SYNTHES (USA) LOW-PROFILE WRIST FIXATOR

K051049 · Synthes (Usa) · KTT · Jul 13, 2005 · Orthopedic

Device Facts

Record IDK051049
Device NameSYNTHES (USA) LOW-PROFILE WRIST FIXATOR
ApplicantSynthes (Usa)
Product CodeKTT · Orthopedic
Decision DateJul 13, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Synthes Low-Profile Wrist Fixator is intended for stabilization of fractures of the distal radius.

Device Story

Synthes Low-Profile Wrist Fixator is an external fixation system for distal radius fractures. Device consists of frame elements including two clamps, a carbon fiber rod, and protective end caps. Stabilization is achieved via Schanz screws inserted proximally and distally to the fracture site, connected by the external bridging frame. The system is provided as a complete sterile assembly. It is intended for use by orthopedic surgeons in a clinical setting to stabilize bone segments during the healing process, facilitating fracture management.

Clinical Evidence

Bench testing only.

Technological Characteristics

External fixation system comprising carbon fiber rod, clamps, and Schanz screws. Provided as a sterile assembly. Mechanical stabilization principle.

Indications for Use

Indicated for stabilization of distal radius fractures in patients requiring external fixation.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the word "SYNTHES" in bold, black letters. To the left of the word is a circular logo with a stylized image inside. A horizontal line runs beneath the word, and a registered trademark symbol is to the right of the word. Image /page/0/Picture/2 description: The image shows the text 'KOS1049' in a handwritten style. The letters and numbers are bold and slightly uneven, suggesting they were written quickly or with a thick marker. The text is the primary focus of the image, with no other discernible background elements. ______________________________________________________________________________________________________________________________________________________________________________ : ## 3.0 510(k) Summary Page ________ of _____________________________________________________________________________________________________________________________________________________________ | Sponsor: | Synthes (USA)<br>1302 Wrights Lane East<br>West Chester, PA 19380<br>(610) 719-5860 | |-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name: | Synthes (USA) Low-Profile Wrist Fixator | | Classification: | Class II, §888.3030 - Single/multiple component bone fixation<br>appliances and accessories | | Predicate Devices: | Orthofix Penning Wrist Fixator<br>Synthes (USA) Articulating Distal Radius System<br>Synthes (USA) 4.0 mm Adjustable Clamp for Distal Radius | | Device Description: | The Synthes Low-Profile Wrist Fixator consists of frame elements<br>that form a construct intended to treat fractures of the distal radius.<br>The Low-Profile Wrist Fixator provides stabilization of fractures<br>via pins (Schanz screws) inserted proximally and distally to a<br>fracture and connected by an external bridging frame consisting of<br>two Low-Profile Wrist Fixator Clamps - MR Safe, a carbon fiber<br>rod and two protective end caps. The Low-Profile Wrist Fixator is<br>available as a complete sterile assembly. | | Intended Use: | The Synthes Low-Profile Wrist Fixator is intended for stabilization<br>of fractures of the distal radius. | | Substantial<br>Equivalence: | Documentation is provided which demonstrates that the Synthes<br>(USA) Low-Profile Wrist Fixator is substantially equivalent to<br>other legally marketed devices.<br>The term "substantial equivalence" as used in this 510(k)<br>notification is limited to the definition of substantial equivalence<br>found in the Federal Food, Drug and Cosmetic Act, as amended<br>and as applied under 21CFR 807, Subpart E under which a device<br>can be marketed without premarket approval or reclassification. A<br>determination of substantial equivalency under this notification is<br>not intended to have any bearing whatsoever on the resolution of<br>patent infringement suits or any other patent matters. No<br>statements related to, or in support of substantial equivalence<br>herein shall be construed as an admission against interest under the<br>US Patent Laws or their application by the courts. | {1}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes representing the three branches of government. The eagle is encircled by the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" in a circular arrangement. The logo is simple and recognizable, representing the department's role in protecting the health of all Americans. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ## JUL 1 3 2005 Ms. Kathy Anderson Regulatory Affairs Manager Synthes (USA) 1302 Wrights Lane East West Chester, Pennsylvania 19380 Re: K051049 K031047 Trade/Device Name: Synthes (USA) Low-Profile Wrist Fixator Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component bone fixation appliances and accessories Regulatory Class: II Product Code: KTT Dated: June 21, 2005 Received: June 22, 2005 Dear Ms. Anderson: We have reviewed your Section 510(k) premarket notification of intent to market the indication We have reviewed your Section 910(x) premainted institution was and starting and in interests a ferenced above and nave determined the arrested predicate devices marketed in interstate for use stated in the encrosule) to regally manated provice Amendments, or to commerce prior to May 28, 1976, the enactment date of the Federal Food. Drug commerce prof to May 26, 1976, the encordance with the provisions of the Federal Food. Drug, devices that have been recure approval of a premarket approval application (PMA). and Cosmetic Act (Act) that do not require approval of a previsions of the Act . The and Cosment Act (Act) that do not require approvine controls provisions of the Act. The You may, therefore, market the device, basjoct to the go general controls provisions of the Act include requirements for annual registration, listing of the general controls provisions of the free labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it II your device is classified (soc aboro) ins Existing major regulations affecting your device can may be subject to such additional controls. Existing major and considerer may be subject to such additional controllar Extrong may on 898. In addition, FDA may be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, be found in the Code of I same concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that TDA S Issualled of a succession complies with other requirements of the Act that FDA has made a determination that your device complies with other must that FDA has made a dolorimiations administered by other Federal agencies. You must of any Federal statutes and regulations anning, but not limited to: registration and listing (21 comply with an the Act 3 requirements, mercially, and manufacturing practice requirements as set and CFK Part 807), labeling (21 CFR Part 807), good and if applicable, the electronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ ## Page 2 - Ms. Kathy Anderson This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to begin mailiering of substantial equivalence of your device to a legally premarket notincation: "The Privating of cation for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acritor 101-12-1120. Also, please note the regulation entitled, Colliact the Office of Compunation in (21CFR Part 807.97). You may obtain Misblanding by reference to promantsvillities under the Act from the Division of Small other general information on your response Assistance at its toll-free number (800) 638-2041 or Manufacturers, International and Octoss http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mirian Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the word "SYNTHES" in bold, black letters. To the left of the word is a logo that appears to be a stylized image of a person. A registered trademark symbol is located to the right of the word. A horizontal line is located underneath the word. Page 1 of 1 2.0 Indications for Use | 510(k) Number (if known): | K051049 | |---------------------------|---------| |---------------------------|---------| Device Name: Synthes (USA) Low-Profile Wrist Fixator Indications for Use: The Synthes Low-Profile Wrist Fixator is intended for stabilization of fractures of the distal radius. X Prescription Use (Per 21 CFR 801.109) AND/OR Over-The-Counter Use_ (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) - 4 - Concurrence of CDRH, Office of Device Evaluation (ODE) Styt Purda (Division Sign-Off) Division of General, Restorative, and Neurological Devices 510(k) Number_________________________________________________________________________________________________________________________________________________________________ KOSTOYG
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...