LYPHOCHEK HEMOSTASIS CONTROL

K040275 · Bio-Rad · GGN · Feb 27, 2004 · Hematology

Device Facts

Record IDK040275
Device NameLYPHOCHEK HEMOSTASIS CONTROL
ApplicantBio-Rad
Product CodeGGN · Hematology
Decision DateFeb 27, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.5425
Device ClassClass 2

Intended Use

Lyphochek® Hemostasis Control is intended for use as an assayed quality control plasma to monitor the precision of laboratory testing procedures for analytes listed in the package insert. [See “B. Analyte(s):” above]

Device Story

Lyphochek® Hemostasis Control is a lyophilized human plasma-based quality control product used to monitor precision of hemostasis laboratory testing. It contains purified biochemicals and preservatives. The device is provided in 12 x 1ml vials across three levels. It is used by laboratory technicians in clinical settings to verify the performance of coagulation assays, including APTT, PT, AT III, TT, FIB, Protein C/S, Plasminogen, D-dimer, and various clotting factors. The control is reconstituted and processed on instruments from manufacturers such as BioMerieux, Chromogenix, Dade Behring, Diagnostica Stago, Instrumentation Laboratories/Hemoliance, and Roche. By comparing observed results against established assay values, clinicians ensure the accuracy and reliability of patient coagulation test results, supporting clinical decision-making in hemostasis management.

Clinical Evidence

No clinical data. Stability studies were performed to establish shelf life (three years at 2-8°C) and reconstituted stability (8 hours at 2-25°C, with Protein S stable for 8 hours at 2-8°C).

Technological Characteristics

Lyophilized human plasma matrix with added purified biochemicals and preservatives. Tri-level control format. Stability: 3 years at 2-8°C; 8 hours reconstituted at 2-25°C. Compatible with various coagulation instrument platforms.

Indications for Use

Indicated for use as an assayed quality control plasma to monitor precision of laboratory coagulation testing procedures. Intended for clinical laboratory use.

Regulatory Classification

Identification

A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.

Special Controls

*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Page 1 of 2 K040275 ### Summary of Safety and Effectiveness FEB 2 7 2004 Lyphochek® Hemostasis Control #### 1.0 Submitter Bio-Rad Laboratories 9500 Jeronimo Road, Irvine, California 92618-2017 (949) 598-1200 Telephone: (949) 598-1557 Fax: ### Contact Person Maria Zeballos Requlatory Affairs Specialist (949) 598-1367 Telephone: ## Date of Summary Preparation January 30, 2004 #### 2.0 Device Identification | Product Trade Name: | Lyphochek® Hemostasis Control | |---------------------|------------------------------------| | Common Name: | Control Plasma Normal and Abnormal | | Classifications: | Class II | | Product Code: | GGC | | Regulation Number: | 21 CFR 864.5425 | #### Device to Which Substantial Equivalence is Claimed 3.0 Lyphochek Hemostasis Control Bio-Rad Laboratories Irvine, California 510(k) Number: K020878 #### Description of Device 4.0 Lyphochek® Hemostasis Control is prepared from human plasma with added purified biochemicals and preservatives. The control is provided in lyophilized form for increased stability. #### Statement of Intended Use 5.0 Lyphochek® Hemostasis Control is intended for use as a quality control plasma to monitor the precision of laboratory analytes listed in the package insert. {1}------------------------------------------------ #### Comparison of the new device with the Predicate Device 6.0 Table 1 (below) contains comparison information of similarities and differences between the new Lyphochek Hemostasis Control and the currently marketed Lyphochek Hemostasis Control ((020878) to which substantial equivalence is claimed. The new Lyphochek Hemostasis Control is a tri-level product (Levels 1, 2 and 3) and contains D-dimer. The Lyphochek Hemostasis Control is a bi-levels 1 and 2) product and does not contain D-dimer. | Characteristics | Bio-Rad<br>Lyphochek® Hemostasis Control<br>(New Device) | Bio-Rad<br>Lyphochek® Hemostasis Control<br>(Predicate DeviceK020878) | |-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | <b>Similarities</b> | | | | Intended Use | Lyphochek® Hemostasis Control is intended for use as<br>a quality control plasma to monitor the precision of<br>laboratory testing procedures for analytes listed in the<br>package insert. | Lyphochek® Hemostasis Control is intended for use as a<br>a quality control plasma to monitor the precision of<br>laboratory testing procedures for analytes listed in the<br>package insert. | | Form | Lyophilized | Lyophilized | | Matrix | Human plasma based | Human plasma based | | Storage (Unopened) | 2-8°C until expiration date | 2-8°C until expiration date | | Reconstituted Vial<br>Claim | 8 hours at 2-25°C with the following exception:<br>Protein S will be stable for 8 hours at 2- 8°C. | 8 hours at 2-25°C with the following exception:<br>Protein S will be stable for 8 hours at 2- 8°C. | | <b>Differences</b> | | | | Levels | Levels 1, 2 and 3 | Levels 1 and 2<br>Does not contain Level 3 | | Analytes | Contains: Prothrombin Time (PT), Activated Partial<br>Thromboplastin Time (APTT), Fibrinogen, Antithrombin<br>III (AT III) Thrombin Time (TT), Factor II, V, VII, VIII, IX,<br>X, XI, XII, Protein S (Functional) Protein C (Functional),<br>Plasminogen and D-dimer | Contains: Prothrombin Time (PT), Activated Partial<br>Thromboplastin Time (APTT), Fibrinogen, Antithrombin III<br>(AT III) Thrombin Time (TT), Factor II, V, VII, VIII, IX, X, XI,<br>XII, Protein S (Functional) and Protein C (Functional),<br>Plasminogen<br>Does not Contain: D-dimer | Table 1. Similarities and Differences between new and predicate device. #### 7.0 Summary of Performance Data Stability studies have been performed to determine the reconstituted stability and shelf life for the Lyphochek® Hemostasis Control. Product claims are as follows: - Reconstituted Stability: All analytes will be stable for 8 hours at 2 -25℃ with the exception of 2.1 Protein S which will be stable for 8 hours at 2 - 8℃. - 2.2 Shelf Life: Three years when stored at 2 8 °C - 2.3 Real time studies will be ongoing to support the shelf life of this product. All supporting data is retained on file at Bio-Rad Laboratories. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines. ## FEB 27 2004 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Ms. Elizabeth Platt Regulatory Affairs/Quality Assurance Manager Bio-Rad Laboratories, QSD 9500 Jeronimo Road Irvine, California 92618-2017 k040275 Rc: Trade/Device Name: Lyphochek Hemostasis Control Regulation Number: 21 CFR § 864.5425 Regulation Name: Multipurpose System for in vitro coagulation studies Regulatory Class: II Product Code: GGN Dated: January 30, 2004 Received: February 5, 2004 Dear Ms. Platt: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. {3}------------------------------------------------ Page 2 If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Joseph L. Arallett Joseph L. Hackett, Ph.D. Acting Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510 (k) Number (if known): K040275 Device Name: Lyphochek Hemostasis Control Indications for Use: For use as a quality control plasma to monitor the precision of laboratory testing procedures for the analytes listed in this package in on the Analytes listed in the package insert: - . Antithrombin III (AT III) - Activated Partial Thromboplastin Time (APTT) . - D-dimer ◆ - . Factor II - Factor V . - Factor VII . - Factor Vill . - Factor IX . - Factor XI . - Factor XII ◀ - . Fibrinogen - . Plasminogen - Protein C (Functional) . - Protein S (Functional) . - Prothrombin Time (PT) - Thrombin Time (TT) Division Sign-Off # Office of In Vitro Diagnostic Device Evaluation and Safety (PLEASE DO NOT WRITE BELOW THE LINE-CONINUE ON ANOTHER PAGE IF NEEDED) 510(K)_________________________________________________________________________________________________ Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription use ***_***__ or Over-the Counter use
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...