K041018 · Doctor'S Research Group, Inc. · HHT · Oct 18, 2004 · Obstetrics/Gynecology
Device Facts
Record ID
K041018
Device Name
DRG SOFTRELEASE
Applicant
Doctor'S Research Group, Inc.
Product Code
HHT · Obstetrics/Gynecology
Decision Date
Oct 18, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.4530
Device Class
Class 2
Indications for Use
DRG SoftRelease is intended for the collection of cervical cells for Pap smear analysis and/or for detecting sexually transmitted disease (STD).
Device Story
DRG SoftRelease is a single-use, disposable, manual gynecological instrument for cervical cell collection. Device consists of a handle with a release mechanism and interchangeable attachments: a brush (bristles on a wire shaft) and a plastic, curve-shaped spatula. Physician or clinician operates device in a clinical setting. For brush use, device is inserted into the cervix, rotated once, and removed. For spatula use, the blunt edge is placed against the cervix, rotated, and removed. Collected cytological material is transferred for analysis. Device facilitates sample collection for Pap smears and STD testing.
Indicated for the collection of cervical cells for Pap smear analysis and/or detection of sexually transmitted diseases. Contraindicated for use in pregnant women.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
{0}------------------------------------------------
15 K041018
# OCT 1 8 2004
Image /page/0/Picture/3 description: The image shows the logo for Doctors Research Group, Inc. The logo consists of the letters "DRG" stacked on top of each other inside of a black circle. Below the circle, the words "Doctors Research Group, Inc." are written in a serif font. The text is left-aligned.
Doctors Research Group, Inc.
'physicians working together
### Premarket Notification 510(k) Summary of Safety and Effectiveness DRG SoftRelease
#### Submitter:
Doctors Research Group, Inc. 50 Altair Avenue Plymouth, CT 06782 (tel) 860-283-1479 (fax) 860-283-1589
Establishment Registration Number: 1226001
#### Contact Person:
Edward H. Goldman Doctors Research Group, Inc. 50 Altair Avenue Plymouth, CT 06782 (tel) 860-283-1479 (fax) 860-283-1589
### Summary Preparation Date
April 9, 2004
#### Device Information
| Trade name: | DRG SoftRelease |
|------------------------------|--------------------------------------------------------------|
| Common name: | Cervical Cell Scraper / Sampler |
| Classification name: | Cytological Cervical Spatula |
| Device Classification Panel: | Obstetrical - Gynecological Specialized Manual<br>Instrument |
| Regulation number: | 21CFR Part 884.4530 |
| Class: | II |
| Product Code: | HHT |
{1}------------------------------------------------
## 510(k) Summary (continued)
Predicate Device
| Rovers Spatula | K002520 | Rovers Medical Devices BV |
|-----------------------|---------|---------------------------|
| Digene Cervical Brush | K971586 | Digene Corp. |
### DRG Device Description
The DRG SoftRelease is a single use, disposable, manual gynecological device intended for the The DNO of cervical cells for Pap smear analysis and/or for detecting sexually transmitted concellon of on noal oulle for a handle with a release mechanism, a spatula discuse (o r b). The donto common handle / release combination allows the attachment, of a brush attach how h for use in cervical or vaginal wall cell collection. The attachnent of oknor the opatals of the cell collector attachments from the device handle.
The brush is composed of bristles joined to a wire shaft. For collecting cervical cells, the brush is inserted into the cervix until only the last row of bristles are visible, rotated once via the handle, and removed from the site for transfer of the cytological material.
The spatula is made of plastic and is curve-shaped, similar to the widely uses Ayre spatula. For r ric opatilia is made of the spatula's blunt edge is placed against the cervix, rotated with gentle oonooting on the handle, and removed from the site for transfer of the cytological material.
#### Intended Use
DRG SoftRelease is intended for the collection of cervical cells for Pap smear analysis and/or for detecting sexually transmitted disease (STD).
#### Basis of Equivalence
The DRG SoftRelease is substantially equivalent to the Rovers Spatula and Digene Cervical Brush based on the indications for use, design, materials, and principles of operation.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird-like figure.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
### OCT 1 8 2004
Mr. Edward H. Goldman Director, Quality Assurance Doctors Research Group, Inc. 50 Altair Avenue :PLYMOUTH CT 06782
Re: K041018
Trade/Device Name: DRG SoftRelease Regulation Number: 21 CFR 884.4530 Regulation Name: Obstetric-gynecologic specialized manual instrument
Regulatory Class: II Product Code: 85 HHT Dated: August 11, 2004 Received: August 16, 2004
·Dear Mr. Goldman:
We have reviewed your Section 510(k) premarket notification of intent to market the device in We have reviewed your Section 910(t) prohentially equivalent (for the indications for use stated in above and have determined the devices marketed in interstate commerce prior to the enclosure) to regally marketed predical Device Amendments, or to devices that have been May 28, 1970, the clacultion date of the Federal Food, Drug, and Cosmetic Act (Act) that reclassified in accordance with the provisions of application (PMA). You may, therefore, market the do not require approval of a prematics approvide approvide (The general controls provisions of the Act. device, subject to the general controls provinsities, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket If your device is classifica (Sec above) into existing major regulations affecting your Approval), It may be subject to such additions, Title 21, Parts 800 to 898. In addition, FDA device can of tound in the coments concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that Please be advised that FDA s issualled of a succeannality with other requirements of the Act of any FDA has made a delemination adal 10. 10. 10. 10. 1. 0. must comply with all the Federal statues and regulations administered of occistration and listing (21 CFR Part 807); labeling Act's requirements, including, but not minted to regulations as set forth in the quality systems (QS) (21 CFR Fart 801), good manatalating product radiation control product radiation control provisions (Sections 531-542 of the Act). 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to begin harketing your aver of your device of your device to a legally premarket notification. THE FDA Intention of Sactian vija
marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire spective advice for your device of our has lastering organ.
contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Also, please note the regulation emilities, "Thisotanas on your responsibilities under the Act from the 807.97). You may outer general international and Consumer Assistance at its toll-free number (800) DVNSon of 301) 443-6597 or at its Internet address http://www.fda.gov/cdrb/dsmaldsmamain.html
Sincerely yours,
Nancy C. brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Doctors Research Group, Inc. 50 Altair Avenue Plymouth, CT 06782 (860) 283-1479
# Statement of Indications For Use
510(k) Number (if Known): 7
Device Name: DRG SoftRelease
Indications for use:
DRG SoftRelease is intended for the collection of cervical cells for Pap smear analysis and/or DRG SoftRelease is intended for the oblicolion of convice is not intended for use in pregnant women.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Рег 21 CFR 801.109)
OR
Over-The-Counter Use
Nancy C Broadon
(Division Sign-Off) Division of Reproductive. Abdominal and Radiological Devi 510(k) Number_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.