8 mHz or 8 mHz The 500P device consists of a battly operation and of and of anHz of 8 mHz or 8 mHz transducers.
adapter) Doppler and display unit with a choice of 2 mHz of 8 mHz or 8 mHz t adapter) Doppler and display und while in insultation of the adult intracranial arteries. The 2 mHz probe is a pulsed Doppic, used for insonating the peripheral arteries. The The 4mHz , 5 mHz of 8 mirrz , w proves are ace activer, system electronics, a memory Doppler and and a USB interface for an external computer.
All of the transducers are connected to the system using quick disconnect connectors.
The 500P Transcranial and Vascular Doppler provides the healthcare professional with The 500P Transcratial and Vascular Dopped province the state of blood flow in the intracranial and extracranial vascular arteries.
{5}------------------------------------------------
# 500P 510k Submission
500P Transcranial and Vascular Doppler Multigon Industries, Inc.
# 5. PERFORMANCE STANDARDS
of I was a more and the been astablished for the 500P Transcranial and Vacanter Doppler No performance standards have been established tot do 3001 * 1225000
under section 514 of the Federal Food and Drug Act. However the 500P Transcranial and
ers and the became under section 514 of the Federal Food and Drag - Vascular Doppler has been designed to meet the following standards:
UL 2601-1 Safety Requirements for Medical Equipment UL 2601-1 Safety Requirements for Medical Equipment
ATUM/NEMA UD 2 Standard for Real Time Dsplay of Thermal and Mechanical Output Indices
Childer Standard For Real Time Depla on Diagnostic Ultrasound Equipment on Diagnostic Ultrasound Equipment
AIUM/NEMA UD 3 Standard for Real Time Display of Thermal and
AIUM/NEMA UD 3 Standard Discount Four and Fouipment AIUM/NEMA OD 3 Standard Ioi resostic Ultrasound Equipment
Mechanical Output Indices on the Rever Mechamedi Claration of Acoustic Power IEC60601-1-2 IEC60601-2-37
# 6. INDICATIONS FOR USE
The 500P Transcranial and Vascular Doppler is to be used for the assessment of circulation in
t is to be use a the support is to be used for the ral vascular vessels. It is t The 500P Transcranial and Vascular Dopler is to be ascular vessels. It is to be used
the adult cephalic, (intractania) and estacemail) and prophens vessels. It is to be used the adult cephalic, (intracranial and extracramar) and performers and connections offices by prescriptions or doctors orders.
# 7. CONTRA-INDICATIONS
None known at this time.
# 8. COMPARISON TO PREDICATE DEVICES
The 500P Transcranial and Vascular Doppler has the same device characteristics as the The 500P Transcranial and Vascular Doppics mas uit the commonality of ultrasound
approved predicate devices listed in item 3 above with the commonslity of ultrasound
ing the approved predicate devices insted in tiem 3 "door of blood flow waveforms.
#### 9. TEST DATA
The Model 500P Transcranial and Vascular Doppler has been subjected to extensive safety, performance testing, and validation before release. Inal testing of the SOOP included writive performance tests designed to ensure that the device met all of its finactional specifications.
Performance tests designed to ensure that the device comples with and performance tests designed to ensure that the device complies with applicable industry and safety standards.
The Model 500P Transcranial and Vascular Doppler device labeling includes instructions for safe and effective use, wernings, cantions and guidance for use. It has therefore shown to be safe and effective.
{6}------------------------------------------------
#### 10. LITERATURE REVIEW
A review of the literature pertaining to the safety of the 500P Transcranial and Vascular Doppler has been conducted and appropriate safeguards have been incorporated in the design of the 500P Transcranial and Vascular Doppler.
### 11. CONCLUSIONS
The conclusion drawn from these tests is that the 500P Transcranial and Vascular Doppler with transducers is substantially equivalent in safety and efficacy to the predicate devices listed in item 3 above.
{7}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol with three curved lines representing its body and wings. The bird is positioned to the right of a circular text element that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The text is arranged around the circumference of the circle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 2 8 2005
Mr. William Stern Official Correspondent Multigon Industries, Inc. One Odell Plaza YONKERS NY 10701
44
Re: K051739
Trade Name: Model 500P Pocket Transcranial and Vascular Doppler Spectrum Analyzer Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: IYN and ITX Dated: June 24, 2005 Received: June 28, 2005
Dear Mr. Stern:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Model 500P Pocket Transcranial and Vascular Doppler Spectrum Analyzer, as described in your premarket notification:
Transducer Model Number
2 MHz Model 2MPB 4 MHz Model 4MPB 5 MHz Model 5MPB 8 MHz Model 8MPB
{8}------------------------------------------------
Page 2 - Mr. Stern
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), If your device is classified (660 acover) inrols. Existing major regulations affecting your device it may be subject to such additional Regulations, Title 21, Parts 800 to 898. In addition, FDA can be found in the Oode of Feactas concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r reast be advised that 1 DTC 5 losantly evice complies with other requirements of the Act that I DA has made a determinentisions administered by other Federal agencies. You must or any I catal stutures and regarments, including, but not limited to: registration and listing (21 Comply with an the Hot 8 requirements, 01); good manufacturing practice requirements as set CITY art 607), adomig (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping I ins uctimination of Submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested information, meruding acoused of the Center's September 30, 1997 "Information for Manufacturers In Appendix Of (cherosou) of the Some stic Ultrasound Systems and Transducers." If the special Secking Marketing Clearantee of Diagnosable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and Prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket I mis letter will unow you to begin manage quivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific as Compliance at (240) 276-0120. Also, please note the regulation entitled, Contact the Office of oremarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general mionnation on Jour Corpumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
{9}------------------------------------------------
Page 3 - Mr. Stern
If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.
Sincerely yours,
Daniel R. Lynn
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{10}------------------------------------------------
50()P Transcranial and
。
Vascular Doppler Multigon Industries, Inc.
# Indications for Use
510(k) Number (if known):
Model 500P Pocket Transcranial and Vascular Doppler Device Name: Spectrum Analyzer
The 500P Transcranial and Vaseular Doppler is to Indications For Use: The 500r framberainsal of circulation in the be used for che abberanial and extracranial) and peripheral vascular vessels. It is to be used and perfoncral vabource in hospitals, clinics by crained medical persoriptions or doctors and physician not to be used for Obstetrics.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Segura
Page 1 of
{11}------------------------------------------------
### 500P 510K Submission
# 500P Transcranial and Vascular Doppler Multigon Industries, Inc.
#### Appendix F
#### Diagnostic Ultrasound Indications for Use Form
#### Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | A | - B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) | Other (specify) |
|-------------------------------|---|-----|---|-----|-----|---------------|-------------------|------------------------|--------------------|-----------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | N | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: 2 MHz Transducer Model 2MPB for the Model 500P Transcranial and Vascular Doppler
(PLEASE DO NOT WRITE BELOW THIS LINE . CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
David A. Seymour
F-3
(Division Sig Division of Reprodu and Radiological Devic 510(k) Nur
{12}------------------------------------------------
### 500P 510K Submission
# 5()()P Transcranial and Vascular Doppler Multigon Industries, Inc.
#### Appendix F
#### Diagnostic Ultrasound Indications for Use Form
### Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------------------------------------------------|-------------------|---|----|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M' | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | N | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal-Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
| N= new indication; P= previously cleared by FDA; E= added under Appendix E | | | | | | | | | | |
| Additional Comments: 4 MHz Transducer Model 4MPB for the | | | | | | | | | | |
| Model 500P Transcranial and Vascular | | | | | | | | | | |
| Doppler | | | | | | | | | | |
| | | | | | | | | | | |
| | | | | | | | | | | |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Daniel G. Seymore ,
(Division Sign-Off) and Hadle 510(k) Nu
22
{13}------------------------------------------------
# 500P Transcranial and Vascular Doppler Multigon Industries, Inc.
Appendix F
# Diagnostic Ultrasound Indications for Use Form
# Fill out one form for each ultrasound system and each transducer.
| | | | | | | | Mode of Operation | | | |
|-------------------------------|---|---|---|-----|-----|---------------|-------------------|------------------------|--------------------|-----------------|
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) | Other (specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | N | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: 5 MHz Transducer Model 5MPB for the Model 500P Transcranial and Vascular Doppler
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
David A. Lypson
(Division Sign-Off) Division of Reproductive, Abdo and Radiological Devices 510(k) Number __
F-3
{14}------------------------------------------------
# 500P Transcranial and Vascular Doppler Multigon Industries, Inc.
Appendix F .
### Diagnostic Ultrasound Indications for Use Form
# Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|-------------------------------------------------------------------------------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | N | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
| N= new indication; P= previously cleared by FDA; E= added under Appendix E | | | | | | | | | | |
| Additional Comments: 8 MHz Transducer Model 8MPB for the Model 500P Transcranial and Vascular Doppler | | | | | | | | | | |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
David A. Lyman
| (Division Sign-Off) | |
|--------------------------------------|---------|
| Division of Reproductive, Abdominal, | |
| and Radiological Devices | |
| 510(k) Number | K051739 |
E-3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.