PUMP CONNECTIVITY INTERFACE

K040985 · Baxter Healthcare Corp · FRN · Nov 30, 2004 · General Hospital

Device Facts

Record IDK040985
Device NamePUMP CONNECTIVITY INTERFACE
ApplicantBaxter Healthcare Corp
Product CodeFRN · General Hospital
Decision DateNov 30, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5725
Device ClassClass 2

Intended Use

The Pump Connectivity Interface is a computerized wireless patient information and medication management tool that is integrated into an existing hospital network infrastructure and allows communications with external devices, including personal computers (PCs), Personal Digital Assistants (PDAs), hospital monitoring systems and Hospital Information Management Systems (HIMS). Communication of data includes infusion parameters, system configuration, history, events, trending, alarms and status. It is intended for use in professional healthcare environments that use electronic, external infusion devices. The Pump Connectivity Interface is intended to provide trained healthcare practitioners access to additional information needed to more safely manage infusion therapy via infusion pumps. It enables remote viewing of pump status within the hospital and enables comparisons of pharmacy-entered prescription information to programmed pump settings. All data entry and validation of infusion parameters is performed by the trained healthcare professional according to a physician's order. The Pump Connectivity Interface also enables secondary notification of pump alerts and alarms to remote users via a personal digital assistant (PDA) or similar device. The Pump Connectivity Interface is intended for use by trained healthcare practitioners in accordance with the instructions provided in Pump Connectivity Interface labeling.

Device Story

System of hardware and software integrating external infusion pumps into hospital network infrastructure; enables wireless communication between pumps and external devices (PCs, PDAs, HIMS, monitoring systems). Inputs: infusion parameters, system configuration, history, events, trending, alarms, status. Outputs: remote display of pump status, secondary alarm notifications to mobile devices. Used in professional healthcare environments by trained practitioners. Facilitates remote monitoring of infusion therapy and comparison of pharmacy-entered prescriptions against pump settings. Does not perform automated data entry; all validation remains with the clinician. Benefits: improved safety through remote visibility and secondary alarm notification.

Clinical Evidence

Bench testing only. Performance data indicate the device meets specified requirements and is substantially equivalent to predicate devices.

Technological Characteristics

Computerized wireless interface system; integrates with hospital network infrastructure; supports communication with PCs, PDAs, and HIMS. Connectivity: wireless network integration. Software-based management tool for infusion pump data.

Indications for Use

Indicated for use by trained healthcare practitioners in professional healthcare environments utilizing electronic, external infusion devices to manage infusion therapy, monitor pump status, and receive secondary notifications of pump alerts/alarms.

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## NOV 3 0 2004 # 510(k) Summary of Safety and Effectiveness #### Submitted by: Jennifer M. Paine Manager, Global Regulatory Affairs Baxter Healthcare Corporation Medication Delivery Rte. 120 and Wilson Road Round Lake, IL 60073 #### Name/Classification of Device: Infusion Pump/ Class II, 80FRN/80MEA - 21 CFR 880.5725 #### Trade Name: Pump Connectivity Interface #### Predicate Devices: Alaris Medley System with MMS B. Braun Horizon Outlook with DoseCom #### Statement of Intended Use: The Pump Connectivity Interface is a computerized wireless patient information and medication management tool that is integrated into an existing hospital network infrastructure and allows communications with external devices, including personal computers (PCs), Personal Digital Assistants (PDAs), hospital monitoring systems and Hospital Information Management Systems (HIMS). Communication of data includes infusion parameters, system configuration, history, events, trending, alarms and status. It is intended for use in professional healthcare environments that use electronic, external infusion devices. The Pump Connectivity Interface is intended to provide trained healthcare practitioners access to additional information needed to more safely manage infusion therapy via infusion pumps. It enables remote viewing of pump status within the hospital and enables comparisons of pharmacy-entered prescription information to programmed pump settings. All data entry and validation of infusion parameters is performed by the trained healthcare professional according to a physician's order. The Pump Connectivity Interface also enables secondary notification of pump alerts and alarms to remote users via a personal digital assistant (PDA) or similar device. The Pump Connectivity Interface is intended for use by trained healthcare practitioners in accordance with the instructions provided in Pump Connectivity Interface labeling. #### Device Description: The Pump Connectivity Interface has been developed to expand upon previously cleared external monitoring features of legally marketed infusion pumps. As described above, the Pump Connectivity Interface is a system of hardware and software that enables external infusion pumps to be integrated into existing hospital network infrastructure. It is intended {1}------------------------------------------------ to provide trained healthcare practitioners with access to additional information needed to more safely manage infusion therapy via external infusion pumps. ### Summary of Technological Characteristics of New Device to Predicate Devices: A comparison of the technological characteristics of the Pump Connectivity Interface to the predicate devices has been performed. The results of this comparison demonstrate that the Pump Connectivity Interface is equivalent to the marketed predicate device in technological characteristics. #### Performance Data: The performance data indicate that the device will meet specified requirements and is substantially equivalent to the predicate devices. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with three stripes representing the agency's mission to protect the health of all Americans and provide essential human services. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 NOV 3 0 2004 Mr. Robert L. Wilkinson, RAC Director, Regulatory Affairs Baxter Healthcare Corporation 1620 Waukegan Road McGaw Park, Illinois 60085 Re: K040985 Trade/Device Name: Pump Connectivity Interface Regulation Number: 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN Dated: November 23, 2004 Received: November 24, 2004 Dear Mr. Wilkinson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Ms. Paine Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours. Susan Runner Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): Device Name: Pump Connectivity Interface #### Indications For Use: The Pump Connectivity Interface is a computerized wireless patient information and medication management tool that is integrated into an existing hospital network infrastructure and allows communications with external devices, including personal computers (PCs), Personal Digital Assistants (PDAs) or similar display/computing devices, hospital monitoring systems, and Hospital Information Management Systems (HIMS). Communication of data includes infusion parameters, system configuration, history, events, trending, alarms and status. It is intended for use in professional healthcare environments that use electronic, external infusion devices. The Pump Connectivity Interface is intended to provide trained healthcare practitioners access to additional information needed to more safely manage infusion therapy via infusion pumps. lt enables remote viewing of pump status within the hospital and enables comparisons of pharmacyentered prescription information to programmed pump settings. All data entry and validation of infusion parameters is performed by the trained healthcare professional according to a physician's order. The Pump Connectivity Interface also enables secondary notification of pump alerts and alarms to remote users via a personal digital assistant (PDA) or similar display/computing device. The Pump Connectivity Interface is intended for use by trained healthcare practitioners in accordance with the instructions provided in Pump Connectivity Interface labeling. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) J. Thm. D.m. (Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices Page 1 of ____________________________________________________________________________________________________________________________________________________________________ 510(k) Number:_长中华ゆみぷ
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...