The intended uses of the SonoSite TITAN™ high-resolution ultrasound system as defined by FDA guidance documents, are: Fetal - OB/GYN, Abdominal, Intra-operative (Abdominal organs and vascular), Laparoscopic, Pediatric, Small Organ (breast, thyroid, testicles.), Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Trans-urethral, Musculo-skel. (Conventional), Musculo-skel. (Superficial), Cardiac Adult, Cardiac Pediatric, Peripheral vessel.
Device Story
Portable, software-controlled ultrasound system; acquires/displays real-time 2D, M-mode, PW/CW Doppler, Color Power Doppler, Velocity Color Doppler, and Directional Color Power Doppler. Inputs: ultrasound signals from curved/linear/intracavitary/static transducers (2.0-10.0 MHz); 3-lead ECG for M-mode timing. Outputs: ultrasound images, Doppler audio, measurements, calculations, DICOM-compliant storage/transfer. Used in clinical settings by healthcare professionals for anatomical assessment, fetal biometry, and interventional guidance (biopsy, drainage, vascular access). ECG feature assists in timing diastolic/systolic function; not for arrhythmia diagnosis. Benefits: portable diagnostic imaging and procedural guidance.
Clinical Evidence
Bench testing only. Device performance verified through compliance with international standards for electrical safety (IEC 60601-1), electromagnetic compatibility (EN 60601-1-2), and acoustic output measurement (NEMA UD2/UD3, IEC 61157). Biocompatibility evaluated per ISO 10993.
Technological Characteristics
Piezoelectric transducer technology; 2.0-10.0 MHz frequency range. Modes: 2D, M-mode, PW/CW Doppler, Color Power/Velocity/Directional Doppler. Connectivity: DICOM-compliant. Safety: Real-time thermal/mechanical index display per AIUM/NEMA 1998. Materials: Biocompatible patient-contact materials. Standards: IEC 60601-1, EN 60601-1-2, ISO 10993, NEMA PS3.15.
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis in fetal, abdominal, intra-operative, laparoscopic, pediatric, small organ, neonatal/adult cephalic, trans-rectal, trans-vaginal, trans-urethral, musculoskeletal, cardiac, and peripheral vascular applications. Contraindicated for long-term cardiac rhythm monitoring.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Philips Ultrasound HDI® 5000 Ultrasound System (K011224)
{0}------------------------------------------------
KO30949
D02576A-510(K) Submission-SonoSite TITAN high-resolution ultrasound system 06-Mar-2003
# APR 0 7 2003 510(K) Summary Of Safety And Effectiveness
This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.
#### Submitter's name, address, telephone number, contact person: 1)
SonoSite, Inc. 21919 30th Drive SE Bothell, WA 98021-3904
| Corresponding Official: | Michael A. Hoffman<br>Director – Quality Assurance and Regulatory Affairs |
|-------------------------|---------------------------------------------------------------------------|
| E-mail: | michael.hoffman@sonosite.com |
| Telephone: | (425) 951-1297 |
| Facsimile: | (425) 951-1201 |
| Date prepared: | March 6, 2003 |
- Name of the device, including the trade or proprietary name if applicable, 2) the common or usual name, and the classification name, if known:
Common/ Usual Name
Diagnostic Ultrasound System with Accessories
### Proprietary Name
SonoSite TITAN™ high-resolution ultrasound system (subject to change)
### Classification Names
| Name | FR Number | Product Code |
|---------------------------------------------|-----------|--------------|
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | 90-IYN |
| Ultrasonic Pulsed Echo Imaging System | 892.1560 | 90-IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
| Picture Archiving And Communications System | 892.2050 | 90-LLZ |
#### 3) Identification of the predicate or legally marketed device:
SonoSite, Inc. believes that the System described in this Submission is substantially equivalent to a combination of the Philips Ultrasound HDI® 5000 Ultrasound System (K011224), the GE LOGIQ 9 Diagnostic Ultrasound System (K011188) and the Siemens OMNIA X/XS Diagnostic Ultrasound System (K020353).
{1}------------------------------------------------
#### 4) Device Description:
The SonoSite TITAN™ high-resolution ultrasound systemis a highly portable. software controlled ultrasound system used to acquire and display highresolution, real-time ultrasound data in 2D, Pulsed Wave (PW) Doppler, Continuous Wave (CW) Doppler, Color Power Doppler, Velocity Color Doppler (VCD) and Directional Color Power Doppler, or in a combination of these modes.
The System has an electrocardiography (ECG) display feature and supports a 3lead ECG cable assembly to collect data for M-mode measurements. The System provides measurement capabilities for anatomical structures and fetal biometry that provide information used for clinical diagnostic purposes. The System has a PW and CW Doppler audio output feature and cine review, image zoom, labeling, biopsy, measurements and calculations, image storage and review, printing, and recording capabilities. The system includes Digital Imaging and Communications (DICOM) to capabilities to provide the acceptance, transfer, display, storage, and digital processing of ultrasound images.
The SonoSite TITAN™ high-resolution ultrasound system is designed to accept curved or linear transducers of the types and frequency listed in the table below. All actions affecting the performance of the transducer are activated from the main system control panel.
| Frequency Range: | 2.0 - 10.0 MHz |
|-------------------|----------------------------------------------------------------------|
| Transducer Types: | Linear array<br>Curved array<br>Intracavitary array<br>Static probes |
The SonoSite TITAN™ high-resolution ultrasound system is designed to comply with the standards listed below.
- a. ANSI/AAMI EC 53: 1995 + Amendments, ECG Cables and Electrodes except for sections 4.4 and 4.5.9.
- b. CAN/CSA-C22.2, No. 601.1:1998, Canadian Standards Association, Medical Electrical Equipment-Part 1. General Requirements for Safety
- c. EN 60529, European Norm, Degrees of Protection Provided by Enclosures (I.P. Code)
- d. EN 60601-1:1997, European Norm, Medical Electrical Equipment-Part 1. General Requirements for Safety
- e. EN 60601-1-1:1993, European Norm, Medical Electrical Equipment Part 1. General Requirements for Safety-Section 1-1. Collateral Standard. Safety Requirements for Medical Electrical Systems
510(K) Summary Of Safety And Effectiveness
{2}------------------------------------------------
- f. EN 60601-1-2:2001, European Norm, Medical Electrical Equipment. General Requirements for Safety-Collateral Standard. Electromagnetic Compatibility. Requirements and Tests, Second Edition
- g. EN 60601-2-25:1996, European Norm, Medical Electrical Equipment-Part 2. Requirements for for Safety-Section 25. Specification Particular for Electrocardiographs
- h. IEC 61157:1992. International Electrotechnical Commission, Requirements for the Declaration of the Acoustic Output of Medical Diagnostic Ultrasonic Equipment
- ISO 10993 Biological Evaluation of Medical Devices and Related Tests i.
- JIS-T-100x Series, Japanese Standards for Medical Electrical Equipment i.
- Medical Ultrasound Safety. American Institute of Ultrasound in Medicine k. (AIUM), 1994
- NEMA PS3.15 2000, Digital Imaging and Communications in Medicine . (DICOM) Part 15: Security Profile.
- m. NEMA UD2-1998, Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment
- n. NEMA UD3-1998. Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
- o. RTCA/DO160D:1997, Radio Technical Commission for Aeronautics, Environmental Conditions and Test Procedures for Airborne Equipment, Section 21.0 Emission of Radio Frequency Energy, Category B
- p. UL 2601-1:1999, Underwriters Laboratories, Inc., Medical Electrical Equipment-Part 1: General Requirements for Safety
- g. UL 94, Fifth Edition, Underwriters Laboratory, Inc., Tests for Flammability of Plastic Materials for Parts in Devices and Appliances
#### 5) Intended Use:
The intended uses of the SonoSite TITAN™ high-resolution ultrasound system as defined by FDA guidance documents, are:
| Fetal - OB/GYN | Trans-rectal |
|---------------------------------|------------------------------|
| Abdominal | Trans-vaginal |
| Intra-operative | |
| (Abdominal organs and vascular) | Trans-urethral |
| Laparoscopic | Musculo-skel. (Conventional) |
| Pediatric | Musculo-skel. (Superficial) |
| Small Organ | |
| (breast, thyroid, testicles.) | Cardiac Adult |
| Neonatal Cephalic | Cardiac Pediatric |
| Adult Cephalic | Peripheral vessel |
{3}------------------------------------------------
Typical examinations performed using the SonoSite TITAN™ high-resolution ultrasound system are:
# Abdominal Imaging Applications:
This system transmits ultrasound energy into the abdomen of patients using 2D, M-mode, color power Doppler (CPD), directional color power Doppler (DCPD), velocity color Doppler (VCD), Tissue Harmonic Imaging (THI), pulsed wave (PW) Doppler, and continuous wave (CW) Doppler to obtain ultrasound images. The liver, kidneys, pancreas, spleen, gallbladder, bile ducts, transplanted organs, abdominal vessels, and surrounding anatomical structures can be assessed for the presence or absence of pathology transabdominally.
# Cardiac Imaging Applications:
This system transmits ultrasound energy into the thorax of patients using 2D. M-mode, directional color power Doppler (DCPD), velocity color Doppler (VCD), color Doppler, Tissue Harmonic Imaging (THI), pulsed wave (PW) Doppler, and continuous wave (CW) Doppler to obtain ultrasound images. The heart, cardiac valves, great vessels, surrounding anatomical structures, overall cardiac performance, and heart size can be assessed for the presence or absence of pathology. The heart can be imaged transthoracic. The patient's electrocardiogram (ECG) may be obtained and is used for accurate timing of diastolic and systolic function.
The ECG is not used to diagnose cardiac arrhythmias and is not designed for Warning: long term cardiac rhythm monitoring.
# Cephalic Imaging Applications:
This system transmits ultrasound energy into the heads of adults and neonates using 2D, M-mode, color power Doppler (CPD), directional color power Doppler (DCPD), velocity color Doppler (VCD), and pulsed wave (PW) Doppler to obtain ultrasound images. The patient's blood flow can be evaluated transcranially.
# Gynecology and Infertility Imaging Applications:
This system transmits ultrasound energy in the pelvis and lower abdomen using 2D, Mmode, color power Doppler (CPD), directional color power Doppler (DCPD), velocity color Doppler (VCD), Tissue Harmonic Imaging (THI), and pulsed wave (PW) Doppler to obtain ultrasound images. The uterus, ovaries, and surrounding anatomical structures can be assessed for the presence or absence of pathology transabdominally or transvaqinally.
## Interventional and Intraoperative Imaging Applications:
This system transmit ultrasound energy into the various parts of the body using 2D. color power Doppler (CPD), directional color power Doppler (DCPD), Tissue Harmonic Imaging (THI), and pulsed wave (PW) Doppler to obtain ultrasound images that provide guidance during interventional and intraoperative procedures. This system can be used to provide ultrasound quidance for biopsy and drainage procedures. vascular line placement, ova harvesting, amniocentesis and other obstetrical procedures, and provide assistance during abdominal and vascular intraoperative procedures.
{4}------------------------------------------------
# Obstetrical Imaging Applications:
This system transmits ultrasound energy into the pelvis of pregnant women using 2D, M-mode, color power Doppler (CPD), directional color power Doppler (DCPD), velocity color Doppler (VCD), color Doppler, Tissue Harmonic Imaging (THI), and pulsed wave (PW) Doppler to obtain ultrasound images. The fetal anatomy, viability, estimated fetal weight, gestational age, amniotic fluid, and surrounding anatomical structures can be assessed for the presence or absence of pathology trans-abdominally or transvaqinally. CPD and DCPD imaging is intended for high-risk pregnant women. High-risk pregnancy indications include, but are not limited to, multiple pregnancy, fetal hydrops, placental abnormalities, as well as maternal hypertension, diabetes, and lupus.
CPD or DCPD images can be used as an adjunctive method, not as a screening tool, Warning: for the detection of structural anomalies of the fetal heart and as an adjunctive method, not as a screening tool for the diagnosis of Intrauterine Growth Retardation (IUGR).
# Pediatric and Neonatal Imaging Applications:
This system transmits ultrasound energy into the pediatric or neonatal patients using 2D, M-mode, color power Doppler (CPD), directional color power Doppler (DCPD), velocity color Doppler (VCD), pulsed wave (PW) and continuous wave (CW) Doppler to obtain ultrasound images. The pediatric abdominal, pelvic and cardiac anatomy, pediatric hips, neonatal heads, and surrounding anatomical structures can be assessed for the presence or absence of pathology.
# Prostate Imaging Applications:
This system transmits ultrasound energy into the prostate of an adult male using 2D. Mmode, color power Doppler (CPD), velocity color Doppler (VCD), and pulsed wave (PW) Doppler to obtain ultrasound images. The prostate gland can be assessed for the presence or absence of pathology transrectally.
# Superficial Imaging Applications:
This system transmits ultrasound energy into various parts of the body using 2D, M-mode, color power Doppler (CPD), velocity color Doppler (VCD), and pulsed wave (PW) Doppler to obtain ultrasound images. The breast, thyroid, testicle, lymph nodes, hernias, musculoskeletal structures, soft tissue structures, and surrounding anatomical structures can be assessed for the presence or absence of pathology.
## Vascular Imaging Applications:
This system transmits ultrasound energy into the various parts of the body using 2D. Mmode, color power Doppler (CPD), velocity color Doppler (VCD), Tissue Harmonic Imaging (THI), and pulsed wave (PW) Doppler to obtain ultrasound images. The carotid arteries, deep veins in the arms and legs, superficial veins in the arms and legs, great vessels in the abdomen, and various small vessels feeding organs can be assessed for the presence or absence of pathology.
{5}------------------------------------------------
#### Technological Characteristics: 6)
This device operates identically to the predicate devices in that piezoelectric material in the transducer is used as an ultrasound source to transmit sound waves into the body. Sound waves are reflected back to the transducer and converted to electrical signals that are processed and displayed as 2D or Mmode images. Doppler shift caused by blood flow is displayed as Color Flow, or as spectrum analysis. The modes of this device (2D, PW Doppler, CW Doppler, velocity color Doppler, Color Power Doppler, and directional color power Doppler, and duplex imaging) are the same as a combination of the predicate devices identified in item 3. Transducer patient contact materials are biocompatible.
This device conforms to the Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment (AIUM/NEMA, 1998) for an on-screen display feature that provides information on potential thermal and cavitation bioeffect mechanisms. A user education program provides additional information so users may moderate the system's acoustic output in accordance with the ALARA (as low as reasonably achievable) principle.
The device's acoustic output limits are:
All applications:
| ISPTA (d) | 720 mW/cm2 (Maximum) |
|-----------------------|-----------------------|
| TIs/TIb/TIc | 0.1 - 4.0 (Range) |
| Mechanical Index (MI) | 1.9 (Maximum) |
| ISPPA (d) | 0 - 700 W/cm2 (Range) |
The limits are the same as predicate Track 3 devices.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized depiction of three human profiles facing to the right, resembling an eagle or bird-like shape.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 0 7 2003
SonoSite, Inc. % Ms. Laura Danielson TUV Product Service 1775 Old Highway 8 NW Suite 104 NEW BRIGHTON MN 55112-1891
Re: K030949
Trade Name: TITANTM High-Resolution Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 IYN and LLZ Dated: March 25, 2003 Received: March 26, 2003
Dear Ms. Danielson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the TITAN™ High-Resolution Ultrasound System, as described in your premarket notification:
### Transducer Model Number
| C11/7-4 |
|----------|
| C11/8-5 |
| C15/4-2 |
| C60/5-2 |
| HST/10-5 |
| ICT/7-4 |
| C8/8-5 |
{7}------------------------------------------------
ICT/8-5 L25/10-5 L38/10-5 L52/10-5 L52-S/10-5 2MHz Doppler Pencil
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801, please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.
Sincerely yours,
David Le. Hyman
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{8}------------------------------------------------
| System: | SonoSite TITAN™ high-resolution ultrasound system | | | | | | | |
|---------------------------|------------------------------------------------------------------------------------|-------------------|---|-----|-----|------------------|-------------------------------|------------------|
| Transducer: | N/A | | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | |
| Clinical Application | | Mode of Operation | | | | | | |
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 |
| | Abdominal<br>Intra-operative<br>(Abdominal organs and<br>vascular) | N | N | N | N | N | B+M; B+PWD;<br>B+CWD;<br>B+CD | Note 1 |
| Fetal Imaging | Intra-operative (Neuro.)<br>Laparoscopic | N | N | N | | | B+M; B+PWD | Note 1 |
| | | | | | | | | |
| & Other | Pediatric | N | N | N | N | N | B+M; B+PWD;<br>B+CWD;<br>B+CD | Note 1 |
| | Small Organ (breast,<br>thyroid, testicles.) | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 |
| | Neonatal Cephalic | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 |
| | Adult Cephalic | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 |
| | Trans-rectal | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 |
| | Trans-vaginal<br>Trans-urethral | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 |
| | Trans-esoph. (non-Card.)<br>Musculo-skel. (Convent.) | N | N | N | | | B+M; B+PWD | Note 1 |
| | Musculo-skel. (Superfic.)<br>Intra-luminal<br>Other (spec.) | P | P | P | | P | B+M; B+PWD;<br>B+CD | Note 1 |
| | Cardiac Adult | N | N | N | N | N | B+M; B+PWD;<br>B+CWD;<br>B+CD | Note 1 |
| Cardiac | Cardiac Pediatric<br>Trans-esophageal (card.)<br>Other (spec.) | N | N | N | N | N | B+M; B+PWD;<br>B+CWD;<br>B+CD | Note 1 |
| | | | | | | | | |
| Peripheral | Peripheral vessel | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 |
# Table 4.3- 1 Diagnostic Ultrasound Indications for Use Form - TITAN System
N= new indication; P= previously cleared by FDA; E= added under Appendix E
### Additional Comments:
Note 1: Other includes Color Power Doppler, combined B and Color Power Doppler, combined B and Directional Color Power Doppler, 3-D lmaging, Tissue Harmonic Imaging for guidance of biopsy. Included with this 510k are imaging to assist in the placement of needles and cathelers in vascular or other anatomical structures. Musculo-skel. (Superfic.) imaging previously cleared in 510(k) K003399. Included with this 510(k) is picture programs and storage functionality previously cleared in 510(k) K023957.
Color Doppler includes Velocity Color Doppler. Division
Prescription Use (Per 21 CFR 801.109) Indications for Use
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K030949
Section 4.3
25 of 961 pages
{9}------------------------------------------------
# Table 4.3- 2 Diagnostic Ultrasound Indications for Use Form - C11/7-4 Transducer
| System: | SonoSite TITAN™ high-resolution ultrasound system | | | | | | | | | |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---|-------------------|-----|-----|------------------|----------------------------|------------------|--|--|
| Transducer: | C11/7-4 7.0 - 4.0 MHz Curved Array | | | | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | | |
| Clinical Application | | | Mode of Operation | | | | | | | |
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) | | |
| Ophthalmic | Ophthalmic | | | | | | | | | |
| | Fetal | | | | | | | | | |
| | Abdominal<br>Intra-operative (Abdominal<br>organs and vascular) | N | N | N | N | N | B+M; B+PWD;<br>B+CWD; B+CD | Note 1 | | |
| | Intra-operative (Neuro.)<br>Laparoscopic | | | | | | | | | |
| Fetal Imaging<br>& Other | Pediatric<br>Small Organ (breast, thyroid, testicles.) | N | N | N | N | N | B+M; B+PWD;<br>B+CWD; B+CD | Note 1 | | |
| | Neonatal Cephalic | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 | | |
| | Adult Cephalic<br>Trans-rectal<br>Trans-vaginal<br>Trans-urethral<br>Trans-esoph. (non-Card.)<br>Musculo-skel. (Convent.)<br>Musculo-skel. (Superfic.)<br>Intra-luminal<br>Other (spec.) | | | | | | | | | |
| | Cardiac Adult | | | | | | | | | |
| Cardiac | Cardiac Pediatric<br>Trans-esophageal (card.)<br>Other (spec.) | N | N | N | N | N | B+M; B+PWD;<br>B+CWD; B+CD | Note 1 | | |
| Peripheral<br>Vessel | Peripheral vessel<br>Other (spec.) | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
### Additional Comments:
Note 1: Other includes Color Power Dopler, combined B and Color Power Doppler, combined B and Directional Color Power Doppler, 3-D Imaging, Tissue Harmonic Inaging for guidance of biopsy. Included with this 510k are imaging to assist in the placement of needles and catheters in vascular or other anatomical structures.
Color Doppler includes Velocity Color Doppler.
Prescription Use (Per 21 CFR 801.109)
David A. Bergman
(Division Sign-C Division of Reproductive, and Radiological D 510(k) Numbe
{10}------------------------------------------------
| System: | SonoSite TITAN™ high-resolution ultrasound system | | | | | | | | |
|---------------------------|------------------------------------------------------------------------------------|--|-------------------|---|-----|-----|---------------|----------------------------|------------------|
| Transducer: | C11/8-5 8.0 – 5.0 MHz Curved Array | | | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | |
| Clinical Application | | | Mode of Operation | | | | | | |
| General<br>(Track I only) | Specific<br>(Tracks I & III) | | B | M | PWD | CWD | Color Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal | | | | | | | | |
| | Abdominal<br>Intra-operative (Abdominal<br>organs and vascular) | | N | N | N | N | N | B+M; B+PWD;<br>B+CWD; B+CD | Note 1 |
| | Intra-operative (Neuro.) | | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | |
| | Pediatric<br>Small Organ (breast,<br>thyroid, testicles.) | | N | N | N | N | N | B+M; B+PWD;<br>B+CWD; B+CD | Note 1 |
| | Neonatal Cephalic<br>Adult Cephalic | | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | | |
| | Intra-luminal | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | | |
| | Cardiac Pediatric | | N | N | N | N | N | B+M; B+PWD;<br>B+CWD; B+CD | Note 1 |
| | Trans-esophageal (card.)<br>Other (spec.) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel<br>Other (spec.) | | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 |
# Table 4.3- 3 Diagnostic Ultrasound Indications for Use Form - C11/8-5 Transducer
N= new indication; P= previously cleared by FDA; E= added under Appendix E
### Additional Comments:
Note 1: Other includes Color Power Dopler, combined B and Color Power Doppler, combined B and Directional Color Power Doppler, 3-D Imaging, Tissue Harmonic Imaging for guidance of biopsy. Included with this 510k are imaging to assist in the placement of needles and catheters in vascular or other anatomical structures,
Color Doppler includes Velocity Color Doppler.
Prescription Use (Per 21 CFR 801.109)
Daniel R. Lyman
(Division Sign-Off) Division of Reproductive, A and Radiological D 510(k) Number
Indications for Use
{11}------------------------------------------------
| System: | SonoSite TITAN™ high-resolution ultrasound system | | | | | | | |
|---------------------------|---------------------------------------------------------------------------------------|-------------------|---|-----|-----|------------------|----------------------------|------------------|
| Transducer: | C15/4-2 4.0 - 2.0 MHz Curved Array | | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as<br>follows: | | | | | | | |
| Clinical Application | | Mode of Operation | | | | | | |
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | N | N | N | | | B+M; B+PWD | Note 1 |
| | Abdominal | N | N | N | N | N | B+M; B+PWD;<br>B+CWD; B+CD | Note 1 |
| | Intra-operative (Abdominal<br>organs and vascular) | N | N | N | | | B+M; B+PWD | Note 1 |
| | Intra-operative (Neuro.) | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | B+M; B+PWD | Note 1 |
| | Pediatric | N | N | N | | | | |
| | Small Organ (breast,<br>thyroid, testicles.) | N | N | N | | | B+M; B+PWD | Note 1 |
| | Neonatal Cephalic | N | N | N | | | B+M; B+PWD | Note 1 |
| | Adult Cephalic | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | N | N | N | | | B+M; B+PWD | Note 1 |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| | Cardiac Adult | N | N | N | N | N | B+M; B+PWD;<br>B+CWD; B+CD | Note 1 |
| Cardiac | Cardiac Pediatric | N | N | N | N | N | B+M; B+PWD;<br>B+CWD; B+CD | Note 1 |
| | Trans-esophageal (card.) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 |
| | Other (spec.) | | | | | | | |
# Table 4.3- 4 Diagnostic Ultrasound Indications for Use Form - C15/4-2 Transducer
N= new indication; P= previously cleared by FDA; E= added under Appendix E
# Additional Comments:
Note 1: Other includes Color Power Doppler, combined B and Color Power Doppler, combined B and Directional Color Power Doppler, 3-D Imaging, Tissue Harmonic Imaging for guidance of biopsy. Included with this 510k are imaging to assist in the placement of needles and catheters in vascular or other anatomical structures.
28
Color Doppler includes Velocity Color Doppler.
Prescription Use (Per 21 CFR 801.109)
David R. Legrown
(Division Division c productive. Abdo and Radiological Dev 510(k) Numbe
Section 4.3
of 961 pages
{12}------------------------------------------------
# Table 4.3- 5 Diagnostic Ultrasound Indications for Use Form - C60/5-2 Transducer
| System: | SonoSite TITAN™ high-resolution ultrasound system | | | | | | | | | |
|---------------------------|------------------------------------------------------------------------------------|----------------------------------------------------------------|-------------------|---|-----|-----|------------------|---------------------|------------------|--|
| Transducer: | C60/5-2 5.0-2.0 MHz Curved Array | | | | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | | |
| Clinical Application | | | Mode of Operation | | | | | | | |
| General<br>(Track I only) | | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) | |
| Ophthalmic | | Ophthalmic | | | | | | | | |
| | | Fetal | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 | |
| | | Abdominal | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 | |
| | | Intra-operative (Abdominal<br>organs and vascular) | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 | |
| | | Intra-operative (Neuro.) | | | | | | | | |
| Fetal Imaging | | Laparoscopic | | | | | | | | |
| | | Pediatric | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 | |
| & Other | | Small Organ (breast,<br>thyroid, testicles.) | | | | | | | | |
| | | Neonatal Cephalic | | | | | | | | |
| | | Adult Cephalic | | | | | | | | |
| | | Trans-rectal | | | | | | | | |
| | | Trans-vaginal | | | | | | | | |
| | | Trans-urethral | | | | | | | | |
| | | Trans-esoph. (non-Card.) | | | | | | | | |
| | | Musculo-skel. (Convent.) | | | | | | | | |
| | | Musculo-skel. (Superfic.) | | | | | | | | |
| | | Intra-luminal | | | | | | | | |
| | | Other (spec.) | | | | | | | | |
| | | Cardiac Adult | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 | |
| Cardiac | | Cardiac Pediatric<br>Trans-esophageal (card.)<br>Other (spec.) | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 | |
| Peripheral<br>Vessel | | Peripheral vessel<br>Other (spec.) | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
## Additional Comments:
Note 1: Other includes Color Power Dopler, combined B and Color Power Doppler, combined B and Directional Color Power Doppler, 3-D Imaging, Tissue Harmonic Imaging for guidance of biopsy. Included with this 510k are imaging to assist in the placement of needles and catheters in vascular or other anatomical structures.
Color Doppler includes Velocity Color Doppler.
Prescription Use (Per 21 CFR 801.109)
David A. Leymm
(Division Sign-Off) Division of Reproductive, Abdomina and Radiological De 510(k) Number
Indications for Use
{13}------------------------------------------------
# Table 4.3- 6 Diagnostic Ultrasound Indications for Use Form - HST/10-5 Transducer
| System: | SonoSite TITAN™ high-resolution ultrasound system | | | | | | | |
|---------------------------|------------------------------------------------------------------------------------|-------------------|---|-----|-----|------------------|---------------------|------------------|
| Transducer: | HST/10-5 10.0-5.0 MHz Linear Array Transducer | | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | |
| Clinical Application | | Mode of Operation | | | | | | |
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Abdominal<br>organs and vascular) | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 |
| | Intra-operative (Neuro.) | | | | | | | |
| Fetal Imaging | Laparoscopic | | | | | | | |
| & Other | Pediatric | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 |
| | Small Organ (breast,<br>thyroid, testicles.) | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | N | N | N | | N | B+M; B+PWD | Note 1 |
| | Musculo-skel. (Superfic.) | P | P | P | | P | B+M; B+PWD;<br>B+CD | Note 1 |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (card.) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral | Peripheral vessel | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
### Additional Comments:
Note 1: Other includes Color Power Dopler, combined B and Color Power Doppler, combined B and Directional Color Power Doppler, 3-D Imaging, Tissue Harmonic Imaging for guidance of biopsy. Included with this 510k are imaging to assist in the placement of needles and catheters in vascular or other anatomical structures. Musculo-skel. (Superfic.) imaging previously cleared in 510(k) K003399.
Color Doppler includes Velocity Color Doppler.
Prescription Use (Per 21 CFR 801.109)
David R. Segmann
(Division Sign-Off) Division of Reproductive, A and Radiologi 510(k) Number
{14}------------------------------------------------
# Table 4.3- 7 Diagnostic Ultrasound Indications for Use Form - ICT/7-4 Transducer
| System: | SonoSite TITAN™ high-resolution ultrasound system | | | | | | | | | |
|----------------|-----------------------------------------------------------|---------------------------------------------------------------------------|---|-----|-----|---------|---------------------|---------|--|--|
| Transducer: | ICT/7-4 7.0-4.0 MHz Intracavitary Transducer | | | | | | | | | |
| Intended Use: | | Diagnostic ultrasound imaging or fluid flow analysis of the human body as | | | | | | | | |
| | follows: | | | | | | | | | |
| | Clinical Application | Mode of Operation | | | | | | | | |
| General | Specific | | | | | Color | Combined | Other | | |
| (Track I only) | (Tracks I & III) | B | M | PWD | CWD | Doppler | (Spec.) | (Spec.) | | |
| Ophthalmic | Ophthalmic | | | | | | | | | |
| | | | | | | | | | | |
| | Fetal | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 | | |
| | Abdominal | | | | | | | | | |
| | Intra-operative (Abdominal | | | | | | | | | |
| | organs and vascular) | | | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | | | |
| Fetal Imaging | Laparoscopic | | | | | | | | | |
| & Other | Pediatric<br>Small Organ (breast,<br>thyroid, testicles.) | | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | | |
| | Adult Cephalic | | | | | | | | | |
| | | | | | | | | | | |
| | Trans-rectal | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 | | |
| | | | | | | | | | | |
| | Trans-vaginal | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 | | |
| | Trans-urethral | | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | | | |
| | Intra-luminal | | | | | | | | | |
| | Other (spec.) | | | | | | | | | |
| | Cardiac Adult | | | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | | | |
| | Trans-esophageal (card.) | | | | | | | | | |
| | Other (spec.) | | | | | | | | | |
| Peripheral | Peripheral vessel | | | | | | | | | |
| Vessel | Other (spec.) | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
### Additional Comments:
Note 1: Other includes Color Power Dopler, combined B and Color Power Doppler, combined B and Directional Color Power Doppler, 3-D Imaging, Tissue Harmonic Inaging for guidance of biopsy. Included with this 510k are imaging to assist in the placement of needles and catheters in vascular or other anatomical structures.
Color Doppler includes Velocity Color Doppler.
Prescription Use (Per 21 CFR 801.109)
David M. Segmon
(Division Sic Division of Reproductive, A and Radi 510(k) Numb
Indications for Use
{15}------------------------------------------------
# Table 4.3- 8 Diagnostic Ultrasound Indications for Use Form -- C8/8-5 Transducer
| System: | | SonoSite TITAN™ high-resolution ultrasound system | | | | | | | |
|---------------------------|----------------------------------------------------|------------------------------------------------------------------------------------|---|-----|-----|------------------|---------------------|------------------|--|
| Transducer: | | C8/8-5 8.0- 5.0 MHz Prostate Transducer | | | | | | | |
| Intended Use: | | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | |
| | | Mode of Operation | | | | | | | |
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 | |
| | Abdominal | | | | | | | | |
| | Intra-operative (Abdominal<br>organs and vascular) | | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ (breast,<br>thyroid, testicles.) | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 | |
| | Trans-vaginal | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | | |
| | Intra-luminal | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| | Cardiac Adult | | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | | |
| | Trans-esophageal (card.) | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | | |
| | Other (spec.) | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
### Additional Comments:
Note 1: Other includes Color Power Dopler, combined B and Color Power Doppler, combined B and Directional Color Power Doppler, 3-D Imaging, Tissue Harmonic Imaging for guidance of biopsy. Included with this 510k are imaging to assist in the placement of needles and catheters in vascular or other anatomical structures.
Color Doppler includes Velocity Color Doppler.
Prescription Use (Per 21 CFR 801.109)
Daniel A. Hyman
(Division Sign-Off) Division of Reproductive, A and Radiological Devic 510(k) Number
{16}------------------------------------------------
| System: | | SonoSite TITAN™ high-resolution ultrasound system | | | | | | | |
|---------------------------|-------------------------------------------------|------------------------------------------------------------------------------------|-------------------|---|-----|-----|------------------|---------------------|------------------|
| Transducer: | | ICT/8-5 10.0-5.0 MHz Intracavitary Transducer | | | | | | | |
| Intended Use: | | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | |
| Clinical Application | | | Mode of Operation | | | | | | |
| General<br>(Track I only) | Specific<br>(Tracks I & III) | | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | | |
| Fetal Imaging<br>& Other | Fetal | | N | N | N | | N | B+M; B+PWD;<br>B+CD | Note 1 |
| | Abdominal | | | | | | | | |
| | Intra-operative (Abdominal organs and vascular) | | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.