K031953 · Focus Technologies, Inc. · NOP · Oct 22, 2003 · Microbiology
Device Facts
Record ID
K031953
Device Name
WEST NILE VIRUS ELISA IGG, MODEL EL0300G
Applicant
Focus Technologies, Inc.
Product Code
NOP · Microbiology
Decision Date
Oct 22, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3940
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K031953 · Oct 22, 2003
WEST NILE VIRUS ELISA IGG, MODEL EL0300G
Focus Technologies, Inc.
Archived clinical serum samples from state departments of health; Archived clinical serum samples from clinical laboratories; Retrospective clinical samples from patients suspected of encephalitis/meningitis; Samples submitted for non-flavivirus testing
Retrospective clinical samples were used to assess the device's clinical reactivity, cross-reactivity with other pathogens, and performance agreement against reference methods (CDC IgG ELISA and PRNT) in populations suspected of West Nile virus infection or undergoing routine clinical testing.
Reactivity with Encephalitis/Meningitis Patients; Retrospective analysis of archived clinical samples
Patients suspected of viral encephalitis or viral meningitis; Sample Size: 300; Number of Sites: 1
CDC IgG ELISA and plaque reduction neutralization test (PRNT)
Clinical sensitivity and agreement with presumptive CDC IgG ELISA
Reactivity with WNV PRNT Positives; Retrospective analysis of archived clinical samples
Patients with samples screened positive by native antigen ELISA and confirmed by PRNT; Sample Size: 75; Number of Sites: 1
Plaque reduction neutralization test (PRNT)
Serological sensitivity
Reactivity with Non-Flavivirus Test Samples; Retrospective analysis of clinical samples; Study Period: August 2003
Samples submitted for non-flavivirus testing; Sample Size: 476; Number of Sites: 1
CDC West Nile IgG ELISA
Positive and negative agreement
Indications for Use
The Focus Technologies West Nile Virus ELISA IgG is intended for qualitatively detecting IgG antibodies to West Nile virus in human serum. In conjunction with the Focus Technologies West Nile Virus IgM Capture ELISA, the test is indicated for testing persons having symptoms of meningioencephalitis, as an aid in the presumptive laboratory diagnosis of West Nile virus infection. Positive results must be confirmed by neutralization test, or by using the current CDC guidelines for diagnosing West Nile encephalitis. This test is not intended for self-testing, and this test is not FDA cleared nor approved for testing blood or plasma donors. Assay performance characteristics have not been established for automated instruments.
Device Story
Indirect IgG ELISA; detects IgG antibodies to West Nile virus in human serum. Polystyrene microwells coated with recombinant West Nile virus antigen; patient serum incubated; specific IgG binds to antigen; peroxidase-conjugated anti-human IgG added; enzyme substrate/chromogen added for color development. Color change quantified via spectrophotometric optical density (OD) reading; compared to reference cut-off OD. Used in clinical laboratories; performed by trained laboratory personnel. Results aid presumptive diagnosis of West Nile virus infection in symptomatic patients; positive results require confirmation via neutralization test or CDC guidelines. Not for blood/plasma donor screening or self-testing.
Clinical Evidence
Clinical performance evaluated across four study sites. In 300 patients with encephalitis/meningitis, clinical sensitivity was 97.3% (36/37) against confirmed WNV PRNT-positive cases. Negative agreement with presumptive CDC WNV IgG ELISA negative samples was 99.0% (203/205). Cross-reactivity assessed against various pathogens (Dengue, SLE, Japanese encephalitis, etc.). Reproducibility studies (inter-lot, intra/inter-assay, inter-laboratory) performed with CVs ranging from 1.0% to 22.3%.
Technological Characteristics
Indirect IgG ELISA; polystyrene microwells coated with recombinant West Nile virus antigen. Detection via peroxidase-conjugated anti-human IgG and spectrophotometric OD measurement. Manual assay; no automated instrument performance established.
Indications for Use
Indicated for testing human serum in patients with symptoms of meningioencephalitis to aid in the presumptive laboratory diagnosis of West Nile virus infection. Not for self-testing or blood/plasma donor screening.
Regulatory Classification
Identification
West Nile virus serological reagents are devices that consist of antigens and antisera for the detection of anti-West Nile virus IgM antibodies, in human serum, from individuals who have signs and symptoms consistent with viral meningitis/encephalitis. The detection aids in the clinical laboratory diagnosis of viral meningitis/encephalitis caused by West Nile virus.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance entitled “Class II Special Controls Guidance Document: Serological Reagents for the Laboratory Diagnosis of West Nile Virus.” See § 866.1(e) for the availability of this guidance document.
Predicate Devices
PanBio West Nile Virus IgM Capture ELISA (K031703)
Submission Summary (Full Text)
{1}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE
A. 510(k) Number:
k031953
B. Analyte:
West Nile Virus IgG Antibody
C. Type of Test:
Qualitative, ELISA
D. Applicant:
Focus Technologies, Inc
E. Proprietary and Established Names:
West Nile Virus ELISA IgG
F. Regulatory Information:
1. Regulation section:
West Nile Virus, serological reagents (21 CFR 866.3940).
2. Classification:
Class II
3. Product Code:
NOP
4. Panel:
Microbiology (83)
G. Intended Use:
1. Intended use(s):
The Focus Technologies West Nile Virus ELISA IgG is intended for qualitatively detecting IgG antibodies to West Nile virus in human serum. In conjunction with the Focus Technologies West Nile Virus IgM Capture ELISA, the test is indicated for testing persons having symptoms of meningioencephalitis, as an aid in the presumptive laboratory diagnosis of West Nile virus infection. Positive results must be confirmed by neutralization test, or by using the current CDC guidelines for diagnosing West Nile encephalitis. This test is not intended for self-testing, and this test is not FDA cleared nor approved for testing blood or plasma donors. Assay performance characteristics have not been established for automated instruments.
2. Indication(s) for use:
The West Nile Virus ELISA IgG is for the laboratory diagnosis of West Nile Virus infection in patients with clinical symptoms consistent with meningitis /encephalitis.
3. Special condition for use statement(s):
Not Applicable
4. Special instrument Requirements:
Not Applicable
{2}
Page 2 of 9
H. Device Description:
Indirect IgG ELISA
I. Substantial Equivalence Information:
1. Predicate device name(s):
PanBio West Nile Virus IgM Capture ELISA
2. Predicate K number(s):
K031703
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Same indications for use. | Focus West Nile Virus ELISA IgG | PanBio West Nile Virus IgM Capture ELISA |
| Same target population. | The test is indicated for testing persons having symptoms of meningioencephalitis | For the laboratory diagnosis of West Nile virus infection in patients with clinical symptoms consistent with meningitis/encephalitis. |
| Differences | | |
| Item | Device | Predicate |
| Different antigens used in the assay | Focus West Nile Virus IgM Capture ELISA | PanBio West Nile Virus IgM ELISA |
| Different ELISA methodology | Recombinant antigen | Inactivated native virus |
| | Indirect ELISA | IgM Capture ELISA |
J. Standard/Guidance Document Referenced (if applicable):
Not Applicable
K. Test Principle:
In the Focus Technologies West Nile Virus ELISA IgG assay, the polystyrene microwells are coated with recombinant West Nile virus antigen. Diluted serum samples and controls are incubated in the wells to allow specific antibody present in the samples to react with the antigen. Nonspecific reactants are removed by washing, and peroxidase-conjugated anti-human IgG is added and reacts with specific IgG. Excess conjugate is removed by washing. Enzyme substrate and chromogen are added, and the color is allowed to develop. After adding the Stop Reagent, the resultant color change is quantified by a
{3}
Page 3 of 9
spectrophotometric reading of optical density (OD). Sample optical density readings are compared with reference cut-off OD readings to determine results.
## L. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
Reproducibility studies included Inter-lot Reproducibility, Inter/Intra-assay Reproducibility, and Inter-laboratory Reproducibility. In each study, two sets of samples were masked duplicates. Focus (Study Site 4) assessed the device's Inter-lot Reproducibility by testing five samples on three separate days with three separate lots. For one lot, the samples were run in triplicate, and run in duplicate with the other two lots. Each of the three lots had a different lot of Antigen and Capture Wells. Focus (Study Site 4) assessed the device's Inter/Intra-assay Reproducibility by testing seven samples in triplicate, once a day, for three days, for a total of 63 data points. A state department of health laboratory located in the northeastern U.S. (Study Site 1), and a clinical laboratory located in the mid-western U.S. (Study Site 2), Focus (Study Site 4), assessed the device's Inter-laboratory Reproducibility. Each of the three laboratories tested seven samples in triplicate on three different days.
| Sample | Inter- & Intra-assay | | | Inter-lot | | Inter-Lab | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | Index Mean | Intra-assay %CV | Inter-assay %CV | Index Mean | Index %CV | Index Mean | Index %CV |
| G6* | 0.23 | 12.2 | 18.2 | 0.30 | 13.1 | 0.32 | 11.6 |
| G2* | 0.29 | 21.2 | 17.3 | 0.34 | 7.5 | 0.35 | 22.5 |
| G5 | 0.65 | 7.9 | 21.3 | 0.73 | 7.2 | 0.69 | 19.0 |
| G7* | 1.14 | 3.5 | 18.2 | 1.30 | 5.7 | 1.21 | 14.1 |
| G1* | 1.22 | 3.2 | 17.1 | 1.36 | 7.0 | 1.25 | 16.4 |
| G4 | 2.44 | 1.0 | 16.2 | 2.79 | 4.1 | 2.47 | 12.8 |
| G3 | 2.98 | 3.9 | 17.3 | 3.37 | 4.1 | 3.10 | 12.6 |
* There were two sets of masked pairs (same sample, different labeled identity): G2 & G6 were one masked pair, and G1 & G7 were the second masked
#### b. Linearity/assay reportable range:
Not Applicable
#### c. Traceability (controls, calibrators, or method):
Not Applicable
#### d. Detection limit:
Not Applicable
#### e. Analytical specificity:
Focus and a state department of health laboratory located in the northeastern U.S. (DOH) (Study Site 1) assessed the device's cross-reactivity with sera that were seropositive to other potentially cross-reactive pathogens (n = 75). The DOH tested the SLE positives, and Focus tested the other sera. The sera were archived and masked. The table below summarizes the data.
{4}
Page 4 of 9
# Cross-reactivity
| Specimens Characterized by Reference Assays | Focus WNV IgG ELISA Results | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | Site | Neg | Eqv | Pos | Total | % Positive |
| Dengue virus (secondary infections) | 4 | 1 | 0 | 19 | 20 | 95.0% (19/20)
95%CI 75.1-99.9% |
| Japanese encephalitis virus | 4 | 14 | 3 | 3 | 20 | 30.0% (6/20)
95%CI 11.9-54.3% |
| St. Louis encephalitis virus | 1 | 8 | 1 | 11 | 21 | 57.1% (12/21)
95%CI 34.0-78.2% |
| Yellow fever virus | 4 | 11 | 4 | 5 | 20 | 45.0% (9/20)
95%CI 23.1-68.5% |
| Alphavirus (Sindbis & Eastern equine viruses) | 4 | 15 | 0 | 2 | 17 | 11.8% (2/17)
95%CI 0.1-36.4% |
| Bunyavirus (Jamestown Canyon & La Crosse) | 1 | 12 | 2 | 1 | 15 | 20.0% (3/15)
95%CI 4.3-48.1% |
| Herpes simplex virus type 1 | 4 | 55 | 0 | 5 | 60 | 8.3% (5/60)
95%CI 2.8-18.4% |
| Epstein-Barr virus | 4 | 11 | 0 | 1 | 12 | 8.3% (1/12)
95%CI 0.2-38.5% |
| Cytomegalovirus | 4 | 16 | 0 | 4 | 20 | 20.0% (4/20)
95%CI 5.7-43.7% |
| Echovirus/Poliovirus | 4 | 18 | 1 | 1 | 20 | 10.0% (2/20)
95%CI 1.2-31.7% |
| Borrelia burgdorferi (Lyme disease) | 4 | 17 | 1 | 2 | 20 | 15.0% (3/20)
95%CI 3.2-37.9% |
Because of the high degree of cross-reactivity with specimens containing antibodies to CMV and bunyaviruses, the following warning has been placed in the package insert.
{5}
Page 5 of 9
Caution: IgG assay cross-reactivity has been noted with some specimens containing antibody to cytomegalovirus (CMV) and bunyaviruses, e.g., LaCrosse virus. Reactive results must be reported with a caution statement regarding possible cross-reactivity with CMV and bunyaviruses, e.g., La Crosse virus.
f. Assay cut-off:
Cut-off Development. In designing the assay, the assay Cut-off was established to slightly favor specificity over sensitivity by using 217 sera consisting of 3 different serum panels: 1) 136 sera submitted for West Nile testing and positive with an in-house WNV IgG ELISA native antigen West Nile ELISA IgG); 2) 61 sera submitted for West Nile testing and negative with an in-house WNV IgG ELISA; and 3) 20 blood donors. The Focus West Nile IgG was positive with 90.4% (122/135) of the in-house WNV IgG ELISA positive samples, negative with 100% (61/61) of the in-house WNV IgG ELISA negative samples, and negative with 100% (20/20) of the blood donor samples.
2. Comparison studies:
a. Method comparison with predicate device:
The Focus West Nile Virus ELISA IgG was compared with two reference assays: The plaque-reduction neutralization test (PRNT) and the CDC IgG ELISA.
b. Matrix comparison:
Not Applicable
3. Clinical studies:
a. Clinical sensitivity:
Not Applicable
b. Clinical specificity:
Not Applicable
c. Other clinical supportive data (when a and b are not applicable):
Study Site 1: Focus Reactivity with Encephalitis/Meningitis Patients (n = 300)
A state department of health laboratory located in the northeastern U.S. assessed the device's reactivity from encephalitis/meningitis patients (n = 300). Patients were suspected of having either viral encephalitis or viral meningitis. Viral encephalitis criteria included: 1) fever; 2) altered mental status and/or other evidence of cortical involvement; and 3) CSF pleocytosis with predominant lymphocytes and/or elevated protein and a negative gram stain and culture. Viral meningitis criteria included: 1) fever; 2) headache, stiff neck and/or other meningeal signs; and 3) CSF pleocytosis with predominant lymphocytes and/or elevated protein and a negative gram stain and culture.
The sera were sequentially submitted to the laboratory, archived, and masked. The reference methods were CDC IgG ELISA, and plaque reduction neutralization test (PRNT) for West Nile virus. Of 300 encephalitis/meningitis patients, 205 were classified as presumed negative patients (CDC IgG ELISA negative), 37 classified as confirmed positive West Nile encephalitis patients (CDC IgG ELISA positive, WNV PRNT
{6}
Page 6 of 9
positive), 4 presumed positive flavivirus encephalitis patients (CDC IgM positive, PRNT negative), one confirmed dengue positive (CDC IgG ELISA positive, dengue PRNT positive), and 53 unclassified because the CDC IgG ELISA results were indeterminant or equivocal. The Focus IgG ELISA was positive with 97.3% (36/37) of the confirmed positive WNV encephalitis patients (including 1 Focus equivocal calculated as negative). The Focus IgG ELISA was positive with 100% (4/4) of the presumed positive WNV encephalitis patients, and positive with the one dengue positive patient. The Focus IgG ELISA was negative with 99.0% (203/205) of the presumed negative patients (including 1 Focus equivocal counted positive). The 53 unclassified patients were excluded from the calculations.
Study Site 1: Focus Reactivity with Encephalitis/Meningitis Patients (n = 300)*
| Specimens Characterized by Reference Assays | Focus WNV IgG ELISA Results | | | | |
| --- | --- | --- | --- | --- | --- |
| | Neg | Eqv | Pos | Total | % |
| Clinical sensitivity (encephalitis or meningitis symptoms, CDC IgG ELISA positive and WNV PRNT positive) | 0 | 1 | 36 | 37 | 97.3% (36/37) 95%CI 85.8-99.9% |
| Agreement with the presumptive CDC IgG ELISA | 203 | 1 | 6 | 210 | Positive† 100% (5/5) 95%CI 47.8-100% Negative 99.0% (203/205) 95%CI 96.5-99.9% |
*Excluding 53 CDC IgG ELISA results (49 indeterminant and 4 equivocal samples).
† One of the presumptive positive samples was dengue PRNT positive and the other 4 presumptive positives were negative with WNV, dengue and SLE PRNT.
# Study Site 2: Focus Reactivity with WNV PRNT Positives (n = 75)
A clinical laboratory located in the mid-western U.S. assessed the device's reactivity with 75 retrospective sera that were screened positive (by Focus) with a West Nile virus native antigen ELISA, and confirmed West Nile positive by plaque reduction neutralization test (PRNT). The sera were sequentially submitted to the laboratory, archived, and masked. The Focus IgG ELISA was positive with 36.0% (27/75) of the 75 PRNT positives (calculating 4 equivocals as negative), equivocal with four samples, and negative with 44 samples.
{7}
Page 7 of 9
## Study Site 2: Focus Reactivity with WNV PRNT Positives (n = 75)
| Specimens Characterized by Reference Assays | Focus WNV IgG ELISA Results | | | | |
| --- | --- | --- | --- | --- | --- |
| | Neg | Eqv | Pos | Total | % |
| Serological sensitivity positive (WNV PRNT positive) | 44 | 4 | 27 | 75 | 36.0% (27/75)
95%CI 25.2-92.3% |
## Study Site 3: Focus Reactivity with West Nile IFA Negatives (n=157)
A clinical laboratory located in the southwestern U.S. assessed reactivity with 157 retrospective West Nile IFA negative samples. The Focus IgG ELISA was 96.8% (152/157) negative with WNV IgG IFA negative samples (including two equivocals calculated as positive), and positive with three samples.
## Study Site 3: Focus Reactivity with West Nile IFA Negatives (n=157)
| Specimens Characterized by Reference Assays | Focus WNV IgG ELISA Results | | | | |
| --- | --- | --- | --- | --- | --- |
| | Neg | Eqv | Pos | Total | % |
| Negative agreement with presumptive WNV IFA negative | 152 | 2 | 3 | 157 | 95.6% (152/157)
95% CI 91.1-98.2% |
## Study Site 4: Focus Reactivity with Suspected Encephalitis/Meningitis Patients (n = 50)
Focus assessed the device's reactivity with 50 sera from patients suspected of encephalitis/meningitis. A U.S. federal government laboratory provided the retrospective and masked sera. One sample was confirmed positive by West Nile PRNT, and the other 49 were presumptively negative (CDC ELISA) for arboviruses present in North America (La Crosse virus, Eastern Equine encephalitis virus, Saint Louis encephalitis virus and WNV). The Focus IgG ELISA was negative with 95.6% (47/49) of the WNV negative samples, and positive with the one positive confirmed by West Nile PRNT.
## Study Site 4: Focus Reactivity with Suspected Encephalitis/Meningitis Patients (n = 50)
| Specimens Characterized by Reference Assays | Focus WNV IgG ELISA Results | | | | |
| --- | --- | --- | --- | --- | --- |
| | Neg | Eqv | Pos | Total | % |
| Serological sensitivity (CDC IgG ELISA positive and WNV PRNT positive) | 0 | 0 | 1 | 1 | 100% (1/1)
95% CI NA |
| Negative agreement with presumptive CDC IgG ELISA negative | 47 | 0 | 2 | 49 | 95.9%
(47/49)
95% CI 86.0-99.5% |
{8}
Page 8 of 9
# Study Site 4: Focus Reactivity with Non-Flavivirus Test Samples (n = 476)
Focus assessed the device's reactivity with 476 samples prospectively collected from North America during August 2003. The samples had been submitted to a clinical laboratory located in Southern California for non-flavivirus tests (e.g., test for other infectious diseases). Positive samples were tested with a CDC West Nile IgG ELISA. The Focus West Nile ELISA IgG was negative with 96.8% (426/440) of the CDC ELISA negative samples (including 14 Focus equivocals calculated as positive), and was positive with 100% (21/21) of the CDC ELISA positive samples. Fifteen CDC ELISA IgG indeterminant samples were excluded from performance calculations.
Study Site 4: Focus Reactivity with Non-Flavivirus Test Samples (n = 476)*
| Specimens Characterized by Reference Assays | Focus WNV IgG ELISA Results | | | | |
| --- | --- | --- | --- | --- | --- |
| | Neg | Eqv | Pos | Total | % |
| Positive agreement with presumptive CDC IgG ELISA positive | 0 | 0 | 21 | 21 | 100% (21/21) 95%CI 83.9-100% |
| Negative agreement with presumptive CDC IgG ELISA negative | 426 | 14 | 0 | 440 | 96.8% (426/440) 95% CI 94.7-98.3% |
*Excludes 15 CDC IgG ELISA indeterminant samples.
4. Clinical cut-off:
Not Applicable
5. Expected values/Reference range:
The prevalence of West Nile antibodies varies depending on age, geographic location, testing method used, and other factors. A community based serosurvey for West Nile infection conducted in New York in 2000 found that 0.2% (5/2433) of persons tested overall had antibodies indicating recent West Nile infection, and that 1.1% (2/176) of persons reporting a recent headache and fever had antibodies indicating a recent West Nile infection. Two serosurveys conducted in New York City (NYC) in 1999 and 2000 showed that approximately 1 in 150 infections (<1%) resulted in meningitis or encephalitis. The NYC results are consistent with a 1996 Romanian serosurvey indicating that 1:140 to 1:320 infections resulted in meningitis or encephalitis.
# Prevalence in Samples Submitted for Non-Flavivirus Testing (n=476)
Focus assessed reactivity with 476 samples prospectively collected from North America during August 2003. The samples had been submitted to a clinical laboratory located in Southern California for non-flavivirus tests (e.g., tests for other infectious diseases). The samples consisted of 64.1% females, 30.5% males, and 1.5% samples from persons of unspecified gender.
{9}
Page 9 of 9
IgG Prevalence with Samples Submitted for Non-Flavivirus Testing (n = 476)
| Age | Neg | Eqv | Pos | % Positive | 95% CI |
| --- | --- | --- | --- | --- | --- |
| 0 to 9 | 20 | 0 | 4 | 16.7% (4/24) | 4.7-37.4% |
| 10 to 19 | 25 | 2 | 2 | 6.9% (2/29) | 0.9-22.8% |
| 20 to 29 | 63 | 4 | 3 | 4.3% (3/70) | 0.9-12.0% |
| 30 to 39 | 76 | 1 | 5 | 6.1% (5/82) | 2.0-13.7% |
| 40 to 49 | 69 | 1 | 8 | 10.3% (8/78) | 4.5-19.2% |
| 50 to 59 | 47 | 1 | 3 | 5.9% (3/51) | 1.2-16.2% |
| 60 to 69 | 32 | 1 | 6 | 15.4% (6/39) | 5.9-30.5% |
| 70 to 79 | 28 | 1 | 5 | 14.7% (5/34) | 5.0-31.1% |
| 80+ | 15 | 1 | 2 | 11.1% (2/18) | 1.4-34.7% |
| Unknown | 41 | 2 | 8 | 15.7% (8/51) | 7.0-28.6% |
| Overall | 416 | 14 | 46 | 9.7% (46/476) | 7.2-12.7% |
M. Conclusion:
The data demonstrated that there was very good agreement between the reference assays and the Focus West Nile Virus ELISA IgG assay. It is believed that the above information demonstrates that the Focus West Nile Virus ELISA IgG is substantially equivalent to the PanBio West Nile Virus IgM ELISA. When the Focus West Nile Virus ELISA IgG is used according to its directions for use, it should be safe and effective for the indications for use claimed.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.