TITAN HIGH RESOLUTION ULTRASOUND SYSTEM AND SONOSITE ULTRASOUND SYSTEM
Applicant
Sonosite,Inc.
Product Code
IYN · Radiology
Decision Date
Dec 21, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The SonoSite Ultrasound Systems are intended for use for ultrasound evaluation of fetal - OB/GYN, abdominal, intra-operative (abdominal organs, neurological, and vascular), laparoscopic, pediatric, small organ (breast, thyroid, testicles), neonatal cephalic, trans-rectal, transvaginal, musculoskeletal (conventional and superficial), cardiac (adult and pediatric), and peripheral vessel applications. The TITAN™ Ultrasound System has been additionally cleared for adult cephalic applications. The systems provide imaging for biopsy guidance, imaging to assist in the placement of needles and catheters in vascular or other anatomical structures, and imaging guidance for peripheral nerve block procedures.
Device Story
Highly portable, software-controlled diagnostic ultrasound systems; used for real-time imaging and fluid flow analysis. Inputs: ultrasound signals via transducers (curved array). Processing: system performs B-mode, M-mode, Pulsed Wave Doppler (PWD), Continuous Wave Doppler (CWD), and Color Doppler imaging; includes tissue harmonic imaging and 3D imaging capabilities. Outputs: visual ultrasound images/data displayed for clinician. Used in clinical settings (OR, clinic) by physicians/technicians for diagnostic evaluation, biopsy guidance, needle/catheter placement, and peripheral nerve blocks. Benefits: provides non-invasive anatomical visualization and procedural guidance.
Clinical Evidence
Bench testing only. No clinical data required as the added indication is not novel and is supported by predicate devices. Systems previously evaluated for acoustic output, biocompatibility, cleaning/disinfection, and thermal/electrical/mechanical safety per recognized standards.
Technological Characteristics
Portable, software-controlled ultrasound systems. Transducers: Curved Array (C11/8-5, C11/7-4). Imaging modes: B, M, PWD, CWD, Color Doppler. Standards: IEC 60601-1, IEC 60601-2-37, ISO 10993, NEMA PS3.15, NEMA UD2/UD3. Connectivity: DICOM (NEMA PS3.15).
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis of the human body, including fetal, OB/GYN, abdominal, intra-operative (abdominal, neurological, vascular), laparoscopic, pediatric, small organ, neonatal/adult cephalic, trans-rectal, transvaginal, musculoskeletal, cardiac, and peripheral vessel applications. Prescription use only.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
TITAN™ High-Resolution Ultrasound System (K030949)
Siemens SONOLINE Omnia X/XS Diagnostic Ultrasound System (K020353)
Advanced Technology Laboratories HDI 5000 Ultrasound System (K011224)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for SonoSite. The logo consists of a black square with a white, stylized "S" shape inside, followed by the word "SonoSite" in a bold, sans-serif font. A period is placed after the word "SonoSite".
K043452
SonoSite, Inc. 21919 30th Drive SE Botheil, WA 98021-3904 USA
Telephone: 1.425.951.1200 Facsimile: www.sonosite.com
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.
### Submitter's name, address, telephone number, contact person; 1)
Daina L. Graham Vice-President - Regulatory Affairs and Quality Assurance SonoSite, Inc. 21919 30th Drive SE Bothell, WA 98021-3904
(425) 951 - 1275
E-mail: daina.graham@sonosite.com
Date prepared: December 3, 2004
- Name of the device, including the trade or proprietary name if applicable, 2) the common or usual name, and the classification name, if known:
Common/ Usual Name
Diagnostic Ultrasound System with Accessories
Proprietary Names
TITAN™ High-Resolution Ultrasound System SonoSite® Ultrasound System
{1}------------------------------------------------
## Classification Names
| Name | FR Number | Product Code |
|---------------------------------------------|-----------|--------------|
| Ultrasonic Pulsed Doppler Imaging<br>System | 892.1550 | 90-IYN |
| Ultrasonic Pulsed Echo Imaging<br>System | 892.1560 | 90-IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
### ldentification of the predicate or legally marketed device: 3)
The SonoSite Ultrasound System that is the subject of this Submission is the collective term for the legally marketed TITAN™ High-Resolution Ultrasound System (K030949),and the SonoSite® Ultrasound System (K014116).
This 510(k) adds the clinical application 'intra-operative (neurological)' as a indication to the SonoSite Ultrasound Systems (K030949, K014116). No other change, including technological change were made to these systems. Predicate devices that have 'intra-operative (neurological)' as a clinical application or indication are Siemens SONOLINE Omnia X/XS Diagnostic Ultrasound System (K020353) and Advanced Technology Laboratories HDI 5000 Ultrasound System (K011224)
### 4) Device Description:
The devices referenced in this Submission are highly portable. software-controlled, diagnostic ultrasound systems with accessories. This Submission does not include any technological or feature changes from the previously cleared SonoSite devices or transducers.
By this Submission, the clinical application 'intra-operative (neurological)' is being added to previously cleared indications for use for each of these systems and to the following transducers:
{2}------------------------------------------------
| System | Transducer | Transducer<br>Type | Frequency<br>Range |
|---------------------------------------------|------------|--------------------|--------------------|
| TITAN™ High-Resolution<br>Ultrasound System | C11/8-5 | Curved Array | 8.0 - 5.0 MHz |
| SonoSite® Ultrasound System | C11/7-4 | Curved Array | 7.0 - 4.0 MHz |
SonoSite ultrasound systems are designed, as applicable to their features, to comply with the standards listed below.
| EN 980 A1:2003 | IEC 60601-2-37 |
|------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| AAMI/ANSI/ISO 10993-1:1997 | CAN/CSA C22.2, No. 601.1:1998 |
| AAMI/ANSI/ISO 10993-4:1992 | UL 2601-1:1999 |
| AAMI/ANSI/ISO 10993-5:1999 | UL 94, 5th ed. |
| AAMI/ANSI/ISO 10993-10:1995 | EN ISO 13485:1996 |
| AAMI/ANSI/ISO 10993-11:1993 | CISPR 11:2003 |
| AAMI/ANSI/ISO 10993-12:1996 | JIS-T-0601-1 |
| IEC 60601-1:1988 | RTCA/DO160D:1997 |
| IEC 60601-1/A1:1991 | ANSI/AAMI EC53:1995 |
| IEC 60601-1/A2:1995 | ASTM D5276-98 |
| IEC 60601-1-1:2000 | ASTM D999-96 |
| IEC 60601-1-2:2001 | NEMA PS3.15 2000 |
| IEC 60601-1-4:1996 | NEMA UD2-1998 |
| IEC 60601-2-25:1996 | NEMA UD3:1998 |
| AIUM Medical Ultrasound Safety,<br>American Institute of Ultrasound in | Acoustic Output Measurement and<br>Labeling Standard for Diagnostic |
| Medicine (1994) | Ultrasound Equipment, American Institute<br>of Ultrasound in Medicine. 1993 |
#### 5) Intended Use:
As defined by FDA guidance documents, the intended uses of the SonoSite ultrasound systems referenced herein remain unchanged from previously cleared indications, except for the addition of 'intra-operative (neurological)' as a new intended use.
The SonoSite Ultrasound Systems are intended for use for ultrasound evaluation of fetal - OB/GYN, abdominal, intra-operative (abdominal organs, neurological, and vascular), laparoscopic, pediatric, small organ (breast, thyroid, testicles), neonatal cephalic, trans-rectal, transvaginal, musculoskeletal (conventional and superficial), cardiac (adult and pediatric), and peripheral vessel applications. The TITAN™ Ultrasound System has been additionally cleared for adult cephalic applications. The systems provide imaging for biopsy guidance, imaging to assist in the placement of needles and catheters in vascular or other anatomical structures, and imaging guidance for
{3}------------------------------------------------
peripheral nerve block procedures.
#### 6) Technological Characteristics:
There are no technological or feature changes in this Submission to any of the legally marketed ultrasound systems, transducers, or accessories identified in Section 3 of this Summary.
#### 7) Testing:
Each of the referenced SonoSite systems has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness, as well as thermal, electrical and mechanical safety, and has been found to conform to applicable medical device safety standards, as referenced in Section 4. Reports were previously included in the referenced predicate submissions. No additional clinical testing is required, as the added indication for use is not a novel indication as shown by the predicate devices in Section 3. The anatomical site is amenable to current transducer and post-processing ultrasound technology available with the SonoSite systems and predicate devices. Additionally, the modes of operation that are indicated with this clinical application for the SonoSite systems is consistent with those identified with the predicate devices. SonoSite has incorporated the specific labeling required by the FDA for the neurological intra-operative indication in its user guides to mitigate risk to that defined level.
### 8) Conclusion:
SonoSite believes that the testing and analysis described in Section 7 demonstrates that the TITAN™ High-Resolution Ultrasound System (K030949),and the SonoSite® Ultrasound System (K014116), incorporating the 'intra-operative (neurological) indication, are substantially equivalent with respect to safety and effectiveness to the predicate devices identified in Section 3.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure with three horizontal lines above it, possibly representing arms or outstretched hands.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 2 1 2004
SonoSite, Inc. % Mr. Mark Job Responsible Third Party Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313
Re: K043452
Trade Name: TITAN™ High-Resolution Ultrasound System and SonoSite® Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: 90 IYN, IYO, and ITX Dated: December 10, 2004 Received: December 15, 2004
## Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the TITAN™ High-Resolution Ultrasound System and SonoSite® Ultrasound System, as described in your premarket notification:
{5}------------------------------------------------
## Transducer Model Number
# C11/8-5 8.0 - 5.0MHz Curved Array (TITANTM) C11/7-4 7.0 - 4.0MHz Curved Array (SonoSite®
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or unly with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Or the rate of 7) its systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain
{6}------------------------------------------------
other general information on your responsibilities under the Act from the Division of Small other goleral international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.
Sincerely yours,
David C. Lyman
Nancy C. Brogdon W Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{7}------------------------------------------------
| System: | SonoSite TITAN TM high-resolution ultrasound system | | | | | | | | |
|---------------------------|---------------------------------------------------------|-------------------|------------------------------------------------------------------------------------|-----|-----|------------------|----------------------------|------------------|--|
| Transducer: | N/A | | | | | | | | |
| | Intended Use: | | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | |
| | Clinical Application | Mode of Operation | | | | | | | |
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal | P | P | P | | P | B+M; B+PWD;<br>B+CD | Note 1 | |
| | Abdominal | P | P | P | P | P | B+M; B+PWD;<br>B+CWD; B+CD | Note 1 | |
| | Intra-operative (Abdominal<br>organs and vascular) | P | P | P | | P | B+M; B+PWD;<br>B+CD | Note 1 | |
| | Intra-operative (Neuro.) | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 | |
| Fetal Imaging | Laparoscopic | P | P | P | | | B+M; B+PWD | Note 1 | |
| | | | | | | | B+M; B+PWD; | | |
| & Other | Pediatric | P | P | P | P | P | B+CWD; B+CD | Note 1 | |
| | Small Organ (breast,<br>thyroid, testicles) | P | P | P | | P | B+M; B+PWD;<br>B+CD | Note 1 | |
| | | | | | | | B+M: B+PWD; | | |
| | Neonatal Cephalic | P | P | P | | P | B+M; B+PWD;<br>B+CD | Note 1 | |
| | | | | | | | B+M; B+PWD; | | |
| | Adult Cephalic | P | P | P | | P | B+CD | Note 1 | |
| | | | | | | | B+M; B+PWD; | | |
| | Trans-rectal | P | P | P | | P | B+CD | Note 1 | |
| | | | | | | | B+M; B+PWD; | | |
| | Trans-vaginal | P | P | P | | P | B+CD | Note 1 | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skel. (Convent.) | P | P | P | | | B+M; B+PWD | Note 1 | |
| | | | | | | | B+M; B+PWD; | | |
| | Musculo-skel. (Superfic.) | P | P | P | | P | B+CD | Note 1 | |
| | Intra-luminal | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| | | | | | | | B+M; B+PWD; | | |
| | Cardiac Adult | P | P | P | P | P | B+CWD; B+CD | Note 1 | |
| | | | | | | | B+M; B+PWD; | | |
| Cardiac | Cardiac Pediatric | P | P | P | P | P | B+CWD; B+CD | Note 1 | |
| | Trans-esophageal (card.) | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| | | | | | | | B+M; B+PWD; | | |
| Peripheral<br>Vessel | Peripheral vessel<br>Other (spec.) | P | P | P | | P | B+CD | Note 1 | |
# Table 4.3- 1 Diagnostic Ultrasound Indications for Use Form
N= new indication; P= previously cleared by FDA; E= added under Appendix E
### Additional Comments:
Note 1: Other includes Color Power Doppler, combined B and Color Power Doppler, combined B and Directional Color Power Doppler, 3-D Imaging, Tissue Harmonic Imaging for guidance of biopsy. Color Doppler includes Velocity Color Doppler. Imaging to assist in the placement of needles and catheters in vascular or other anatomical structures, and picture archiving, communications and storage functionality were previously cleared in K030949. Imaging guidance for peripheral nerve block procedures was previously cleared in K033367.
Prescription Use (Per 21 CFR 801.109)
David A. Ingram
(Division Sign-Off) Division of Reproductive, Abdor and Radiological Device 510(k) Number
{8}------------------------------------------------
| | Table 4.3- 2 Diagnostic Ultrasound Indications for Use Forn. | | |
|--|--------------------------------------------------------------|--|--|
|--|--------------------------------------------------------------|--|--|
| System: | SonoSite TITAN TM high-resolution ultrasound system | | | | | | | | |
|---------------------------|------------------------------------------------------------------------------------|--|-------------------|---|-----|-----|------------------|----------------------------|------------------|
| Transducer: | C11/8-5 8.0 – 5.0 MHz Curved Array | | | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | |
| Clinical Application | | | Mode of Operation | | | | | | |
| General<br>(Track I only) | Specific<br>(Tracks I & III) | | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal | | | | | | | | |
| | Abdominal | | P | P | P | P | P | B+M; B+PWD;<br>B+CWD; B+CD | Note 1 |
| | Intra-operative (Abdominal<br>organs and vascular) | | P | P | P | P | P | B+M; B+PWD;<br>B+CD | Note 1 |
| | Intra-operative (Neuro.) | | N | N | N | N | N | B+M; B+PWD;<br>B+CD | Note 1 |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | |
| | Pediatric<br>Small Organ (breast,<br>thyroid, testicles) | | P | P | P | P | P | B+M; B+PWD;<br>B+CWD; B+CD | Note 1 |
| | Neonatal Cephalic | | P | P | P | | P | B+M; B+PWD;<br>B+CD | Note 1 |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | | |
| | Intra-luminal | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| | Cardiac Adult | | | | | | | | |
| Cardiac | Cardiac Pediatric | | P | P | P | P | P | B+M; B+PWD;<br>B+CWD; B+CD | Note 1 |
| | Trans-esophageal (card.) | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel<br>Other (spec.) | | P | P | P | | P | B+M; B+PWD;<br>B+CD | Note 1 |
| | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
### Additional Comments:
Note 1: Other includes Color Power Doppler, combined B and Color Power Doppler, combined B and Directional Color Power Noter : Only moldes Oolor Form Boppion, onlinem in the guidance of biopsy. Color Doppler includes Velocity Color Dopper, 3-D maging, Tissue hamono maging, and maging for ground of the range in the cludes, and picking and picking and picking a Dopper. This is a storage functionality were previously cleared in K030949. Imaging guidance for periotheral nerve block procedures was previously cleared in K033367.
Prescription Use (Per 21 CFR 801.109)
David A. Lyman
(Division Sign-Off) Division of Reproductive, Abdomina and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________
{9}------------------------------------------------
| System: | SonoSite® Ultrasound System | | | | | | | |
|---------------------------|------------------------------------------------------------------------------------|-------------------|---|-----|-----|------------------|----------------------|------------------|
| Transducer: | N/A | | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | |
| Clinical Application | | Mode of Operation | | | | | | |
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | P | P | P | | | B+M; B+PWD | Note 1 |
| | Abdominal | P | P | P | P | | B+M; B+PWD;<br>B+CWD | Note 1 |
| | Intra-operative<br>(Abdominal organs and<br>vascular) | P | P | P | | | B+M; B+PWD | Note 1 |
| Fetal Imaging | Intra-operative (Neuro.) | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 |
| | Laparoscopic | P | P | P | | | B+M; B+PWD | Note 1 |
| | & Other | P | P | P | P | | B+M; B+PWD;<br>B+CWD | Note 1 |
| & Other | Pediatric | | | | | | | |
| | Small Organ (breast,<br>thyroid, testicles) | P | P | P | | | B+M; B+PWD | Note 1 |
| | Neonatal Cephalic | P | P | P | | | B+M; B+PWD | Note 1 |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | P | P | P | | | B+M; B+PWD | Note 1 |
| | Trans-vaginal | P | P | P | | | B+M; B+PWD | Note 1 |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | P | P | P | | | B+M; B+PWD | Note 1 |
| | Musculo-skel. (Superfic.) | P | P | P | | | B+M; B+PWD | Note 1 |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| Cardiac | Cardiac Adult | P | P | P | P | | B+M; B+PWD;<br>B+CWD | Note 1 |
| | Cardiac Pediatric | P | P | P | P | | B+M; B+PWD;<br>B+CWD | Note 1 |
| | Trans-esophageal (card.) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | | | B+M; B+PWD | Note 1 |
| | Other (spec.) | | | | | | | |
# Table 4.3- 3 Diagnostic Ultrasound Indications for Use Form
N= new indication; P= previously cleared by FDA; E= added under Appendix E
## Additional Comments:
Note 1: Other includes Color Power Doppler, combined B and Color Power Doppler, combined B and PowerMap™ Directional Color Now 1. One Tireducs Ooler Forwar Deption of Street of biopsy cleared in K014116. Imaging guidance Fower Dopper, o-d maging, Tissue Tiamonial in the placement of beedes and catheres in 1970 and cather anatomical structures was previously cleared in K033367.
Prescription Use (Per 21 CFR 801.109)
David R. Lyynn
(Division Sign-Off) Division of Reproductive, Abdom and Radiological Device 510(k) Number
000019
{10}------------------------------------------------
| System: | SonoSite® Ultrasound System | | | | | | | |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|---|-----|-----|------------------|----------------------|------------------|
| Transducer: | C11/7-4 7.0 - 4.0 MHz Curved Array | | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | |
| Clinical Application | | Mode of Operation | | | | | | |
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | | | | | | | |
| | Abdominal | P | P | P | P | | B+M; B+PWD;<br>B+CWD | Note 1 |
| | Intra-operative (Abdominal<br>organs and vascular) | P | P | P | | | B+M; B+PWD | Note 1 |
| | Intra-operative (Neuro.)<br>Laparoscopic | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 |
| Fetal Imaging<br>& Other | Pediatric<br>Small Organ (breast,<br>thyroid, testicles)<br>Neonatal Cephalic | P | P | P | P | | B+M; B+PWD;<br>B+CWD | Note 1 |
| | Adult Cephalic<br>Trans-rectal<br>Trans-vaginal<br>Trans-urethral<br>Trans-esoph. (non-Card.)<br>Musculo-skel. (Convent.)<br>Musculo-skel. (Superfic.)<br>Intra-luminal<br>Other (spec.)<br>Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric<br>Trans-esophageal (card.)<br>Other (spec.) | P | P | P | P | | B+M; B+PWD;<br>B+CWD | Note 1 |
| Peripheral<br>Vessel | Peripheral vessel<br>Other (spec.) | P | P | P | | | B+M; B+PWD | Note 1 |
# Table 4.3- 4 Diagnostic Ultrasound Indications for Use Form
previously cleared by FDA; E= added under Appendix E ew indication; F
### Additional Comments:
Note 1: Other includes Color Power Doppler, combined B and Color Power Doppler, combined B and PowerMap™ Directional Color Power Doppler, 3-D Imaging, Tissue Harmonic Imaging for guidance of biopsy cleared in K014116. Imaging guidance for peripheral nerve block procedures and imaging to assist in the placement of needles and catheters in vascular or other anatomical structures was previously cleared in K033367.
Prescription Use (Per 21 CFR 801.109)
David A. Segerman
(Division Sign-Off) Division of Reproductive, Abdomina and Radiological Devices 510(k) Number _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.