K040680 · Smith & Nephew, Inc. · MEH · Jun 14, 2004 · Orthopedic
Device Facts
Record ID
K040680
Device Name
CONTOUR HA COATED RECONSTRUCTION RING
Applicant
Smith & Nephew, Inc.
Product Code
MEH · Orthopedic
Decision Date
Jun 14, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Total hip components are indicated for individuals undergoing primary or revision surgery where other treatments have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis; or noninflammatory degenerative joint disease such as osteoarthritis; avascular necrosis; slipped capital epiphysis; fused hip; fracture of the pelvis and diastrophic, with an extended drainage-free period, nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; fracture dislocation of the hip and correction of deformity. Acetabular reinforcement and reconstruction rings are intended to be used in primary and revision surgeries where the acetabular roof, anterior or posterior column has deficiencies as a result of the indications listed previously. The device is intended for single use.
Device Story
Contour HA Coated Reconstruction Ring is a surgical implant used in primary or revision hip arthroplasty. Device features a partial cup shape with multiple screw holes and three flanges. Implant is secured to host or graft bone using bone screws; a Reflection All Poly Cup is cemented into the ring using PMMA bone cement. Used by orthopedic surgeons in an operating room setting to address acetabular deficiencies, including roof, anterior, or posterior column defects, or protrusion. Device provides structural reinforcement for the acetabulum, facilitating stable hip joint reconstruction. Single-use device.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Partial cup-shaped reconstruction ring with multiple screw holes and three flanges. Features hydroxyapatite (HA) coating. Designed for use with PMMA bone cement for acetabular cup fixation. Single-use, non-active implant.
Indications for Use
Indicated for patients undergoing primary or revision hip surgery due to trauma, inflammatory joint disease (e.g., rheumatoid arthritis), noninflammatory degenerative joint disease (e.g., osteoarthritis), avascular necrosis, slipped capital epiphysis, fused hip, pelvic fracture, nonunion, femoral neck/trochanteric fractures, or hip fracture dislocation requiring deformity correction and acetabular reinforcement.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
Smith & Nephew HA Ti-nua and Acetabular Components (K922621)
Smith & Nephew InterFit HA Coated Acetabular Shell (K990666)
Submission Summary (Full Text)
{1}------------------------------------------------
K040680
/tf/
JUN 1 4 2004
# 510(K) SUMMARY contour ha coated reconstruction ring
- SUBMITTER'S NAME: SUBMITTER'S ADDRESS: SUBMITTER'S TELEPHONE NUMBER: CONTACT PERSON: DATE SUMMARY PREPARED: TRADE OR PROPRIETARY DEVICE NAME: COMMON OR USUAL NAME: CLASSIFICATION NAME: DEVICE CLASS: PANEL CODE:
Smith & Nephew, Inc., Orthopaedic Division 1450 Brooks Road, Memphis, TN 38116 901-399-6670 John Reabe March 8, 2004 Contour HA Coated Reconstruction Ring Surgical Mesh Surgical Mesh Class II General and Plastic Surgery/79
## DEVICE INFORMATION:
#### INTENDED USE:
Notel boo.
Total hip components are indicated for individuals undergoing primary or revision surgery where Than hip components are indication for in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis; or noninflammatory degenerative joint imaninatory John disease Such as mountations such as osteoarthritis; avascular necrosis; disease (Nibbia) of thritis; slipped capital epiphysis; fused hip; fracture of the pelvis and diastrophic tradination of this , shpped daptial opply , with an extended drainage-free period, nonunion, femoral neck vanure and trochanteric fractures of the proximal femur with head involvement that are fracture and troonantono fractures; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip and correction of deformity.
Acetabular reinforcement and reconstruction rings are intended to be used in primary and revision Acelabular romore mort and reserved as the deficiencies of the acetabular roof, anterior or posterior Surgenes witcre the doctabularin has aresult of the indications listed previously. The device is intended for single use.
#### DEVICE DESCRIPTION:
The Contour HA Coated Reconstruction Ring features a partial cup shape, multiple screw holes and three flanges. The reconstruction ring is secured to host bone or graft bone with bone screws. A Reflection All Poly Cup may be cemented in the reconstruction ring with polymethylmethacrylate (PMMA) bone cement.
#### SUBSTANTIAL EQUIVALENCE INFORMATION:
The Contour HA Coated Reconstruction Ring is substantially equivalent to the Smith & Nephew The Oonlour TH Oodtod Rooment Rings (K962541), Osteonics Restoration Acetabular Ring Rencellen Acolubalar Romersonione & K962007). The Contour HA Coated Reconstruction Ring has a HA coating that is substantially equivalent to the HA coating on the Smith & Nephew HA Tinua and Acetabular Components (K922621) and the Smith & Nephew InterFit HA Coated Acetabular Shell (K990666).
### SUMMARY OF TECHNOLOGICAL COMPARISON:
The Contour HA Coated Reconstruction Ring is substantially equivalent to the predicate devices I rro our the previous section in terms of material, indications for use and design.
{2}------------------------------------------------
Image /page/1/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of three wavy lines, which are likely meant to represent the department's mission of promoting health and well-being. The seal is black and white.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 1 4 2004
John Reabe Smith & Nephew, Inc. 1450 Brooks Road Memphis, Tennessee 38116
Re: K040680
Trade/Device Name: Contour HA Coated Reconstruction Ring Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: II Product Code: MEH Dated: March 10, 2004 Received: March 16, 2004
Dear Mr. Reabe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for association to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, de reco that hat t (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may) ateres, mains of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - John Reabe
This letter will allow you to begin marketing your device as described in your Section 510(k)
The start of the market of a local starticle and worlds of your device to Jegal This letter will anow you to begin marketing your artist equivalence of your device to a legally
premarket notification. The FDA finding of substantial equivalence of thus, premarket notification. The PDA inding of substanted vice and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
rs the regulation and 160 m 160 m the results and are not the rogulation en If you desire specific advice for your act 100 mease note the regulation entitled, the regulation entitled, contact the Office of Complaned at (301) 64 - 17 - 4 - 1 807.97). You may obtain "Misbranding by relerence to premarket notification of Set from the Division of Small
other general information on your responsibilities under the Act from worker (800) 638, other general Information on your responsion.com its toll-free number (800) 638-2041 or
Manufacturers, International and Consumer Assistance of Armanian http://www.ain.html Manufacturers, International and Combailers http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
\$\Lambda\$
# 510(k) number (if known): _ K 6 40 (80
Device Name: Contour HA Coated Reconstruction Ring
## Indications for Use:
Total hip components are indicated for individuals undergoing primary or revision surgery Total hip components are indicated fore failed in rehabilitating hips damaged as a result of where other treatments of donose nuch as rheumatoid arthritis; or noninflammatory trauma, inniammatory joint disease (NIDJD) or any of its composite diagnoses such as degencrative joint diodoo (KTBCB) enthritis; slipped capital epiphysis; fused hip; Usteoanthilis, avasodial Hoorono, travant; old, remote osteomyelitis with an extended fracture of the period; nonunion; femoral neck fracture and trochanteric fractures of the uranage-free perfou, nonanien, remat that are unmanageable using other techniques; proximal lemur with houd involvement. fracture dislocation of the hip and correction of deformity.
Acetabular reinforcement and reconstruction rings are intended to be used in primary and Acetabular reinforoomont and rooshulum has the deficiencies of the acetabular roof, revision surgenos while the acolable leficiency, and/or protrusion as a result of the indications listed previously. The device is intended for single use.
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-the-Counter Use
(Optional Format 1-2-96)
uriani C. Provost
(Division Sign-Off) Division of General, Restorative. and Neurological Devices
**510(k) Number** K040680
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.