ENVITEC REUSABLE MULTI-SITE Y SPO2 SENSORS

K043463 · Envitec-Wismar GmbH · DQA · Feb 11, 2005 · Cardiovascular

Device Facts

Record IDK043463
Device NameENVITEC REUSABLE MULTI-SITE Y SPO2 SENSORS
ApplicantEnvitec-Wismar GmbH
Product CodeDQA · Cardiovascular
Decision DateFeb 11, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2700
Device ClassClass 2
AttributesPediatric

Indications for Use

EnviteC Reusable Multi-site Y SpO2 Sensors are indicated for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for adult, pediatric, infant, and neonatal patients in hospital-type facilities, mobile, and home environments. Not recommended for use on the ear. Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.

Device Story

Reusable multi-site SpO2 sensors; consist of connector, cable, and two housings (one with dual LED light source, one with photodetector). Sensors attach to finger, thumb, toe, or foot via medical grade tape. Operates by illuminating arterially perfused tissue with two wavelengths of light; photodetector measures time-varying absorbance to determine oxygen saturation. Used in anesthesia, recovery, critical care, transport, and home environments. Operated by clinicians or patients. Output (SpO2 and pulse rate) displayed on compatible pulse oximeter monitors. Enables continuous monitoring to assist clinical decision-making regarding patient oxygenation status.

Clinical Evidence

Clinical testing performed under controlled hypoxia versus co-oximetry. Results support accuracy claims for the 70% to 100% SaO2 range. Bench testing included electrical safety, electromagnetic compatibility, and pulse rate accuracy using a simulator.

Technological Characteristics

Reusable multi-site Y-sensor design; dual-wavelength LED light source and photodiode detector. Patient contact materials (housing and tape) are biocompatible. Connectivity via manufacturer-specific connector pin-outs. Operates as a passive sensor component for compatible pulse oximeter monitors.

Indications for Use

Indicated for continuous non-invasive monitoring of SpO2 and pulse rate in adult, pediatric, infant, and neonatal patients. Environments include hospitals, mobile, and home settings. Contraindicated for use on the ear.

Regulatory Classification

Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ FEB 11 2005 K043463 Page 1 of 3 # 510(k) Summary of Safety and Effectiveness in accordance with 21 CFR 807.92 | (1) | Submitted by: | EnviteC-Wismar GmbH<br>Alter Holzhafen 18<br>Wismar 23966<br>Germany<br>Tel.: +49-38 41 360 221<br>Fax: +49-38 41 360 222<br>E-mail: b.lindner@envitec.com | | |-----|--------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------| | | Contact Person: | Bernd Lindner | | | | Position/Title: | Managing Director | | | | Date of Preparation: | November 1, 2004 | | | (2) | Trade Name: | EnviteC Reusable Multi-site Y SpO2 Sensors | | | | Common/Classification Name: OXIMETER | | | | | Product Code(s): | DQA; 21 CFR §870.2700 | | | | Class: | Class II | | | (3) | Predicate Device(s): Substantial Equivalence to: | | | | | K Number | Model | Manufacturer | | | K944760 | Nellcor Dura-Y Oxygen<br>Transducer, Ear Clip | Nellcor Puritan Bennett, Inc. | | | K962127 | Ohmeda 3800 Pulse Oximeter | Datex-Ohmeda, Inc. | | | K893877 | Biochem 3100 Oximeter | BCI-Biochem International, Inc | | | K910770 | CSI 504-US Pulse Oximeter | CSI-Criticare Systems, Inc. | | | K010463 | Datex-Ohmeda S/5 Oxygen<br>Saturation Module, M-OSAT and<br>Accessories | Datex-Ohmeda, Inc. | | | K032979 | Philips Reusable SpO2 Sensors | Philips Medizin Ayateme<br>Böeblingen GmbH (formerly HP) | | | K890033 | Model 8700 Pulse Oximeter | Nonin Medical, Inc. | | | K993979 | Model 2001MARSpO2 Pulse<br>Oximeter | Novametrix Medical Systems Inc. | ### Reason for Submission: New Device(s) {2}------------------------------------------------ #### Description of Device: (4) The EnviteC Reusable Multi-site Y SpO2 Sensors are a family of oximeter THE ENTROS NOWBLOG Walidated for compatibility with the predicate oximeter manufacturers listed above. EnviteC's Reusable Multi-site Y SpO₂ Sensors consist of a connector and a cable which divides at the distal end into two sensor housings. One a ouble white a dual LED light source, and the other contains a light sensitive photodetector. The housings may be positioned on various seristive proceded.org finger, thumb, toe, or foot, and are affixed by a medical grade tape. A unique Y sensor exists for each manufacturer above, and each Y sensor has unique labeling and specifications designed for compatibility with the specific monitor manufacturer (Nellcor, CSI, etc.). Each sensor type includes the following unique features: - Connector pin-outs specific for the manufacturer type . - Component specifications specific for the manufacture type 에 Each sensor also specifies the manufacturer type with two compatibility statements: - One printed on or attached to the sensor . - One on the instructions for use. l #### Intended use: (5) The measurement of functional oxygen saturation of arterial hemoglobin (SpO2) has been a standard of care in the USA for 20 years. Applications (opez) no both de monitoring in the anesthesia, recovery, and critical care environments, as well as transport monitoring and home care. # Indications for Use: EnviteC Reusable Multi-site Y SpO2 Sensors are indicated for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for adult, pediatric, infant, and neonatal patients in hospitals, hospital-type facilities, mobile, and home environments. Not recommended for use on the ear. Prescription device. {3}------------------------------------------------ #### Technological Characteristics: (6) The EnviteC Reusable Multi-site Y SpO2 Sensors employ the same The "Envited" Nousable "Mail" the predicate devices to determine arterial oxygen saturation: arterially perfused tissue is illuminated sequentially by two wavelengths of light emitted by light emitting diodes (LED's), and the two wavelenging absorbance of the tissue is measured from a photodiode light time varying absorbands characteristic of all Y sensors which are the subject of this submission as well as the predicate devices. #### Non-Clinical Tests Submitted: (1) (b) The sensors were tested in accordance with applicable standards for medical device Electrical Safety and Electromagnetic Compatibility. The medical device monitors) were tested for pulse rate with a sensors (with prodicate dones ro-optical parameters were measured and Insted Simmator. " Conton" Store Y sensors passed all of the tests. Sensor patient contact materials (housing and attachment tape) meet applicable standards for biocompatibility. #### Clinical Tests Submitted: (2) Clinical testing was performed to validate the performance and accuracy Official testing was perfect Multi-site Y SpO2 Sensors under controlled hypoxia versus arterial oxygen saturation as determined by co-oximetry. Trypoxia versus arterial oxygen-can institutionally approved protocol with All testing was porrormed consent. Clinical test results support the stated accuracy claims for the specified range of 70% to 100% SaO2. #### Conclusions from Tests: (3) As described in (b)(1) and (b)(2) above, EnviteC Reusable Multi-site Y SpO2 Sensors are equivalent to predicate sensors as substantiated by Spoz Sensoro and clinical testing. Device safety is substantiated by parameter, bother, and by biocompatibility of patient contact materials. {4}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is a stylized image of an eagle. FEB 1 1 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Envitec-Wismar GMBH Mr. Stephen Gorski Imagenix, Incorporated S65 W35739 Piper Road Eagle, Wisconsin 53119 Re: K043463 Trade/Device Name: EnviteC Reusable Multi-Site Y SpO2 Sensors Regulation Number: 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: December 12, 2004 Received: December 15, 2004 Dear Mr. Gorski: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ # Indications for Use # 510(k) Number (if known): #### EnviteC Reusable Multi-site Y SpO2 Sensors Device Name: # Indications for use: EnviteC Reusable Multi-site Y SpO2 Sensors are indicated for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for adult, pediatric, infant, and neonatal patients in hospital-type facilities, mobile, and home environments. Not recommended for use on the ear. Caution: Federal law (USA) restricts this device to sale by or on the order of a physician. Prescription Use × (Part 21 CFR 801 Subpart D) AND / OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) | Concurrence of CDRH, Office of Device Evaluation (ODE) | |--------------------------------------------------------| |--------------------------------------------------------| Cure Suleom ion of Annsthesiology, General Hospital, Jution Control. Dental Devices Number K043463 Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...