ORAVIVE TOOTH REVITALIZING PASTE

K040473 · Novamin Technology, Inc. · LBH · Mar 10, 2004 · Dental

Device Facts

Record IDK040473
Device NameORAVIVE TOOTH REVITALIZING PASTE
ApplicantNovamin Technology, Inc.
Product CodeLBH · Dental
Decision DateMar 10, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3260
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Oravive™ is a fluoride-free toothpaste product intended for cleaning the tooth surface on a daily basis. Clinical studies have shown that Oravive™ also provides relief from tooth sensitivity due to cold, heat, acids, sweets, or contact by the occlusion of dentin tubules.

Device Story

Oravive™ is a fluoride-free, non-aqueous toothpaste containing NovaMin® (calcium sodium phosphosilicate). Designed for daily home use by patients to clean teeth and manage dentin hypersensitivity. Upon exposure to aqueous oral environment, NovaMin® particles undergo surface reaction to form hydroxycarbonate apatite (HCA), physically occluding exposed dentin tubules and blocking hydrodynamic flow. This mechanism reduces patient discomfort from thermal, chemical, or tactile stimuli. Healthcare providers recommend the paste for daily use; efficacy is monitored via patient-reported sensitivity reduction. Benefits include improved quality of life through decreased pain associated with sensitive teeth.

Clinical Evidence

Randomized, double-blind, placebo-controlled 6-week clinical trial (n=75). Evaluated sensitivity reduction using 10cm visual analog scale (VAS) following air blast and cold water stimulation. Results showed significant reduction (p < .01) in VAS scores for Oravive™ compared to placebo, with 36-39% reduction in sensitivity. Performed equivalently to SrCl2-containing control. Bench testing included in vitro demin block model (statistically significant tubule occlusion) and RDA abrasion testing (mean 125.55, within safe limits).

Technological Characteristics

Non-aqueous toothpaste containing calcium sodium phosphosilicate (NovaMin®). Mechanism: surface reaction forming hydroxycarbonate apatite (HCA) to occlude dentinal tubules. Relative Dentin Abrasivity (RDA) value: 125.55. Daily-use formulation.

Indications for Use

Indicated for daily cleaning of tooth surfaces and relief of tooth sensitivity caused by cold, heat, acids, sweets, or contact in patients experiencing dentin hypersensitivity.

Regulatory Classification

Identification

Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.

Special Controls

*Classification.* Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ MAY 1 0 2004 **510(k) Premarket Notification** NovaMin Technology, Inc. Oravive™ Tooth Revitalizing Paste the first the state of the states . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . #### 1. SUBMITTER INFORMATION: | Name: | NovaMin Technology, Inc. | |------------|---------------------------------------------------------| | Address: | 13709 Progress Blvd., #23<br>Alachua, Florida 32615 USA | | Phone: | (386) 418-1551 | | Facsimile: | (386) 418-1465 | | Contact: | David C. Greenspan, Ph.D. | February 23, 2004 Preparation Date: ### 2. DEVICE NOMENCLATURE: | Trade Name: | Oravive™ Tooth Revitalizing Paste | |----------------------|-----------------------------------------| | Common Name: | Dentifrice, Toothpaste | | Classification Name: | Agent, Polishing, Abrasive, Oral Cavity | ### 3. LEGALLY MARKETED PREDICATE DEVICE: | Device Name: | Butler GUM® Prophylaxis Paste with NovaMin® | |----------------|---------------------------------------------| | 510(k) Number: | K024343 | | Applicant: | USBiomaterials Corporation | ### 4. DEVICE DESCRIPTION: Oravive™ is a daily-use, fluoride-free toothpaste device that incorporates NovaMin® as its active ingredient. The non-aqueous formulation is designed to clean your teeth as well as give your whole mouth a fresh feeling we call REVITALIZING! Oravive™ is also designed to physically occlude dentin tubules for the management of sensitive teeth. NovaMin® (calcium sodium phosphossilicate) is composed of elements that occur naturally in the body (Ca, Na, Si, P, and O). When exposed to an aquoous environment, NovaMin® undergoes a rapid surface reaction, allowing it to physically occlude tubules. Within a short period of time, essentially all of the NovaMin® reacts to form hydroxycarbonate apatite (HCA), which is chemically and structurally similar to natural tooth mineral. ## S. INTENDED USE: ﺮ Oravive™ is a fluoride-free toothpaste product intended for cleaning the tooth surface on a daily basis. Clinical studies have shown that Oravive™ also provides relief from tooth sensitivity due to cold, heat, acids, sweets, or contact by the occlusion of dentin tubules. {2}------------------------------------------------ #### 6. TECHNOLOGICAL CHARACTERISTICS: The technological characteristics of Oravive™ and Butler GUM® Prophylaxis Paste with NovaMine are very similar. Both devices are designed to relieve hypersensitivity associated with exposed dentin by the deposition of a calcium phosphate layer onto the tooth surface. Both devices use NovaMia" to produce a calcium phosphate layer that occludes dentinal tubules and blocks hydrodynamic flow. The primary difference between the two devices is that Oravive™ supplies the calcium and phosphate ions on a daily basis, whereas Butler GUM® Prophylaxis Paste with NovaMin supplies the ions once every professional tooth cleaning. In addition, there are some differences in the proportion of the ingredients in the two products. #### 7. SAFETY AND PERFORMANCE DATA: Many different biocompatibility tests have been performed on NovaMin®, the active ingredient in Oravive The results of these tests indicate that there is no evidence of any hazardous effects to the patient if the product is used as directed. The tubule occlusion efficacy of Oravive™ was evaluated using an in vitro demin block model. The results indicate that Oravive™ occludes a statistically significant number of tubules when compared to controls. The relative abrasion level of Oravive™ was evaluated at Indiana University School of Dentistry. The procedure used was the ADA recommended procedure for determination of toothpaste abrasivity. The result was a mean Relative Dentin Abrasivity (RDA) value of 125.55, This RDA value is well under the limit considered safe for daily-use (RDA value < 250). A randomized, double-blind, placebo-controlled, six-week clinical trial was conducted to evaluate the efficacy of Oravive™ Tooth Revitalizing Paste with a placebo and a commercially available StCl2 desensitizing toothpaste. The study was approved by the Medical Ethics Committee of the School of Stomatology, Wuhan University, Wuhan, China. Written informed consem and medical history were obtained prior to entering patients into the study. A total of seventy-five (75) patients were accepted into the study; twenty-five (25) in each group. There were no significant differences in the average ages between the test groups. Brushing regimen was twice daily for a six-week period. Tooth sensitivity was recorded by marking the degree of discomfort experienced by the patient on a 10cm visual analog scale (VAS) after stimulation of selected teeth by cold water and a metered air blast. The results were analyzed for statistical significance using ANOVA and Bonferonni post hoc tests. Sixty-seven (67) patients completed the six-week study. No adverse events in any of the groups were reported. The results showed that the mean values of the VAS scores resulting from the air stimulus and cold water were reduced significantly (p < . 01) in both the test and positive control groups compared with the placebo group after six weeks of tooth brushing. Reduction in VAS for the NovaMin" containing toothpaste at six weeks was 36% for the air stimulus and 39% for the cold water stimulus. The results of this study showed that the NovaMin® containing toothpaste reduced dentin sensitivity significantly over a six-week time period, and that it performed equivalently to a SrCl2 containing, commercially available desensitizing toothpaste. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a bird in flight, composed of three curved lines. MAY 1 0 2004 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Dr. David C. Greenspan Vice President & Chief Technology Officer NovaMin Technology, Incorporated 13709 Progress Boulevard #23 Alachua, Florida 32615 USA Re: K040473 Trade/Device Name: Oravive™ Tooth Revitalizing Paste Regulation Number: Unclassified Regulation Name: None Regulatory Class: None Product Code: NRE Dated: February 23, 2004 Received: February 24, 2004 Dear Dr. Greenspan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially cquivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the gencral controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 -Dr. Greenspan Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours. Ques Chiu Lin, Ph.D Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ 510(k) Premarket Notification NovaMin Technology, Inc. Oravive™ Tooth Revitalizing P # OF INDICATIONS FOR USE 510(k) Number (if known): K040473 Oravive™ Tooth Revitalizing Paste Device Name: -- INDICATIONS FOR USE: Oravive™ is a fluoride-free toothpaste product intended for cleaning the tooth surface on a Oravive - 15 & incelides have shown that Oravive™ also provides relief from tooth sensitivity due to cold, heat, acids, sweets, or contact by the occlusion of dentin tubules. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Suser (Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices 510(k) Number: Prescription Use ------- OR (Per 21 CFR 801.109) Over-The-Counter Use X SECTION D -- STATEMENT OF INDICATIONS FOR USE
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...