SONOSURG ULTRASONIC SURGICAL SYSTEM

K041566 · Aomori Olympus Co., Ltd. · LFL · Sep 8, 2004 · SU

Device Facts

Record IDK041566
Device NameSONOSURG ULTRASONIC SURGICAL SYSTEM
ApplicantAomori Olympus Co., Ltd.
Product CodeLFL · SU
Decision DateSep 8, 2004
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

The SonoSurg Ultrasonic Surgical System is indicated for use in surgical procedure in neurosurgery where dissection, fragmentation, emulsification and aspiration of soft tissue are desirable. This system consists of the SonoSurg Generator (SonoSurg-G2 set), SonoSurg Irrigation Unit (SonoSurg-IU), SonoSurg Transducer (SonoSurg-T2L-MS), and SonoSurg ultrasonic surgical instruments, and is designed to be used with a separate surgical suction source. Do not use this system for any purpose other than its intended use.

Device Story

SonoSurg Ultrasonic Surgical System is an electromechanical surgical tool for neurosurgery. System components include a generator (SonoSurg-G2), irrigation unit (SonoSurg-IU), transducer (SonoSurg-T2L-MS), and ultrasonic surgical instruments. Device operates by converting electrical energy into ultrasonic vibrations at the instrument tip to dissect, fragment, and emulsify soft tissue; integrated irrigation and external suction source facilitate tissue removal. Used in clinical surgical settings by neurosurgeons. Output is mechanical tissue disruption and aspiration, allowing precise tissue removal during surgery. Benefits include controlled tissue fragmentation and aspiration, aiding surgical access and resection.

Clinical Evidence

Bench testing only; no clinical data required for evaluation of safety and efficacy.

Technological Characteristics

Ultrasonic surgical system; components include generator, irrigation unit, transducer, and instruments. Meets IEC 60601-1, IEC 60601-1-1, and IEC 60601-1-2 safety standards. Patient-contacting materials verified via ISO 10993 biocompatibility testing.

Indications for Use

Indicated for neurosurgical procedures requiring dissection, fragmentation, emulsification, and aspiration of soft tissue.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K041566 # SEP - 8 2004 ## SMDA 510(k) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21CFR, Part 807, Subpart E, Section 807.92. ## A. GENERAL INFORMATION ## 1. Applicant Name & Address: Registration Number: #### 2. Initial Importer Name & Address: Registration Number: ## 3. Submission Correspondent Name, Address, Tel & Fax: Aomori Olympus Co., Ltd. 2-248-1 Okkonoki Kuroishi-shi, Aomori-ken, Japan 036-0357 9614641 OLYMPUS AMERICA Inc. Two Corporate Center Drive, Melville, NY 11747-3157 2429304 OLYMPUS AMERICA Inc. Two Corporate Center Drive, Melville, NY 11747-3157 Tina Steffanie-Oak Associate Manager, Regulatory Affairs/Clinical Monitor 631-844-5477 631-844-5416 2429304 Registration Number: ## B. DEVICE IDENTIFICATION - 1. Common/Usual Name Ultrasonic Surgical System - 2. Device Name SonoSurg Ultrasonic Surgical System - 3. Class, Classification Number and Classification Name No classification, Ultrasonic Surgical Instrument, 80LFL ## C. IDENTIFICATION OF LEGALLY MARKETED DEVICES WHICH WE CLAIM SUBSTANTIAL EQUIVALENCE The following list identifies the predicate devices: {1}------------------------------------------------ | Model | 510(k)# | Manufacturer | Class | Product<br>Code | |----------------------------------------------------|----------------------------------------------|------------------------|-------------------|------------------| | SonoSurg System | #K032066 | Olympus<br>Corporation | No Classification | 80-LFL | | SonoSurg Trocar XT3900 System | #K000095 | Olympus<br>Corporation | II | 80-LFL | | Ultrasonic Surgical System | #K021962<br>#K031710<br>#K031523<br>#K031305 | Olympus<br>Corporation | No Classification | 80-LFL | | CUSA Excel Ultrasonic Surgical<br>Aspirator System | #K981262 | Valleylab | No Classification | 80-LFL<br>84-LBK | | SONOPET UST-2001 Ultra<br>Surgical Aspirator | #K010309 | Miwatec | No Classification | 80-LFL | | Ultrasonic Surgical System USU | #K962952 | Olympus<br>Corporation | No Classification | 84-LBK | ## D. DEVICE DESCRIPTION ## 1. Summary The subject SonoSurg is an ultrasonic surgical device designed to be used with Generator Irrigation Unit and Transducer and ultrasonic surgical instruments to dissect, fragment, enulsify and aspirate tissue in the neurosurgical field. This system consists of the SonoSurg Senerator (SonoSurg-G2 set), SonoSurg Imigation Unit (SonoSurg-IU), SonoSurg-Transducer(SonoSurg-T2L-MS), and SonoSurg ultrasonic surgical instruments, and is designed to be used with a separate surgical suction source. #### 2. Design This System has been designed, manufactured and tested in compliance with voluntary safety standards. It meets the requirement of IEC 60601-1, IEC 60601-1-1 and IEC 60601-1-2. ## 3. Materials Certificate of Identical Materials is included in Attachment 5. Concerning the patient contacting part, some materials are identical with the predicate device. While the others are not identical with the predicate device, biocompatibility testing was performed in accordance with ISO 1093. 1. The test data has shown in Attachment 5-A. # 4. Intended Use of the device: Indications for Use of the SonoSurg Ultrasonic Surgical System The SonoSurg Ultrasonic Surgical System is indicated for use in surgical System neurosurgery where dissection, fragmentation, emulsification and aspiration of soft tissue are desirable. This system consists of the SonoSurg Generator (SonoSurg-G2 set), SonoSurg Irrigation Unit (SonoSurg-IU), SonoSurg Transducer (SonoSurg-T2L-MS), and SonoSurg ultrasonic surgical instruments, and is designed to be used with a separate surgical suction source. Do not use this system for any purpose other than its intended use. ## 5. Summary including conclusion drawn from Non-clinical Tests When compared to the Olympus SonoSurg System (#K032066), CUSA EXcel Ultrasonic Surgical Aspirator System (#K681262), SONOPET UST-2001 Ultra Surgical Aspirator (#K010309) and Ultrasonic System (#K062952, #K000095, #K021962, #K031710, {2}------------------------------------------------ #K031523, #K031305), SonoSurg Ultrasonic Surgical System does not incorporate any significant changes in the intended use, method of operation, material, or design that could affect safety or effectiveness. Therefore clinical data is not necessary for its evaluation of safety and efficacy. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing health, human services, and well-being. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP = 8 2004 Aomori Olympus Co., Ltd. c/o Ms. Tina Steffanie-Oak Associate Manager, Regulatory Affairs/Clinical Monitor Olympus America, Inc. Two Corporate Center Drive Melville, New York 11747-3157 Re: K041566 Trade/Device Name: SonoSurg Ultrasonic Surgical System Regulatory Class: Unclassified Product Code: LFL Dated: June 8, 2004 Received: June 10, 2004 Dear Ms. Steffanie-Oak: We have reviewed your Section 510(k) premarket notification of intent to market the device we nave roviewed your becemined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmetic Free, increating subject to the general controls provisions of the Act. The r ou may, aterere, mains of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or uny 1 with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ Page 2 - Ms. Tina Steffanie-Oak This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Sincerely yours, Mark N. Milkersen Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ 510(k) Number (if known): K041566 Device Name: SonoSurg Ultrasonic Surgical System Indications for Use: Indications for Use of SonoSurg Ultrasonic Surgical System The SonoSurg Olympus Ultrasonic Surgical System is indicated for use in surgical procedure in neurosurgery where dissection, fragmentation, emulsification and aspiration of soft tissue are desirable. This system consists of the SonoSurg Generator (SonoSurg-G2 set), SonoSurg Irrigation Unit (SonoSurg-IU), SonoSurg Transducer (SonoSurg-T2L-MS), and SonoSurg ultrasonic surgical instruments, and is designed to be used with a separate surgical suction source. Do not use this system for any purpose other than its intended use. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use_ 1 (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96) Mark A Milkeusen vision of General, Restorative, and Neurological Devices 510(k) Number K041566
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...