K040709 · Aktina Medical Physics Corp. · IYE · Apr 28, 2004 · Radiology
Device Facts
Record ID
K040709
Device Name
PROTOS SOFTWARE
Applicant
Aktina Medical Physics Corp.
Product Code
IYE · Radiology
Decision Date
Apr 28, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Software as a Medical Device
Intended Use
The intended purpose of the Protos Software is to provide a Radiation Therapy Quality Control (QC) tool that allows for the automated and rapid comparison of predicted and measured radiation dose values. The software is intended for the OC of external beam radiation treatments. Measured data will be acquired from radiographic films. Predicted data will be calculated in a Radiotherapy Treatment Planning System (RTPS) and transferred to the software via a Digital Imaging and Communications in Medicine (DICOM) protocol.
Device Story
Protos Software serves as a quality control tool for external beam radiation therapy. It inputs predicted radiation dose data from a Radiotherapy Treatment Planning System (RTPS) via DICOM and measured dose data from radiographic films scanned from an IMRT phantom. The software performs automated, rapid comparisons between these predicted and measured values. Operated by trained radiation therapy personnel on a Microsoft Windows-based computer, the output allows clinicians to verify the accuracy of the radiation delivery process prior to patient treatment. This verification helps ensure the precision of radiotherapy, potentially benefiting patients by confirming that the delivered dose matches the planned dose.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Dosimetric analysis software; operates on Microsoft Windows or equivalent environment. Inputs: DICOM data from RTPS and scanned radiographic film data. Function: Automated comparison of predicted vs. measured radiation dose values. No patient contact; no biocompatibility testing required.
Indications for Use
Indicated for quality control of external beam radiation therapy delivery. Used by trained radiation therapy personnel to compare predicted dose data from treatment planning systems against measured dose data obtained from radiographic films (IMRT phantoms) prior to patient treatment.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
K162577 — Dosimetry Check Version 5 Release 1 · Math Resolutions, LLC · Nov 18, 2016
K132605 — DOSIMETRY CHECK VERSION 4 RELEASE 1 · Math Resolutions, LLC · Nov 14, 2013
K141230 — DOSELAB PRO · Mobius Medical Systems, LP · Aug 27, 2014
K150848 — Dose Calculator · Sun Nuclear Corporation · May 20, 2015
K201798 — myQA iON · Iba Dosimetry GmbH · Jul 17, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
KD40709
APR 2 8 2004
# 510(k) Summary Safety and Effectiveness Summary In Compliance with 513 (i) of the Federal Food, Drug and Cosmetic Act
#### General Provisions 1.
Device Trade Name: Protos Software Common Name: Dosimetry Analysis Software
Applicant Name and Address:
AKTINA Medical Physics Corporation 360 North Route 9 W Congers, New York, 10920 Phone: 914-268-0101 FAX: 914-268-1700 Registration Number: 2436865
#### 2. Name of Predicate Device
Radiological Imaging Technology, Inc., RIT 113 Film Analysis System, K935928
Any statement made in conjunction with this submission regarding substantial equivalence to any other product only relates to whether the product can be lawfully marketed without pre-market approval or reclassification and is not to be interpreted as an admission or used as evidence in patent infringement litigation. As the Commissioner of the FDA has indicated, "...a determination of substantial equivalence under the Federal Food, Drug, and Cosmetic Act relates to the fact that the product can be lawfully marketed without pre-market approval or reclassification. This determination is not intended to have any bearing whatsoever on the resolution of patent infringement suits." 42 Fed. Reg. 42,520 et seg. (1977),
# 3. Classification
This device is classified as a class II device according to 21 CFR 892.5050
#### 4. Performance Standards
The FDA under Section 514 of the Food, Drug and Cosmetic Act has not established performance standards for this product.
{1}------------------------------------------------
# Intended Use and Device Description 5.
The intended purpose of the Protos Software is to provide a Radiation Therapy Quality Control (QC) tool that allows for the automated and rapid comparison of predicted and measured radiation dose values. The software is intended for the OC of external beam radiation treatments. Measured data will be acquired from radiographic films. Predicted data will be calculated in a Radiotherapy Treatment Planning System (RTPS) and transferred to the software via a Digital Imaging and Communications in Medicine (DICOM) protocol.
# Technological Characteristics 6.
The Aktina Medical Physics' Protos Software has the same technological characteristics as the RIT 113 Film Analysis System (Radiological Imaging Technology, Inc., K935928). Both systems:
- Consist of Dosimetric Analysis software a.
- b. Import predicted radiation dosimetry data from treatment planning systems
- 0Import measured radiation dosimetry data imported from a film scanner; C. radiographic films are obtained from an IMRT phantom.
- d. Are used as Radiation Therapy Quality Control (OC) tools that allow for the automated and rapid comparison of predicted and measured radiation dose values
- Require personnel trained in the use of Radiation Therapy equipment. c.
- f. Require a Microsoft Windows or equivalent computer environment.
## 7. Biocompatibility
The Aktina Medical Corporation Protos Software does not come into contact with the patient and therefore does not require biocompatibility testing.
## 8. Summary of Substantial Equivalence
This device is similar in design and intended use and performance characteristics to the predicate device. No new issues of safety or effectiveness are introduced by using this device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure with three heads, representing the department's focus on health and human well-being. The figure is positioned to the right of the department's name, which is arranged in a circular pattern around the logo. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 2 8 2004
Mr. Tony Spaccarotella Director. OA/RA Aktina Medical Corporation 360 North Route 9W CONGERS NY 10920
Re: K040709 Trade/Device Name: Protos Software (Model 11-860) Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle Radiation therapy system Regulatory Class: II Product Code: 90 IYE Dated: March 15, 2004 Received: March 18, 2004
Dear Mr. Spaccarotella:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number: にの中の709
Device Name: Protos Software
Indications for Use:
External Beam Radiation Therapy is typically delivered to patients from a linear External Beatli Rathanon Therapy is typically over from a radioactive source (high and accelerator with entire paoton of encircal Physics Corporation has designed and low dose rate valiants). "AIX FINAL Privation of the delivery process prior to manufactured software to and in the vermission is a compares the predicted data from a
radiotherapy treatment. The software analyzes and compares the pressurer and radiotherapy treatment. The Software und.7200 and volupered dose measurements.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use: V or
Over-The Counter Use:
(Per 21 CFR 801.109)
David R. Segmen
(Division Sign-Off) (Division of Reproductive, Abdon and Radiological Devices 510(k) Number _
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.