DIGENE CERVICAL BRUSH
Device Facts
| Record ID | K971586 |
|---|---|
| Device Name | DIGENE CERVICAL BRUSH |
| Applicant | Digene Corp. |
| Product Code | HHT · Obstetrics/Gynecology |
| Decision Date | Sep 19, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 884.4530 |
| Device Class | Class 2 |
Indications for Use
The Digene Cervical Brush is intended for the collection of cervical cells for Pap smear analysis and/or for sexually transmitted disease (STD) testing. The Digene Cervical Brush is not intended for use in pregnant women.
Device Story
Digene Cervical Brush is a manual specimen collection tool; features truncated cone-shaped brush head with bristles on wire shaft bonded to plastic handle. Used by clinicians to collect cervical cells; inserted 1.0 to 1.5 cm into cervical os; rotated three full turns counter-clockwise; withdrawn. Collected specimen is spread on glass slide for Pap smear or placed in transport tube for STD testing. Device facilitates diagnostic sampling; benefits patient by enabling screening for cervical abnormalities and STDs.
Clinical Evidence
No clinical data provided; substantial equivalence based on design and material comparison to predicate.
Technological Characteristics
Truncated cone-shaped brush head; bristles attached to wire shaft; plastic handle; distal wire end embedded in solidified glue. Manual operation; non-powered.
Indications for Use
Indicated for collection of cervical cells for Pap smear analysis and/or STD testing in non-pregnant women.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
Predicate Devices
- Medscand Cytobrush (K832986)