WEST NILE VIRUS IGG INDIRECT ELISA

K041068 · Panbio Limited · NOP · Oct 20, 2004 · Microbiology

Device Facts

Record IDK041068
Device NameWEST NILE VIRUS IGG INDIRECT ELISA
ApplicantPanbio Limited
Product CodeNOP · Microbiology
Decision DateOct 20, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3940
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K041068 · Oct 20, 2004WEST NILE VIRUS IGG INDIRECT ELISAPanbio LimitedRetrospective clinical serum samples from routine laboratory testing; State health laboratory records; Private reference laboratory recordsRetrospective clinical samples were used to assess the performance (sensitivity and specificity) of the PANBIO West Nile Virus IgG Indirect ELISA against established clinical and serological characterizations.Retrospective study; Routine clinical testing; Clinical performance; Serological characterization

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Retrospective analysis of clinical sera; Study Period: 2002-2003Individuals of various ages and genders; includes 100 PRNT-confirmed WNV positive samples and 200 randomly selected normal specimens from routine laboratory testing.; Sample Size: 300; Number of Sites: 1PRNT and WNV IgG IFA ASRPositive and negative presumptive agreement
Retrospective analysis of clinical seraIndividuals of various ages and genders; includes 166 WNV positive and 159 WNV negative samples characterized by IFA slides (ASR).; Sample Size: 325; Number of Sites: 1WNV IgG IFA ASRPositive and negative presumptive agreement

Indications for Use

The PANBIO West Nile Virus IgG Indirect ELISA is for the qualitative presumptive detection of IgG antibodies to West Nile virus in serum. In conjunction with the PANBIO West Nile Virus IgM Capture ELISA, this test is intended as an aid in the clinical laboratory diagnosis of West Nile virus infection in patients with clinical symptoms consistent with encephalitis / meningitis. Positive results must be confirmed by plaque reduction neutralization test (PRNT), or by using the current Centers for Disease Control and Prevention (CDC) guidelines for diagnosis of this disease.

Device Story

The PANBIO West Nile Virus IgG Indirect ELISA is an in vitro diagnostic test used in clinical laboratories to detect IgG antibodies to West Nile virus (WNV) in human serum. The device utilizes microwell test strips coated with purified, inactivated WNV antigen. Patient serum is added to the wells; if WNV-specific IgG antibodies are present, they bind to the antigen. After washing, HRP-conjugated anti-human IgG monoclonal antibody is added, followed by a TMB/H2O2 substrate system. The enzymatic reaction produces a color change (blue, turning yellow upon acidification) proportional to the amount of bound IgG. The intensity is measured to provide a qualitative result. The test is intended to be used alongside the PANBIO West Nile Virus IgM Capture ELISA as an aid in diagnosing WNV infection in symptomatic patients. Positive results require confirmation via Plaque Reduction Neutralization Test (PRNT) or CDC-approved guidelines. The device assists clinicians in identifying WNV infection, facilitating appropriate patient management and public health monitoring.

Clinical Evidence

Clinical performance evaluated using retrospective sera from symptomatic patients and endemic normal populations. Study Site 1 (n=300) showed 79.0% sensitivity (PRNT confirmed) and 90.5% negative presumptive agreement. Study Site 2 (n=325) evaluated encephalitis/meningitis patients (n=34) and WNV IFA-characterized specimens, showing 81.3% positive presumptive agreement in symptomatic patients and 88.0% overall positive presumptive agreement. Analytical specificity tested against 314 specimens from patients with other diseases; cross-reactivity noted with CMV, EBV, and rheumatoid factor.

Technological Characteristics

Indirect IgG ELISA; polystyrene microwell test strips coated with purified/inactivated WNV antigen; HRP-conjugated anti-human IgG monoclonal antibody; TMB/H2O2 substrate; colorimetric detection. Manual or automated assay procedure (excluding background subtraction).

Indications for Use

Indicated for qualitative presumptive detection of IgG antibodies to West Nile virus in serum of patients with clinical symptoms consistent with encephalitis/meningitis. Not for cord blood, neonates, prenatal screening, or general population screening.

Regulatory Classification

Identification

West Nile virus serological reagents are devices that consist of antigens and antisera for the detection of anti-West Nile virus IgM antibodies, in human serum, from individuals who have signs and symptoms consistent with viral meningitis/encephalitis. The detection aids in the clinical laboratory diagnosis of viral meningitis/encephalitis caused by West Nile virus.

Special Controls

*Classification.* Class II (special controls). The special control is FDA's guidance entitled “Class II Special Controls Guidance Document: Serological Reagents for the Laboratory Diagnosis of West Nile Virus.” See § 866.1(e) for the availability of this guidance document.

Submission Summary (Full Text)

{1} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: k041068 B. Purpose for Submission: New device clearance C. Analyte: West Nile Virus IgG Antibody D. Type of Test: Qualitative, ELISA E. Applicant: PANBIO Limited F. Proprietary and Established Names: West Nile Virus IgG Indirect ELISA G. Regulatory Information: a) Regulation section: West Nile Virus, serological reagents (21 CFR 866.3940). b) Classification: Class II Product Code: NOP c) Panel: 83 Microbiology H. Intended Use: a) Intended use(s): The PANBIO West Nile Virus IgG Indirect ELISA is for the qualitative presumptive detection of IgG antibodies to West Nile virus in serum. In conjunction with the PANBIO West Nile Virus IgM Capture ELISA, this test is intended as an aid in the clinical laboratory diagnosis of West Nile virus infection in patients with clinical symptoms consistent with encephalitis / meningitis. Positive results must be confirmed by plaque reduction neutralization test (PRNT), or by using the current Centers for Disease Control and Prevention (CDC) guidelines for diagnosis of this disease. {2} Page 2 of 9 Assay performance characteristics have not been established for testing cord blood, neonate, prenatal screening, general population screening without symptoms of meningioencephalitis or automated instruments b) Indication(s) for use: The PANBIO West Nile Virus IgG Indirect ELISA is for the laboratory diagnosis of West Nile Virus infection in patients with clinical symptoms consistent with meningitis/encephalitis c) Special condition for use statement(s): The device is for prescription use only d) Special instrument Requirements: NA I. Device Description: Indirect IgG ELISA J. Substantial Equivalence Information: a) Predicate device name(s): Focus Technologies West Nile Virus IgG ELISA b) Predicate K number(s): K031953 Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Same indications for use. | PanBio West Nile Virus IgG Indirect ELISA (K041068) | Focus West Nile Virus IgG ELISA (K031953) | | Same target population. | Test persons having symptoms of meningioencephalitis | Test persons having symptoms of meningioencephalitis | | Same ELISA methodology | IgG Indirect ELISA | IgG Indirect ELISA | | Differences | | | | Item | Device | Predicate | | Assay Procedure | PanBio West Nile Virus IgG Indirect ELISA (K041068) | Focus West Nile Virus IgG ELISA (K031953) | | | Does not use Background Subtraction method in the Assay Procedure | Does use Background Subtraction method in the Assay Procedure | {3} # K. Standard/Guidance Document Referenced (if applicable): Class II Special Controls Guidance Document: Serological Reagents for the Laboratory Diagnosis of West Nile Virus. October 30, 2003 # L. Test Principle: Specific antibodies combine with purified and inactivated WNV antigen coated on the polystyrene surface of the microwell test strips (assay plate). Residual serum is removed from the assay plate by washing. HRP-conjugated anti-human IgG monoclonal antibody (Mab) is added to the assay plate. After incubation, the microwells are washed and a colorless substrate system, tetramethylbenzidine/hydrogen peroxide $(\mathrm{TMB} / \mathrm{H}_2\mathrm{O}_2)$ , is added. The substrate is hydrolyzed by the HRP, if present, and the chromogen changes to a blue color. After stopping the reaction with acid, the TMB becomes yellow. Color development is indicative of the presence of WNV antibodies in the test sample. # M. Performance Characteristics (if/when applicable): Analytical performance: # a) Precision/Reproducibility: The reproducibility of the PANBIO West Nile Virus IgG Indirect ELISA was determined by testing 8 sera 3 times each on 3 different assays at one Australian study site and two study sites in the USA. Within-run, between day, between site and total precision were estimated by Analysis of Variance (ANOVA) Type II. The results are presented in the Table below. PANBIO West Nile Virus IgG Indirect ELISA | Sample | n | *Mean | Within | | Between Day | | Between Site | | Total | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | *SD | CV | *SD | CV | *SD | CV | *SD | CV | | Reactive | 27 | 5.83 | 0.25 | 4.3% | 0.00 | 0.0% | 0.00 | 0.0% | 0.24 | 4.2% | | Negative | 27 | 0.19 | 0.02 | 8.8% | 0.00 | 0.0% | 0.08 | 45.5% | 0.07 | 38.8% | | #1 | 27 | 2.24 | 0.16 | 7.0% | 0.12 | 5.1% | 1.08 | 48.3% | 0.92 | 41.0% | | #2 | 27 | 2.94 | 0.20 | 6.9% | 0.16 | 5.5% | 0.13 | 4.3% | 0.26 | 9.0% | | #3 | 27 | 4.07 | 0.32 | 7.9% | 0.00 | 0.0% | 0.52 | 12.8% | 0.53 | 13.1% | | #4 | 27 | 1.48 | 0.11 | 7.7% | 0.11 | 7.6% | 0.08 | 5.6% | 0.16 | 11.0% | | #5 | 27 | 1.53 | 0.06 | 4.2% | 0.06 | 3.9% | 0.18 | 11.9% | 0.17 | 11.2% | | #6 | 27 | 1.23 | 0.09 | 7.4% | 0.03 | 2.3% | 0.16 | 13.3% | 0.17 | 13.5% | | #7 | 27 | 0.82 | 0.05 | 6.2% | 0.02 | 1.9% | 0.13 | 16.0% | 0.12 | 14.8% | | #8 | 27 | 0.84 | 0.07 | 7.9% | 0.00 | 0.0% | 0.12 | 14.7% | 0.12 | 14.4% | All values are calculated from Ratios * $\mathrm{SD} =$ Standard Deviation; $\mathrm{CV} =$ Coefficient of Variation $(\%)$ a. Linearity/assay reportable range: NA b. Traceability, Stability, Expected values (controls, calibrators, or method): NA {4} Page 4 of 9 c. Detection limit: NA d. Analytical specificity: This study consisted of a panel of 314 specimens from patients with confirmed diseases other than WNV. The purpose of this study was to establish the analytical specificity of the PANBIO West Nile Virus IgG Indirect ELISA through the analysis of specimens from patients with diseases that have the potential for cross-reactivity. Each of the specimens included in the study was characterized with respect to disease state prior to analysis of the specimens with the PANBIO West Nile Virus IgG Indirect ELISA. The Table below provides a summary of specimens in the disease panel. | Disease State | Total Specimens^{a} | PANBIO IgG ELISA Results | | | | --- | --- | --- | --- | --- | | | | Pos | Eqv | Pos and Eqv | | Dengue virus | 15 | 14 | 0 | 14/15 | | St. Louis encephalitis | 35 | 25 | 3 | 28/35 | | Japanese encephalitis | 3 | 3 | 0 | 3/3 | | La Crosse encephalitis | 26 | 2 | 0 | 2/26 | | California encephalitis | 11 | 0 | 1 | 1/11 | | Eastern Equine encephalitis | 1 | 0 | 0 | 0/1 | | Varicella-Zoster virus | 15 | 0 | 0 | 0/15 | | Cytomegalovirus | 48 | 7 | 0 | 7/48 | | Epstein-Barr virus | 40 | 7 | 0 | 7/40 | | Enterovirus | 15 | 1 | 0 | 1/15 | | Ross River virus | 39 | 10 | 1 | 11/39^{b} | | Barmah Forest virus | 36 | 10 | 3 | 13/36^{c} | | Rheumatoid Factor | 15 | 4 | 0 | 4/15 | | Anti-Nuclear Antibody | 15 | 2 | 0 | 2/15 | a Sample testing was conducted at PANBIO (Site 5) except as follows: Site 1: Testing on 25 Saint Louis encephalitis, 9 California encephalitis and 1 Eastern Equine encephalitis specimens was conducted at a state health laboratory in Louisiana, USA. Site 2: Testing on 5 Dengue virus, 10 Saint Louis encephalitis and 2 California encephalitis specimens was conducted at a private reference laboratory in Utah, USA. Site 3: Testing on 26 La Crosse encephalitis specimens was conducted at a hospital laboratory in Ohio, USA. Site 6: Testing on 25 Epstein-Barr virus and 33 Cytomegalovirus specimens was conducted at a private research laboratory in Maryland, USA. b 11 of 11 samples that were reactive on PANBIO West Nile virus IgG Indirect ELISA were confirmed flavivirus positive by Western blot. {5} Page 5 of 9 © 9 of 13 samples that were reactive on PANBIO West Nile virus IgG Indirect ELISA were confirmed flavivirus positive by Western blot. Caution: Cross-reactivity has been noted with the PANBIO West Nile Virus IgG assay in specimens containing antibody to cytomegalovirus (CMV), Epstein-Barr Virus (EBV) and rheumatoid factor (RF). Reactive results must be reported with a caution statement regarding possible cross-reactivity with CMV, EBV and RF. ## a. Assay cut-off: The cut-off was determined using a total of 379 characterized positive and 593 characterized negative specimens from three external clinical trials conducted within the USA. Specimens were characterized by immunofluorescence assay (IFA). The cut-off was determined by two-graph receiver operating characteristic analysis (TG-ROC). ## b. Comparison studies: ## a. Method comparison with predicate device: The PanBio Indirect IgG ELISA was compared to reference assays: The plaque-reduction neutralization test (PRNT) and the WN IgG IFA ASR ## b. Matrix comparison: NA ## c. Clinical studies: ## a. Clinical sensitivity: NA ## b. Clinical specificity: NA ## c. Other clinical supportive data (when a and b are not applicable): ## PERFORMANCE CHARACTERISTICS ## Study Site 1: Three hundred (300) retrospective sera from individuals of various ages and both genders were tested at a state health laboratory in Louisiana, USA. The sera include samples from the following groups: 100 samples characterized as positive for WN virus by PRNT and 200 randomly selected normal specimens from routine laboratory testing not known to have a flavivirus related illness. These samples were masked and tested on the PANBIO West Nile Virus IgG Indirect ELISA and the results were compared to the clinical and serological characterization of the samples to determine performance of the assay. The data is summarized in Table 1. {6} Page 6 of 9 Table 1 – Study Site 1 PANBIO West Nile Virus IgG Indirect ELISA Reactivity with Endemic Normal and WNV PRNT Confirmed Specimens | Specimen Characterisation | PANBIO IgG ELISA Results | | | | | --- | --- | --- | --- | --- | | | Pos | Eqv a | Neg | Total | | Endemic normal specimens (randomly selected) b | 18 d | 1 | 181 | 200 | | West Nile virus Positive c (PRNT confirmed) | 79 | 0 | 21 e | 100 | | Total | 97 | 1 | 202 | 300 | a Retesting of equivocals was not conducted as samples were unavailable. b Randomly selected normal specimens from routine laboratory testing from 2002-2003. Not known to have a flavivirus related illness. Further tested by WNV IgG IFA. c West Nile virus PRNT positive. Collected in 2002. Further tested by WNV IgG IFA ASR. d Nine of the 18 endemic specimens that were positive by PANBIO ELISA were also positive by IgG IFA ASR. Refer to note below. e Ten of the 21 PRNT confirmed positive specimens that were negative by PANBIO ELISA were negative by IgG IFA and positive by PANBIO WNV IgM Capture ELISA. Endemic normal specimens 95% CI* Negative presumptive agreement = 181/200 90.5% 85.6 – 94.2% West Nile virus positive specimens Serological sensitivity (PRNT) = 79/100 79.0% 69.7 – 86.5% *CI = Confidence Study Site 2: Three hundred and twenty-five (325) retrospective sera from individuals of various ages and both genders were tested at a private reference laboratory in Utah, USA. The serum panel was comprised of 166 samples that were characterized positive and 159 samples that were characterized negative for WNV by IFA slides (ASR). Thirty-four of these samples were from patients with clinical symptoms consistent with encephalitis / meningitis, of which 32 were characterized positive and 2 were characterized negative for WNV by IFA. The samples were not masked and were tested by the PANBIO West Nile {7} Page 7 of 9 Virus IgG Indirect ELISA. Assay performance was determined by comparing PANBIO West Nile Virus IgG Indirect ELISA results with the clinical and serological characterization of the samples. The data is summarized in Table 2 and Table 3. Table 2 – Study Site 2 PANBIO West Nile Virus IgG Indirect ELISA Reactivity with Encephalitis / Meningitis Patients | Specimen Characterization | PANBIO IgG ELISA Results | | | | | --- | --- | --- | --- | --- | | | Pos | Eqv a | Neg | Total | | Encephalitis/ meningitis patients (IgG IFA positive) | 26 | 2 | 4 | 32 | | Encephalitis/meningitis patients (IgG IFA negative) | 0 | 0 | 2 | 2 | | Total | 26 | 2 | 6 | 34 | a Retesting of equivocals was not conducted as cut-off was modified following clinical trials. Encephalitic symptoms (IgG IFA positive) 95% CI* Positive Presumptive Agreement = 26/32 = 81.3% 63.6 – 92.8% WNV (presumptive by (IgG IFA negative) Negative Presumptive Agreement = 2/2 = 100.0% 15.8 – 100.0% *CI = Confidence {8} Page 8 of 9 Table 3 – Study Site 2 PANBIO West Nile Virus IgG Indirect ELISA Reactivity with WNV IFA Characterized Specimens | Specimen Characterization | PANBIO IgG ELISA Results | | | | | --- | --- | --- | --- | --- | | | Pos | Eqv a | Neg | Total | | WNV positive (IgG IFA positive) | 146 | 5 | 15 | 166 | | WNV negative (IgG IFA negative) | 15 | 4 | 140 | 159 | | Total | 161 | 9 | 155 | 325 | a Retesting of equivocals was not conducted as cut-off was modified following clinical trials. WNV IFA positive (presumptive) 95% CI* Positive Presumptive Agreement = 286/325 = 88.0% 84.5 – 91.5% Negative Presumptive Agreement = 140/159 = 88.1% 83.0 – 93.1% *CI = Confidence interval d. Clinical cut-off: NA e. Expected values/Reference range: {9} # Expected Values A total of 203 normal prospective sera collected during 2002 from patients of various ages in Ohio and Maryland, USA, were assayed on the PANBIO West Nile Virus IgG Indirect ELISA. The distribution of females was $51\%$ (103/203), and males $49\%$ (100/203). The data in Table 1 illustrates the prevalence of IgG antibodies in different age groups when using the PANBIO West Nile Virus IgG Indirect ELISA. Prevalence of West Nile virus by Age | Age | Total | Equivocala | Positive | Prevalence | FlavivirusReactiveb | | --- | --- | --- | --- | --- | --- | | 0 – 9 | 29 | 0 | 4 | 13.8% | 1 | | 10 – 19 | 24 | 1 | 3 | 16.7% | 0 | | 20 – 29 | 25 | 0 | 4 | 16.0% | 2 | | 30 – 39 | 25 | 0 | 7 | 28.0% | 4 | | 40 – 49 | 25 | 0 | 3 | 12.0% | 2 | | 50 – 59 | 25 | 0 | 3 | 12.0% | 2 | | 60 – 69 | 25 | 0 | 3 | 12.0% | 2 | | ≥ 70 | 25 | 0 | 3 | 12.0% | 1 | | Total | 203 | 1 | 30 | 15.3% | 14 | ${}^{a}$ Retesting of equivocals was not conducted as cut-off was modified following clinical trials. b Number of West Nile virus reactive samples that were positive by Dengue IFA ASR # N. Conclusion: The submitted material in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...