APAPRO Desensitizer Homecare is a fluoride-free daily-use cleaning toothpaste that also provides relief from tooth sensitivity due to cold, heat, acids, sweets or contact, through its action of occluding dentin tubules through surface coating, promoting an environment conducive to remineralization.
Device Story
Fluoride-free toothpaste; contains hydroxyapatite as active ingredient. Used daily at home; applied with toothbrush. Mechanism: hydroxyapatite deposits on tooth surface; acts as template for calcium/phosphate ion deposition; promotes remineralization; physically occludes exposed dentin tubules. Occlusion blocks hydrodynamic flow; reduces tooth hypersensitivity. Healthcare provider/patient uses output (reduced sensitivity) to manage oral comfort. Benefits: daily relief from sensitivity; restoration of tooth hydroxyapatite structure.
Clinical Evidence
Bench testing only. SEM micrographs compared dentin tubule occlusion between subject and predicate. Quantitative testing showed 93.1% occlusion rate for subject device vs 91.5%-93.2% for predicate. Relative abrasivity (RDA) measured at 23.42 ± 0.59, well below 250 threshold for daily use.
Technological Characteristics
Active ingredient: hydroxyapatite. Non-sterile. Form factor: toothpaste in aluminum-laminate tube. Mechanism: physical occlusion of dentin tubules via hydroxyapatite surface coating. Biocompatible per ISO 10993. RDA value 23.42.
Indications for Use
Indicated for adult and pediatric populations for relief of tooth sensitivity due to cold, heat, acids, sweets, or contact via dentin tubule occlusion.
Regulatory Classification
Identification
Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.
Special Controls
*Classification.* Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.
Oralief OTC Toothpaste for Sensitive Teeth (K080228)
Submission Summary (Full Text)
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 18, 2026
Sangi Co., Ltd.
% Dave Yungvirt
CEO
Third Party Review Group, LLC
7 Giralda Farms, Suite 120a
Madison, New Jersey 07940
Re: K260830
Trade/Device Name: APAPRO Desensitizer Homecare (Remineralizing Anti-sensitivity Toothpaste)
Regulation Number: 21 CFR 872.3260
Regulation Name: Cavity varnish
Regulatory Class: Class II
Product Code: LBH
Dated: March 13, 2026
Received: March 13, 2026
Dear Dave Yungvirt:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K260830 - Dave Yungvirt
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{3}
K260830 - Dave Yungvirt
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K260830 | |
| Device Name APAPRO Desensitizer Homecare | |
| Indications for Use (Describe) APAPRO Desensitizer Homecare is a fluoride-free daily-use cleaning toothpaste that also provides relief from tooth sensitivity due to cold, heat, acids, sweets or contact, through its action of occluding dentin tubules through surface coating, promoting an environment conducive to remineralization. Intended Population: For use in Adult and Pediatric Population | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{5}
Sangi Co., Ltd.
Traditional 510(k)
APAPRO Desensitizer Homecare
APAPRO Desensitizer Homecare
K260830
Submitter Information:
Name: Sangi Co., Ltd.
Address: 3-11-6 TSUKIJI, CHUO-KU, TOKYO 104-8440, JAPAN
Phone: +81 03-3545-6013
Email: regulatory@sangi-co.com
Facsimile: +81 3-3543-3651
Contact Person: Tetsuya Yoshioka
Preparation Date: 18 Mar 2026
1. Device Nomenclature:
Trade Name: APAPRO Desensitizer Homecare
Common Name: Remineralizing Anti-sensitivity Toothpaste
Classification Name: Cavity Varnish (21 CFR 872.326 Product code LBH)
2. Legally Marketed Predicate Device:
Device Name: APAPRO Desensitizer
510(k) Number: (K220419)
Applicant: Sangi Co., Ltd.
3. Legally Marketed Reference Device:
Device Name: Oralief OTC Toothpaste for Sensitive Teeth
510(k) Number: (K080228)
Applicant: Novamin Technology, Inc.
4. Device Description:
APAPRO Desensitizer Homecare is a fluoride-free daily-use toothpaste device that contains hydroxyapatite, the main component of teeth, as its active ingredient. The
Page1
{6}
Sangi Co., Ltd.
Traditional 510(k)
APAPRO Desensitizer Homecare
formulation is designed to clean teeth as well as to help restore a tooth's hydroxyapatite structure through surface coating, promoting an environment conducive to remineralization, which results in physical occlusion of exposed dentin tubules, providing relief from tooth hypersensitivity. The hydroxyapatite deposited into the dentin tubules by APAPRO Desensitizer Homecare acts as a template for the further deposition of large amounts of calcium and phosphate ions, promoting crystal integrity and growth.
## 5. Indication/ Intended Use:
APAPRO Desensitizer Homecare is a fluoride-free daily-use cleaning toothpaste that also provides relief from tooth sensitivity due to cold, heat, acids, sweets or contact, through its action of occluding dentin tubules through surface coating, promoting an environment conducive to remineralization.
## 6. Technological Comparison:
- Mechanism of Action: Both the subject (APAPRO Desensitizer Homecare) and the predicate (APAPRO Desensitizer) depend on deposition of hydroxyapatite to occlude dentinal tubules.
- Concentration: APAPRO Desensitizer Homecare uses a somewhat lower concentration of hydroxyapatite, but the mechanism remains the same.
- Usage Context: APAPRO Desensitizer Homecare is designed for repeated, everyday OTC use with a toothbrush; the predicate is for professional in-office application using rubber cup or cotton swab. This difference is reflected to some extent in the formulation but there is no difference in the fundamental technology.
## 7. Safety and Performance Data:
The ability of APAPRO Desensitizer Homecare to reduce tooth hypersensitivity was evaluated through non-clinical testing. The mechanism of action is based on restoration of the tooth's hydroxyapatite structure through surface coating, promoting an environment conducive to remineralization, resulting in occlusion of exposed dentin tubules.
Non-clinical testing demonstrated that APAPRO Desensitizer Homecare physically occludes exposed dentin tubules. Scanning electron microscopy (SEM) micrographs of dentin samples treated with APAPRO Desensitizer Homecare were compared with those treated with the predicate device, APAPRO Desensitizer. Both products demonstrated comparable dentin tubule occlusion, primarily through deposition of a hydroxyapatite layer on the dentin surface.
Quantitative dentinal tubule occlusion testing showed an average occlusion rate of 93.1% for APAPRO Desensitizer Homecare. In comparison, the predicate device demonstrated an average dentin tubule occlusion rate of 91.5% when applied using a rubber cup and 93.2% when applied using a cotton swab. These results indicate that APAPRO Desensitizer Homecare performs comparably to the predicate device with respect to dentin tubule occlusion.
The relative abrasivity of APAPRO Desensitizer Homecare was evaluated at the Indiana University School of Dentistry using the American Dental Association (ADA) recommended procedure for determination of toothpaste abrasivity. The product demonstrated a mean Radioactive Dentin Abrasion (RDA) value of 23.42 ± 0.59. This RDA value is well below the generally accepted safety threshold for daily use (RDA < 250), indicating that APAPRO Desensitizer Homecare is safe for routine oral care use.
Page2
{7}
Sangi Co., Ltd.
Traditional 510(k)
APAPRO Desensitizer Homecare
8. Device Comparison Table:
| Descriptive Information | Subject Device: APAPRO Desensitizer Homecare | Predicate Device: APAPRO Desensitizer (K220419) | Reference Device: Oralief OTC Toothpaste for Sensitive Teeth (K080228) | Differences |
| --- | --- | --- | --- | --- |
| Classification Name | Cavity Varnish | Cavity Varnish | Cavity Varnish | Same |
| Regulatory Class | Class II | Class II | Class II | Same |
| Product Code | LBH | LBH | LBH | Same |
| Device Description | APAPRO Desensitizer Homecare is a fluoride-free daily-use toothpaste device that contains hydroxyapatite, the main component of teeth, as its active ingredient. The formulation is designed to clean teeth as well as to help restore a tooth's hydroxyapatite structure through surface coating, promoting an environment conducive to remineralization, which results in physical occlusion of exposed dentin tubules, providing relief from tooth hypersensitivity. The hydroxyapatite deposited into the dentin tubules by APAPRO Desensitizer Homecare acts as a template for the further deposition of large amounts of calcium and phosphate ions, promoting crystal integrity and growth. | APAPRO Desensitizer is a fluoride-free tooth desensitizing paste that utilizes hydroxyapatite, the main component of teeth, as its active ingredient to relieve tooth sensitivity by physically occluding exposed dentin tubules. The hydroxyapatite deposited into dentin tubules by APAPRO Desensitizer then acts as a template for the further deposition of large amounts of calcium and phosphate ions, promoting crystal integrity and growth. | Oralief™ OTC Toothpaste for Sensitive Teeth is a daily-use, fluoride-free toothpaste device that incorporates NovaMin® (calcium sodium phosphosilicate) as its active ingredient. The non-aqueous formulation is designed to clean teeth as well as to physically occlude dentin tubules for the reduction of tooth sensitivity. When exposed to an aqueous environment, NovaMin® undergoes a rapid surface reaction, allowing it to physically occlude tubules. Within a short period of time, essentially all of the NovaMin reacts to form hydroxycarbonate apatite (HCA), which is chemically and structurally similar to natural tooth mineral. | Similar / Similar |
| Indication/Intended Use | APAPRO Desensitizer Homecare is a fluoride-free daily-use toothpaste that also provides relief from tooth sensitivity due to cold, heat, acids, sweets or contact, through its action of occluding dentin tubules through surface coating, promoting an environment conducive to remineralization. | APAPRO Desensitizer is a fluoride-free paste that is indicated to provide relief from tooth hypersensitivity resulting from cold, heat, acids, sweets, or contact, through its action of dentin tubule occlusion. | Oralief OTC Toothpaste for Sensitive Teeth is a fluoride-free daily use cleaning toothpaste that also provides rapid and continual relief from tooth sensitivity due to cold, heat, acids, sweets or contact, through its action of the occlusion of dentin tubules. | Similar / Similar |
Page3
{8}
Sangi Co., Ltd.
Traditional 510(k)
APAPRO Desensitizer Homecare
| Descriptive Information | Subject Device: APAPRO Desensitizer Homecare | Predicate Device: APAPRO Desensitizer (K220419) | Reference Device: Oralie/OTC Toothpaste for Sensitive Teeth (K080228) | Difference |
| --- | --- | --- | --- | --- |
| Packaging | 4.23oz (120g) aluminum-laminate tube. The tube is enclosed in an individual paperboard box and placed in a cardboard carton for shipment. | 1.94oz (55g) aluminum-laminate tube. The tube is enclosed in an individual paperboard box, which is then shrink-wrapped in PE, and placed in a cardboard carton for shipment. | Not disclosed | Different |
| Intended Population | For use in Adult and Pediatric Population | For use in Adult Population | Not disclosed | Different |
| Delivery Form | Daily-use over-the-counter toothpaste for application with a toothbrush. | Applied with a cotton bud or swab, in a customized dental tray, or using a rubber cup attached to a dental handpiece at low (>1000 rpm) revolution speed. | Daily-use over-the-counter toothpaste for application with a toothbrush. | Similar / Same |
| Mode of Action | APAPRO Desensitizer Homecare directly occludes dentin tubules by depositing a crystalline layer of synthetic hydroxyapatite directly onto the tooth surface, promoting an environment conducive to remineralization, which physically occludes exposed dentin tubules, resulting in desensitization. | APAPRO Desensitizer directly occludes dentin tubules by depositing a crystalline calcium phosphate layer, i.e. introducing synthetic hydroxyapatite directly onto the tooth surface, resulting in the physical occlusion of exposed dentin tubules, which results in desensitization | Calcium phosphosilicate particles react with the user's saliva in an aqueous environment to release calcium and phosphate ions, which forms a layer of hydroxycarbonate apatite (calcium phosphate) over time. The calcium phosphate layer physically occludes exposed dentinal tubules and blocks hydrodynamic flow to achieve the desensitization effect. | Same / Similar |
| Anatomical Site | Oral Cavity | Oral Cavity | Oral Cavity | Same |
| Application Time | Twice per day: 2 minutes brushing, spit out, light rinse (or no rinse) | Single application: 30 seconds per tooth | Daily frequency not disclosed | Same |
| Nature of Body Contact | Long term (>30 days) | Limited (<24 hours) | Long term (>30 days) | Different / Same |
| Composition | Calcium phosphate (Hydroxyapatite) + Inactive Ingredients | Calcium phosphate (Hydroxyapatite) + Inactive Ingredients | NovaMin® (calcium sodium phosphosilicate) + Inactive Ingredients | Same / Similar |
| Biocompatibility | Biocompatible, non-antigenic | Biocompatible, non-antigenic | Biocompatible, non-antigenic | Same |
| Sterility | Non-sterile | Non-sterile | Non-sterile | Same |
| Shelf-life | 3 years | 3 years | Not disclosed | Same |
| Prescription / OTC | OTC | Prescription | OTC | Different / Same |
Page4
{9}
Sangi Co., Ltd.
Traditional 510(k)
APAPRO Desensitizer Homecare
## 9. Biocompatibility:
Biocompatibility testing in accordance with ISO 10993 was previously performed on APAPRO Desensitizer, the predicate device. The formulation of APAPRO Desensitizer Homecare is similar to that of the predicate device, with the addition of certain ingredients that are commonly used in commercially marketed toothpaste products.
Based on the similarity in formulation, the established history of safe use of the additional ingredients, and the biocompatibility data available for the predicate device, no additional biocompatibility testing was conducted for APAPRO Desensitizer Homecare. A biocompatibility justification, including the predicate device test reports and rationale for not performing additional testing on the added ingredients, has been provided in the 510(k) submission.
Based on this information, APAPRO Desensitizer Homecare is considered biocompatible and does not pose an unacceptable biological risk to users when used as intended in both adult and pediatric populations.
## 10. Conclusion:
APAPRO Desensitizer Homecare is considered substantially equivalent to the legally marketed predicate device, APAPRO Desensitizer. Both products are indicated for the relief of tooth hypersensitivity through the formation of a calcium hydroxyapatite layer on the tooth surface, promoting an environment conducive to remineralization, which physically occludes dentin tubules.
Additionally, both products have demonstrated comparable effectiveness in dentin tubule occlusion. No additional safety concerns were identified when comparing APAPRO Desensitizer Homecare to APAPRO Desensitizer.
Page5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.