The PMP4 Oxy Pro digital pulse oximeter system is intended for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patients in different settings: a) At hospitals, homes and alternate-sites as well as for EMS environment and for transport use. b) At remote locations, where patients can obtain, display and transmit their pulse oximeter vital signs data to a medical expert via a communication device, such as a PDA or a remote server.
Device Story
Photoelectric pulse oximeter; measures arterial hemoglobin oxygen saturation (SpO2) and pulse rate. Manually activated by patient; displays real-time data on integrated LCD. Features Bluetooth connectivity for data transmission to compatible PDAs, phones, or PCs for remote medical expert review. Used in hospitals, homes, EMS, and transport environments. Provides vital signs monitoring; enables remote patient management; facilitates clinical decision-making by transmitting data to healthcare providers. Internally powered; type CF applied parts; includes low battery detection.
Clinical Evidence
Bench testing only. Design verification and validation conducted to confirm compliance with specifications. Biocompatibility evaluated per ISO 10993.
Technological Characteristics
Photoelectric pulse oximeter; internally powered; type CF applied parts. Connectivity via Bluetooth. Standards: ISO 14971, IEC 601-1-1, IEC 601-1-2, IEC 601-1-4, ISO 10993, ISO 13485, ISO 9001. Moderate level of concern system.
Indications for Use
Indicated for spot checking and/or continuous monitoring of SpO2 and pulse rate in adult and pediatric patients in clinical, home, EMS, transport, and remote settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Nonin Avant® model 4000 Digital Pulse Oximetry System (K041156)
Submission Summary (Full Text)
{1}------------------------------------------------
DEC 29 2004
CARD GUARD
Scientific Survival Ltd
K043162
PMP4 Oxy Pro
510(k) Premarket Notification
Chapter 16
# Summary of Safety and Effectiveness
{2}------------------------------------------------
| | PMP4 Oxy Pro<br>510(k) Summary of Safety and Effectiveness |
|-----------------|--------------------------------------------------------------------------------------------------------|
| Submitter: | Card Guard Scientific Survival Ltd., |
| | 2 Pekeris St. P.O.B. 527<br>Rehovot 76100, Israel |
| | Tel: 972-8-9484000 |
| | Fax: 972-8-9484044 |
| Contact Person: | Boris Aradovsky,<br>VP QA and Regulatory Affairs<br>Tel: 972-8-9484000<br>E-mail: borisa@cardguard.com |
| Date Prepared: | October 12, 2004 |
#### Definition 1.
The PMP4 Oxy Pro is a photoelectric pulse oximeter device to be used for self testing upon prescription of an authorized health care provider. Specifically, it is designed for patients who wish to measure the oxygen saturation levels in the blood and pulse rate.
The PMP4 OxyPro is manually activated by the patient. It functions as a stand-alone device as patient can see "real time" data on its LCD. In addition, the PMP4 OxyPro enables transmission of real time data to the patient's PMP4 PDA or compatible generic PDA phone or personal computer supporting Bluetooth (BT) protocol.
#### 2. Intended Use
The PMP4 Oxy Pro digital pulse oximeter system is intended for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patients in different settings:
- a) At hospitals, homes and alternate-sites as well as for EMS environment and for transport use.
- b) At remote locations, where patients can obtain, display and transmit their pulse oximeter vital signs data to a medical expert via a communication device, such as a PDA or a remote server.
The PMP4 Oxy Pro is indicated for spot checking and/or continuous monitoring.
UNCONTROLLED 1 1 NOV 2004 COPY
| This document contains proprietary information of<br>CARD GUARD Scientific Survival Ltd. | Document No. | Rev. | Page | of |
|------------------------------------------------------------------------------------------|--------------|------|------|----|
| | --- | 01 | 1 | 4 |
{3}------------------------------------------------
# 3. Applicable Standards, Regulations, Guidances
PMP4 Oxy Pro meets the requirements of the following Standards, Regulations and Guidances:
- . ISO 14971:2000 , Medical devices - Application of risk management to medical devices
- . IEC 1025: 1990 Fault tree analysis (FTA)
- IEC 801-1, 1984, General Introduction .
- . IEC 601-1-1, 1996, Safety Requirements for Medical Electrical Systems
- IEC 601-1-2, 2001, Part 2: Electromagnetic compatibility-Requirements and Tests .
- IEC 601-1-4, 1996, Part 1-4, Programmable Electrical Medical Systems
- IEC 801-2, 1991, Electrostatic Discharge Requirements' .
- IEC 801-3, 1992, Immunity to Radiated Radiofrequency electromagnetic fields .
- . IEC 801-4, 1988, Electrical Fast Transient Burst Requirements
- . CISPR 11 1990 Limits and Methods of Measurement of Electromagnetic Disturbance Characteristics of Industrial, Scientific and Medical (ISM) Radio frequency Equipment 2nd Edition
- Reviewer Guidance for Computer Controlled Medical Devices, FDA Aug 29, 1991 .
- ISO 13485 (2003), Medical Devices Quality Management Systems .
- ISO 9001:2000, Quality Management Systems Requirements .
- ISO 10993:2003, parts 1, 5, 10 Biological evaluation of medical devices. . Evaluation and testing MIL-STD 810E, product environmental testing
- FDA's New 510(k) Paradigm, Alternate Approaches to Demonstrating Substantial . equivalence in Premarket Notifications Final Guidance, CDRH, March 20, 1998.
- 21 CFR part 820 subchapter H medical devices , quality system regulations .
#### 4. Features
- . Two control buttons
- . Transmission via Bluetooth an IR (future optional)
- The device is internally-powered with applied parts of type CF, suitable for . continuous operation.
- . Low Battery detection.
| This document contains proprietary information of<br>CARD GUARD Scientific Survival Ltd. | Document No. | Rev. | Page | of |
|------------------------------------------------------------------------------------------|--------------|------|------|----|
| --- | 01 | 2 | 4 | |
{4}------------------------------------------------
Image /page/3/Figure/0 description: The image shows the text "CARD GUARD" in bold, black letters. Below this text is the phrase "Scientific Survival LTD" in a smaller, lighter font. To the right of the text, there is a stylized graphic resembling an electrocardiogram (ECG) waveform, which is commonly associated with heart monitoring.
#### 5. User Interface
The PMP4 Oxy Pro user interface incorporates the following controls and signals:
- . On/off control button
- . Low battery warning
# 6. Substantial Equivalence
The substantially equivalence to the following predicate devices is claimed:
| Nonin<br>Avant®<br>model<br>4000 | Digital Pulse Oximetry System | K041156 | Decision Date<br>06/09/2004 |
|----------------------------------|-------------------------------|---------|-----------------------------|
|----------------------------------|-------------------------------|---------|-----------------------------|
The Nonin Avant® 4000 was cleared as a prescription device on 06/09/04.
The PMP4 Oxy Pro, subject of this application, is substantially equivalent to the predicate device.
- . Intended use:
Both devices are indicated for spot checking by patients and/or continuous monitoring. Both devices are digital pulse oximeter systems intended for measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. They both intended for adults and pediatrics.
The correlation between the comparable parameters and features of the devices, for the purpose of proving their substantial equivalency is herby provided in the comparison table on chapter 7.
# 7. Design Controls and Hazard Analysis
The Card Guard's product design procedure, and QA and QC policy, formalize the design and production process and assure that all respective requirements are met. In the framework of the Design Controls laboratory testing were conducted to verify and validate the PMP4 Oxy Pro compliance with all design specifications.
The device biocompatibility was evaluated and found to be satisfactory.
The device Level of Concern criteria were evaluated and PMP4 Oxy Pro was characterized as a moderate level of concern system.
| This document contains proprietary information of<br>CARD GUARD Scientific Survival td. | Document No. | Rev. | Page | of |
|-------------------------------------------------------------------------------------------|--------------|------|------|----|
| | - | 01 | 3 | 4 |
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Card Guard Scientific Survival LTD. The words "CARD GUARD" are in bold, black letters, with a line underneath. Below the line, the words "Scientific Survival LTD" are in a smaller, lighter font. To the right of the text is a graphic that resembles an EKG readout.
The system safety and risk analysis conducted for PMP4 Oxy Pro provided rigorous design and structural evaluation aimed at revealing potential failures or possible system flaws which could directly or indirectly affect the patient.
### 8. Conclusions
1.4
PMP4 Oxy Pro constitutes a safe and reliable means for measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patients. Its material composition and operation present no adverse health effect or safety risks to patients when used as intended.
| This document contains proprietary information of | | | | |
|---------------------------------------------------|--------------|------|------|--|
| | Document No. | Rev. | Page | |
| CARD GUARD Scientific Survival Ltd. | | | | |
| | | | | |
| | | | | |
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 2 9 2004
Mr. Boris Aradovsky Vice President Quality Assurance and Regulatory Affairs Card Guard Scientific Survival, Limited 2 Pekeris Street P.O.B 527 Rehovot, 76100 ISRAEL
Re: K043162
Trade/Device Name: PMP4 Oxy Pro Regulation Number: 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: November 16, 2004 Received: November 16, 2004
Dear Mr. Aradovsky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 - Mr. Aradovsky
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours,
Chu Lin, Ph.D.
hiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
510(k) Number: K043162
Device Name: PMP4 Oxy Pro
Indications For Use:
The PMP4 Oxy Pro digital pulse oximeter system is intended for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patients in different settings:
- a) At hospitals, homes and alternate-sites as well as for EMS environment and for transport use.
- b) At remote locations, where patients can obtain, display and transmit their pulse oximeter vital signs data to a medical expert via a communication device, such as a PDA or a remote server.
Prescription Use ___ √______ AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON NOTHERPAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
e
(Division Sign-Off)
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices
510(k) Number: K043162
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.