Pefakit® APC-R Factor V Leiden is a plasma based functional assay for the determination of resistance to activated protein C caused by the factor V Leiden mutation (FV:Q506). Pefakit® APC-R Factor V Leiden is a plasma based functional assay for the determination of resistance to activated protein C caused by the factor V Leiden mutation (FV:Q506) on automated and semi-automated blood coagulation analyzers using mechanical or optical detection principle.
Device Story
Pefakit® APC-R Factor V Leiden is an in vitro diagnostic test kit for detecting Factor V Leiden mutation-induced resistance to activated protein C (APC-R). The device uses citrated human plasma as input. The assay principle involves a factor V-dependent prothrombin activator derived from snake venom, which triggers coagulation in the absence of calcium. The device measures clotting times in the presence and absence of APC; the ratio of these times indicates the presence of the mutation. It is used in clinical laboratories on automated or semi-automated coagulation analyzers. Healthcare providers use the calculated ratio to identify patients with Factor V Leiden mutation, aiding in the assessment of thrombotic risk. The assay is designed to be independent of calcium and upstream coagulation factors, and includes polybrene to inhibit heparin interference.
Clinical Evidence
Clinical studies conducted at two central hospitals (CIMCLD/AKH, Vienna; Duke University Medical Center, USA) compared Pefakit to the predicate. Results showed comparable precision (CV < 6%) and no interference from oral anticoagulant treatment. Pefakit demonstrated superior diagnostic sensitivity and specificity, particularly in patients with lupus anticoagulants, where the predicate showed interference. The device successfully discriminated between wild-type, heterozygous, and homozygous carriers of the FV Leiden mutation.
Technological Characteristics
Plasma-based functional clotting assay. Reagents: lyophilized APC/RVV-V, PTA (prothrombin activator), and dilution plasma. Sensing principle: mechanical or optical detection of clot formation on automated/semi-automated analyzers. Connectivity: standalone reagent kit for use on third-party coagulation analyzers. Sterilization: N/A (in vitro diagnostic reagents).
Indications for Use
Indicated for the determination of resistance to activated protein C caused by the factor V Leiden mutation (FV:Q506) in human plasma samples using automated or semi-automated coagulation analyzers. For prescription use only.
Regulatory Classification
Identification
A partial thromboplastin time test is a device used for primary screening for coagulation abnormalities, for evaluation of the effect of therapy on procoagulant disorders, and as an assay for coagulation factor deficiencies of the intrinsic coagulation pathway.
Predicate Devices
COATEST® APC™ RESISTANCE V/ COATEST® APC™ RESISTANCE VS (K963111)
Submission Summary (Full Text)
{1}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE
A. 510(k) Number:
K042762
B. Purpose for Submission:
To obtain clearance for the Pefakit® APC-R Factor V Leiden assay, a plasma-based functional assay for the determination of resistance to activated protein C caused by the factor V Leiden mutation.
C. Analyte:
Resistance to activated protein C (APC-R)
D. Type of Test:
Plasma-based functional clotting assay
E. Applicant:
Pentapharm Ltd.
F. Proprietary and Established Names:
Pefakit® APC-R Factor V Leiden
G. Regulatory Information:
1. Regulation section:
21 CFR 864.7925
2. Classification:
Class II
3. Product Code:
GGW
4. Panel:
81 Hematology
H. Intended Use:
1. Intended use(s):
Pefakit® APC-R Factor V Leiden is a plasma based functional assay for the determination of resistance to activated protein C caused by the factor V Leiden mutation (FV:Q506).
2. Indication(s) for use:
Pefakit® APC-R Factor V Leiden is a plasma based functional assay for the determination of resistance to activated protein C caused by the factor V Leiden mutation (FV:Q506) on automated and semi-automated blood coagulation analyzers using mechanical or optical detection principle.
{2}
Page 2 of 11
3. Special condition for use statement(s):
Not applicable
4. Special instrument Requirements:
Not applicable
I. Device Description:
Pefakit® APC-R Factor V Leiden is an in vitro diagnostic test kit containing 3 vials each of the following 4 lyophilized plasmas:
R1: APC/RVV-V (+APC) Reagent (APC, RVV-V, Polybrene, Hepes, BSA)
R2: APC/RVV-V (-APC) Reagent (RVV-V, Polybrene, Hepes, BSA)
R3: PTA Reagent (Prothrombin Activator, EDTA, Hepes, BSA)
R4: Dilution Plasma (human plasma, processed)
For quality assurance/quality control the corresponding control kit Pefakit® APC-R Factor V Leiden Controls has to be used.
J. Substantial Equivalence Information:
1. Predicate device name(s):
(a) COATEST® APC™ RESISTANCE V/ COATEST® APC™ RESISTANCE VS
2. Predicate K number(s):
(a) K963111
3. Comparison with predicate:
| Similarities | | | |
| --- | --- | --- | --- |
| Item | | Device | Predicate |
| Sample Type: | Pefakit® APC-R Factor V Leiden | | Coatest APC Resistance V/Coatest APC Resistance VS |
| | Lypophilized citrated human plasma | | Lypophilized citrated human plasma |
| Reagent components: | FV deficient plasma and polybrene | | FV deficient plasma and polybrene |
| Differences | | | |
| Item | | Device | Predicate |
| Reagent components: | | Snake venom | Phospholipids |
{3}
Page 3 of 11
| Stability:
Plasma reagent at 15-25° C (on board, once opened or after reconstitution) | 8 – 24 hours | 6 - 8 hours |
| --- | --- | --- |
| Other reagents at 15-25° C (on board, once opened or after reconstitution) | 24 hours | 1 week |
| Plasma reagent at 2 – 8° C (once opened or after reconstitution) | 14 days | 24 hours (3 mos. at -20°C) |
| Interference: Lupus anticoagulant | No influence | Abnormal clotting time, possible misleading results |
## K. Standard/Guidance Document Referenced (if applicable):
### L. Test Principle:
Pefakit® APC-R Factor V Leiden is a plasma-based functional clotting assay and differs from other functional APC resistance tests by acting specifically on the prothrombinase complex level. It is based on a factor V-dependent prothrombin activator isolated from snake venom. Robustness and specificity of the assay is enhanced by elimination of possible disturbing influences by factors upstream of the coagulation cascade and independency from calcium.
Sample plasma is prediluted in dilution plasma and incubated at 37° C with factor V activator. Coagulation is triggered by the addition of a FV-dependent prothrombin activator in the absence of calcium. The ratio of the clotting time in the presence of APC and the clotting time in the absence of APC is determined.
### M. Performance Characteristics (if/when applicable):
#### 1. Analytical performance:
##### a. Precision/Reproducibility:
The coefficient of variation (CV) for all factor V Leiden genotype groups (homozygous, heterozygous, wild-type) and for the analytical instruments tested so far (BCS®, CA-500, STA®-R) within the series and from day to day determined by in-house and external validation studies was below 6%.
Two plasma samples genotyped with PCR, one factor V Leiden heterozygous and one normal wild-type, were tested in a series of 25 measurements on the same day on two different fully automated analytical systems (BCS®, CA-500). The CV of the ratios for the plasma genotypes on both instruments was below 5%.
{4}
In a further study on the STA®-R analyzer, a series of 10 tests on each of 5 consecutive days were done using the same factor V Leiden heterozygous and normal wild-type plasmas. The CV of the clotting times and the ratios within each day and over all 5 days was $\leq 5.0$ for the negative control and $< 3.0$ for the heterozygous control.
The values obtained during the 5 days were within the following ranges:
| | C1 FVL Negative Control | | | C2 FVL Heterozygous Control | | |
| --- | --- | --- | --- | --- | --- | --- |
| | +APC(s) | -APC(s) | Ratio | +APC(s) | -APC(s) | Ratio |
| min | 132.9 | 24.2 | 5.3 | 39.2 | 25.4 | 1.5 |
| max | 166.3 | 25.7 | 6.7 | 42.6 | 27.7 | 1.6 |
b. Linearity/assay reportable range: Not applicable
c. Traceability (controls, calibrators, or method): Not applicable
d. Detection limit: Not applicable
e. Analytical specificity:
Factor deficiency:
The influence of factor deficiencies was tested in spiking experiments, where deficient plasmas were supplemented with increasing concentrations (0%, 25%, 50%, 75%, 100%) of the factor for which they were deficient. +APC and -APC clotting times were determined on an Amelung KC-10 ATM.
No significant influence on ratio or test sensitivity in case of fibrinogen, prothrombin, FVIII, FX, ATIII, protein C, or protein S deficiency was found up to $100\%$ . A deficiency of FV $(< 50\%)$ may lead to strongly elevated clotting times and thus may lead to loss of discrimination performance.
Factor excess:
The influence of factor excess was tested in spiking experiments, where normal plasmas were supplemented with increasing concentrations (100%, 200%, 300%, 400%, 500%) of factor.
No significant influence on ratio or test sensitivity in case of excess of fibrinogen, FVIII, ATIII, or TPFI up to 5 times the normal value.
{5}
Page 5 of 11
# Anticoagulants:
To demonstrate that unfractionated heparin (UFH), low molecular weight heparins (LMWH), and pentasaccharides (PES) are inhibited by the addition of polybrene to the reagents, control plasmas were spiked with increasing amounts of UFH (Liquemin, 0.0-3.0 IU/ml), LMWH (Fraxiparin, 0.0-3.0 IU/ml) and PED (Arixtra, 0.0-2.0 $\mu$g/ml). Up to a concentration of $\leq 2.0$ IU/ml for UFH and LMWH and a concentration of $\leq 2.0$ $\mu$g/ml for PES, +APC and –APC clotting times for both controls were not or only slightly increasing and the ratios were within the predefined acceptable range.
Hirudin and argatroban are not inhibited by polybrene and thus preclude proper measurement at concentrations as low as $0.5\ \mu\mathrm{g/ml}$.
The presence of lupus anticoagulant antibodies does not influence this test because there is no use of phospholipids in an APTT system. Due to the test principle, oral anticoagulant therapy (with warfarin or marcoumar) does not influence test outcome.
The influence of patient treatment with aprotinin or protamine was tested by spiking control plasma with increasing concentrations of the treatments. The presence of aprotinin (which inhibits the APC used in this test) and protamine in the patient's blood can result in decreasing ratios which were outside the acceptable range.
The effect of direct thrombin inhibitors (DTI) such as hirudin or argatroban are not inhibited by polybrene. Hirudin in the patient plasma has a strong effect on clotting times and thus precludes proper discrimination of the different genotypes.
# Additional interferences:
15 citrated blood samples were centrifuged and the plasmas tested with the Pefakit assay on a BCS® coagulation analyzer, once freshly after collection and again after storage in a freezer at $-80^{\circ}\mathrm{C}$ for 3 weeks and rethawing (5 minutes at $37^{\circ}\mathrm{C}$). Clotting times were measured for +APC and –APC reagent and the ratio calculated according to the instructions given in the box insert. No relevant differences were observed between clotting times and ratios measured with fresh plasma and with frozen and re-thawed plasma respectively.
In order to test for influence of hemolytic plasma on test performance, a freshly drawn blood sample was frozen at $-80^{\circ}\mathrm{C}$ and rapidly thawed at $37^{\circ}\mathrm{C}$. The sample was centrifuged for 20 minutes, which gave hemolyzed plasma to be used for the dilution series. To both the negative and heterozygous control, $0 - 5\%$ hemolyzed plasma was added and the resulting samples tested using
{6}
Page 6 of 11
the submitted device. The results showed no influence on clotting time and ratio even at the highest level of 5% hemolyzed plasma.
In order to test for influence of platelet residues in the test plasma a platelet-rich plasma (PRP) was prepared by centrifugation of a freshly drawn blood sample. The supernatant of this sample was used to prepare a dilution series of 0-100% with normal pool plasma used a negative control. It showed that there was no influence on clotting time and ratio even at the highest level of 100% PRP.
| | wt | het | hom | NC | Total |
| --- | --- | --- | --- | --- | --- |
| Wild-type (normal) | 137 | 0 | 0 | 0 | 137 |
| FV Leiden het | 0 | 92 | 0 | 0 | 92 |
| FV Leiden hom | 0 | 0 | 7 | 0 | 7 |
| Total | 137 | 92 | 7 | 0 | 236 |
Sensitivity: 100%
Specificity: 100%
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
The comparative studies were performed at two sites: The Clinical Institute for Medical and Chemical Laboratory Diagnostics (CIMCLD), University of Austria and Duke University Medical Center (DUMC).
Side by side comparison studies were conducted between Pefakit® APC-R Factor V Leiden and COATEST® APC™ Resistance V as the predicate device on the same analyzer (STA® R) with the same batches of test kits.
Study design:
The studies were divided into three phases: familiarization, precision study and main clinical study. During the familiarization phase the investigator and the technicians acquired experience with the test method and its adaptation to the analytical device. In this phase, the investigator did some blind studies with samples provided by Pentapharm Ltd (FV-L status known by Pentapharm only) on three subsequent days. Decision on continuation of the study was made based on the reported results.
The precision study was performed in order to determine the intra- (20 consecutive tests with identical samples) and inter-assay (tests performed with identical sample on 10 consecutive days) precision of the test method on the analytical device. Precision testing was
{7}
Page 7 of 11
performed with control plasmas provided in the Pefakit® APC-R Factor V Leiden Controls kit.
The data obtained in the inter-assay testing were used to establish a confidence range (mean ±2SD) for the use of these controls during phase 3. In phase 3, citrated samples (3.8% or 3.2%) were analyzed with Pefakit® APC-R Factor V Leiden and the predicate device, COATEST® APC™ Resistance V.
The samples included the following:
- At least 100 healthy subjects (50 females, no oral contraceptives, 50 males) with wild-type factor V confirmed by the routine functional test used at the DUMC, for determination of the normal range
- 50-100 subjects with heterozygous FV Leiden mutation confirmed by PCR method
- 10-25 subjects with homozygous FV Leiden mutation confirmed by PCR method
- 5-10 patients with confirmed deficiency of protein S
- 5-10 patients with confirmed deficiency of protein C
- At least 10 patients with confirmed lupus anticoagulants
- At least 10 patients with F VIII > 150%
- 5-10 patients with PT20210G/A mutation, if available
- 10 pregnant women (preferably 2nd and 3rd trimenon)
The samples excluded from the test group had the following conditions:
- Treatment with a direct thrombin inhibitor (DTI) during the last 2 days
- Treatment with UFH, LMWH, or PES leading to a blood concentration > 2 U/ml
- FV deficiency (FV<30%)
Treatment with oral anticoagulants (vitamin K antagonists) was expected to have no influence on both tests and was neither an inclusion nor an exclusion criterion. This condition applied for up to about ¾ of all samples tested.
All tests were done in duplicate, analyzing with the COATEST APC assay followed by the Pefakit APC-R assay. Both test series were separated by a rigorous washing step. The test runs were only started once the ratio and time values obtained with the quality controls fell into the confidence range as established in phase 2 of the study. The QC testing was repeated if a test series contained more than 25 samples.
{8}
Page 8 of 11
Precision results of CIMCLD Study
| Intra-assay precision - Pefakit | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | Control C1 | | | Control C2 | | |
| | +APC | -APC | Ratio | +APC | -APC | Ratio |
| Mean (n=20) | 145.35 | 26.03 | 5.58 | 37.69 | 27.74 | 1.37 |
| SD | 8.09 | 0.61 | 0.24 | 0.44 | 0.49 | 0.05 |
| CV | 5.57 | 2.34 | 4.32 | 1.16 | 1.76 | 3.50 |
| Intra-assay precision - COATEST | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | Level 1 | | | Level 2 | | |
| | +APC | -APC | Ratio | +APC | -APC | Ratio |
| Mean (n=20) | 37.7 | 83.7 | 2.2 | 38.6 | 62.5 | 1.62 |
| SD | 0.3 | 1.9 | 0.1 | 0.2 | 0.4 | 0.01 |
| CV | 0.8 | 2.3 | 2.3 | 0.58 | 0.70 | 0.82 |
| Inter-assay precision - Pefakit | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | Control C1 | | | Control C2 | | |
| | +APC | -APC | Ratio | +APC | -APC | Ratio |
| Mean (n=10) | 154.86 | 25.91 | 5.97 | 43.58 | 27.27 | 1.59 |
| SD | 10.82 | 0.90 | 0.35 | 2.15 | 1.42 | 0.07 |
| CV | 6.99 | 3.49 | 5.88 | 4.93 | 5.21 | 4.67 |
| Inter-assay precision - COATEST | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | Level 1 | | | Level 2 | | |
| | +APC | -APC | Ratio | +APC | -APC | Ratio |
| Mean (n=10) | 37.7 | 88.7 | 2.4 | 39.2 | 66.4 | 1.69 |
| SD | 1.3 | 4.6 | 0.1 | 1.4 | 4.1 | 0.10 |
| CV | 3.3 | 5.2 | 3.8 | 3.60 | 6.21 | 5.76 |
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical sensitivity:
The diagnostic sensitivity (defined as the percentage of patient plasmas having a heterozygous or homozygous factor V Leiden mutation correctly classified as positive for this mutation) in all in-house and external validation studies was 100%. The 100% sensitivity claim was confirmed by both clinical studies for the device.
{9}
Page 9 of 11
b. Clinical specificity:
The diagnostic specificity (defined as the percentage of patient plasmas not having a factor V Leiden mutation correctly classified as negative for this mutation) in all in-house and external validation studies was 100%. The 100% specificity claim was confirmed by both studies.
c. Other clinical supportive data (when a and b are not applicable):
Stability:
The Pefakit® APC-R Factor V Leiden Controls were used to determine clotting times and ratios for stability verification. All tests were performed on an Amelung KC-10 A™ micro equipment.
Stability of lyophilized reagents:
In preliminary stability tests on the sample lots 1-4, the reagents and controls showed to be stable at 2-8°C for more than a year.
In stress tests, the lyophilized reagents were exposed to 37°C for up to 14 days. Samples of the test kit reagents, R1-R4, were reconstituted on the first and second day, on day 10, day 12 and day 14. Samples of the controls C1 and C2 were also reconstituted on the first and second day, on day 6, day 8 and day 12. The +APC and -APC clotting times were measured. The clotting times were slightly rising over this time period, but the ratio remained stable during 14 days for the test kit reagent and 8 days for the controls.
Stability of reconstituted reagents:
After reconstitution, the open reagent R1-R4 were placed on the analytical instrument (15-25°C) and were tested using freshly prepared controls immediately after reconstitution, at 2 hour intervals over an 8 hour period, and again after 24 h, 32 h, and 96 h. During this period, the clotting times for the +APC measurement especially for the normal wild-type control (C1) were slightly rising, but were still within predefined acceptable range. This rise of clotting times had no relevant influence on the ration and on the power of the test to discriminate the different FV Leiden genotypes. It was concluded that the reagents have an on-board stability of at least 24 hours.
Reconstituted samples of reagents and controls were frozen and stored at -20°C. After a storage time of 6 months, the reagents and controls were thawed and clotting times and rations for C1 and C2 were determined. Clotting times and ratios were within the predefined acceptable ranges. No shift was observed. It was
{10}
Page 10 of 11
concluded that the device and its reagent are stable when frozen at -20° C for at least 6 months.
Reagent R1-R4 were reconstituted and stored at 4° C. Over a period of 32 days clotting times and ratios were determined all 3 days. During the first 15 days the ratio remained stable within the predefined acceptable range. Beginning with day 18, the ratio began to drop. It was concluded that the reconstituted reagents can be stored at 2-8° C for at least 14 days.
Stability of plasma samples:
Blood samples were tested immediately after blood collection and plasma preparation and again after 24 hours storage at room temperature. Only a minimal rise of both clotting times and ratios was observed. It was concluded that plasma samples can stay as long as 12 hours on the analytical instrument with no influence on the discrimination power of the test.
Long term stability:
Long term stability studies were performed according to written standard protocols on sample lots 1-4 and on the pilot lots 1-3. Three test and three control kits released by QC were transferred to a controlled cool storage room at +5 C. Clotting times were determined with these test and control kits after a storage period of 3, 6, 9, 12, 18, 24, 36, 48, and 60 months. The study is ongoing and the longest storage time analyzed so far is 24 months. The ratios measured with these kits were within the predefined ranges.
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Typical ratio ranges for PCR-genotyped patient plasmas on different devices are shown in the tables below:
| KC-4/-10 A micro | | |
| --- | --- | --- |
| Genotype FV:Q506 | n | Ratio range (min/max) |
| Negative | 25 | 4.2 – 6.9 |
| heterozygous | 104 | 1.3 – 1.9 |
| homozygous | 15 | 1.0 – 1.1 |
{11}
| BCS (Behring Coagulation System) | | |
| --- | --- | --- |
| Genotype FV:Q506 | n | Ratio range (min/max) |
| Negative | 69 | 4.0 – 5.5 |
| heterozygous | 108 | 1.4 – 2.2 |
| homozygous | 17 | 0.9 – 1.1 |
| CA-500/540 | | |
| --- | --- | --- |
| Genotype FV:Q506 | n | Ratio range (min/max) |
| Negative | 25 | 4.2 – 6.0 |
| heterozygous | 105 | 1.4 – 1.8 |
| homozygous | 15 | 1.0 – 1.1 |
| ACL 900 | | |
| --- | --- | --- |
| Genotype FV:Q506 | n | Ratio range (min/max) |
| Negative | 53 | 3.2 – 5.4 |
| heterozygous | 57 | 1.4 – 2.1 |
| homozygous | 14 | 0.9 – 1.2 |
# N. Proposed Labeling:
The labeling is sufficient and satisfies the requirements of 21 CFR section 809.10.
# O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.